K020144 · Siemens Medical Solutions USA, Inc. · DRT · Feb 15, 2002 · Cardiovascular
Device Facts
Record ID
K020144
Device Name
SIEMENS INFINITY SC 6002XL MODIFICATIONS
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
DRT · Cardiovascular
Decision Date
Feb 15, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The intended use of this device is to monitor heart rate, respiration rate, invasive pressure, noninvasive pressure, arrhythmia, temperature, arterial oxygen saturation and pulse rate, central apnea, end tidal carbon dioxide, and ST Segment Analysis. This device will produce visual and audible end tidal of these parameters vary beyond preset limits and produce timed or alarm recordings. This device will connect to Siemens R50 recorders, either directly or via the INFINITY network.
Device Story
The Siemens INFINITY SC 6002XL is a physiological patient monitor. This modification adds a wireless ethernet adapter (PCMCIA) for data transmission to the INFINITY Network and MultiView WorkStation, enabling remote viewing. It also includes a 'Full Arrhythmia Option' with eight additional arrhythmia calls and a fourth channel display option. The device processes physiological signals (ECG, pressure, SpO2, CO2, etc.) to provide real-time monitoring, ST segment analysis, and arrhythmia detection. It generates visual and audible alarms when parameters exceed preset limits. Used in clinical settings by trained healthcare professionals to assess patient status and support clinical decision-making. The device is not MRI compatible.
Clinical Evidence
No clinical data provided. Equivalence is based on non-clinical performance data and the use of established, previously cleared algorithms.
Technological Characteristics
Physiological monitor with PCMCIA wireless ethernet adapter (IEEE 802.11). Supports invasive/non-invasive pressure, ECG, SpO2, CO2, temperature, and respiration monitoring. Features include a fourth channel display and full arrhythmia detection software. Connectivity via INFINITY network to MultiView WorkStation. Not MRI compatible.
Indications for Use
Indicated for monitoring heart rate, respiration, invasive/non-invasive pressure, arrhythmia, temperature, SpO2, pulse rate, central apnea, end-tidal CO2, and ST segment in adult, pediatric, and neonatal populations. Contraindicated for use in MRI magnetic fields. Intended for use by trained healthcare professionals (physicians, nurses, technicians) in clinical environments.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
# FEB 1 5 2002
| 1. 510(k) SUMMARY |
|--------------------------------------------------------|
| as required per 807.92(c) |
| Submitters Name, Address: |
| Siemens Medical Solutions, Inc. |
| Electromedical Systems Group, PCS |
| Danvers, MA 01923 |
| Tel: (978) 907-7500 |
| Fax: (978) 750-6879 |
| Official Correspondent: Connie Hertel, Director |
| Quality Assurance & Regulatory Affairs |
| Contact person for this submission: Penelope H. Greco |
| Regulatory Submissions Manager |
| Date submission was prepared: January 14, 2002 |
# 2. Trade Name, Common Name and Classification Name:
- A. Trade Name:
1.
Siemens INFINITY SC 6002XL Modifications
- Common Name, Classification Name, Class and Regulation Number: B.
| Common Name | Product Code | Class | Regulation Number |
|--------------------------------------------------------------------------|--------------|-------|-------------------|
| Monitor, Physiological, Patient (with<br>Arrhythmia Detection or Alarms) | MHX | III | 870.1025 |
| Arrhythmia Detector & Alarm | 74DSI | III | 870.1025 |
| System, Network and Communication,<br>Physiological Monitors | MSX | | |
- 3. Legally Marketed Device: Siemens INFINITY SC 6002XL K993974, K002105
- 4. Device Description:
## SC 6002XL Wireless Option
A wireless ethernet adapter installed in the PCMCIA port communicates with the INFINITY Network (K955059) through installed access points hardwired to the network. All patient related and network data is transmitted to the pre-installed access points allowing the transmitted data from the monitor to be viewed at an assigned MultiView WorkStation (K955059). This configuration provides the network with the same functionality as a hardwired system.
The MultiView WorkStation supports wireless operation by providing information to the SC6002XL allowing the monitor to have a unique network address. Once the SC6002XL has established itself on the network, the MultiView WorkStation has no new functionality. MultiView WorkStation hazards identified with the MVWS support of the wireless functionality have been addressed in MVWS VF2 software.
#### 1/2
## COMPANY CONFIDENTIAL
#### Siemens Medical Solutions, USA
Electromedical Systems Group, PCS
16 Electronics Avenue Danvers, MA 01923
Tel: (978) 907-7500 Fax: (978) 750-6879
{1}------------------------------------------------
K020144
p. 2/2
# SC 6002XL with Full Arrhythmia Option
The SC 6002XL VF1 release includes the Full Arrhythmia Option. With this option eight additional arrhythmia calls are enabled. The arrhythmia algorithm is the same algorithm found in Siemens MVWS INFINITY Telemetry (K003179), as well as Siemens patient monitors, including: SC 7000 / SC 9000XL (K980882) and SC 8000 (K983632). Previous releases of the SC 6002XL (K993974 DC 900012 (1196002) the same arrhythmia algorithm, however, in the first two releases of the SC and I would the Basic Arrhythmia functionality, with fewer arrhythmia calls was made available.
#### Fourth Channel Option
I ourth Chamer Option
In addition, to provide customers with more readily available information, a fourth channel is being made available. The fourth channel is also a locked option.
#### 5. Intended Use:
The intended use of this device is to monitor heart rate, respiration rate, invasive pressure, noninvasive pressure, arrhythmia, temperature, arterial oxygen saturation and pulse rate, central apnea, end tidal carbon dioxide, and ST Segment Analysis. This device will produce visual and audible end tidal of these parameters vary beyond preset limits and produce timed or alarm recordings. This device will connect to Siemens R50 recorders, either directly or via the INFINITY network.
