pHoenix ISE Reagents FOR Olympus AU Chemistry Systems
K020148 · Phoenix Diagnostics, Inc. · CGZ · Feb 4, 2002 · Clinical Chemistry
Device Facts
Record ID
K020148
Device Name
pHoenix ISE Reagents FOR Olympus AU Chemistry Systems
Applicant
Phoenix Diagnostics, Inc.
Product Code
CGZ · Clinical Chemistry
Decision Date
Feb 4, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1170
Device Class
Class 2
Indications for Use
The pHoenix ISE Reagents for Olympus® AU Chemistry Systems are intended for use as ISE Reagents for the determination of Nat, K+, and Cli for the Olympus® AU Clinical Chemistry Systems. The ISE Buffer is intended for use as a diluent for patient samples for the quantitative determination of Nat, K+ and Cl in serum by ISE. The ISE Internal Reference Solution is intended as a means of compensating for calibration drift in the quantitative determination of Na * and Cl in serum samples on the Olympus® AU Chemistry Systems. The ISE High and Low Standards are intended to provide calibration points for the Na , K+ and Cli Electrodes on the Olympus® AU ISE systems.
Device Story
pHoenix ISE Reagents serve as direct replacements for Olympus Diagnostics reagents on Olympus AU Clinical Chemistry Systems. System uses ISE Buffer to dilute patient serum samples; ISE Internal Reference Solution to compensate for calibration drift; and ISE High/Low Standards to calibrate Na+, K+, and Cl- electrodes. Used in clinical laboratory settings by trained technicians. Output consists of quantitative electrolyte concentration values displayed on the analyzer interface. Healthcare providers use these results to assess patient electrolyte status and guide clinical management. Benefits include maintaining continuity of testing using existing analyzer infrastructure with equivalent performance.
Clinical Evidence
Bench testing only. Precision study (N=80 per level, 20 days) evaluated total and run-to-run CV% for Na+, K+, and Cl- (Na+ CV% 0.87-1.00; K+ CV% 0.96-2.26; Cl- CV% 0.68-1.08). Method comparison (N=50) against Olympus reagents showed high correlation (r=0.997-0.999) across clinical ranges.
Technological Characteristics
In vitro diagnostic chemical reagents for ion-selective electrode (ISE) analysis. Includes buffer, reference solution, and calibration standards. Designed for use on Olympus AU Chemistry Systems. No specific materials of construction or software algorithms described; device functions as chemical consumables for existing hardware.
Indications for Use
Indicated for the quantitative determination of sodium (Na+), potassium (K+), and chloride (Cl-) in human serum samples using Olympus AU Clinical Chemistry Systems.
Regulatory Classification
Identification
A chloride test system is a device intended to measure the level of chloride in plasma, serum, sweat, and urine. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
Predicate Devices
Olympus Diagnostics ISE Buffer (AUH1011)
Olympus Diagnostics ISE Internal Reference Solution (AUH1017)
Olympus Diagnostics ISE Low Standard (AUH1014)
Olympus Diagnostics ISE High Standard (AUH1015)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human profile, composed of three overlapping shapes.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 9, 2016
Phoenix Diagnostics, Inc. Mr. Ram Nunna President 8 Tech Circle Natick, MA 01760
Re: K020148
Trade/Device Name: pHoenix ISE Reagents for Olympus® AU Chemistry Systems Regulation Number: 21 CFR 862.1170 Regulation Name: Chloride test system Regulatory Class: II Product Code: CGZ, JGS, CEM, JIX Dated: January 14, 2002 Received: January 16, 2001
Dear Mr. Ram Nunna:
This letter corrects our substantially equivalent letter of February 4, 2002.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{1}------------------------------------------------
medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image shows the name "Katherine Serrano -S" in a large, bold font. The text is black and appears to be centered on a white background. The letters are clear and easy to read, with a simple and modern typeface.
For: Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K 0 20148
Device Name: pHoenix ISE Reagents for Olympus® AU Chemistry Systems
Indications For Use:
Intended Use :-
The pHoenix ISE Reagents for Olympus® AU Chemistry Systems are intended for use as ISE Reagents for the determination of Nat, K+, and Cli for the Olympus® AU Clinical Chemistry Systems.
The ISE Buffer is intended for use as a diluent for patient samples for the quantitative determination of Nat, K+ and Cl in serum by ISE.
The ISE Internal Reference Solution is intended as a means of compensating for calibration drift in the quantitative determination of Na * and Cl in serum samples on the Olympus® AU Chemistry Systems.
The ISE High and Low Standards are intended to provide calibration points for the Na , K+ and Cli Electrodes on the Olympus® AU ISE systems.
Image /page/2/Figure/9 description: The image shows a handwritten signature of "Alan Cooper" along with the text "Divisio" and "510(k)". Below the signature, there is a handwritten code "K020148". The text and code appear to be part of a document or form, possibly related to a division or section identified by the code.
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Image /page/2/Figure/11 description: The image shows the concurrence of CDRH, Office of Device Evaluation (ODE). There is an arrow pointing to the left side of the image, which says "Prescription Use (Per 21 CFR 801.109)". On the right side of the image, it says "OR Over-The-Counter Use (Optional Format 1-2-96)".
