K020771 · ACON Laboratories, Inc. · DIO · May 15, 2002 · Clinical Toxicology
Device Facts
Record ID
K020771
Device Name
ACON ONE STEP DRUG SCREEN TEST CARD
Applicant
ACON Laboratories, Inc.
Product Code
DIO · Clinical Toxicology
Decision Date
May 15, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3250
Device Class
Class 2
Indications for Use
The ACON One Step Drug Screen Test Card and ACON One Step Drug Screen Test Card with Integrated Cup are rapid chromatographic immunoassay devices for the qualitative and simultaneous detection of Marijuana, Cocaine, Methamphetamine, Amphetamine, Opiates, Phencyclidine, Benzodiazepine, Methadone, and/or barbiturate in human urine. The test is calibrated using analytes with their respective cutoff concentrations listed in the following table. [Table omitted for brevity]. The configurations of the ACON One Step Multi-Drug Screen Test Card and ACON One Step Multi-Drug Screen Test Card with Integrated Cup come with any combination of the above listed drug analytes. They are intended for the healthcare professionals including professionals at point-of-care sites.
Device Story
Rapid chromatographic immunoassay for qualitative drug screening in human urine. Device utilizes competitive binding lateral flow technology with mouse monoclonal antibodies to detect specific drug analytes. Input is human urine specimen; output is visual colored-line presence/absence in test region. No instrumentation required. Used by healthcare professionals at point-of-care sites. Positive result indicated by absence of colored line in test region; negative result indicated by colored line. Procedural control line confirms proper specimen volume and membrane wicking. Provides preliminary analytical results to assist in clinical decision-making regarding drug presence.
Clinical Evidence
No clinical data provided; device relies on bench testing of analytical performance (cutoff concentrations and qualitative detection).
Indicated for healthcare professionals at point-of-care sites for qualitative, simultaneous detection of Marijuana, Cocaine, Methamphetamine, Amphetamine, Opiates, Phencyclidine, Benzodiazepine, Methadone, and barbiturates in human urine at specified cutoff concentrations.
Regulatory Classification
Identification
A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.
Special Controls
*Classification.* Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Submission Summary (Full Text)
{0}------------------------------------------------
# MAY 1 5 2002
K020771
#### 510(k) SUMMARY 7.
This 510 (k) summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The Assigned 510(k) number is K020771.
#### Submitter:
ACON Laboratories, Inc. 4108 Sorrento Valley Boulevard San Diego, California 92121
Tel.: 858-535-2030 Fax: 858-535-2038
#### Date:
March 4, 2002
### Contact Person:
Edward Tung, Ph.D.
#### Product Names:
ACON® One Step Drug Screen Test Card ACON® One Step Drug Screen Test Card with Integrated Cup
#### Common Name:
test for the qualitative detection of of Marijuana. Cocaine. Immunochromatographic Methamphetamine, Amphetamine, Opiates, Phencyclidine, Benzodiazepine, Methadone, and barbiturate in human urine.
#### Device Classification:
The ACON One Step Drug Screen Test Card ACON One Step Drug Screen Test Card with Integrated Cup is similar to other FDA-cleared devices for the qualitative and simultaneous detection of drugs in urine specimens. These drug test devices are used only to provide a preliminary analytical result (21 CFR 862.3650). The test systems have been classified as Class II devices with moderate complexity. Product codes LDJ, DIO, LAF, DKZ, DJG, LCM, JXM, DJR, and DIS have been assigned for the test system.
{1}------------------------------------------------
# Classification Name:
Marijuana, Cocaine, Methamphetamine, Amphetamine, Opiates, Phencyclidine, Benzodiazepine, Methadone, and barbiturate test systems
# Intended Use:
The ACON One Step Drug Screen Test Card and ACON One Step Drug Screen Test Card with Integrated Cup are rapid chromatographic immunoassay devices for the qualitative and simultaneous detection of Marijuana, Cocaine, Methamphetamine, Amphetamine, Opiates, Phencyclidine, Benzodiazepine, Methadone, and/or barbiturate in human urine. The test is calibrated using analytes with their respective cutoff concentrations listed in the following table.
| Test | Calibrator | Cut-off |
|-------------------------------|--------------------------------|-------------|
| Amphetamine (AMP) | D-Amphetamine | 1,000 ng/mL |
| Barbiturates (BAR) | Secobarbital | 300 ng/mL |
| Benzodiazepines (BZO) | Oxazepam | 300 ng/mL |
| Cocaine (COC) | Benzoylecgonine | 300 ng/mL |
| Methamphetamine (mAMP) | D-Methamphetamine | 1,000 ng/mL |
| Morphine (MOP 300 or OPI 300) | Morphine | 300 ng/mL |
| Methadone (MTD) | Methadone | 300 ng/mL |
| Opiates (OPI 2000) | Morphine | 2,000 ng/mL |
| Phencyclidine (PCP) | Phencyclidine | 25 ng/mL |
| Marijuana (THC) | 11-nor- $\Delta^9$ -THC-9 COOH | 50 ng/mL |
The configurations of the ACON One Step Multi-Drug Screen Test Card and ACON One Step Multi-Drug Screen Test Card with Integrated Cup come with any combination of the above listed drug analytes. They are intended for the healthcare professionals including professionals at point-of-care sites.
