ORTHOPAEDIC INNOVATIONS RENEWAL ACETABULAR CUP SYSTEM

K020145 · Orthopaedic Innovations, Inc. · LPH · Apr 15, 2002 · Orthopedic

Device Facts

Record IDK020145
Device NameORTHOPAEDIC INNOVATIONS RENEWAL ACETABULAR CUP SYSTEM
ApplicantOrthopaedic Innovations, Inc.
Product CodeLPH · Orthopedic
Decision DateApr 15, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3358
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Renewal™ Acetabular Cup System is intended for primary or revision reconstruction, with or without bone cement, of the acetabular portion of severely disabled and/or very painful hip joints, where radiographic evidence of sufficient sound bone is present. Clinical Indications for Use are: 1. Patient conditions of noninflammatory degenerative joint disease (NIDJD), e.g., avascular necrosis, osteoarthritis, ankylosis, protrusio acetabuli, and painful hip dysplasia; 2. inflammatory joint disease (IJD), e.g., rheumatoid arthritis; 3. patients with failed previous surgery where pain, deformity, or dysfunction persists; 4. revision of previously failed hip arthroplasty; and 5. treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.

Device Story

Renewal™ Acetabular Cup System is a hip prosthesis component for total hip arthroplasty. System comprises titanium alloy hemispherical shells (with/without peripheral rim flare, porous bead coating, optional hole configurations) and UHMWPE liners. Liners feature snap-lock assembly, various overhang angles (0°, 10°, 15°, 20°), and optional lateralized shifts. Device is used by orthopedic surgeons in clinical settings for hip reconstruction. It articulates with existing metal or ceramic femoral heads. Benefits include restoration of hip function and pain relief for patients with degenerative or inflammatory joint disease, or failed prior surgeries.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Materials: Titanium alloy shell, commercially pure titanium (CPTi) porous bead coating, UHMWPE liners. Design: Hemispherical shell, snap-lock liner interface, 0-20° overhang options, 6+mm lateralized shift options. Articulation: Metal/polymer interface. Sterilization: Not specified. Connectivity: None (mechanical device).

Indications for Use

Indicated for patients with severe hip joint disability/pain requiring primary or revision reconstruction. Conditions include NIDJD (avascular necrosis, osteoarthritis, ankylosis, protrusio acetabuli, hip dysplasia), IJD (rheumatoid arthritis), failed previous surgery, revision of failed arthroplasty, and nonunion/fractures of the proximal femur with head involvement.

Regulatory Classification

Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

Submission Summary (Full Text)

