The SCPC System is a software-controlled cardiopulmonary bypass device containing an uninterruptable power supply (UPS) and S3 Sensor Modules that is designed for operation of the Stöckert Centrifugal Pump and monitoring the cardiopulmonary bypass circuit. The SCPC System, in combination with the Stöckert Centrifugal Pump and the COBE Revolution Pump Head, is indicated for speed-controlled pumping through the cardiopulmonary bypass circuit for typical durations of six hours or less. The SCPC System has been qualified only for durations appropriate to cardiopulmonary bypass procedures and has not been qualified through in vitro, in vivo, or clinical studies, for long term use as a bridge to transplant, pending recovery of the natural heart, or extracorporeal membrane oxygenation (ECMO) procedures.
Device Story
The Stöckert Centrifugal Pump Console (SCPC) is a software-controlled system providing electrical power and operational control for the Stöckert Centrifugal Pump. It integrates S3 Sensor Modules to monitor cardiopulmonary bypass circuit parameters, including pressure, temperature, level, bubble detection, and elapsed time. Used in clinical settings during cardiopulmonary bypass procedures, the device is operated by perfusionists or clinicians. It functions as a console component for heart-lung machines, allowing speed-controlled blood pumping. The system includes an uninterruptable power supply (UPS) to ensure continuous operation. Clinicians use the console's interface to set pump parameters and monitor real-time circuit data, facilitating safe extracorporeal circulation. The device benefits patients by providing reliable, controlled blood flow during cardiac surgery.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included electrical safety, electromagnetic compatibility, functional acceptance, and hardware/software validation to confirm performance and compatibility with Stöckert CAPS, Stöckert S3, and COBE Century heart-lung machine consoles.
Technological Characteristics
Software-controlled console with integrated uninterruptable power supply (UPS) and S3 Sensor Modules. Provides electrical power and control for centrifugal pumps. Monitors pressure, temperature, level, bubble detection, and time. Compatible with Stöckert and COBE heart-lung machine consoles. Connectivity includes sensor module integration for circuit monitoring.
Indications for Use
Indicated for speed-controlled pumping through cardiopulmonary bypass circuits for patients undergoing cardiopulmonary bypass procedures for durations of six hours or less. Not indicated for long-term use, bridge to transplant, or ECMO.
Regulatory Classification
Identification
A cardiopulmonary bypass heart-lung machine console is a device that consists of a control panel and the electrical power and control circuitry for a heart-lung machine. The console is designed to interface with the basic units used in a gas exchange system, including the pumps, oxygenator, and heat exchanger.
{0}------------------------------------------------
## SEP 2 3 2002
K020571
# 510(k) Summary for Stöckert Centrifugal Pump Console
#### 1. SPONSOR
Stöckert Instrumente GmbH Lindberghstrasse 25 80939 Munich, Germany
Contact: Helmut Hofl, Director, Quality Assurance and Regulatory Affairs 011 49 89 323 010 Telephone: 011 49 89 323 01100 Facsimile:
Date Prepared: July 9, 2002
#### 2. Device Name
Proprietary Name: Stöckert Centrifugal Pump Console Common/Usual Name: Cardiopulmonary bypass device Classification Name: Multiple (see Table 1)
### Table G-1. SCPC System Classifications
| Classification Name | 21 CFR | ProCode | Classification |
|-------------------------------------------------------------------------|----------|---------|----------------|
| Cardiopulmonary bypass heart-lung console | 870.4220 | 74DTQ | Class II |
| Cardiopulmonary bypass bubble detector | 870.4205 | 74KRL | Class II |
| Cardiopulmonary bypass level sensing monitor/control | 870.4340 | 74DTW | Class II |
| Accessory to the cardiopulmonary bypass console:<br>Temperature Monitor | 870.4220 | 74DTQ | Class II |
| Accessory to the cardiopulmonary bypass console: Timer | 870.4220 | 74DTQ | Class II |
| Cardiopulmonary bypass coronary pressure gauge | 870.4310 | 74DXS | Class II |
#### Predicate Devices 3.
- . Medtronic Bio-Console 540/550 (Multiple)
- RotaFlow Centrifugal Pump System (K991864) ●
- Himex Centriflow Centrifugal Perfusion Pump System (K972740) ●
- Stöckert Compact System (K982014) ●
{1}------------------------------------------------
#### Device Description 4.
The SCPC System provides electrical power and sets specific operational parameters for the Stöckert Centrifugal Pump Control Panel and Drive Unit described in K011838. The SCPC System also monitors the following parameters of the cardiopulmonary bypass circuit:
- . Pressure
- Temperature .
