PRE-LOADED BIOSTINGER HORNET

K020377 · Linvatec Corp. · HWC · May 6, 2002 · Orthopedic

Device Facts

Record IDK020377
Device NamePRE-LOADED BIOSTINGER HORNET
ApplicantLinvatec Corp.
Product CodeHWC · Orthopedic
Decision DateMay 6, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

BioStinger™ Bioabsorbable Implant provides fixation of The longitudinal vertical meniscus lesions (bucket handle tears) located in the vascular area of the meniscus.

Device Story

BioStinger Hornet™ is a sterile, single-use, cannulated meniscal repair device; composed of absorbable Poly (L-lactic) Acid homopolymer; colored with D&C Violet #2. Implant is pre-loaded on a sterile, disposable, re-loadable inserter for single-patient use. Device is radiotranslucent under fluoroscopy; visible via MRI and CAT scan. Used by surgeons in clinical settings for soft tissue fixation. Implant gradually absorbs into the body post-procedure.

Clinical Evidence

Bench testing only.

Technological Characteristics

Material: Poly (L-lactic) Acid homopolymer; D&C Violet #2 colorant. Form factor: Cannulated, pre-loaded on disposable inserter. Radiotranslucency: Fluoroscopy-translucent; MRI/CAT visible. Sterilization: Sterile, single-use.

Indications for Use

Indicated for fixation of longitudinal vertical meniscus lesions (bucket handle tears) located in the vascular area of the meniscus.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MAY 0 6 2002 Page 1 of 2 February 4, 2002 ## SUMMARY OF SAFETY AND EFFECTIVENESS In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, Linvatec Corporation is hereby submitting the 510(k) Summary of Safety and Effectiveness for the Pre-Loaded BioStinger Hornet™. 510(k) Number Kosu377 #### A. Submitter Linvatec Corporation 11311 Concept Boulevard Largo, Florida 33773-4908 #### Company Contact B. Laura D. Seneff, RAC Manager, Regulatory Affairs (727) 399-5234 Telephone (727) 399-5264 FAX ### C. Device Name Trade Name: Pre-Loaded BioStinger Hornet™ Common Name: Meniscal Repair Device Classification Names: Fastener, Fixation, Biodegradable, soft tissue Proposed Class/Device: Class II, MAFHWC & MAI Product Code {1}------------------------------------------------ page 2 of 2 Summary of Safety and Effectiveness Pre-Loaded BioStinger Hornet TM 510(k) # Kod'U377 February 4, 2002 Page 2 of 2 Predicate/Legally Marketed Devices D. BioStinger-V™ K991715 Linvatec Corporation #### ய் Device Description The BioStinger™ Bioabsorbable Meniscal Repair Device is a cannulated, sterile, single-use fixation device made from an absorbable homopolymer derived from Poly (L-lactic) Acid, and will gradually be absorbed into the body. The implant is colored with D&C Violet #2. The implant is preloaded on a single-patient use, sterile, disposable, re-loadable inserter. The implant is radiotranslucent with regard to intraoperative fluoroscopy, but it can be visualized with MRI and CAT scan. #### F. Intended Use BioStinger™ Bioabsorbable Implant provides fixation of The longitudinal vertical meniscus lesions (bucket handle tears) located in the vascular area of the meniscus. #### G. Substantial Equivalence The Pre-Loaded BioStinger Hornet™ is substantially equivalent in design, technology and intended use to Linvatec's existing BioStinger-V™ implant. Performance testing has been conducted to show that pre-loading the implant does not raise any new issues regarding safety and effectiveness. {2}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # MAY 0 6 2002 Ms. Laura D. Seneff. RAC Manager. Regulatory Affairs Linvatec Corp. 11311 Concept Boulevard Largo, Florida 33773-4908 Re: K020377 Trade/Device Name: Pre-Loaded BioStinger Hornet™ Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or Threaded Metallic Bone Fixation Fastner Regulatory Class: Class II Product Code: HWC Dated: February 4, 2002 Received: February 5, 2002 Dear Ms. Seneff: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Ms. Laura D. Seneff, RAC This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Mark A Milliken Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ page 1 of 1 February 4, 2002 510(k) Number (if known): Ko >0377 Device Name: Pre-Loaded BioStinger Hornet™ Indications for Use: The BioStinger™ Bioabsorbable Implant provides fixation of longitudinal vertical meniscus lesions (bucket handle tears) located in the vascular area of the meniscus. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE If NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use OR Over-the-Counter Use No (Per 21 CFR 801.109) (Optional Format 1-2-96) Mark N. Melbern Division Sion-Off Division of General, Restorative and Neurological Devices Page 1.4 510(k) Number K020377
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...