HD/HDT, (POLYMACON) SOFT CONTACT LENS FOR DAILY WEAR (CLEAR AND TINTED, FULLY CAST-MOLDED LENS)
K020608 · Poly Vue Technolgies, Inc. · LPL · May 16, 2002 · Ophthalmic
Device Facts
Record ID
K020608
Device Name
HD/HDT, (POLYMACON) SOFT CONTACT LENS FOR DAILY WEAR (CLEAR AND TINTED, FULLY CAST-MOLDED LENS)
Applicant
Poly Vue Technolgies, Inc.
Product Code
LPL · Ophthalmic
Decision Date
May 16, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
The HD (polymacon) Aspheric Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and early visual acuity in not aphakic persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eye. The HDT (polymacon) Aspheric Toric Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in not aphakic persons with non-diseased eyes with myopia or hyperopia, not exceeding 5.00 diopters and early presbyopia up to 1.25 diopters. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eye. Eyecare practitioners may prescribe any of the above lenses for frequent/planned replacement wear, with replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system.
Device Story
Hemispherical soft contact lenses; molded spherical base curves; molded front surfaces. Material: polymacon (hydrophilic polymer of 2-HEMA cross-linked with EGDMA); 38% water content. Available clear or tinted with pigments (phthalocyanine blue/green, titanium dioxide). Used for daily wear; prescribed by eyecare practitioners for vision correction. Lens conforms to eye curvature; requires hydration in compatible saline solution. If dried, lens warps but recovers upon rehydration. Benefits include vision correction and cosmetic eye color alteration.
Clinical Evidence
No clinical data. Pre-clinical performance (cytotoxicity, systemic injection, ocular irritation) and manufacturing verification data are referenced from K013220. Compatibility with lens care products and preservative uptake/release studies are not required as the material (polymacon) is not new or modified.
Technological Characteristics
Material: polymacon (62% polymer, 38% water). Refractive index: 1.445 (hydrated). Light transmission: >90%. Oxygen permeability: 8.4 x 10^-11 (cm^2/sec)(ml O2/ml x mmHg @ 35C). Form factor: hemispherical shell, fully cast-molded. Connectivity: N/A. Sterilization: referenced in K013220.
Indications for Use
Indicated for aphakic and non-aphakic persons with non-diseased eyes for correction of myopia, hyperopia, and refractive astigmatism (up to 1.50D for HD, up to 5.00D for HDT), and early presbyopia (up to 1.25D for HDT).
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
PolyVue Silver Chord, Unisoft (methafilcon A) (K980818)
Submission Summary (Full Text)
{0}------------------------------------------------
## MAY 1 6 2002
# 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K020608
Applicant information:
| Date Prepared: | February 19, 2002 |
|-----------------|------------------------------------------------|
| Name: | PolyVue Technologies, Inc. |
| Address | Seven Hazel Avenue<br>Larkspur, CA 94939 |
| Contact Person: | Mr. Harold E. Johnson<br>President/CEO |
| Phone number: | (415) 945-9043 |
| USA Consultant: | Med-Vice Consulting, Inc.<br>Ms. Deanna Werber |
| Phone number: | (970) 243-5490 |
| Fax number: | (970) 243-5501 |
| Email address: | dwerber@FDApproval.com |
Device Information:
| Device Classification: | Class II |
|------------------------|----------------------------------------------------------------------------------------------------|
| Classification Number: | LPL |
| Classification Name: | Lenses, Soft Contact, Daily Wear |
| Trade Name: | HD/HDT, (polymacon) Soft Contact Lens for Daily Wear<br>(clear and tinted, fully cast-molded lens) |
{1}------------------------------------------------
### Purpose of 510(k) Submission:
### NEW INDICATION ~
A new indication of "early presbyopia" up to 1.25 diopters is being added to the previously cleared HD/HDT, (polymacon) Soft Contact Lens for Daily Wear reference K013220.
### Equivalent Devices:
The HD/HDT, (polymacon) Soft Contact Lenses are substantially equivalent to the following predicate devices:
#### Predicate devices:
- "HD/HDT" (polymacon) K013220, approval by Poly Vue Technologies, Inc. .
- "PolyVue" (polymacon) K982110, approval by Optech, Inc. .
