K020529 · A-Med Systems, Inc. · KFM · Aug 26, 2002 · Cardiovascular
Device Facts
Record ID
K020529
Device Name
LARGE CENTRIFUGAL BLOOD PUMP, MODEL CP-2
Applicant
A-Med Systems, Inc.
Product Code
KFM · Cardiovascular
Decision Date
Aug 26, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4360
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The A-Med Large Centrifugal Blood Pump is indicated for pumping blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve operations, surgery of the vena cava or aorta, liver transplants, etc). The A-Med Large Centrifugal Blood Pump is indicated for use only with the A-Med Blood Pump Controller.
Device Story
A-Med Large Centrifugal Blood Pump is a sterile, disposable, non-pulsatile, non-roller pump. It uses an impeller to impart energy to blood via centrifugal forces. The pump is demand-responsive, automatically adjusting flow and pressure based on resistance and fluid return. It features a hermetically sealed drive cable and magnetic coupling. The device is used in clinical settings for extracorporeal circulatory support. It requires a separately available motor, ultrasonic flow sensor, and microcomputer-based control console. The healthcare provider operates the system to maintain blood circulation during surgical procedures. The device benefits patients by providing temporary circulatory support during cardiac or vascular surgeries.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on design, material, and performance similarities to predicate devices.
Technological Characteristics
Sterile, disposable, non-pulsatile, non-roller centrifugal pump. Utilizes impeller for energy transfer. Features hermetically sealed drive cable and magnetic coupling. Designed for use with external motor, ultrasonic flow sensor, and microcomputer-based control console. Identical materials and design to predicates, with increased size and pressure capacity.
Indications for Use
Indicated for patients requiring extracorporeal circulatory support during cardiopulmonary bypass or procedures not requiring complete bypass (e.g., valvuloplasty, mitral valve surgery, vena cava/aorta surgery, liver transplants) for up to six hours. Must be used with A-Med Blood Pump Controller.
Regulatory Classification
Identification
A nonroller-type cardiopulmonary and circulatory bypass blood pump is a prescription device that uses a method other than revolving rollers to pump the blood through an extracorporeal circuit for periods lasting less than 6 hours for the purpose of providing either:(i) Full or partial cardiopulmonary bypass (i.e., circuit includes an oxygenator) during open surgical procedures on the heart or great vessels; or(ii) Temporary circulatory bypass for diversion of flow around a planned disruption of the circulatory pathway necessary for open surgical procedures on the aorta or vena cava.
Special Controls
*Classification* —Class II (special controls). The special controls for this device are:(i) Non-clinical performance testing must perform as intended over the intended duration of use and demonstrate the following: Operating parameters, dynamic blood damage, heat generation, air entrapment, mechanical integrity, and durability/reliability;
(ii) The patient-contacting components of the device must be demonstrated to be biocompatible;
(iii) Sterility and shelf life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components; and
(iv) Labeling must include information regarding the duration of use, and a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(b)
*Nonroller-type temporary ventricular support blood pump* —(1)*Identification.* A nonroller-type temporary ventricular support blood pump is a prescription device that uses any method resulting in blood propulsion to provide the temporary ventricular assistance required for support of the systemic and/or pulmonary circulations during periods when there is ongoing or anticipated hemodynamic instability due to immediately reversible alterations in ventricular myocardial function resulting from mechanical or physiologic causes. Duration of use would be less than 6 hours.(2)
*Classification.* Class III (premarket approval).(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with FDA on or before September 8, 2015, for any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976, or that has, on or before September 8, 2015, been found to be substantially equivalent to any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976. Any other nonroller-type temporary ventricular support blood pump shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
A-Med Miniature Centrifugal Blood Pump System (K992592)
{0}------------------------------------------------
K020529 p.113. "
## 510(k) SUMMARY OF SAFETY & EFFECTIVENESS
AUG 2 6 2002
This 510(k) Summary details sufficient information to provide an understanding of the basis for a determination of substantial equivalence. For convenience, the summary is formatted pursuant to 21 CFR \$807.92. This section may be used, as presented, to provide a substantial equivalence summary to anyone requesting it from the Agency.
#### 21 CFR §807.92 a (1)
A-Med Systems, Inc 2491 Boatman Avenue West Sacramento, CA 95691-3817 (916) 375-7400 ext. 5316 Contact person: Cynthia G. Royster Date prepared: 13 February 2002
#### 21 CFR §807.92 a (2)
| Trade name: | A-Med Large Centrifugal Blood Pump |
|-------------|------------------------------------|
|-------------|------------------------------------|
| Common name: | Centrifugal Pump |
|----------------------|-----------------------------------------------------------------|
| Classification name: | "Non-roller type Cardiopulmonary Bypass Blood<br>Pumps"870.4360 |
#### 21 CFR §807.92 a (3)
Identification of predicate(s): Substantial equivalence for the A-Med Large Centrifugal Blood Pump is based on its similarities to the predicate devices, A-Med Miniature Centrifugal Blood Pump System (K992592) and Medtronic Bio-Medicus Centrifugal Blood Pump (K973011). The line extension A-Med Large Centrifugal Blood Pump is substantially equivalent to the predicates in intended use, material, design, performance and physical characteristics. The line extension included a larger motor to the existing larger housing to enable it to produce more pressure.
