The Opiate Enzyme Immunoassay is a homogeneous enzyme immunoassay with a 300 ng/mL cutoff. The assay is intended for use in the qualitative and semi-quantitative analyses of opiates in human urine. The Opiate Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgement should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.
Device Story
Ready-to-use, liquid reagent, homogeneous enzyme immunoassay; detects opiates in human urine. Principle: competition between G6PDH-labeled morphine and free drug in urine for specific antibody binding sites. Absence of free drug allows antibody binding to labeled enzyme, decreasing enzyme activity. Activity measured spectrophotometrically at 340 nm via NAD to NADH conversion. Used in clinical laboratory settings by technicians/professionals. Output: preliminary qualitative or semi-quantitative opiate concentration. Results assist clinicians in identifying potential opiate presence; requires confirmation by GC/MS before clinical decision-making.
Clinical Evidence
Bench testing only. Precision studies (within-run and run-to-run) performed at various concentrations (225, 300, 375, 1000 ng/mL) showing CVs < 5%. Accuracy evaluated against predicate device (n=216) with 97.1% agreement for positive samples and 93.8% for negative samples. Analytical recovery demonstrated within ±10% of nominal concentration between 60-900 ng/mL.
Technological Characteristics
Homogeneous enzyme immunoassay; liquid reagent format. Sensing principle: spectrophotometric measurement of G6PDH enzyme activity at 340 nm. Cutoff: 300 ng/mL. Designed for use on automated clinical chemistry analyzers.
Indications for Use
Indicated for qualitative and semi-quantitative analysis of opiates in human urine. Provides preliminary analytical results; requires confirmatory testing (e.g., GC/MS). Intended for professional use.
Regulatory Classification
Identification
An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
{0}------------------------------------------------
K020368
## JUN 0 4 2002
# 510(k) Summary of Safety and Effectiveness
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
## Introduction
According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.
#### Submitter name, Address, and Contact
Lin-Zhi International, Inc. 2391 Zanker Road, Suite 340 San Jose, CA 95131-1124 Phone: (408) 944-0360 Fax: (408) 944-0359
Contact: Chiu Chin Chang, Ph.D. VP, R&D
## Device Name and Classification
| Classification Name: | Opiate test system, Class II, DJG (91 Toxicology),<br>21CFR 862.3650 |
|----------------------|-------------------------------------------------------------------------------------|
| Common Name: | Homogeneous enzyme immunoassay for the determination<br>of opiates levels in urine. |
| Proprietary Name: | None |
## Legally Marketed Predicate Device(s)
Lin-Zhi International, Inc.' Opiate Enzyme Immunoassay is substantially equivalent to the DRI® DAU Opiate Assay (By Microgenics Corp.), cleared under premarket notification K011150.
LZI's Opiate Enzyme Immunoassay is identical or similar to its predicate in terms of intended use, method principle, device components, and clinical performance.
#### Device Description
LZI's Opiate Enzyme Immunoassay is a ready-to-use, liquid reagent, homogeneous enzyme immunoassay. The assay uses specific antibody that can detect opiates in human urine with minimal cross-reactivity to various, common prescription drugs and abused drugs.
{1}------------------------------------------------
The assay is based on competition between morphine labeled with glucose-6-phosphate dehydrogenase (G6PDH) enzyme, and free drug from the urine sample for a fixed amount of specific antibody. In the absence of free drug from the urine sample the specific antibody binds to the drug labeled with G6PDH enzyme causing a decrease in enzyme activity. The G6PDH enzyme activity is determined spectrophotometrically at 340 nm by measuring its ability to covert nicotinamide adenine dinucleotide (NAD) to NADH.
## Intended Use
The Opiate Enzyme Immunoassay is a homogeneous enzyme immunoassay with a 300 ng/mL cutoff. The assay is intended for use in the qualitative and semi-quantitative analyses of opiates in human urine.
## Comparison to Predicate Device
LZI's Opiate Enzyme Immunoassay is substantially equivalent to other products in commercially distribution intended for similar use. Most notably it is substantially equivalent to the currently, commercially marketed DR1® Opiate Enzyme Immunoassay (K011150) by Microgenics Corporation.
The following table compares LZI's Opiate Enzyme Immunoassay with the predicate device, DRI® DAU Opiate Enzyme Immunoassay:
Similarities:
- . Both assays are for qualitative and semi-quantitative determination of opiates in human urine.
- Both assays use the same method principle, and device components. ●
- Both assays use 300 ng/mL as cutoff level per recommendations of The ● Substance Abuse and Metal Health Services Administration (SAMHSA).
Differences:
DRI® Opiate Enzyme Immunoassay can also be used as a 2000 ng/mL cutoff assay when different size of sample is used. Current LZI assay is for use as an assay with 300 ng/mL cutoff only.