- "PolyVue" Silver Chord, Unisoft (methafilcon A) K980818, approval by Steven A. Dunn, Inc. .
### Device Description:
The HD/HDT, (polymacon) Soft Contact Lenses are hemispherical shells with molded spherical base curves and molded front surfaces.
The nonionic lens material, (polymacon) is a hydrophilic polymer of 2- Hydroxyethyl methacrylate (2-HEMA) and cross-linked with ethylene glycol dimethacrylate (EGDMA), plus an initiator. The copolymer consists of 62% polymacon and 38% water by weight when immersed in normal buffered saline solution. The lenses are available clear or tinted. Lenses that contain a unique tinting pattern are subsequently processed to incorporate the 'listed' color additives and contain only the amount of color additive needed to accomplish the intended coloring effect. Lenses are tinted with one or a combination of one or more of the following pigments, 'listed' color additives.
Phthalocyanine blue (21 CFR § 74.3045), Phthalocyanine green (21 CFR § 73.3124) and Titanium dioxide (21 CFR § 73.3126)
In the hydrated state, the lens conforms to the curvature of the eye covering the cornea and extending slightly beyond the limbus forming a transparent optical surface. The (polymacon) soft hydrophilic contact lens has a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard and appear somewhat warped however, it will return to its proper configuration when completely rehydrated in the proper storage solution.
The hydrophilic characteristics allow aqueous solutions to enter the lens and in its fully hydrated state the lens is approximately 38% water by weight. The physical properties of the lens are:
1.445 (hydrated) Refractive Index greater than 90% Light Transmission (clear) Light Transmission (tinted) greater than 90%
{2}------------------------------------------------
Water Content Oxygen Permeability 38 % ± 2% 8.4 X 10-1 (cm /sec) (ml O /ml x mm Hg @ 35 C), (revised Fatt method).
### Intended Use:
The HD (polymacon) Aspheric Soft Contact Lenses for daily wear are indicated for the correction of ITIC \porymated.h} I spiritorio is with non-diseased eyes with myopia or hyperopia and early visual actiry in not apinatio persons who new by persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eye.
The HDT (polymacon) Aspheric Toric Soft Contact Lenses for daily wear are indicated for the The IID I (polymaton) Issphoric roskic persons with non-diseased eyes with myopia or hyperopia, contection of visual would in not exceeding 5.00 diopters and early presbyopia up to 1.25 diopters. The lens is available clear or timed and may be used to enhance or alter the apparent color of the eye.
Eyecare practitioners may prescribe any of the above lenses for frequent/planned replacement wear, with Dyour prociment may prescribed replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system.
#### Pre-Clinical Performance Data:
Pre-clinical performance data addressing the cytotoxicity test, systemic injection test, and ocular eye irritation test can be referenced in the 510(k) # K013220. In addition, data for all relevant manufacturing information, including verification data can be referenced in the 510(k) # K013220.
Concerning compatibility testing, the recommended lens care products (cleaning, rinsing and disinfection) have been approved for use with lens of the same lens group. Therefore, no additional compatibility testing is included.
With regard to preservative uptake and release studies, since the subject contact lens material (polymacon) is not a new and/or modified lens material, no additional studies need be conducted.
{3}------------------------------------------------
### Substantial Equivalence:
The HD/HDT, Soft Contact Lens will be manufactured according to specified process controls and a The more box , box of arogram currently in place and referenced in 510(k) K002099.
Reference K013220 for he final lens packaging and sterilization of the lenses.
The established safety profile (pre-clinical toxicology and manufacturing/chemistry data) of the THE ESTADISICU Salery promo (pro onlined to the previously. Being similar with respect to indications for use, target population, hydrophilic materials, physical construction and safety & to mucations for use, target population, areas the requirements per section 510(k) of the act regarding substantial equivalence and does not raise different questions of safety and effectiveness than the predicate device identified above.
