The CARESIDE LDH cartridge is intended for in vitro diagnostic use in conjunction with the CARESIDE Analyzer to quantitatively measure LDH activity in anti-coagulated whole blood, serum or plasma. For in vitro diagnostic use with the CARESIDE Analyzer to measure LDH activity from anti-coagulated whole blood, serum or plasma specimens to aid in the diagnosis and treatment of patients with certain liver diseases, heart diseases, and tumors of the lung, kidneys, and liver.
Device Story
CARESIDE LDH cartridge is a single-use disposable test used with the CARESIDE Analyzer. Input: anti-coagulated whole blood, serum, or plasma. Operation: user deposits specimen into cartridge well, closes lid, and inserts into analyzer. Analyzer heats sample to 37°C, spins to distribute sample into channels, and dispenses onto multi-layer reagent film. Reaction: lactate converted to pyruvate; NADH produced reacts with nitrotetrazolium blue (NTB) via diaphorase to form bluish formazan dye. Output: rate of change of dye color intensity measured via reflectance at 570 nm. Analyzer uses lot-specific standard curve to calculate LDH activity (U/L). Used in clinical settings; provides quantitative results to assist healthcare providers in diagnosing/treating liver, heart, and oncological conditions. Benefits patient through rapid, point-of-care assessment of tissue injury.
Clinical Evidence
Bench testing only. Performance characteristics include: reportable range 50-650 U/L; mean recovery 103%; total CV of 7.2% at 335 U/L. Interference testing showed no significant impact from ascorbic acid (20 mg/dL), bilirubin (20 mg/dL), or triglycerides (3000 mg/dL). Relative accuracy compared to BM/Hitachi 902 yielded correlation coefficient r=0.99.
Technological Characteristics
Dry film-based reflectance assay. Reagents: lithium lactate, NAD+, nitrotetrazolium blue, diaphorase. Sensing: reflectance at 570 nm via photodiode/LED. Form factor: single-use cartridge with multi-layer film. Connectivity: barcode-based lot-specific calibration. Reaction temperature: 37°C. Sample volume: 8.5 μL (metered).
Indications for Use
Indicated for in vitro diagnostic measurement of LDH activity in anti-coagulated whole blood, serum, or plasma to aid in diagnosis and treatment of patients with liver diseases, heart diseases, and tumors of the lung, kidneys, and liver.
Regulatory Classification
Identification
A lactate dehydrogenase test system is a device intended to measure the activity of the enzyme lactate dehydrogenase in serum. Lactate dehydrogenase measurements are used in the diagnosis and treatment of liver diseases such as acute viral hepatitis, cirrhosis, and metastatic carcinoma of the liver, cardiac diseases such as myocardial infarction, and tumors of the lung or kidneys.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 862.9.
Predicate Devices
Johnson and Johnson's Vitros LDH DT Slides for Johnson and Johnson's Vitros DT 60 (K912844/A)
Submission Summary (Full Text)
{0}------------------------------------------------
C.
D.
E.
F.
Page 1
3 200-JUL
May 14, 2002
K020484
## 510(K) SUMMARY: CARESIDE LDH SAFETY AND EFFECTIVENESS
#### I. Applicant Information
- Applicant Name A. B. Applicant/Manufacturer Address
Telephone Number
# CARESIDE, Inc.
6100 Bristol Parkway Culver City, CA 90230 310-338-6767 Renate A. MacLaren, Ph.D. 310-670-6986 rmaclaren@CARESIDE.com May 14, 2002
- G. Date 510(k) Summary prepared
Contact Person
e-Mail Address
FAX Number
#### II. Device Information
- A. Device Name (Trade) CARESIDE LDH Device Name B. LDH test system (Classification) C. Device Classification Clinical chemistry panel LDH test system Regulation Number: 21 CFR 862.1440 Regulatory Class II Classification Number: 75CFH D. Special controls and None applicable performance standards
#### III. Substantial Equivalence Claim
#### A. General Equivalency Claim
The ability to monitor analyte-specific biochemical reactions in dry film and other formats is widely recognized and has gained widespread acceptance for use in chemistry assays.
LDH in vitro diagnostic products, in both dry film and other formats, are legally marketed in the United States.
