EASYTEM EAR THERMOMETER, MODEL BT-020

K020520 · Metatech Co., Ltd. · FLL · Aug 6, 2002 · General Hospital

Device Facts

Record IDK020520
Device NameEASYTEM EAR THERMOMETER, MODEL BT-020
ApplicantMetatech Co., Ltd.
Product CodeFLL · General Hospital
Decision DateAug 6, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2910
Device ClassClass 2
AttributesPediatric

Indications for Use

The easytem Ear Thermometer is intended for the intermittent measurement and monitoring of human body temperature in the home. It is intended for use on people of all ages.

Device Story

Handheld electronic ear thermometer; measures natural thermal radiation emitted from tympanic membrane and adjacent surfaces via infrared sensor. Used in home setting by patients of all ages. Provides body temperature readings to assist in monitoring. Device design, materials, and components are similar to predicate infrared ear thermometers.

Clinical Evidence

No clinical data provided; substantial equivalence demonstrated via design and performance characteristics comparison to predicate devices.

Technological Characteristics

Handheld infrared ear thermometer; electronic sensing of thermal radiation; intended for home use. Design and materials are similar to predicate devices.

Indications for Use

Indicated for intermittent measurement and monitoring of human body temperature in neonatal, pediatric, and adult populations in a home setting.

Regulatory Classification

Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Ko26520 AUG 0 6 2002 MTC Media Trade Corporation 11820 Red Hibiscus Drive - Bonita Springs, FL 34135 Tel (941) 948-2001 – Fax (941) 948-2002 E-mail: GG(a)mediatradecorp.com Web: www.mediatradecorp.com ## 510(k) Summary | Submitter's Name: | Guenter Ginsberg<br>Media Trade Corporation | |--------------------|--------------------------------------------------------------------------------------------------------------------------------| | Address: | 11820 Red Hibiscus Drive<br>Bonita Springs, FL 34135 | | Phone: | (941) 948-2001 | | Fax: | (941) 948-2002 | | E-mail: | gg@mediatradecorp.com | | Contact: | Guenter Ginsberg | | Date of Summary: | February 12, 2002 | | Trade Name: | easytem Ear Thermometer, Model BT-020 | | Classification: | Thermometer, Clinical, Electronic<br>Product Code: FLL<br>Regulation No. 880.2910<br>Class: II<br>Panel: 80 (General Hospital) | | Predicate Devices: | Braun Thermoscan, IRT-3520<br>K 983295 (Predicate #1)<br>Omron Gentle Temp, MC-509<br>K922344 (Predicate #2) | {1}------------------------------------------------ | Page -2- (510k Summary) | | |--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Description: | The easytem Ear Thermometer is a hand held<br>instrument that measures body temperature throu<br>the opening of the auditory canal. Operation is<br>based on measuring the natural thermal radiation<br>emitted from the tympanic membrane and adjacent<br>surfaces. | | Intended Use: | The easytem Ear Thermometer is intended for the<br>intermittent measurement and monitoring of human<br>body temperature in the home. It is intended for use<br>on people of all ages. | | Technological Characteristics: | The easytem Ear Thermometer has the same general<br>design and performance characteristics as the<br>predicate devices from Braun and Omron. The main<br>difference is the physical size, shape and weight.<br>The easytem Ear Thermometer has the same<br>intended use, general design and incorporates<br>similar materials and components, hence should<br>therefore raise no new questions of safety and<br>effectiveness.<br>This submitter concludes that the easytem<br>Thermometer is therefore substantially equivalent to<br>the predicate devices "Braun Thermoscan<br>IRT3020"and the "Omron Gentletemp Instant Ear | {2}------------------------------------------------ Page -2- | Purpose of submission: | The Easytem Ear Thermometer, manufactured by<br>Metatech Co., Ltd., Korea, is a new device intended to be<br>marketed in the USA. | |------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | The Easytem Ear Thermometer is similar to other Ear Thermometers, approved and<br>marketed in the USA, such as the predicate devices listed. | | Predicate Devices: | Braun Thermoscan, IRT-3520<br>K 983295 (Predicate #1) | | | Omron Gentle Temp, MC-509<br>K922344 (Predicate #2) | | U.S. Contact: | Guenter Ginsberg (Official Correspondent)<br>Media Trade Corporation<br>Reg. No. 9023800<br>11820 Red Hibiscus Drive<br>Bonita Springs, FL 34135<br>Tel: 941 948-2001 Fax: 941 948-2002<br>E-mail: gg@mediatradecorp.com | This application was prepared according to FDA Guidance Documents and includes all required data to demonstrate substantial equivalence to legally marketed predicate devices. Sincerely yours, J-C Guenter Ginsberg President MTC Notes about the attachments: The attachments may include data and information not required by the FDA, but by European Institutions, and were left in the binder for convenience only. {3}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle or bird with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the bird image. The logo is simple and iconic, representing the department's role in health and human services. ## Public Health Service AUG 0 6 2002 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Metatech Company, limited C/O Mr. Guenter Ginsberg President Media Trade Corporation 11820 Red Hibiscus Drive Bonita Springs, Florida 34135 Re: K020520 > Trade/Device Name: Easytem Ear Thermometer, Model BT-020 Regulation Number: 21 CFR 880.2910 Regulation Name: Ear Thermometer Regulatory Class: II Product Code: FLL Dated: June 13, 2002 Received: June 17, 2002 Dear Mr. Ginsberg: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ ## Page 2 - Mr. Ginsberg Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrly/dsma/dsmamain.html Sincerely, Timothy A. Ulatowski Timothy Directo Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Page 510(k) NUMBER (IF KNOWN): ドバスク520 DEVICE NAME : METATECH Co. Ltd., Easytem Ear Thermometer Model BT-020 INDICATIONS FOR USE: This device is an electronic clinical thermometer using an infrared sensor to detect body I his devioo is an electrolitory canal in the neonatal, pediatric and adult population used at home. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter-Use \checkmark (Optional Format 1-2-9 Patrice Cucenik Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices 510(k) Number: K080520
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...