The following matrix illustrates the equivalencies between the HD/HDT, Soft Contact Lens and the substantial equivalent predicate devices.
| Substantial<br>Equivalency | HD/HDT<br>'Subject Device' | HD/HDT,<br>'Predicate Device' | Poly Vue<br>'Predicate Device' | Poly Vue Silver Chord<br>'Predicate Device' |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacture | Poly Vue Technologies | Poly Vue Technologies | Optech, Inc. | Steven A. Dunn, Inc. |
| INDICATION | Soft Contact Lenses for daily<br>wear are indicated for the<br>correction of visual acuity in<br>aphakic and not aphakic<br>persons with non-diseased eyes<br>with myopia or hyperopia<br>and/or possesses refractive<br>astigmatism not exceeding 5.00<br>diopters, and early presbyopia<br>up to 1.25 diopters. | Soft Contact Lenses for daily<br>wear are indicated for the<br>correction of visual acuity in<br>aphakic and not aphakic<br>persons with non-diseased eyes<br>with myopia or hyperopia<br>and/or possesses refractive<br>astigmatism not exceeding 5.00<br>diopters, and/or are presbyopic. | Soft Contact lenses for daily<br>wear are indicated for the<br>correction of refractive<br>ametropia (myopia and<br>hyperopia), presbyopia and<br>astisgmatism in aphakic and not<br>aphakic persons with non-<br>diseased eyes. | Soft Contact lenses for daily<br>wear are indicated for the<br>correction of refractive<br>ametropia (myopia and<br>hyperopia), and astisgmatism in<br>aphakic and not aphakic<br>persons with non-diseased eyes. |
| INTENDED<br>USE | Daily Wear, Soft<br>(hydrophilic) Contact<br>Lens | Daily Wear, Soft<br>(hydrophilic) Contact<br>Lens | Daily Wear, Soft<br>(hydrophilic) Contact<br>Lens | Daily Wear, Soft<br>(hydrophilic) Contact<br>Lens |
| Manufacturing<br>Method | Fully-molded | Fully-molded | Lathe-cut<br>(Semi-Mold) | Lathe-cut<br>(Semi-Mold) |
| USAN name<br>Material name | polymacon | polymacon | polymacon | methafilcon A |
| Water Content<br>(%) | 38.0% | 38.0% | 38.0% | 55.0% |
| Toxicity<br>(safety) | Non-Toxic | Non-Toxic | Non-Toxic | Non-Toxic |
### Substantial Equivalence Matrix
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure, composed of three curved lines that suggest wings and a body.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## MAY 1 6 2002
Poly Vue Technologies, Inc. c/o Deanna Werber Medvice Consulting, Inc. 623 Glacier Drive Grand Junction, CO 81503
Re: K020608
Trade/Device Name: HD/HDT, (polymacon) Soft Contact Lens for Daily Wear (clear and tinted, fully cast-molded lens) Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lens Regulatory Class: Class II Product Code: LPL Dated: February 19, 2002 Received: February 25, 2002
#### Dear Ms. Werber:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter will anow you to oegill inding of substantial equivalence of your device to a legally promatication. "The PDF interior in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of additionally 21 OF F F ar 3613. Additionally, for questions on the promotion and advertising of Compliance at (301) 591 - 1010 - 12 - 101 - 12 - 101 - 10 - 4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division Other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
HD/HDT, (polymacon) Soft Contact Lens for Daily Wear (clear and tinted, Device Name: fully cast-molded lens)
### INDICATIONS FOR USE:
The HD (polymacon) Aspheric Soft Contact Lenses for daily wear are indicated for the correction of The IID (porymacon) I sopherrous with non-diseased eyes with myopia or hyperopia and early visual active in not apliance persons who note by persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eye.
The HDT (polymacon) Aspheric Toric Soft Contact Lenses for daily wear are indicated for the The MD I (porymacom) in not aphakic persons with non-diseased eyes with myopia or hyperopia, Concellon of visual actity in hot uphane peing 5.00 diopters and early presbyopia up to 1.25 diopters. possesses it cractive astiginatism not oneoouhs very to enhance or alter the apparent color of the eye.
Eyecare practitioners may prescribe any of the above lenses for frequent/planned replacement wear, with Eyecan oplaintoners may problish raily of any of except bed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system.
(PLEASE DO OT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel W. C. Brown, PHD
(Division Sign-Off)
Division of Ophthalmic Ear,
Nose and Throat Devises
510(k) Number K020608
Prescription Use ✗
(Per 21 CFR 801.109)
or
(Optional Format 1-2-96)
Over-The-Counter Use *_*_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.