MICROSCAN SYNERGIS PLUS GRAM NEGATIVE MIC/COMBO PANELS WITH TOBRAMYCIN (0.12-32 UG/ML)

K020168 · Dade Behring, Inc. · LON · Feb 6, 2002 · Microbiology

Device Facts

Record IDK020168
Device NameMICROSCAN SYNERGIS PLUS GRAM NEGATIVE MIC/COMBO PANELS WITH TOBRAMYCIN (0.12-32 UG/ML)
ApplicantDade Behring, Inc.
Product CodeLON · Microbiology
Decision DateFeb 6, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2

Indications for Use

MicroScan® Synergies plus™ Gram-Negative MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on sold media of rapidly growing aerobic and facultative anaerobic Gram-Negative bacilli (Enterobacteriaceae, glucose non-fermenters, and non-Enterobacteriaceae glucose fermenters. After inoculation, panels are read on the WalkAway® SI System or equivalent (upgraded WalkAway® 40 or WalkAway® 96) according to the Package Insert

Device Story

Device consists of micro-titer panels containing dried antimicrobial agents (Tobramycin) in Mueller-Hinton broth. Clinical laboratory personnel inoculate and rehydrate panels with standardized bacterial suspensions from solid media cultures. Panels are incubated (35°C) and processed in WalkAway® SI, 40, or 96 automated systems. System measures bacterial growth inhibition across antimicrobial concentrations to determine Minimum Inhibitory Concentration (MIC). Output provides quantitative/qualitative susceptibility data to clinicians for guiding antibiotic therapy. Benefits include standardized, automated susceptibility testing for rapid identification of effective antimicrobial agents for Gram-negative infections.

Clinical Evidence

Performance evaluated via external studies using fresh/stock clinical isolates and challenge strains. Compared against NCCLS frozen reference panels. Essential Agreement for Tobramycin was 96.6%. Reproducibility and precision testing performed using turbidity inoculum method on WalkAway® SI/40/96 instruments. Quality control testing met acceptance criteria.

Technological Characteristics

Micro-titer panel format; dried antimicrobial agents in Mueller-Hinton broth; automated incubation and optical reading via WalkAway® SI/40/96 systems; turbidity-based inoculum preparation; 35°C incubation temperature; 4.5-18 hour incubation duration.

