GEMS is intended to be used as a data management tool for cardiologists, general practitioners, cardiac or ECG technicians, nurses, monitoring service technicians, and other cardiac related institutions or care givers to store, retrieve, communicate and report ECG and ECG data acquired from a variety of ECG sources including single and multi-lead ECG devices, including diagnostic 12-lead devices. Users will be able to purchase specific modules for managing other patient cardiac related data such as pacemaker and rehabilitation data that fit their patients' needs. The ECG module may be licensed to other software developers as an ECG viewer for their products. GEMS is intended for use in clinics, hospitals, physician's offices, or anywhere a medical doctor deems appropriate. GEMS does not offer diagnosis or medical alarms. It is intended that competent human intervention be involved before any impact on health occurs. Clinical judgment and experience are used to check and interpret the data.
Device Story
GEMS is a modular software-based ECG data management system for Windows environments. It functions as a central repository for cardiology records, including ECGs, physician notes, and pacemaker/ICD data. Inputs are received via keyboard, mouse, barcode reader, serial port, or IrDA port. The system enables storage, retrieval, communication, and reporting of cardiac data. It operates in clinics, hospitals, and physician offices, managed by healthcare professionals. The software supports multi-tasking and linking of information objects. It does not provide automated diagnosis or medical alarms; clinical judgment is required for all interpretations. The system benefits patients by centralizing cardiac history and facilitating efficient data access for clinicians to support decision-making.
Clinical Evidence
Bench testing only. No clinical data provided. Software heart rate calculation functionality was verified against AAMI EC38 standards.
Technological Characteristics
Software-based ECG management system for Windows 98/NT/2000. Modular architecture. Supports client-server database management. Connectivity via serial port, IrDA, and network (LAN/WAN). Heart rate calculation algorithm compliant with AAMI EC38.
Indications for Use
Indicated for cardiologists, GPs, cardiac/ECG technicians, nurses, and monitoring service technicians to manage ECG data from single/multi-lead and 12-lead devices. Used in clinics, hospitals, and physician offices. No specific age or gender restrictions. Contraindications: Not for life-supporting/sustaining use; does not provide automated diagnosis or medical alarms.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
APR 3 0 2002
Image /page/0/Picture/1 description: The image shows the logo for CardioComm Solutions Inc. The logo features a stylized heart shape above the company name. The heart shape is formed by two curved lines that meet at the bottom and top, creating a heart-like silhouette. Below the heart shape, the words "CARDIOCOMM" and "SOLUTIONS INC" are printed in a bold, sans-serif font.
## 11020366 01/2
## Submitter
| Name: | CardioComm Solutions Inc. |
|----------|-----------------------------------------------------------------|
| Address: | 201 – 3060 Cedar Hill Road<br>Victoria, B.C., Canada<br>V8T 3J5 |
| Phone: | (250) 744-1822 |
| Fax: | (250) 744-1866 |
| Contact: | Angela Halwas |
| Date: | January 30, 2002 |
510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K982384.
## Device
Global ECG Management System II (GEMS) Name:
Brentwood PC-ECG (K955023) Substantial Equivalence is claimed to:
## Description:
GEMS is a software product designed for Microsoft Windows 98, Windows NT, and Windows 2000 operating systems running on an IBM compatible platform. GEMS is compatible with industry standard client server database management systems and as such will operate standalone or on a local or wide area networks. GEMS consists of a user interface that enables health care professionals to input, store, and output data from a relational database.
The product consists of a set of modules that can be installed as required to customize the application to individual users' needs. Modular design has the benefit of encapsulating changes within the overall application. If a module changes, then in most cases the change will affect only that module and will not affect the rest of the program. This permits more efficient and controlled development and testing, and prevents new and unknown safety issues and anomalies from being introduced. This stability contributes to the safety and reliability of the product.