- B. Specific equivalency claim
The CARESIDE LDH test is substantially equivalent in principle, intended use, and clinical performance to the currently marketed Vitros slides for the quantitative measurement of LDH on the Vitros DT 60 II.
| Name of Predicate Device: | Johnson and Johnson's (formerly Eastman Kodak, Inc.) Vitros LDH DT Slides for<br>Johnson and Johnson's Vitros DT 60<br>(formerly Eastman Kodak's DT 60 II). | |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Predicate Device 510K number: | K912844/A | |
| Product Code: | 75CFJ | |
{1}------------------------------------------------
#### IV. Device Description
CARESIDE LDH cartridges are used with the CARESIDE, Inc. CARESIDE Analyzer to measure LDH activity in anti-coagulated whole blood, serum or plasma specimens. The CARESIDE LDH cartridge, a single use disposable in vitro diagnostic test cartridge, delivers a measured volume of serum or plasma to a dry film to initiate the measurement of LDH activity. The patented film cartridge contains all reagents necessary to measure LDH activity.
#### Explanation of Device Function A.
The activity of the CARESIDE LDH test measures LDH activity in the direction of the conversion of lactate to pyruvate. Some other LDH tests measure the activity of the enzyme in the direction of the conversion of pyruvate to lactate. The two different methods yield clinically equivalent results when interpreted with respect to their own reference ranges although the results are different quantitatively.
Each CARESIDE LDH cartridge consists of an LDH-specific multi-layer reagent film mounted in a plastic base with a hinged lid. The user introduces the specimen into the cartridge sample deposition well, closes the lid and inserts the cartridge into the CARESIDE Analyzer.
Once loaded, the CARESIDE Analyzer scans the cartridge barcode, brings the cartridge and the contained specimen to 37℃, and spins the cartridge to move the sample from the sample deposition well into the cartridge channels and chambers. Approximately 8.5 microliters of sample remains in the metering passage. Any excess sample flows into an overflow well.
The sample is automatically dispensed onto the multi-layer reagent film. The spreading and substrate layer distributes the LDH containing specimen uniformly. The sample moves through a reagent layer where the NADH formed in the LDH catalyzed reaction of lactate and NAD reacts with nitrotetrazolium blue (NTB) in a diaphorase catalyzed reaction to produce a bluish formazan dve. The rate of change of the dye's color intensity, as measured by the amount of reflected light at 570 nanometers, directly relates to the specimen LDH activity.
```
Test Reaction Sequence:
L-Lactate + NAD+ -------------------------------------------------------------------------------------------------------------------------------------------------------------
NTB + NADH -------------------------------------------------------------------------------------------------------------------------------------------------------------------
```
As the cartridges spin, a photodiode measures reflectance of light emitted by a wavelength-specific light emitting diode (LED) over a fixed time period. The instrument uses the reflectance measurements and the lot-specific standard curve to calculate LDH activity.
{2}------------------------------------------------
#### B. Test Summary
Lactate dehydrogenase (LDH) is distributed very widely in the body and is found in the in very high activities in the cytoplasm of cells in the heart, liver, kidney, and skeletal muscle, and in lesser amounts in lung, smooth muscle, erythrocytes, brain and pancreas. Because activities of LDH are much higher in tissues than in plasma, injury to tissue, with accompanying leakage of the cytoplasm into the peripheral blood, can increase the blood LDH dramatically. At least five forms of LDH are separable by electrophoresis. The predominant form in the blood varies with the tissue of origin, and therefore, LDH sub-typing may have diagnostic value.
Above-normal LDH activities in blood (> 220 U/L) are seen in several hematologic, neoplastic, cardiac, hepatic, skeletomuscular, and renal diseases. Very high elevations (> 500 U/L) of LDH activity have been seen in megaloblastic anemia, extensive carcinomatosis, viral hepatitis, shock, hypoxia, and extreme hyperthermia. Somewhat lower elevations (>300 U/L) occur after myocardial or pulmonary infarction, leukemia, and hemolytic anemia. Moderate elevations occur in cirrhosis, obstructive jaundice, and neoplastic diseases. Thus, an elevated LDH activity is only a nonspecific finding.
#### V. Intended Use
- Intended Use A.
The CARESIDE LDH cartridge is intended for in vitro diagnostic use in conjunction with the CARESIDE Analyzer to quantitatively measure LDH activity in anti-coagulated whole blood, serum or plasma.
- B. Indications for Use
For in vitro diagnostic use with the CARESIDE Analyzer to measure LDH activity from anti-coagulated whole blood, serum or plasma specimens to aid in the diagnosis and treatment of patients with certain liver diseases, heart diseases, and tumors of the lung, kidneys, and liver.
