← Product Code [GEI](/productcode/GEI) · K020711

# WALLACH LOOP ELECTRODE (K020711)

_Wallach Surgical Devices, Inc. · GEI · Jun 3, 2002 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K020711

## Device Facts

- **Applicant:** Wallach Surgical Devices, Inc.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Jun 3, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Used to excise target tissue, perform biopsies and control bleeding through a standard Monopolar Electrosurgical Generator.

## Device Story

Wallach LOOP Electrode is a sterile, single-use, disposable electrosurgical accessory. Device connects to standard monopolar electrosurgical generator (ESU) to deliver energy for tissue cutting and coagulation. Used by clinicians during surgical procedures to excise target tissue or perform biopsies. Device consists of tungsten wire loops or stainless steel balls mounted on stainless steel shafts with polyolefin insulation. Operation relies on standard electrosurgical principles where current flows through electrode to tissue to achieve desired thermal effect. Benefits include precise tissue removal and effective bleeding control during procedures.

## Clinical Evidence

Bench testing only. No clinical data provided. Substantial equivalence supported by design, material, and performance testing, including conformance with ANSI/AMMI HF18-1993.

## Technological Characteristics

Materials: Tungsten wire, stainless steel shaft, stainless steel ball, polyolefin insulation. Energy: Monopolar electrosurgical. Form factor: Various loop and ball configurations. Standards: Conforms to ANSI/AMMI HF18-1993. Sterilization: Sterile, single-use.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Unimed 'Surgi-Link Eletrosurgical Electrode' ([K944433](/device/K944433.md))
- Megadyne 'Electrosurgical Electrode' ([K932102](/device/K932102.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JUN 0 3 2002

510(k) Premarket Notification Submittal 'Wallach LOOP Electrode' By Wallach Surgical Devices, Inc. SUMMARY CERTIFCIATION page of SECTION 2 March 4, 2002 -Page 2 of 3

## 510(k) Safety and Effectiveness Summary

| Applicant:            | Wallach Surgical Devices, Inc,<br>235 Edison Road<br>Orange, CT 06477                                |
|-----------------------|------------------------------------------------------------------------------------------------------|
| Registration:         | 1219739                                                                                              |
| Contact:              | Michael Malis                                                                                        |
| Phone:                | 203-799-2000                                                                                         |
| Fax:                  | 203-799-2002                                                                                         |
| Trade Name:           | Wallach LOOP Electrode                                                                               |
| Devices Generic Name: | Electrosurgical Electrode                                                                            |
| Classification Name:  | Electrode, Electrosurgical                                                                           |
| Classification:       | Currently classified as a Class II, under Product<br>Code 79 GEI, Regulation Number 878.4400, 21 CFR |

### Predicate Devices to which we are claiming substantial equivalence:

- Unimed 'Surgi-Link Eletrosurgical Electrode', K944433, 1
- 2 Megadyne 'Electrosurgical Electrode, K932102,

#### Product Description:

The Wallach LOOP Electrode is a Sterile, Disposable, Single Use Electrode. Used in conjunction with an ESU (electrosurtical unit or generator) to CUT or COAG during an electorsurgical procedure.

#### Indications for Use:

Used to excise target tissue, perform biopsies and control bleeding through a standard Monopolar Electrosurgical Generator.

#### Safety and Performance:

Substantial equivalence for this device is based on design, operation, intended use, materials, and performance claims. Testing that was performed on the Wallach LOOP Electrodes indicates that the devices are substantially equivalent in the performance and design of operation.

Hazard analysis evaluations performed on the Wallach LOOP Electrodes indicated that there were no new hazards presented with the use of the Wallach LOOP Electrodes as compared to the predicated devices.

{1}------------------------------------------------

K020711

510(k) Premarket Notification Submittal 'Wallach LOOP Electrode' By Wallach Surgical Devices, Inc. SUMMARY CERTIFCIATION page of SECTION 2

March 4, 2002 -- Page 3 of 3

# Comparison Chart:

| Feature:                                   | Wallach Surgical Devices,<br>Inc.<br>Wallach LOOP Electrodes<br>K<br>(pending)<br>Class II, Product Code: GEI<br>Regulation #: 878.4400                                                                     | UNIMED Surgical<br>Products, Inc.<br>Surgi-Link Electrosurgcial<br>Electrode<br>K944433<br>Class II, Product Code: GEI<br>Regulation #: 878.4400 | Megadyne Medical<br>Products, Inc.<br>Electrosurgical Electrodes<br>K932102<br>Class II, Product Code: GEI<br>Regulation #: 878.4400 |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
|                                            |                                                                                                                                                                                                             | Equivalent                                                                                                                                       | Equivalent                                                                                                                           |
| Intended<br>Use:                           | With an ESU for CUT and<br>COAG                                                                                                                                                                             | Equivalent                                                                                                                                       | Equivalent                                                                                                                           |
| Sterile,<br>Disposable<br>Single Use       | Yes                                                                                                                                                                                                         | Yes                                                                                                                                              | Yes                                                                                                                                  |
| Design:                                    | Tungsten Wire<br>Curved Loop<br>Square Loop<br>Stainless Steel Ball                                                                                                                                         | Equivalent                                                                                                                                       | Equivalent                                                                                                                           |
|                                            |                                                                                                                                                                                                             |                                                                                                                                                  |                                                                                                                                      |
| Material                                   | Tungsten Wire<br>Stainless Steel Shaft<br>Stainless Steel Ball<br>Polyolefin Insulation                                                                                                                     | Equivalent                                                                                                                                       | Equivalent                                                                                                                           |
|                                            |                                                                                                                                                                                                             |                                                                                                                                                  |                                                                                                                                      |
| Conforms<br>with<br>ANSI/AMMI<br>HF18-1993 | Yes                                                                                                                                                                                                         | Unknown                                                                                                                                          | Unknown                                                                                                                              |
| Sizes:                                     | 3mm Ball<br>5mm Ball<br>5mm Ball 6cm Shaft<br>10mmX7mm Loop<br>10mmX10mm Loop<br>15mmX5mm Loop<br>15mmX8mm Loop<br>20mmX8mm Loop<br>20mmX10mm Loop<br>20mmX15mm Loop<br>10mmX8mm Square<br>10mmX10mm Square | Similar                                                                                                                                          | Similar                                                                                                                              |

#### Conclusion:

Based on the indications for use, technological characteristics and comparison to currently marketed devices, the Wallach LOOP Electrode has been shown to be safe and effective for its intended use.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle are three stylized human profiles facing to the right. The profiles are stacked on top of each other and are connected by a flowing line.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 0 3 2002

Mr. Michael Malis General Manager Wallach Surgical Devices, Inc. 235 Edison Road Orange, CT 06477

Re: K020711

Trade/Device Name: Wallach LOOP Electrode Regulation Number: 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: March 4, 2002 Received: March 5, 2002

Dear Mr. Malis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. Michael Malis

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Stypt Rhodes

1. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Wallach Surgical Devices, Inc. Section 5

Page 1 of 1 __

510(k) Number (if known): _ 0207 / /

Wallach LOOP Electrode Device Name: _________________________________________________________________________________________________________________________________________________________________

Indications For Use:

Used to excise target tissue, perform biopsies and control bleeding through a standard Monopolar Electrosurgical Generator.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of General, Restorative
and Neurological Devices

510(k) Number K020711

Prescription Use ✓
(Per 21 CFR 801.109)
OR
Over-The-Counter Use *_*_
(Optional Format 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K020711](https://fda-staging.innolitics.com/device/K020711)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