GEMS is capable of multi-tasking and supports the linking and embedding of related information objects in the ECG. The software stores all aspects of a patient's cardiology record including: arrhythmia diagnosis, pathological
cV
{1}------------------------------------------------
#### GEMS (K982384) -- 510(k) Summary
diagnosis, ECGs, ECG information, physician notes, clinical history, pacemaker/ICD data and associated reports.
Data can be entered via keyboard, mouse, bar code reader, serial port, or IrDA port, and stored to and retrieved from any computer media. Information can be displayed on the computer monitor or printed.
GEMS is not a life-supporting or life-sustaining system. It is intended that competent human intervention be involved before any impact on health occurs. Clinical judgment and experience are used to check and interpret the data.
#### Intended Use:
GEMS is intended to be used as a data management tool for cardiologists, general practitioners, cardiac or ECG technicians, nurses, monitoring service technicians, and other cardiac related institutions or care givers to store, retrieve, communicate and report ECG and ECG data acquired from a variety of ECG sources including single and multi-lead ECG devices, including diagnostic 12lead devices. Users will be able to purchase specific modules for managing other patient cardiac related data such as pacemaker and rehabilitation data that fit their patients' needs. The ECG module may be licensed to other software developers as an ECG viewer for their products. GEMS is intended for use in clinics, hospitals, physician's offices, or anywhere a medical doctor deems appropriate. GEMS does not offer diagnosis or medical alarms. It is intended that competent human intervention be involved before any impact on health occurs. Clinical judgment and experience are used to check and interpret the data.
#### Technological Characteristics:
Only one feature of GEMS cannot be found in the predicate device: heart rate calculation in software. Testing for calculation of heart rate in software has been completed and submitted according to the relevant sections of AAMI EC38.
All other features of GEMS are present in the predicate. GEMS is therefore substantially equivalent to the predicate device.
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# APR 3 0 2002
Ms. Angela Halwas Quality System Manager CardioComm Solutions, Inc. 201 - 3060 Cedar Hill Road Victoria, B.C., Canada V8T 3J5
Re: K020366
Trade Name: Global ECG Management System II (GEMS) Regulation Name: Electrocardiograph Regulation Number: 21 CFR 870.2340 Regulatory Class: Class II (two) Product Code: 74 DPS Dated: January 31, 2002 Received: February 4, 2002
Dear Ms. Halwas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
# Page 2 - Ms. Angela Halwas
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I lease oc advised that I Dri brisaantes vour device complies with other requirements of the Act that I Dr Has Intact and regulations administered by other Federal agencies. You must or any 1 coolar bakated that egirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI IT rat 6077, idoomig (DI CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to ough finding of substantial equivalence of your device to a legally premainst notification " results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of additionally 21 Of ICP article of the stime on the promotion and advertising of Compliance at (301) 594-4639 of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). regaration on information on your responsibilities under the Act may be obtained from the Oiner general miormation on your nernational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
V. Craig Tiller, Ph.D.
Donna-Bea Tillman, Ph.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of
KOZO368 510(k) Number (if known):
Global ECG Management System (GEMS) Device Name:
Indications For Use:
GEMS is intended to be used as a data management tool for cardiologists, general practitioners, cardiac or ECG technicians, nurses, monitoring service technicians, and other cardiac related institutions or care givers to store, retrieve, communicate and report ECG and ECG data acquired from a variety of ECG sources including single and multi-lead ECG devices, including diagnostic 12-lead devices. Users will be able to purchase specific modules for managing other patient cardiac related data such as pacemaker and rehabilitation data that fit their patients' needs. The ECG module may be licensed to other software developers as an ECG viewer for their products. GEMS is intended for use in clinics, hospitals, physician's offices, or anywhere a medical doctor deems appropriate. GEMS does not offer diagnosis or medical alarms. It is intended that competent human intervention be involved before any impact on health occurs. Clinical judgment and experience are used to check and interpret the data.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Division of Cardiovascular & Respiratory Devices
510(k) Number K020362
(Optional Format 3-10-98)
**Prescription Use**
**(Per 21 CFR 801.109)**
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.