K020291 · Exactech, Inc. · JDI · Feb 27, 2002 · Orthopedic
Device Facts
Record ID
K020291
Device Name
EXACTECH, MODEL PMMA FEMORAL STEM CENTRALIZER
Applicant
Exactech, Inc.
Product Code
JDI · Orthopedic
Decision Date
Feb 27, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also indicated for ankylosing spondylitis, congenital hip dysplasia, revision of potentially failed reconstructions where sufficient bone stock is present, and to restore mobility resulting from previous fusion.
Device Story
Femoral stem centralizers designed to provide central placement of cemented femoral stems within femoral canal during hip arthroplasty. Device features three-fin design with tapered post for attachment to cemented femoral components. Used by orthopedic surgeons in clinical/OR settings. Surgeon selects component based on patient requirements and surgical technique. Device aids in proper positioning of prosthesis, potentially improving long-term stability and reducing risk of early failure.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Material: Polymethylmethacrylate (PMMA). Design: Three-fin configuration with tapered post. Dimensions: 7mm to 17mm diameter. Function: Mechanical centralizer for cemented femoral stems. No software or electronic components.
Indications for Use
Indicated for skeletally mature individuals undergoing primary hip replacement or revision surgery due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degeneration, proximal fractures, ankylosing spondylitis, or congenital hip dysplasia. Contraindicated in patients with active infection, insufficient bone stock, neuromuscular disorders preventing hip joint control, or patients whose weight, age, or activity level suggests early system failure. L-Series unipolar contraindicated for acetabular degenerative changes or pelvic fractures.
Regulatory Classification
Identification
A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Exactech
K020291
2320 NW 66TH COUR GAINESVILLE, FL 3265
352-377-1140 FAX 352-378-2617
# Exactech® Femoral Stem Centralizers
# 510(k) Summary of Safety and Effectiveness Special 510(k)
# FEB 2 7 2002
Exactech® Inc. Sponsor: 2320 N.W. 66th Court Gainesville, Florida 32653
| Phone: | (352) - 377 - 1140 |
|--------|--------------------|
| Fax: | (352) - 378 - 2617 |
FDA Establishment Number 1038671
Contact: Robert Paxson Director of Engineering & Development
January 24, 2002 Date:
> Section 4 Page 1 of 3
EXACTECH exists to inquove the quility of the advity ond intependence. We to this through innovative ideos, high-quoting probection and commitment to server.
{1}------------------------------------------------
020291
# Exactech® Femoral Stem Centralizers
# 510(k) Summary of Safety and Effectiveness Special 510(k)
# Classifications / Proprietary Names:
| Name: | Prosthesis, Hip, Semi-Constrained, Metal/Polymer,<br>Cemented (Centralizer Component) |
|-----------------------|---------------------------------------------------------------------------------------|
| Product Code: | JDI |
| C.F.R. Section: | 888.3350 |
| Device Class: | II |
| Classification Panel: | Orthopedic |
# Legally Marketed Devices for Substantial Equivalence Comparison:
| Model | Manufacturer |
|------------------|----------------|
| Total Hip System | Exactech, Inc. |
| VerSys | Zimmer |
| Answer | Biomet |
## Device Description:
## INDICATIONS
All Exactech Hip Systems are indicated for use in skeletally mature individuals Andergoing primary surgery for hip replacement due to osteoarthritis, theumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also the surgeon as the preferrod a oning spondylitis, congenital hip dysplasia, revision of potentially marcated reconstructions where sufficient bone stock is present, and to restore mobility resulting from previous fusion.
{2}------------------------------------------------
020291
## Exactech® Femoral Stem Centralizers
## 510(k) Summary of Safety and Effectiveness Special 510(k)
## CONTRAINDICATIONS
Exactech Hip Systems are contraindicated in patients with active infection, patients without sufficient bone stock to allow appropriate insertion and fixation of the prosthesis, in neuromuscular disorders that do not allow control of the hip joint, and in patients whose weight, age, or activity level would cause the surgeon to expect early failure of the system. The L-Series unipolar is also contraindicated for use in patients with evidence of degenerative changes in the acetabulum and/or pelvic fractures.