#### 21 CFR §807.92 a (4)
Device Description-parts and function/concept: The A-Med Large Centrifugal Blood Pump is a sterile, disposable, non-pulsatile, non-roller pump that utilizes an impeller to impart energy to the blood through centrifugal forces. The flow of the pump is "demand responsive" by
{1}------------------------------------------------
K020529 p.2-13
automatically responding to the resistance against which it is pumping and to the amount of fluid returned to the large pump with the appropriate changes in flow and pressure. A drive cable and magnetic coupling are hermetically sealed components of the pump.
A motor ultrasonic flow sensor and a microcomputer-based control console are available separately.
#### 21 CFR 8807.92 a (5)
Intended use and relationship to predicate(s): The A-Med Large Centrifugal Blood Pump is indicated for pumping blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve operations, surgery of the vena cava or aorta, liver transplants, etc). The A-Med Large Centrifugal Blood Pump is indicated for use only with the A-Med Blood Pump Controller.
#### CFR 8807.92 a (6)
Technological characteristics and relationship to predicate(s): The A-Med Large Centrifugal Blood Pump is identical in design, material, intended use and technological characteristics to the predicate devices. The difference is size and pressure only.
#### 21 CFR §807.92 b
This substantial equivalence is based on similarities to the predicate devices in terms of intended uses and technological characteristics.
#### 21 CFR §807.92 c
In accordance with the specifications of this subsection, this summary (two pages) is its own section, and has been clearly identified as such.
{2}------------------------------------------------
# K020529 p 3 /3
### Literature (cited)
- 1. Driessen JJ. Dhaese H. Fransen G. Verrelst P. Rondelez L. Gevaert L. van Becelaere M, Schelstraete E. "Pulsatile Compared with Nonpulsatile Perfusion Using a Centrifugal pump for Cardiopulmonary Bypass during Coronary Artery Bypass Grafting. Effect on Systemic Haemodynamics. Oxygenation, and Inflammatory Response Parameters", Perfusion 10 No. 1: 3-12 1995.
- 2. Glauber M, Szefner J, Senni M, Gamba A, Mamprin F, Fiocchi R, Somachini M, Ferrazzi P. "Reduction of Haemorrhagic Complications during Mechanically Assisted Circulation with the Use of a Multi-System Anticoagulation Protocol", Artificial Organs Vol. 18 No. 10: 649-55 Oct. 1995.
- 3. Noon GP, Lafuente JA, Irwin S. " Acute and Temporary Ventricular Support with Bio-Medicus Centrifugal Pump", Ann Thorac Surg. Vol 68 2: 650-4.68 Aug 1999.
- 4. Ricardo Grossi Dantas, Eduardo Tavares Costa. "Ultrasonic Pulsed Doppler Blood Flowmeter for Use in Extracorporeal Circulation," Artificial Organs Vol. 24 No.3: 198-201 March 2000.
- 5. Takami Y, Makinoushi K, Nakazawa T, Benkowski R, Glueck J, Ohara Y, Nose Y . "Hemolytic characteristics of a Pivot Bearing Supported Gyro Centrifugal Pump (C1E3) Simulating Various Clinical Applications", Artif Organs Vol. 20 No.9: 1042-9 Sept. 1996.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" around the perimeter. Inside the circle is an abstract symbol that resembles three stylized human profiles facing to the right, stacked on top of each other.
**AUG 26 2002**
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
A-Med Systems, Inc. Ms. Cynthia G. Royster Director, Regulatory Affairs 2491 Boatman Avenue West Sacramento, CA 95691-3817
Re: K020529
Trade/Device Name: A-Med Large Centrifugal Blood Pump Regulation Number: 21 CFR 870.4360 Regulation Name: Nonroller-type cardiopulmonary bypass blood pump. Regulatory Class: Class III Product Code: KFM Dated: June 3, 2002 Received: June 21, 2002
Dear Ms. Royster:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Ms. Cynthia G. Royster
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4586. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Elia Mallis
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indication Statement
Page
510(k) Number (if known): N/A K 020529
Device Name: A-Med Large Centrifugal Blood Pump
#### Indications for Use
The A-Med Large Centrifugal Blood Pump is indicated for pumping blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve operations, surgery of the vena cava or aorta, liver transplants, etc). The A-Med Large Centrifugal Blood Pump is indicated for use only with the A-Med Blood Pump Controller.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use Use (Per 21 CFR 801.109)
OR
Over-The-Counter
Optional Format 1-
Elias Mollis
Sign-Off
Fratory Device
imber.
>10(; .
CONFIDENTIAL
7of 101
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.