- Level monitoring/bubble detection .
- Elapsed time .
The SCPC System provides procedural flexibility for use of the Stöckert Centrifugal Pump, allowing the pump to be used as a component of other legally marketed heartlung machine consoles.
#### న. INTENDED USE
The SCPC System is a software-controlled cardiopulmonary bypass device containing an uninterruptable power supply (UPS) and S3 Sensor Modules that is designed for operation of the Stöckert Centrifugal Pump and monitoring the cardiopulmonary bypass circuit. The SCPC System, in combination with the Stöckert Centrifugal Pump and the COBE Revolution Pump Head, is indicated for speed-controlled pumping through the cardiopulmonary bypass circuit for typical durations of six hours or less.
The SCPC System has been qualified only for durations appropriate to cardiopulmonary bypass procedures and has not been qualified through in vitro, in vivo, or clinical studies, for long term use as a bridge to transplant, pending recovery of the natural heart, or extracorporeal membrane oxygenation (ECMO) procedures.
#### 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The proposed SCPC System is identical in intended use and fundamental scientific technology to the cited predicate devices. Both the proposed SCPC System and the predicate devices provide electrical power and set operational parameters of connected pump(s) and monitor critical parameters of the cardiopulmonary bypass procedure. The differences between the proposed and predicate devices are limited to features that raise no new issues of safety or effectiveness.
{2}------------------------------------------------
#### 7. PERFORMANCE TESTING
Electrical safety and electromagnetic compatibility testing was performed to demonstrate conformance with the appropriate standards. Functional acceptance testing, hardware and software testing, and validation testing was performed to confirm that the proposed SCPC System performed as designed and met user requirements, including compatibility with the Stöckert CAPS (K863541), Stöckert S3, and COBE® Century™ (K960974) heart-lung machine (HLM) consoles.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a circular design with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a bird in flight, composed of three curved lines that also suggest human profiles.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 2 3 2002
Stöckert Instrumente GmbH Cynthia J. M. Nolte, Ph.D., RAC Staff Consultant c/o Medical Device Consultants, Inc. 49 Plain Street North Attleboro, MA 02760
Re: K020571 Stöckert Centrifugal Pump Console Regulation Number: 870.4360 Regulation Name: Nonroller-type cardiopulmonary bypass blood pump Regulatory Class: Class II (two) Product Code: DTQ Dated: July 9, 2002 Received: July 11, 2002
Dear Ms. Nolte:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality
{4}------------------------------------------------
Page 2 - Cynthia J. M. Nolte, Ph.D., RAC
systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Bram D/ Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known):
Device Name: Stöckert Centrifugal Pump Console
Indications for Use:
The Stöckert Centrifugal Pump Console (SCPC System) is a software-controlled cardiopulmonary bypass device containing an uninterruptable power supply (UPS) and S3 Sensor Modules that is designed for operation of the Stöckert Centrifugal Pump (SCP) and monitoring the cardiopulmonary bypass circuit. The SCPC System, in combination with the SCP and the COBE Revolution Pump Head, is indicated for speed-controlled pumping through the cardiopulmonary bypass circuit for typical durations of six hours or less.
The SCPC System has been qualified only for durations appropriate to cardiopulmonary bypass procedures and has not been qualified through in vivo, or clinical studies, for long term use as a bridge to transplant, pending recovery of the natural heart, or extracorporeal membrane oxygenation (ECMO) procedures.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division of Cardiovascular & Resr
510(k) Number K02057 devices
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use -------
(Optional Format 1-2-96)
February 20, 2002
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.