← Product Code [JXN](/productcode/JXN) · K020787

# VERDICT-II PROPOXYPHENE (K020787)

_Medtox Diagnostics, Inc. · JXN · May 2, 2002 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K020787

## Device Facts

- **Applicant:** Medtox Diagnostics, Inc.
- **Product Code:** [JXN](/productcode/JXN.md)
- **Decision Date:** May 2, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3700
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology

## Indications for Use

VERDICT® II PROPOXYPHENE is a one-step immunochromatographic test for the rapid, qualitative detection of propoxyphene and its metabolite, norpropoxyphene, in human urine. The test detects the major urinary metabolite of propoxyphene (norpropoxyphene) at 300 ng/mL. It is not for over-the-counter sale. VERDICT®-II PROPOXYPHENE PROVIDES ONLY A PRELIMINARY ANALYTICAL TEST RESULT. A MORE SPECIFIC ALTERNATE CHEMICAL METHOD MUST BE USED IN ORDER TO OBTAIN A CONFIRMED ANALYTICAL RESULT. GAS CHROMATOGRAPHY/ MASS SPECTROMETRY (GC/MS) IS THE PREFERRED CONFIRMATORY METHOD. CLINICAL CONSIDERATION AND PROFESSIONAL JUDGMENT SHOULD BE APPLIED TO ANY DRUG OF ABUSE TEST RESULT, PARTICULARLY WHEN PRELIMINARY POSITIVE RESULTS ARE OBTAINED.

## Device Story

VERDICT®-II PROPOXYPHENE is a one-step immunochromatographic assay for qualitative detection of propoxyphene and norpropoxyphene in human urine. Device utilizes lateral flow immunoassay technology; sample applied to test device; presence of drug metabolite at or above 300 ng/mL threshold inhibits binding, resulting in visual signal change. Intended for professional use in clinical settings to provide preliminary analytical results. Results require confirmation via more specific chemical methods, preferably GC/MS. Clinical judgment required for interpretation of preliminary positive results.

## Clinical Evidence

No clinical data provided. Substantial equivalence established via bench testing of analytical performance, including sensitivity, specificity, and cutoff verification.

## Technological Characteristics

One-step immunochromatographic lateral flow assay. Qualitative visual readout. Detects norpropoxyphene at 300 ng/mL cutoff. In vitro diagnostic device.

## Regulatory Identification

A propoxyphene test system is a device intended to measure propoxyphene, a pain-relieving drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of propoxyphene use or overdose or in monitoring levels of propoxyphene to ensure appropriate therapy.

## Special Controls

*Classification.* Class II (special controls). A propoxyphene test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the words "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" arranged in a circular pattern around a central emblem. The emblem consists of a stylized caduceus, a symbol often associated with medicine and healthcare, with three wavy lines representing the snakes intertwined around a staff.

## Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850.

Mr. Michael Turanchik Director, Research and Development MEDTOX Diagnostics, Inc. 1238 Anthony Road Burlington, NC 27215

Re:

k020787 R020707
Trade/Device Name: VERDICT® -II PROPOXYPHENE Regulation Number: 21 CFR 862.3700 Regulation Name: Propoxphene test system Regulatory Class: Class II Product Code: JXN Dated: March 8, 2002 Received: March 11, 2002

Dear Mr. Turnachik:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

MAY 0 2 2002

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Page:2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## EXHIBIT III VERDICT®-II PROPOXYPHENE 510(k) Submission

## INDICATIONS FOR USE FORM

510(k) Number (if known): _ KO20787

Device Name: VERDICT®-II PROPOXYPHENE

Indications for Use:

VERDICT® II PROPOXYPHENE is a one-step immunochromatographic test for the rapid, qualitative detection of propoxyphene and its metabolite, norpropoxyphene, in human urine. The test detects the major urinary metabolite of propoxyphene (norpropoxyphene) at 300 ng/mL. It is not for over-the-counter sale.

VERDICT®-II PROPOXYPHENE PROVIDES ONLY A PRELIMINARY ANALYTICAL TEST RESULT. A MORE SPECIFIC ALTERNATE CHEMICAL METHOD MUST BE USED IN ORDER TO OBTAIN A CONFIRMED ANALYTICAL RESULT. GAS CHROMATOGRAPHY/ MASS SPECTROMETRY (GC/MS) IS THE PREFERRED CONFIRMATORY METHOD. CLINICAL CONSIDERATION AND PROFESSIONAL JUDGMENT SHOULD BE APPLIED TO ANY DRUG OF ABUSE TEST RESULT, PARTICULARLY WHEN PRELIMINARY POSITIVE RESULTS ARE OBTAINED.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109) the text in the image is:

tan Coopes

(Division Sign-Off)
Division of Clinical Laboratory Levices
510(k) Number. K020787

---

**Source:** [https://fda-staging.innolitics.com/device/K020787](https://fda-staging.innolitics.com/device/K020787)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
