← Product Code [GNP](/productcode/GNP) · K020707

# EBV-EBNA IGG ELISA KIT, MODEL EBG-100 (K020707)

_Pan Bio Pty. , Ltd. · GNP · Jun 13, 2002 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K020707

## Device Facts

- **Applicant:** Pan Bio Pty. , Ltd.
- **Product Code:** [GNP](/productcode/GNP.md)
- **Decision Date:** Jun 13, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3235
- **Device Class:** Class 1
- **Review Panel:** Microbiology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K020707 · Jun 13, 2002 | EBV-EBNA IGG ELISA KIT, MODEL EBG-100 | Pan Bio Pty. , Ltd. | Retrospective clinical serum samples from a state health laboratory | Retrospective clinical samples were used to determine the sensitivity, specificity, and agreement of the EBV-EBNA IgG ELISA kit compared to existing EBV serological status. | Retrospective study; Clinical laboratory samples; Diagnostic performance |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Retrospective clinical study | Patients with various ages and genders undergoing EBV testing at a state health laboratory; Sample Size: 156; Number of Sites: 1 | EBV ELISA assays from an alternate manufacturer | Sensitivity, specificity, and agreement of the EBV serological status |

## Indications for Use

The Epstein Barr Virus Nuclear Antigen (EBNA) IgG ELISA Test is for the qualitative detection of IgG antibodies to EBNA in serum as an aid in the clinical laboratory diagnosis of Epstein barr virus (EBV) infection in patients with clinical symptoms consistent with infectious mononucleosis (IM). The PANBIO EBNA IgG ELISA should be used in conjunction with other EBV serologies.

## Device Story

The EBV-EBNA IgG ELISA Kit is an in vitro diagnostic assay used in clinical laboratories. It detects IgG antibodies to EBNA-1 in human serum samples. The device utilizes microwells coated with E. coli-expressed recombinant EBNA-1 antigen. Patient serum is added to the wells; if EBNA-1 IgG antibodies are present, they bind to the immobilized antigen. After washing, peroxidase-conjugated anti-human IgG is added, followed by a TMB/H2O2 substrate system. The resulting enzymatic reaction produces a color change (blue, turning yellow upon acidification) proportional to the amount of bound antibody. Healthcare providers use the resulting color intensity to determine the presence of EBNA IgG antibodies. This information serves as an aid in diagnosing EBV infection in patients with symptoms of infectious mononucleosis, to be used alongside other EBV serological tests.

## Clinical Evidence

Clinical performance evaluated via two studies. Study 1 (retrospective, n=156): sensitivity 97.1% (91.6-99.4% CI), specificity 92.3% (74.9-99.1% CI) for acute, 96.4% (81.6-99.9% CI) for negative, agreement 96.2%. Study 2 (prospective, n=352): sensitivity 85.1% (80.8-89.4% CI), specificity 100% (91.6-100% CI) for acute, 100% (92.6-100% CI) for negative, agreement 88.9%. Analytical specificity tested against 32 specimens (CMV, VZV, HSV-1, HSV-2, ANA, Rheumatoid Factor) with 0% cross-reactivity. Reproducibility assessed via ANOVA across three sites with CVs ranging from 4.7% to 28.8%.

## Technological Characteristics

Enzyme-linked immunosorbent assay (ELISA) using E. coli-expressed recombinant EBNA-1 antigen coated on polystyrene microwells. Detection system uses peroxidase-conjugated anti-human IgG and TMB/H2O2 substrate. Manual or automated plate processing. No electronic components or software algorithms; results determined by colorimetric measurement.

## Regulatory Identification

Epstein-Barr virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Epstein-Barr virus in serum. The identification aids in the diagnosis of Epstein-Barr virus infections and provides epidemiological information on diseases caused by these viruses. Epstein-Barr viruses are thought to cause infectious mononucleosis and have been associated with Burkitt's lymphoma (a tumor of the jaw in African children and young adults) and postnasal carcinoma (cancer).