{2}------------------------------------------------
## (Comparison to Predicate Device, continued)
| Feature | Microgenics' DRI® Opiate EIA | | | | LZI's Opiate EIA | | | |
|-----------------------|-------------------------------------------------------------|------------|------|--------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|------------|-------|------|
| Within Run Precision: | | | | | | | | |
| Qualitative: | | Mean Rate | SD | % CV | | Mean Rate | SD | % CV |
| | Negative | - | - | - | Negative | 291.6 | 2.19 | 0.75 |
| | 225 ng/mL | 374 | 2.2 | 0.6 | 225 ng/mL | 374.4 | 3.01 | 0.80 |
| | 300 ng/mL | 401 | 2.3 | 0.6 | 300 ng/mL | 399.8 | 3.42 | 0.86 |
| | 375 ng/mL | 421 | 2.4 | 0.6 | 375 ng/mL | 421.8 | 3.20 | 0.76 |
| | 1000 ng/mL | - | - | - | 1000 ng/mL | 530.8 | 5.17 | 0.97 |
| Semi-quantitative: | | Mean Conc. | SD | % CV | | Mean Conc. | SD | % CV |
| | 225 ng/mL | 226 | 6.0 | 2.7 | 225 ng/mL | 218.6 | 5.96 | 2.73 |
| | 300 ng/mL | 303 | 8.1 | 2.7 | 300 ng/mL | 298.0 | 9.84 | 3.30 |
| | 375 ng/mL | 379 | 15.1 | 4.0 | 375 ng/mL | 373.3 | 11.18 | 3.00 |
| Run-To-Run Precision: | (total precision) | | | | | | | |
| Qualitative: | | Mean Rate | SD | % CV | | Mean Rate | SD | % CV |
| | Negative | - | - | - | Negative | 292.8 | 1.81 | 0.62 |
| | 225 ng/mL | 374 | 2.6 | 0.7 | 225 ng/mL | 375.8 | 3.61 | 0.96 |
| | 300 ng/mL | 401 | 3.2 | 0.8 | 300 ng/mL | 400.8 | 3.34 | 0.83 |
| | 375 ng/mL | 421 | 3.0 | 0.7 | 375 ng/mL | 421.1 | 2.87 | 0.68 |
| | 1000 ng/mL | - | - | - | 1000 ng/mL | 528.6 | 4.84 | 0.92 |
| Semi-quantitative: | | Mean Conc. | SD | % CV | | Mean Conc. | SD | % CV |
| | 225 ng/mL | 226 | 8.2 | 3.6 | 225 ng/mL | 223.8 | 9.45 | 4.22 |
| | 300 ng/mL | 303 | 9.4 | 3.1 | 300 ng/mL | 301.0 | 9.15 | 3.04 |
| | 375 ng/mL | 379 | 15.9 | 4.2 | 375 ng/mL | 377.8 | 7.53 | 1.99 |
| Sensitivity: | 6 ng/mL | | | | 15 ng/mL | | | |
| Accuracy: | Vs. GC/MS | | | | Vs. DRI® Opiate EIA (n = 216) | | | |
| | Positive Samples: 100 % agreement | | | | 97.1 % agreement | | | |
| | Negative Samples: 90 % agreement | | | | 93.8 % agreement | | | |
| Analytical Recovery: | | | | | | | | |
| | Qualitative: No data available | | | | 100 % accuracy on positive vs. negative tests | | | |
| | Semi-quantitative: No data available | | | | Quantitates within ±10% of the nominal | | | |
| | concentration between 60ng/mL and 900 | | | | | | | |
| | ng/mL. | | | | | | | |
| | | | | Average 97.7 % recovery at 225 ng/mL level<br>(Cutoff - 25%) | | | | |
| | | | | Average 96.7 % recovery at 375 ng/mL level | | | | |
| | (Cutoff + 25%) | | | | | | | |
| Interference: | See attached Microgenics' DRI® Opiate<br>EIA package insert | | | | Compounds tested are comparable to the<br>predicate device. Results included in the<br>specificity table. | | | |
| Specificity: | See attached Microgenics' DRI® Opiate<br>EIA package insert | | | | Comparable to the predicate device. | | | |
# Performance Characteristics (Comparing the 300 ng/mL cutoff assay performance only)
{3}------------------------------------------------
## Conclusion
LZI's Opiate Enzyme Immunoassay was evaluated for several performance characteristics including precision, sensitivity, accuracy, analytical recovery, and specificity. All the studies showed acceptable results when compared to the predicate device.
We trust the information provided in this Premarket Notification [510(k)] submission will support a determination of substantial equivalence of the LZI's Opaite Enzyme Immunoassay to other opiate test systems currently marketed in the United States.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular emblem with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle with three lines representing its wings and head. The eagle faces to the right.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
# IUN 0 4 2002
Chiu Chin Chang, Ph.D. VP. R&D Lin-Zhi International, Inc. 2391 Zanker Road. Suite 340 San Jose, CA 95131-1124
Re: k020368
> Trade/Device Name: Opiate Enzyme Immunoassay Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate test system Regulatory Class: Class II Product Code: DJG Dated: April 24, 2002 Received: April 26, 2002
Dear Dr. Chang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entifled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".
Sincerely yours.
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Premarket Notification
## Indications for Use Statement
510(k) Number (if known): _ KO20368
## Device Name: Opiate Enzyme Immunoassay
## Indications for Use:
The Opiate Enzyme Immunoassay is a homogeneous enzyme immunoassay with a 300 ng/mL cutoff. The assay is intended for use in the qualitative and semi-quantitative analyses of opiates in human urine.
The Opiate Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgement should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.
Cooper
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K020368
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.