K020655 · Deltec, Inc. · LZG · Aug 13, 2002 · General Hospital
Device Facts
Record ID
K020655
Device Name
DELTEC INSULIN INFUSION PUMP AND ACCESSORIES
Applicant
Deltec, Inc.
Product Code
LZG · General Hospital
Decision Date
Aug 13, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Deltec Cozmo™ Insulin Infusion Pump (Model 1700) is a syringe infusion pump designed for Continuous Subcutaneous Insulin Infusion (CSII) for the control of diabetes. The Deltec Cozmo™ 3-ml Cartridge is designed for use with the Deltec Cozmo™ Insulin Pump for Continuous Subcutaneous Insulin Infusion (CSII). The Deltec Cozmo™ Treatment Assistant™ PC Communications System is designed for use with the Deltec Cozmo™ Insulin Infusion Pump as a tool in optimizing diabetes management. It allows communications between the pump and a computer for accessing data stored in pump memory, and programming of the pump.
Device Story
Ambulatory, battery-powered, microprocessor-controlled external insulin syringe pump; delivers insulin via CSII. Features include LCD, keypad, audible/vibratory alarms, real-time clock, and 3-ml cartridge reservoir. PC-based Treatment Assistant software enables data download and pump programming via infrared (IR) communication. Used by patients for diabetes management; healthcare providers use downloaded data to optimize therapy. Benefits include precise insulin delivery and improved diabetes management through stored memory and programmable settings.
Clinical Evidence
No clinical data; bench testing only. Evidence includes software validation, verification of programming functions, and in-vitro/biocompatibility testing for the 3-ml cartridge.
Technological Characteristics
Microprocessor-controlled syringe pump; AAA-battery powered; backlit LCD; 4-key keypad; IR communication interface. 3-ml cartridge (barrel, plunger, rod) with double o-ring seal. Software-controlled basal/bolus delivery rates. Waterproof design.
Indications for Use
Indicated for patients with diabetes requiring Continuous Subcutaneous Insulin Infusion (CSII) for glycemic control.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
MiniMed Com-Station™ Communication System (K993012)
Submission Summary (Full Text)
{0}------------------------------------------------
# AUG 1 3 2002
KO28655
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
### Deltec Cozmo™ Insulin Infusion Pump (Model 1700) and Accessories
June 3, 2002
### I. GENERAL INFORMATION
| Applicant's Name and Address: | Deltec, Inc.<br>1265 Grey Fox Road<br>St. Paul, MN 55112 |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Lisa J. Stone<br>Manager, Regulatory Affairs |
| Common/Usual Name: | Insulin Infusion Pump and Accessories |
| Proprietary Name: | Deltec Cozmo™ Insulin Infusion Pump (Model<br>1700) and Accessories |
| Equivalence Device Comparison: | MiniMed Model 508 Insulin Pump, MiniMed 3.0 ml Reservoir, Deltec CADD-Diplomat® System,<br>and MiniMed Com-Station™ Communication<br>System |
### DEVICE DESCRIPTION II.
The Deltec Cozmo™ Insulin Infusion Pump is a small, ambulatory, battery-powered, microprocessor controlled, external insulin syringe pump. Key features of the pump include: backlit liquid crystal display (LCD), 4-key keypad, separate audio bolus button, AAA-battery power source, waterproof, audible alarm option, vibratory alarm option, real time clock, belt-clip and interface, various basal and bolus delivery rates, stored memory, PC-downloading and programming capability, IR windows, 3-ml cartridge, cartridge cap and child-proof cartridge cap.
The Deltec Cozmo™ 3-ml Cartridge is a syringe style reservoir and is a proprietary design for use specifically with the Deltec Cozmo™ Insulin Infusion Pump. It is not compatible with other pumps. The 3-ml Cartridge is made of three separate pieces: barrel, plunger and rod. The barrel has an integral luer fitting on one end, volume markings, and a raised ring at the opposite end of the luer fitting that acts as a stop for the plunger. The plunger is conical in shape with a double o-ring seal design. The plunger and rod are detachable with a snap-fit design. The rod is removed from the plunger prior to cartridge insertion into the pump.
The Deltec Cozmo™ Treatment Assistant™ PC Communications System is a software program, which allows communications between a personal computer (PC) and an infusion pump. It allows for downloading of information from the pump to a PC and for programming of the pump via a PC. The "connection" between the pump
{1}------------------------------------------------
and the PC is via infra-red (IR) windows on the back of the pump and the computer's IR port. A standard off-the-shelf IR cable connector for the PC may also be used if the user's computer does not have an integral IR port.
