K020623 · M.K. Meditech Co. , Ltd. · MEG · May 2, 2002 · General Hospital
Device Facts
Record ID
K020623
Device Name
DUOPRO SAFETY SYRINGE (DUOPROSS)
Applicant
M.K. Meditech Co. , Ltd.
Product Code
MEG · General Hospital
Decision Date
May 2, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DuoPro Safety Syringe (DuoProSS™) is a sterile, single-use, disposable and non-reusable, manual, retractable safety syringe which is intended for injection of fluids into the body, while reducing the risk of sharps injuries and the potential for syringe reuse.
Device Story
DuoPro Safety Syringe (DuoProSS™) is a 3 mL manual, retractable safety syringe; sterile, single-use, disposable. Device features two-part plunger; distal part holds hypodermic needle; proximal part contains projection spike. Operation: user performs injection; manually retracts needle-plunger into barrel; breaks off plunger rod to lock needle and prevent reuse. Used by clinicians for fluid administration. Design reduces sharps injury risk via needle retraction mechanism. Bench testing confirms performance equivalence to predicate.
Clinical Evidence
Bench testing only. Testing included biocompatibility, latex protein measurement, packaging integrity, standards conformity, and simulated use testing per FDA guidance.
Technological Characteristics
3 mL piston syringe; manual retractable needle mechanism; two-part plunger with locking spike; Luer slip connector; sterile; single-use. Materials and dimensions vary by configuration. No software or electronic components.
Indications for Use
Indicated for injection of fluids into the body for patients requiring hypodermic delivery; intended to reduce sharps injury risk and prevent syringe reuse.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
# MAY 0 2 2002
K620623
# 510(k) Summary for the DuoPro Safety Syringe (DuoProSSTM) (per 21CFR807.92)
#### SPONSOR 1.
M.K. Meditech Co., Ltd. 4th Floor, No. 1221 Chung Cheng Road Taoyuan City Taiwan 330
Contact Person: I-Ming Shih Telephone: 886-3-3166399, extension 883 Date Prepared: February 25, 2002
#### 2. DEVICE NAME
DuoPro Safety Syringe (DuoProSSTM) Proprietary Name: Common/Usual Name: Hypodermic Syringe (with or without needle) Classification Name: Piston syringe Hypodermic single lumen needle
#### 3. PREDICATE DEVICE
- E.N.S.I. Retractable Safety Syringe (K003348, K000572) .
### DEVICE DESCRIPTION 4.
The DuoPro Safety Syringe (DuoProSS™) is a sterile, single-use, disposable and non-reusable, manual, retractable 3 mL safety syringe, provided with or without needle in various product configurations.
### 5. INTENDED USE
The DuoPro Safety Syringe (DuoProSS™) is a sterile, single-use, disposable and non-reusable, manual, retractable safety syringe which is intended for injection of
{1}------------------------------------------------
fluids into the body, while reducing the risk of sharps injuries and the potential for syringe reuse.
### 6.
# TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
M.K. Meditech Co., Ltd. makes a claim of substantial equivalence of the DuoProSS™ to the E.N.S.I. Retractable Safety Syringe (K003348) based on similarities in intended use, design, technological, and operational characteristics. Both are indicated for injecting fluids into the body, while helping to reduce the risk of sharps injuries. Both syringes are piston syringes that use single lumen hypodermic needles. The DuoProSS™ can be ordered both with and without needles, while the E.N.S.I. is not provided with needles. The DuoProSS™M currently uses a Luer slip needle connector, while the E.N.S.I. uses a Luer lock connector. Both syringes are provided sterile, single-use, and disposable. The major difference between the DuoProSS™ and the E.N.S.I. products is the syringe volume (and associated dimensions). This difference does not affect the performance of the syringe, since syringe size is typically determined by drug volume to be administered and user preference. Bench testing shows that the performance of the DuoProSS™ is equivalent to the E.N.S.I. product. Both syringes have two-part plungers. The distal part holds the hypodermic needle and the proximal part has a projection spike that mates with the distal part, thereby locking the needle to the plunger. Both syringes require the user to manually retract the needle-plunger into the syringe barrel, break off the plunger rod, and discard the pieces. M.K. Meditech Co., Ltd. believes that the differences between the DuoPro Safety Syringe (DuoProSS™) and the predicate device are minor and they raise no new issues of safety or effectiveness.
#### 7. PERFORMANCE TESTING
Testing provided in this premarket notification includes biocompatibility, measurement of latex protein, packaging integrity, standards conformity, and testing according to FDA guidance (including a simulated use test).
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, depicted with three parallel lines forming the shape of the bird.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# MAY 0 2 2002
M. K. Meditech Company Limited C/O Ms. Rosina Robinson Medical Device Consultants, Incorporated 49 Plain Street North Attleboro, Massachusetts 02760
Re: K020623
Trade/Device Name: Duopro Safety Syringe (DuoProSSTM) Regulation Number: 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: MEG and FMI Dated: February 25, 2002 Received: February 26, 2002
Dear Ms. Robinson :
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements
{3}------------------------------------------------
### Page 2 - Ms. Rosina Robinson
of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Timo A. Ulatowski Director Division of Dental. Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
## 510(k) Number (if known):
### M.K. Meditech Co., Ltd., DuoPro Safety Syringe Device Name: (DuoProSS™)
## Indications For Use:
The DuoPro Safety Syringe (DuoProSS™) is a sterile, single-use, disposable and nonreusable, manual, retractable safety syringe which is intended for injection of fluids into the body, while reducing the risk of sharps injuries and the potential for syringe reuse.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Patara Cucente
vision Sian-Off Division of Dental, Infection Control, and General Hospital Device 510(k) Number _
Prescription **Use**
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
M. K. Meditech Co., Ltd. Traditional 510(k) February 25, 2002 DuoPro Safety Syringe (DuoProSSTM)
Page vi
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.