## Predicate Devices

- Incstar EBV-EBNA IgG ELISA

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# JUN 1 3 2002

# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K020707

#### Applicant Information:

| Date Prepared: | 07th June, 2002        |
|----------------|------------------------|
| Name:          | PANBIO Limited         |
| Address:       | 116 Lutwyche Road      |
|                | Windsor 4030 Australia |

| Contact Person: | Helen Jennings    |
|-----------------|-------------------|
| Phone Number.   | 61-(0)7-3357-1177 |
| Fax Number.     | 61-(0)7-3357-1222 |

#### Device Information:

| Trade Name:          | EBV-EBNA IgG ELISA Kit           |
|----------------------|----------------------------------|
| Common Name.         | EBV-EBNA IgG EIA Test            |
| Classification Name: | EBV-EBNA IgG Serological Reagent |

## Equivalent Device:

Incstar EBV-EBNA IgG ELISA

#### Device Description:

The EBV-EBNA IgG ELISA Kit is an enzyme-linked immunosorbent assay (ELISA) for the detection of IgG antibodies to EBV-EBNA antigen in human serum.

#### Intended Use:

The Epstein Barr Virus Nuclear Antigen (EBNA) IgG ELISA Test is for the qualitative detection of IgG antibodies to EBNA in serum as an aid in the clinical laboratory diagnosis of Epstein barr virus (EBV) infection in patients with clinical symptoms consistent with infectious mononucleosis (IM). The PANBIO EBNA IgG ELISA should be used in conjunction with other EBV serologies.

#### Principle of Procedure:

Serum containing antibodies to EBNA-1 IgG, when present, combine with E.coli expressed EBNA-1 recombinant antigen attached to the polystyrene surface of the microwells. Residual serum is removed by washing and peroxidase conjugated anti-human IgG is added. The microwells are washed and a colourless substrate system, tetramethylbenzidine/hydrogen peroxide (TMB/H/O2) is added. The substrate is hydrolysed by the enzyme and the chromogen changes to a blue colour. After stopping the reaction with acid, the TMB becomes yellow. Color development is indicative of the presence of EBNA IgG antibodies in the test sample.

{1}------------------------------------------------

## PERFORMANCE CHARACTERISTICS

### Study Site 2:

156 frozen retrospective sera of various ages and genders were submitted to a state health laboratory in Maryland USA for EBV testing. The sera include the following groups: 28 seronegative samples from patients with acute Infectious Mononucleosis, and 102 samples from patients with past exposure to EBV. These sera were tested on the PANBIO EBV-EBNA IgG kit and EBV ELISA assays from an alternate manufacturer to determine the EBV status of the sera. The PANBIO EBNA IgG results were compared to the EBV status of the sera to determine the sensitivity, specificity, and agreement of the EBV serological status. The data is summarized in Table 1.

| PANBIO ELISA                                                 |          |           |          |       |  |  |  |
|--------------------------------------------------------------|----------|-----------|----------|-------|--|--|--|
| EBV Status                                                   | Positive | Equivocal | Negative | Total |  |  |  |
| Seronegative<br>VCA IgG (-)<br>VCA IgM (-)<br>EBNA IgG (-)   | 0        | 1         | 27       | 28    |  |  |  |
| Acute<br>VCA IgM (+)<br>EBNA IgG (-)                         | 1        | 1         | 24       | 26    |  |  |  |
| Past Infection<br>VCA IgG (+)<br>VCA IgM (-)<br>EBNA IgG (+) | 99       | 2         | 1        | 102   |  |  |  |
| TOTAL                                                        | 100      | 4         | 52       | 156   |  |  |  |

# TABLE 1 EBV-EBNA IgG Serological Sensitivity and Specificity of PANBIO ELISA versus EBV Status

95% Confidence Interval Serological Sensitivity (Past) = 99/102 = 97.1% 91.6 - 99.4% Serological Specificity (Acute) = 24/26 = 92.3% 74.9-99.1% Serological Specificity (Negative) = 27/28 = 96.4% 81.6-99.9% Serological Agreement = 150/156 = 96.2% 91.8 - 98.6%

* Retesting of equivocal samples was not conducted, as the samples were unavailable.