# Description:
The ACON One Step Drug Screen Test card are competitive binding, lateral flow immunochromatographic assays for the qualitative and simultaneous screening of Marijuana, Cocaine, Methamphetamine, Amphetamine, Opiates, Phencyclidine, Benzodiazepine, Methadone, and/or barbiturate in human urine.
The test is based on the principle of antigen-antibody immunochemistry. It utilizes mouse monoclonal antibody to selectively detect elevated levels of Marijuana, Cocaine, Methamphetamine. Amphetamine. Opiates. Phencyclidine, Benzodiazepine, Methadone, and/or barbiturate in urine. These tests can be performed without the use of an instrument.
A drug-positive urine specimen will not generate a colored-line in the designated test region, while a negative urine specimen or a urine specimen containing the drug concentration below the cut-off level will generate a colored-line in the test region. To serve as a procedural control, a colored-line will always appear at the control region, indicating that proper volume of specimen has been applied and membrane wicking has occurred.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Edward Tung, Ph.D. Director of Regulatory Affairs ACON Laboratories, Inc. 4108 Sorrento Valley Blvd. San Diego, CA 92121
MAY 1 5 2002
Re: k020771
Trade/Device Names: ACON® One Step Drug Screen Test Card ACON® One Step Drug Screen Test Card with Integrated Cup Regulation Numbers: 21 CFR 862.3610; 21 CFR 862.3650; 21 CFR 862.3100; 21 CFR 862.3150; 21 CFR 862.3170; 21 CFR 862.3870; 21 CFR 862.3250; 21 CFR 862.3640; 21 CFR 862.3620 Regulation Names: Methamphetamine test system; Opiate test system; Amphetamine test system: Barbiturate test system: Benzodiazephine test system: Cannabinoid test system; Cocaine and cocaine metabolite test system;
Morphine test system; Methadone test system
Regulatory Class: Class II
Product Codes: DJG; LCM; DKZ; DIS; JXM; LDJ; DIO; DPK; DJR Dated: March 6, 2002 Received: March 8, 2002
Dear Dr. Tung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4639. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
#### INDICATIONS FOR USE 10.
K020771 510(k) Number:
ACON One Step Drug Screen Test Card Device Name: ACON One Step Drug Screen Test Card with Integrated Cup
### Indications for Use:
The ACON One Step Drug Screen Test Card and ACON One Step Drug Screen Test Card with Integrated Cup are rapid chromatographic immunoassay devices for the qualitative and simultaneous detection of Marijuana, Cocaine, Methamphetamine, Amphetamine, Opiates, Phencyclidine, Benzodiazepine, Methadone, and/or barbiturate in human urine. The test is calibrated against analytes with their respective cutoff concentrations listed in the following table.
| Test | Calibrator | Cut-off |
|-------------------------------|--------------------------------|-------------|
| Amphetamine (AMP) | D-Amphetamine | 1,000 ng/mL |
| Barbiturates (BAR) | Secobarbital | 300 ng/mL |
| Benzodiazepines (BZO) | Oxazepam | 300 ng/mL |
| Cocaine (COC) | Benzoylecgonine | 300 ng/mL |
| Methamphetamine (mAMP) | D-Methamphetamine | 1,000 ng/mL |
| Morphine (MOP 300 or OPI 300) | Morphine | 300 ng/mL |
| Methadone (MTD) | Methadone | 300 ng/mL |
| Opiates (OPI 2000) | Morphine | 2,000 ng/mL |
| Phencyclidine (PCP) | Phencyclidine | 25 ng/mL |
| Marijuana (THC) | 11-nor- $\Delta^9$ -THC-9 COOH | 50 ng/mL |
The configurations of the ACON One Step Multi-Drug Screen Test Card and ACON One Step Multi-Drug Screen Test Card with Integrated Cup come with any combination of the above listed drug analytes. They are intended for the healthcare professionals including professionals at point-of-care sites. (Division Sign-Off) Division of Clinical Laboratory Devices
510(k) Number K220771
(Please do not write below this point)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
Or
over-the-counter Use **__**
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.