{0}------------------------------------------------ ## APR 1 5 2002 510(k) Notification Orthopaedic Innovations Renewal™ Acetabular Cup System K020145 1 OF 2 # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807, this information serves as a Summary of Safety and Effectiveness for use of the Renewal™ Acetabular Cup System. Submitted By: Date: Contact Person: Orthopaedic Innovations, Inc. Renewal™ Acetabular Cup System Metal/Polymer Acetabular Components 21 CFR 888.3358 Prosthesis, Hip, Semi- January 2, 2002 Gregory M. Mercuri Director of R & D Proprietary Name: Common Name: Classification Name and Reference: Device Product Code and Panel Code: Uncemented - Class II Constrained, metal/polymer, Orthopedics/87/LPH ## DEVICE INFORMATION ### Intended Use: The Renewal™ Acetabular Cup System is intended for primary or revision reconstruction, with or without bone cement, of the acetabular portion of severely disabled and/or very painful hip joints, where radiographic evidence of sufficient sound bone is present. Clinical Indications for Use are: - 1. Patient conditions of noninflammatory degenerative joint disease (NIDJD), e.g., avascular necrosis, osteoarthritis, ankylosis, protrusio acetabuli, and painful hip dysplasia; - 2. inflammatory joint disease (IJD), e.g., rheumatoid arthritis; - patients with failed previous surgery where pain, deformity, or dysfunction 3. persists; - 4. revision of previously failed hip arthroplasty; and - 5. treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques. {1}------------------------------------------------ The Renewal™ Acetabular Cup System components are single use components intended for use in conjunction with associated metal or ceramic femoral heads as part of an uncemented or cemented total hip arthroplasty. K020145 20F2 ## Device Description: The Renewal™ Acetabular Cup System consists of metal acetabular shells and UHMWPE acetabular liners. The Renewal™ Acetabular Shell will be available in a hemispherical titanium alloy shell with and without a 2 mm peripheral rim flare. The shells will be coated with commercially pure titanium porous beads. Design features of the Renewal™ Acetabular shell are summarized below: - Total hemispherical design . - Hemispherical design with a 2 mm peripheral flare . - Coated with CPTi porous beads ● - . No-Hole, 3-hole, 5-Hole, and 9-Hole options - Threaded apical hole plug The Renewal™ Acetabular Liners will be available with 0°, 10°, 15°, 20° overhangs with/without a 6+mm lateralized shift. The liner's internal geometry will be intended to be used with existing femoral heads manufactured from cobalt chrome or ceramic. The Renewal™ liner's external geometry will be designed to accept the Renewal™ Acetabular Shells. Design features of the Renewal™ Acetabular Liner are summarized below: - 360° liner overhang positioning options . - Features an easy-to-assemble snap lock system to lock the liner into the shell . - . The lip of the 10° and 20° liners will be full hemispheres - The lip of the 15° liner will be machined to 180° . - The liners will be offered with 0°, 10°, 15°, and 20° overhangs . - The 0° and 10° liners will also be offered with a 6+mm lateralized shift ● - Internal diameter will have a 2mm chamfer to minimize impingement. . The thinnest part of any UHMWPE articulating insert will be greater than 4mm if attached to a metal backing. ### Substantial Equivalence Information: The intended use, material, type of interface, and design features of the Renewal™ Acetabular Cup System are substantially equivalent to the competitive devices previously cleared for market. The safety and effectiveness of the Renewal™ Acetabular Cup System are adequately supported by the substantial equivalence information, materials data, and testing results provided within the Premarket Notification. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three overlapping profiles, suggesting a sense of community or collaboration. The logo is rendered in black and white. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR 1 5 2002 Mr. Gregory M. Mercuri Director of Research & Development Orthopaedic Innovations, Inc. 6188 Olson Memorial Highway Golden Valley, Minnesota 55422 Re: K020145 Trade Name: Renewal™ Acetabular Cup System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH Dated: January 10, 2002 Received: January 16, 2002 Dear Mr. Mercuri: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Mr. Gregory M. Mercuri This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, for Mark A. Wilkerson Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Device Name: Renewal™ Acetabular Cup System The Renewal™ Acetabular Cup System is intended for primary or revision reconstruction, with or without bone cement, of the acetabular portion of severely disabled and/or very painful hip joints, where radiographic evidence of sufficient sound bone is present. Clinical Indications for Use are: - 1. Patient conditions of noninflammatory degenerative joint disease (NIDJD), e.g., avascular necrosis, osteoarthritis, ankylosis, protrusio acetabuli, and painful hip dysplasia; - 2. inflammatory joint disease (IJD), e.g., rheumatoid arthritis; - 3. patients with failed previous surgery where pain, deformity, or dysfunction persists: - 4. revision of previously failed hip arthroplasty; and AND ITE DE CHI THIC I THE - 5. treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques. | DEADER INC. YANIE DEAZY THE LINE-CONTINUE VIY AND I ODA FA | | | |------------------------------------------------------------|-----------------|----------------------------------------------------------------------------------| | Concurrence of CDRH, Office of Device Explugtion (ODE) | 510(k) Number - | vision Sign-Off)<br>Division of General. Restorative<br>and Neurological Devices | | Prescription Use × | OR | Over-The-Counter Use |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...