6. Assessment of non-clinical performance data for equivalence: Section L
7. Assessment of clinical performance data for equivalence: Not applicable
8. Biocompatability: Not applicable
9. Sterilization: Not applicable
10. Standards and Guidances: IEEE 802.11
Standard for Wireless Medium Access Method (MAC) and Physical Layer (PHY)
2/2
# COMPANY CONFIDENTIAL
#### Siemens Medical Solutions, USA Electromedical Systems Group, PCS
16 Electronics Avenue Danvers, MA 01923
Tel: (978) 907-7500 Fax: (978) 750-6879
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized eagle-like symbol with three parallel, curved lines representing the bird's body and wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 5 2002
Ms. Penelope H. Greco Manager, Regulatory Submissions Siemens Medical Solutions, Inc. Electromedical Systems Group, PCS 16 Electronics Avenue Danvers, MA 01923
Re: K020144
Trade Name: Siemens Medical INFINITY SC 6002XL (VF1) Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector & Alarm Regulatory Class: Class III (three) Product Code: MHX Dated: January 14, 2002 Received: January 16, 2002
Dear Ms. Greco:
We have reviewed your Section 510(k) premarket notification of intent to market the device wt nave reviewed your becamined the device is substantially equivalent (for the indications ferenced above und nave acterimes ally marketed predicate devices marketed in interstate for use stated in the encreated of the enactment date of the Medical Device Americans, or to connicted proc to May 20, 1978, the excordance with the provisions of the Federal Food, Drug, de necs that have been recassion at require approval of a premarket approval application (PMA). and Cosmeter roc (110) that as nov co, subject to the general controls provisions of the Act. The Tou may, dicrorere, mains of the Act include requirements for annual registration, listing of general controls provisions wastice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is elassified (soc as words. Existing major regulations affecting your device can
may be subject to such additional controls. Existing major regulations of the may be subject to back address soulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
# Page 2 - Ms. Penelope H. Greco
Please be advised that FDA's issuance of a substantial equivalence determination does not mean
ts of the Active in the start the see device complies with other requirements o Please be advised that FDA's issuance of a substantial equires with other requirements of the Act
that FDA has made a determination that your device complies with other requi that FDA has made a determination inal your divres on other Federal agencies. You must
or any Federal statutes and regulations administered by other Federal on and listin or any Federal statutes and regulations and limited to: registration and listing (21
comply with all the Act's requirements, including, but not tice requirements as set comply with all the Act's requirements, including, out morning practice requirements as set
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirement CFR Part 807); labeling (21 CFK Part 801); good manufacture policable, the electronic
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the e forth in the quality systems (QS) regulation (21 CFR Part obs) and 11 TT
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k)
The are and the many on the time of rubstantial equivalence of your device to a This letter will allow you to begin marketing your ucvice of your device of your device to a legally
premarket notification. The FDA in a los if section for senits your devi premarket notification. The FDA finding of substantal equivalence by -------------------------------------------------------------------------------------------------------to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire specific advice for your devices on on ademing copy.es contact the Office of
additionally 21 CFR Part 809.10 for in vites diagnostic devices on the promotion an additionally 21 CFR Part 809. ID for m Wird unserved by promotion and advertising of
Compliance at (301) 594-464. Additionally, for questions on the promotion and advertisin Compliance at (301) 594-4646. Additionally, 10. questions of G44-4639. Also, please note the your device, please contact the Office of Complanse are a potification" (21CFR Part 807.97).
regulation entitled, "Misbranding by reference to premarket not may be obtained f regulation entitled, "Misbranding of responsibilities under the Act may be obtained from the Other general information on your responsional and Consumer Assistance at its toll-free number
Division of Small Manufacturers, International and Consumer Assistance at its t Division of Sman Manazassassa or at its Internet address (600) 030x . fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Sincerely yours,
Bram D. Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# 510(k) Number (if known): ٻک کان کا
Device Name: Siemens INFINITY SC 6002XL Modifications
Indications for Use:
This device is capable of monitoring:
- Heart Rate .
- Respiration Rate .
- Invasive Pressure .
- Non-Invasive Pressure ●
- Arrhythmia .
- Temperature .
- Arterial oxygen saturation .
- Pulse rate .
- (central) apnea .
- end-tidal CO2 .
- ST Segment Analysis .
This device will produce visual and audible alarms if any of these parameters vary This device will produce visual and adding arm recordings. This device will connect to
beyond preset limits and produce timed on always were waters Deyond preset mints and proof directly or via the INFINITY network.
The device is intended to be used in the environment where patient care is provided by The device is miended to be used in the environment, and technicans, trained on the use of the Healthcare Froressionals, F.C. pity Sichans, Indicated, based upon their devices, who willement of the patient's medical condition.
The devices are intended to be used in the Adult, Pediatric and Neonatal populations, with the exception of the The devices are intended to be ascu in the Raan, I valuational for the neonatal population.
#### MRI Compatibility Statement:
MRI Compatibunty Statement.
The Siemens INFINITY SC 6002XL is not compatible for use in a MRI magnetic field.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| | | Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|----------------------|---------------------------------------------------------|--------------------------------------------------------|-----------------------------|
| Prescription Use | <div> <img alt="Checkmark" src="checkmark.png"/> </div> | OR | Over-The-Counter Use ______ |
| (Per 21 CFR 801.109) | | | (Optional Format 1-2-96) |
| | <div> <img alt="Signature" src="signature.png"/> </div> | | |
| | Division of Cardiovascular & Respiratory Devices | | |
| 510(k) Number | K200444 | | |
1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.