{3}------------------------------------------------
#### Technological Characteristics VI.
#### Similarities A.
| | CARESIDE LDH | Vitros LDH DT Slides |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Intended Use | For in vitro diagnostic use | Same |
| Indications | Primarily to aid in the diagnosis<br>and treatment of patients with<br>certain liver diseases, heart disease,<br>and tumors of the lung, kidneys,<br>and liver. | Same |
| Measurement | Quantitative | Quantitative |
| Method<br>Principle | Dry film based reflectance<br>measurement of NAD reduction to<br>NADH consequent to conversion of<br>lactate to pyruvate. | Dry film based reflectance<br>measurement of NADH oxidation to<br>NAD consequent to conversion of<br>pyruvate to lactate. |
| Specimen<br>Dilution | Not required | Not required |
| Materials | Lithium lactate, NAD+,<br>nitrotetrazolium blue, and<br>diaphorase | NADH and sodium pyruvate |
| Detector | Reflectance (570 nm) | Reflectance (340 nm) |
| Test time | Approx. 4-minute warm-up (on-<br>board) plus 4 minute test time. | 15 minutes slide warm-up (off-line)<br>plus 5 minutes test time. |
| Sample Type | Anti-coagulated whole<br>blood,<br>serum or plasma | Serum or plasma |
| Specimen<br>Volume | 8.5 μl test volume<br>(90 ± 10 μl applied volume) | 10 μl |
| Calibration | Calibration information bar-coded<br>on each cartridge. Calibration<br>information may change with each<br>lot. | Run Vitros DT II calibrators<br>whenever a new slide lot is used or<br>when necessary. |
| Quality<br>Control | 2 levels | 2 levels |
| Reporting<br>Units | U/L | U/L |
| Reaction<br>Temp. | 37 °C | 37 °C |
| | CARESIDE LDH | Vitros LDH DT Slides |
| Specimen Pre-<br>treatment | Not Required | Not Required |
| Reportable<br>Range | 50 to 650 U/L | 100 to 1750 U/L |
| Accurate<br>pipetting | Not required | Required |
| Reagent pre-<br>warming | Not required | Required |
{4}------------------------------------------------
#### B. Differences
#### C. Comparative Performance Characteristics
| | CARESIDE LDH | Vitros LDH DT Slides |
|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Detection limit | 50 U/L | 100 U/L |
| Reportable range | 50 - 650 U/L | 100 - 1750 U/L |
| Recovery (Mean) | 103% | Not available |
| Linearity | Linearity by dilution yielded slope and correlation coefficient within acceptable limits | |
| Interference | No significant interference observed at tested concentration of interferent:<br>Ascorbic Acid, 20 mg/dL<br>Bilirubin, 20 mg/dL<br>Triglycerides 3000 mg/dL | |
| Precision | Total CV, 335 U/L, 7.2% | Total CV, 649 U/L, 2.9% |
| Rel. Accuracy | Careside = 0.97 (BM/Hitachi 902) + 9.1 U/L r = 0.99 | |
#### D. Conclusion
The nonclinical and clinical data provided demonstrate that the CARESIDE LDH product is substantially equivalent to the legally marketed predicate device.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the top half. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized human profiles facing to the right.
JUL 3 2002 Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Renate A. MacLauren, Ph.D. Clinical Affairs Manager Careside, Inc. 6100 Bristol Parkway Culver City, CA 90230
k020484 Re:
> Trade/Device Name: Careside LDH Regulation Number: 21 CFR 862.1440 Regulation Name: Lactate dehydrogenase test system Regulatory Class: Class II Product Code: CFH Dated: May 15, 2002 Received: May 16, 2002
Dear Dr. MacLauren:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrb/dsma/dsmamain.html".
Sincerely yours.
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# INDICATIONS FOR USE
510(k) Number:
<020484
Device Name:
CARESIDE LDH
Indications for use: For in vitro diagnostic use with the CARESIDE Analyzer to measure LDH activity from anti-coagulated whole blood, serum or plasma specimens to aid in the diagnosis and treatment of patients with certain liver diseases, heart diseases, and tumors of the lung, kidneys, and liver.
Han Cooper
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K020484
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
...
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
Over-The-Counter Use (Optional Format 1-2-96)
OR
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.