Indications for Use

Indicated for determining antimicrobial susceptibility of rapidly growing aerobic and facultative anaerobic Gram-Negative bacilli (e.g., Enterobacteriaceae, Pseudomonas aeruginosa, Citrobacter spp., Klebsiella spp., Proteus spp.) in patients requiring susceptibility testing for Tobramycin.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" arranged in a circular fashion around the emblem. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 AUG 3 1 2004 Mr. Robert Eusebio Manager Regulatory Affairs Dade MicroScan, Inc. 1584 Enterprise Boulevard West Sacramento, CA 95691 > k020168 Trade/Device Name: MicroScan® Synergies Plus Gram Negative MIC/Combo Panels with Tobramycin (0.12 - 32 µg/ml) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System Regulatory Class: II Product Code: LON Dated: July 30, 2004 Received: August 3, 2004 Dear Mr. Eusebio: Re: This letter corrects our substantially equivalent letter of February 6, 2002, regarding the trade name which was changed to MicroScan® Synergies Plus to better reflect the intended use of the device. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent [(for the indications for use stated in the enclosure)] to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). Vou may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set premainet notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-3084. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufactures, International and Consumer Assistance at their toll free number (800) 638-2041 or at (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, fucddite. Rocke for Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use Statement 510(k) Number (if known): K020168 MicroScan® Synergies plus™ Gram-Negative MIC/Combo Panels with Tobramycin Device Name: (0.12 - 32 ug/ml) Indications For Use: MicroScan® Synergies plus™ Gram-Negative MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on sold media of rapidly growing aerobic and facultative anaerobic Gram-Negative bacilli (Enterobacteriaceae, glucose non-fermenters, and non-Enterobacteriaceae glucose fermenters. After inoculation, panels are read on the WalkAway® SI System or equivalent (upgraded WalkAway® 40 or WalkAway® 96) according to the Package Insert This particular submission is for the antimicrobial Tobramycin on the Synergies plus™ Gram-Negative MIC/Combo Panels. The Gram-Negative organisms which may be used for Tobramycin susceptibility testing in this panel are: Citrobacter amalonaticus Citrobacter koseri (diversus) Citrobacter freundit Citrobacter spp Escherichia coli Enterobacter aerogenes Enterobacter agglomerans Enterobacter cloacae Enterobacter sakazakii Enterobacter spp Klebsiella oxytoca Klebsiella ozaenae Klebsiella pneumoniae Klebsiella rhinoscleromatis Klebsiella spp Morganella morganii Proteus mirabilis Proteus vulgaris Proteus spp Providencia alcalifaciens Providencia rettgeri Providencia stuartii Pseudomonas aeruginosa Serratia spp Serratia marcescens Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Lueddin L. Poole Division Sign Off Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety Page 1 of _ 1 510(k) 22-0168 {3}------------------------------------------------ K020168 ## FEB 0 6 2002 ### 510(k) Summary ### 510(k) Submission Information: | Device Manufacturer: | Dade Behring Inc. | |----------------------|------------------------------------------------------------| | Contact name: | Maureen Mende, Group Manager Regulatory Affairs | | Fax: | 916-374-3144 | | Date prepared: | January 15, 2002 | | Product Name: | Microdilution Minimum Inhibitory Concentration (MIC) Panel | | Trade Name: | MicroScan® rapID/S plus™ Gram-Negative MIC/Combo panel | | Intended Use: | To determine antimicrobial agent susceptibility | | 510(k) Notification: | Antimicrobials: Tobramycin | | Predicate device: | MicroScan Dried Gram Negative MIC/Combo Panels | #### 510(k) Summary: MicroScan® rapID/S plus™ Gram-Negative MIC/Combo Panels are designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-negative bacilli. The MicroScan® rapID/S plus™ Gram-Negative MIC/Combo Panels are read on the WalkAway® SI System or equivalent (upgraded WalkAway® 40 or WalkAway® 96 instruments). The antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test that have been diluted in Mueller-Hinton Broth to concentrations bridging the range of clinical interest and are presented in micro-titer wells in dried form. rapID/S plus™ panels are inoculated and rehydrated with a standardized suspension of the organism and incubated at 35℃ in the WalkAway® SY System or equivalent for 4.5 - 18 hours. The minimum inhibitory concentration (MIC) for the test organism is determined by the lowest antimicrobial concentration showing inhibition of growth. The proposed MicroScan® rapID/S plus™ Gram-Negative MIC/Combo Panel demonstrated substantially equivalent performance when compared with an NCCLS frozen Reference Panel, as defined in the FDA DRAFT document "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", dated March 8, 2000. The Premarket Notification (510[k]) presents data in support of the MicroScan® rapID/S plus™ Gram-Negative MIC/Combo Panel Tobramycin. The external evaluation was conducted with fresh and stock Efficacy isolates and stock Challenge strains. The external evaluations were designed to confirm the acceptability of the proposed rapID/S plasm-Negative Panel by comparing its performance with an NCCLS frozen Reference panel. Challenge strains were compared to Expected Results determined prior to the evaluation. The rapID/S plus™ Gram-Negative Panel demonstrated acceptable performance with an overall Essential Agreement of 96.6% for Tobranycin when compared with the frozen Reference panel. Instrument reproducibility testing demonstrated acceptable reproducibility and precision with Tobramycin with Turbidity inoculum preparation method and the WalkAway® SI System or equivalent (upgraded WalkAway® 40 or WalkAway® 96 instruments). Quality Control testing demonstrated acceptable results for Tobramycin.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...