### III. INTENDED USE OF THE DEVICE
The Deltec Cozmo™ Insulin Infusion Pump (Model 1700) is a syringe infusion pump designed for Continuous Subcutaneous Insulin Infusion (CSII) for the control of diabetes.
The Deltec Cozmo™ 3-ml Cartridge is designed for use with the Deltec Cozmo™ Insulin Pump for Continuous Subcutaneous Insulin Infusion (CSII).
The Deltec Cozmo™ Treatment Assistant™ PC Communications System is designed for use with the Deltec Cozmo™ Insulin Infusion Pump as a tool in optimizing diabetes management. It allows communications between the pump and a computer for accessing data stored in pump memory, and programming of the pump.
### IV. DEVICE COMPARISON
The Deltec Cozmo™ Insulin Infusion Pump is similar in design, function, and intended use to the MiniMed Model 508 Insulin Pump (K990801). The pumps are small, ambulatory, battery-powered, microprocessor controlled, external insulin syringe pump delivery systems. Each device is indicated for the delivery of insulin and includes the following features: backlit liquid crystal display (LCD), 4-key keypad, audible alarm option, vibratory alarm option, real time clock, belt-clip and interface, various basal and bolus delivery rates, stored memory, PC-downloading capability, IR windows, and 3-ml medication cartridge.
The Deltec Cozmo™ 3-ml Cartridge is similar in design, function, and intended use to the MiniMed 3-ml Reservoir (K991936). Each device is a syringe style reservoir with a capacity of 3 ml. They are made of the same materials and both are provided with a 22-gauge needle. Each device is indicated for use in the infusion of insulin with an external infusion pump.
The Deltec Cozmo™ Treatment Assistant™ PC Communications System is similar in design, function, and intended use to the Deltec CADD-Diplomat® System (K973917). The products allow communications between a personal computer (PC) and an infusion pump. The software programs allow for downloading of information from the pump to a PC and for programming of the pump via a PC. The Deltec Cozmo™ Treatment Assistant™ PC Communications System is also similar to the MiniMed Com-Station™ Communication System (K993012), which uses IR communication.
{2}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness Page 3 of 3
### V. SUMMARY OF STUDIES
#### A. Functional Testing
Test plans associated with software validation, verification of software controlled programming functions, and software related to proper pump operation were certified for the Deltec Cozmo™ Insulin Infusion Pump and the Deltec Cozmo™ Treatment Assistant™ Communications System.
In-vitro testing and biocompatibility testing was performed on the Deltec Cozmo™ 3-ml Cartridge.
#### B. Clinical Studies
Clinical studies were not deemed necessary regarding the use of the Deltec Cozmo™ Insulin Infusion Pump and Accessories.
### C. Conclusions Drawn from Studies
Based upon the information provided above, the Deltec Cozmo™ Insulin Infusion Pump and Accessories are substantially equivalent to other commercially available devices.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized design of three curved lines that resemble a bird or a wave. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the design. The text is in all capital letters and is evenly spaced around the circle.
AUG 1 3 2002
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Lisa Stone Managing, Regulatory Affairs Deltec, Incorporated 1265 Grey Fox Road Saint Paul, Minnesota 55112
Re: K020655
Trade/Device Name: Deltec Cozmo™ Insulin Infusion Pump (Model 1700), Deltec Cozmo™ 3-ml Cartridge, and Deltec Cozmo™ Treatment Assistant™ PC Communications System Regulation Number: 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: LZG Dated: June 3, 2002 Received: June 4, 2002
Dear Ms. Stone
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
Page 2 - Ms. Stone
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html
Sincerely yours.
JAWithdmer for,
Timothy A. Ulatowski Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Deltec Cozmo™ Insulin Infusion Pump (Model 1700), Deltec Cozmo™ 3-ml Cartridge, and Deltec Cozmo™ Treatment Assistant™ PC Communications System
Indications for Use:
"The Deltec Cozmo™ Insulin Infusion Pump (Model 1700) is a syringe infusion pump designed for Continuous Subcutaneous Insulin Infusion (CSII) for the control of diabetes."
"The Deltec Cozmo™ 3-ml Cartridge is designed for use with the Deltec Cozmo™ Insulin Infusion Pump for Continuous Subcutaneous Insulin Infusion (CSII).
"The Deltec Cozmo™ Treatment Assistant™ PC Communications System is designed for use with Deltec Cozmo™ Insulin Infusion Pump as a tool in optimizing diabetes management. It allows communications between the pump and a computer for accessing data stored in pump memory, and programming the pump."
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109
OR
Over-The Coun
Over-The Counter Use
ion Sianivision of Anesthesiology General Hosnital Infection Control, Dental
510(k) Number: K020655
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.