{3}------------------------------------------------
K 020149
# FEB 0 4 2002
#### 510 (k) Summary
#### pHoenix ISE Reagents for Olympus® AU Chemistry Systems
Olympus® Diagnostics was the original manufacturer of the Olympus® AU Clinical Systems. Olympus® Diagnostics manufactures these products in Ireland and distributes through Olympus America Inc., which is located in Melville, NY.
pHoenix Diagnostics, Inc. is submitting a 510 (k) notification for the following: (1) pHoenix ISE Buffer, (2) ISE Internal Reference Solution and (3) ISE Low and High Standards. These ISE Reagents are intended for use on the ISE Module of the Olympus® AU Chemistry Systems. The ISE Buffer dilutes all measured patient samples for the quantitative determination of Na+, K+ and Cl- in serum by ISE. The ISE Internal Reference Solution is intended as a means of compensating for calibration drift in the quantitative determination of Na+, K+ and CI- in serum samples on the Olympus AU Chemistry Systems. The ISE High and Low Standards are intended to provide calibration points for the Na+, K+ and C1- Electrodes on the ISE system. pHoenix Diagnostics, Inc. is claiming substantial equivalence to predicate devices manufactured by Olympus® Diagnostics Corporation.
The products encompassed by this 510 (k) submission are Class I (75JJG) and Class II (75 JIX) In Vitro Diagnostic Solutions manufactured by pHoenix Diagnostics, Inc., 8 Tech Circle, Natick, MA 01760. These pHoenix ISE Reagents are intended to serve as direct replacements to like named products manufactured by Olympus® Diagnostics. Listed below are pHoenix products and their Olympus® Diagnostics equivalents.
| pHoenix<br>Cat.# | Olympus<br>Cat. # | Description | Models | Class |
|------------------|-------------------|---------------------------------|--------|-------|
| TBD | AUH1011 | ISE Buffer | AU | 1 |
| TBD | AUH1017 | ISE Internal Reference Solution | AU | 1 |
| TBD | AUH1014 | ISE Low Standard | AU | 2 |
| TBD | AUH1015 | ISE High Standard | AU | 2 |
pHoenix uses a similar composition, description and packaging design as that used by Olympus® Diagnostics in its products. pHoenix has shown performance equivalence of its products to Olympus® Diagnostics products in the following manner:
{4}------------------------------------------------
#### 510 (k) Summary cont.
- Through a method comparison where results obtained on a Olympus® AU Chemistry . Systems calibrated with pHoenix products and compared with results obtained on the same analyzer calibrated with Olympus® AU products; and
- Through a precision study where pHoenix products were installed on Olympus® AU . Chemistry Systems and samples were measured over 20 runs.
A summary of the results of these studies follows:
Precision data was collected from the analysis of 2 levels of serum controls measured 2 runs per day, 2 times per run for 20 days on Olympus AU Systems for Nat, K, and Cl calibrated with pHoenix standard reagents. The NCCLS Guideline for precision evaluation, EP5-T, was followed. Typical Results are as follows:
| Analyte | | N | Mean | STD | CV% | Min | Max |
|---------|------------|----|------|-------|------|-----|-----|
| Na+ | Total | 80 | 122 | 1.23 | 1.00 | 120 | 124 |
| | Run to Run | 20 | 122 | 0.67 | 0.55 | 121 | 123 |
| K+ | Total | 80 | 4.6 | 0.105 | 2.26 | 4.4 | 4.8 |
| | Run to Run | 20 | 4.6 | 0.039 | 0.84 | 4.6 | 4.7 |
| Cl- | Total | 80 | 70.5 | 0.76 | 1.08 | 69 | 73 |
| | Run to Run | 20 | 70.5 | 0.30 | 0.42 | 70 | 71 |
Level 2
#### Level 4
| Analyte | | N | Mean | STD | CV% | Min | Max |
|---------|------------|----|------|-------|------|-----|-----|
| Na+ | Total | 80 | 165 | 1.43 | 0.87 | 163 | 167 |
| | Run to Run | 20 | 165 | 0.61 | 0.37 | 164 | 166 |
| K+ | Total | 80 | 6.50 | 0.062 | 0.96 | 6.4 | 6.6 |
| | Run to Run | 20 | 6.50 | 0.028 | 0.44 | 6.4 | 6.6 |
| Cl- | Total | 80 | 121 | 0.82 | 0.68 | 119 | 122 |
| | Run to Run | 20 | 121 | 0.47 | 0.39 | 120 | 122 |
{5}------------------------------------------------
## 510 (k) Summary cont.
Correlation with Olympus Diagnostics Standard Reagents
Correlation data was collected from 50 samples (patient serum samples, control samples and spiked samples) for Nat, K+, and CI measured on Olympus AU Clinical Chemistry Systems installed with pHoenix reagents (ISE Diluent, ISE Internal Reference Solutions and Standards) as compared with Olympus reagents separately. A Linear Regression Analysis was performed using pHoenix data as the independent X Variable and Olympus Data as the Dependent Y Variable in the equation Y = a + bX . Typical results are as follows:
| Analyte | N | Slope | Intercept | Correlation<br>Coefficient | Range |
|---------|----|-------|-----------|----------------------------|-----------|
| Na+ | 50 | 1.06 | -4.9 | 0.998 | 100 - 180 |
| K+ | 50 | 1.03 | 0.11 | 0.999 | 3 - 10 |
| Cl- | 50 | 0.98 | 4.8 | 0.997 | 70 - 150 |
I hope you find this information useful and informative.
Ram Nunna, President
1/14/02
Date
Date
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.