Note: "Serological" sensitivity and specificity refers to the PANBIO assay results to that of other assays normally used to diagnose EBV associated with IM. There was not an attempt to correlate the assay's results with disease presence. No judgement can be made on the comparison's accuracy to predict disease. Since the above studies were performed on a pre-selected, retrospective, population, no calculations for the assay's positive and negative predictive value may be done or inferred.

{2}------------------------------------------------

## Study Site 3:

352 prospective sera of various ages and genders were submitted to a private pathology laboratory in Queensland Australia for EBV testing. The sera include the following groups: 48 seronegative samples, 42 samples from patients with acute Infectious Mononucleosis, and 262 samples from patients with past exposure to EBV. These sera were tested on the PANBIO EBV-EBNA IgG kit and EBV ELISA assays from an alternate manufacturer to determine the EBV status of the sera. The PANBIO EBV EBNA IgG results were compared to the EBV status of the sera to determine the sensitivity, specificity, and agreement of the assay relative to the EBV serological status. The data is summarized in Table 2.

| PANBIO ELISA                                                 |          |            |          |               |
|--------------------------------------------------------------|----------|------------|----------|---------------|
| EBV Status                                                   | Positive | Equivocal* | Negative | Total         |
| Seronegative<br>VCA IgG (-)<br>VCA IgM (-)<br>EBNA IgG (-)   | 0        | 0          | 48       | 48            |
| Acute<br>VCA IgM (+)<br>EBNA IgG (-)                         | 0        | 0          | 42       | 42            |
| Past Infection<br>VCA IgG (+)<br>VCA IgM (-)<br>EBNA IgG (+) | 223      | 4          | 35       | 262           |
| Total                                                        | 223      | 4          | 125      | 352           |
| 95% Confidence Interval                                      |          |            |          |               |
| Serological Sensitivity (Past)                               |          | = 223/262  | = 85.1%  | 80.8 - 89.4%  |
| Serological Specificity (Acute)                              |          | = 42/42    | = 100.0% | 91.6 - 100.0% |
| Serological Specificity (Negative)                           |          | = 48/48    | = 100.0% | 92.6 - 100.0% |
| Serological Agreement                                        |          | = 313/352  | = 88.9%  | 85.6 - 92.2%  |

## TABLE 2 EBV-EBNA IgG Serological Sensitivity and Specificity of PANBIO ELISA versus EBV Status

* These equivocal samples were not tested on an alternative method due to insufficient sample. Collection of a further sample was not possible.

{3}------------------------------------------------

# REPRODUCIBILITY

The reproducibility of the PANBIO EBV-VCA IgG ELISA kit was determined by testing 8 sera 3 times each on three different days at three Australian study sites. Two sites were private pathology laboratories and the third site was PANBIO. Within-run, between day, between site and total precision were estimated by analysis of variance (ANOVA Type II) and are presented in table 3 below.

# TABLE 3 REPRODUCIBILITY DATA PANBIO EBV-EBNA IgG Study Site 1,2 & 3 Precision Measures (Using Cut-Off Ratio)

| Sample   | n  | *Mean | Within |       | Between Day |      | Between Site |       | Total |       |
|----------|----|-------|--------|-------|-------------|------|--------------|-------|-------|-------|
|          |    |       | *S.D   | CV    | *S.D        | CV   | *S.D         | CV    | *S.D  | CV    |
| Positive | 27 | 2.38  | 0.16   | 6.9%  | 0.09        | 3.7% | 0.00         | 0.0%  | 0.18  | 7.4%  |
| Cut-off  | 27 | 1.00  | 0.05   | 5.2%  | 0.00        | 0.0% | 0.00         | 0.0%  | 0.05  | 4.8%  |
| Negative | 27 | 0.10  | 0.01   | 6.8%  | 0.00        | 3.0% | 0.01         | 7.3%  | 0.01  | 9.4%  |
| #1       | 27 | 2.97  | 0.15   | 4.9%  | 0.06        | 1.9% | 0.48         | 16.2% | 0.43  | 14.4% |
| #2       | 27 | 3.20  | 0.18   | 5.7%  | 0.07        | 2.1% | 0.07         | 2.3%  | 0.20  | 6.3%  |
| #3       | 27 | 1.20  | 0.08   | 6.7%  | 0.09        | 7.2% | 0.35         | 29.0% | 0.31  | 25.8% |
| #4       | 27 | 1.28  | 0.09   | 6.9%  | 0.07        | 5.7% | 0.42         | 33.1% | 0.37  | 28.8% |
| #5       | 27 | 0.65  | 0.06   | 9.4%  | 0.04        | 6.5% | 0.00         | 0.0%  | 0.07  | 10.7% |
| #6       | 27 | 0.95  | 0.06   | 6.5%  | 0.06        | 6.6% | 0.12         | 12.1% | 0.13  | 13.2% |
| #7       | 27 | 3.45  | 0.15   | 4.4%  | 0.09        | 2.6% | 0.00         | 0.0%  | 0.16  | 4.7%  |
| #8       | 27 | 1.37  | 0.15   | 11.1% | 0.09        | 6.8% | 0.08         | 5.8%  | 0.18  | 13.3% |