#### DESIGN
The Exactech centralizers are designed to provide central placement of Exactech cemented stems within the femoral canal.
The components have three fins and a tapered post for attachment to Exactech cemented femoral components. The device is composed of polymethylmethacrylate (PMMA) and comes in sizes ranging from 7mm to 17mm in diameter.
#### UTILIZATION
Selection of the centralizer component is made by the surgeon in relationship to the requirements of the patient. The surgeon should become thoroughly familiar with the technique of implantation of the prosthesis by: 1) appropriate reading of the literature and 2) training in the operative skills and techniques required for hip arthroplasty surgeries.
#### SUBSTANTIAL EQUIVALENCY
The proposed Exactech Centralizers are similar in design and material specifications to other products legally marketed in the United States, most notably Exactech's predicate centralizer design cleared through premarket notification #K862234.
> Section 4 Page 3 of 3
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure, with stylized lines representing its wings and body.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 2 7 2002
Mr. Robert Paxson Director of Engineering & Development Exactech® Inc. 2320 N.W. 66th Court Gainesville, Florida 32653
Re: K020291
Trade/Device Name: Exactech® Femoral Stem Centralizer Regulation Number: 21 CFR 888.3350 Regulation Name: Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented Regulatory Class: Class II Product Code: JDI Dated: January 25, 2002 Received: January 28, 2002
Dear Mr. Paxson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreations of the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Erth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
## Page 2 - Mr. Robert Paxson
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter with anow you to obgin manisting of substantial equivalence of your device to a legally prematication. "The PDF Intentify of Cassification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and II you desire spoorte ad 10 for in vitro diagnostic devices), please contact the Office of additionally 21 CFA Fat 607.10 for the raily, for questions on the promotion and advertising of Compliance at (301) 591-1837. First of Compliance at (301) 594-4639. Also, please note the your devices, product commonding by reference to premarket notification" (21CFR Part 807.97). regulation entitied, "Misoranaming of responsibilities under the Act may be obtained from the Division Other general information on your respender Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Sincerely yours,
Mark N. Melkerson
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Exactech® Femoral Stem Centralizers
## Indications for Use
510(k) Number:
K020291
Device Name:
Exactech® Femoral Stem Centralizer
### INDICATIONS
INDICATIONS
All Exactech Hip Systems are indicated for use in skeletally mature individuals All Exactech Hip Systems are manufactures for as a close of the hip, and for undergoing primary surgery for mp roplations of the hip, and for arthritis, osteonecrosis, post-traumance cophere prosthetic replacement is determined by treatment of proximal tractures was was were of Exacted Hip Systems are also
the surgeon as the preferred treatment. Components of Exactes as also the surgeon as the preferred treatment "over its hip dysplasia, revision of
entially indicated for ankylosing spondylitis, congenital his restore potentially indicated for alleytosing spondy may was and is restore mobility resulting from previous fusion,
## CONTRAINDICATIONS
CONTRAINDICATIONS
Exactech Hip Systems are contraindicated in patients with active infection, patients Exactech Hip Systems are contrainciences in patiention and fixation of the prosthesis, without sufficient bone stock to anow appropriation of the hip joint, and in patients.
in neuromuscular disorders that do not allow control of the supert accty failure of in neuromuscular disorders that do not answer the surgeon to expect early failure of the whose weight, age, of activity lever would calls and one of patients with evidence of System. The E Sories in the acetabulum and/or pelvic fractures.
R. Mark N. Mellen
Division Sign-Off) Division of General, Rest rative and Neurological Devices
K020291
Please do not write below this line - use another page if needed.
510(k) Number
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
X
or
Over the Counter Use
Section 3 Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.