All values are calculated from Ratios (Cut-Off using O.D) SD = Standard Deviation; CV = Coefficient of Variation

Note: Standard Deviation results have been rounded to two decimal places for tabulation purposes

{4}------------------------------------------------

## POTENTIAL CROSS-REACTIVITY

## Study Site 5:

A panel of 32 specimens from patients with confirmed diseases other than Epstein Barr Virus was tested to establish the analytical specificity of the EBV-EBNA IgG ELISA Test. The specimens were from patients with diseases that have the potential for cross-reactivity. Each of the specimens included in the study was characterized with respect to disease diagnosis prior to analysis with the EBV-EBNA IgG ELISA Test. Table 4 lists a summary of the results.

### TABLE 4

| Disease Type           | Number of Specimens | Result |
|------------------------|---------------------|--------|
| Cytomegalovirus        | 9                   | (0/9)  |
| Varicella zoster       | 6                   | (0/6)  |
| Herpes simplex virus 1 | 6                   | (0/6)  |
| Herpes simplex virus 2 | 1                   | (0/1)  |
| Anti-Nuclear Antigen   | 5                   | (0/5)  |
| Rheumatoid Factor      | 5                   | (0/5)  |
| Total                  | 32                  | (0/32) |

## PANBIO EBV-EBNA IgG CROSS-REACTIVITY SPECIMEN PANEL

Results indicate that no specimens (0/32) were positive when analysed with the EBV-EBNA IgG ELISA Kit. Refer to 'Study Document - Site 5' for raw data and section 2.3.4.1 for the summary table.

The true negative result of 100% for the above disease panel is consistent with good analytical specificity for the EBV-EBNA IgG ELISA Test.

The PANBIO assay employs an E. coli expressed EBNA-1 protein. The cross-reactivity or interference of human anti-E. coli antibodies is unknown with the assay's results.

{5}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an emblem that resembles a stylized eagle or bird with three wavy lines extending from its head. The logo is black and white.

# JUN 1 3 2002

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Helen Jennings Ouality and Regulatory Affairs Manager PANBIO Limited 116 Lutwyche Road Windsor, Brisbane Queensland, 4030 Australia

Re: k020707

> Trade/Device Name: Epstein Barr Nuclear Antigen IgG ELISA Test Regulation Number: 21 CFR 866.3235 Regulation Name: Epstein - Barr virus Serological Reagents Regulatory Class: Class I Product Code: GNP Dated: May 9, 2002 Received: May 15, 2002

Dear Ms. Jennings:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{6}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device. please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Page 1

510(k) Number (if known): K020707

Device Name: EBV-EBNA IgG ELISA

> Indications For Use: The Epstein Barr Virus Nuclear Antigen (EBNA) IgG ELISA Test is for the qualitative detection of IgG antibodies to EBNA in serum as an aid in the diagnosis of Epstein Barr (EBV) infection in patients with clinical symptoms of infectious mononucleosis (IM). The PANBIO EBNA IgG ELISA should be used in conjunction with other EBV serologies.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Wooly Dubres

Timcal Laboratory Devices KD20 70 510(k) Number

PRESCRIPTION USE X

(Optional Format 3-10-98)

---

**Source:** [https://fda-staging.innolitics.com/device/K020707](https://fda-staging.innolitics.com/device/K020707)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
