K020134 · United States Surgical, A Division of Tyco Healthc · KWP · Apr 15, 2002 · Orthopedic
Device Facts
Record ID
K020134
Device Name
SPIRAL RADIUS 90-D RODDING SYSTEM
Applicant
United States Surgical, A Division of Tyco Healthc
Product Code
KWP · Orthopedic
Decision Date
Apr 15, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When used as an anterolateral/anterior system, the levels of attachment are the lumbar, thoracic, and cervical spine. The points of attachment are screw fixation into the anterolateral vertebral bodies of the lumbar and thoracic spine (T1-L5) and anterior vertebral bodies of the cervical spine. The indications are degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion. When used as a nonpedicle, posterior system consisting of hooks, crosslinks, and sacral/iliac screws, the levels of attachment are the lumbar, thoracic, and cervical spine and the sacrum and ilium. The indications are degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion. When used as a pedicle screw system in the non-cervical spine of skeletally mature patient, the system is indicated for patients: (a) having severe spondylolisthesis (Grades 3 and 4) at L5-S1 joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (d) who are having the device removed after the development of solid fusion mass. In addition, when used as a pedicle screw system in the non-cervical spine of skeletally mature patients, the system is intended to provide immobilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis). The Surgical Dynamics Spiral Radius 90-D Rodding System is to be used in conjunction with the SDRS Spinal Rods, SDRS Spinal Rod with Sacral Eye and the SDRS Locking Screw.
Device Story
Spinal fixation system comprising screws, hooks, rods, crossbars, and connectors; implanted in vertebral bodies to provide immobilization and stabilization as an adjunct to spinal fusion. Operated by surgeons in clinical/OR settings. Components locked via locking caps. System provides mechanical support to treat spinal instabilities, deformities, and fractures. Benefits include structural stabilization of spinal segments to facilitate fusion and patient recovery.
Clinical Evidence
Bench testing only; includes static and dynamic compression testing and static torsion testing. No clinical data provided.
Technological Characteristics
Components composed of Ti-6Al-4V (ASTM F136). System includes screws, hooks, rods, crossbars, and connectors. Mechanical fixation via locking caps. No software or electronic components.
Indications for Use
Indicated for skeletally mature patients requiring spinal immobilization/stabilization as an adjunct to fusion for degenerative disc disease, spondylolisthesis, fracture, spinal stenosis, deformities (scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed fusion. Specific pedicle screw use restricted to severe L5-S1 spondylolisthesis (Grades 3-4) or L3-sacrum stabilization with autogenous bone graft.
Regulatory Classification
Identification
A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.
Predicate Devices
Surgical Dynamics Rodding System
Reference Devices
Spiral Radius 90-D Rodding System Rod to Rod Connector (K970635)
Spiral Radius 90-D Rodding System Multi-Angle Concepts Device (K970635)
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) SUMMARY
# 020134
Surgical Dynamics™, 150 Glover Avenue, Norwalk, CT 06856 USA
| CONTACT PERSON: | Jenny Schuck |
|----------------------|-------------------------------------------------|
| PRODUCT NAME: | Spiral Radius 90-D Rodding System |
| CLASSIFICATION NAME: | Spinal Intervertebral Body Fixation Orthosis |
| | Spinal Interlaminal Fixation Orthosis |
| | Spondylolisthesis Spinal Fixation Device System |
| PREDICATE DEVICE: | Surgical Dynamics Rodding System |
#### DEVICE DESCRIPTION
The Spiral Radius 90-D Rodding System consists mainly of 1) screws and hooks that are implanted in vertebral bodies and 2) rods that fit into and are locked upon the screw heads using a locking cap. The system also includes crossbars that connect 2 parallel-running rods, offset screws and clips that allow a rod to be placed in certain situations where a direct connection to the vertebrae would cause excessive bending of the rod, and rod-to-rod connectors that allow a construct to be extended. All components of the system are composed of Ti-6Al-4V which conforms to ASTM F136.
#### INDICATIONS FOR USE
When used as an anterolateral/anterior system, the levels of attachment are the lumbar, thoracic, and cervical spine. The points of attachment are screw fixation into the anterolateral vertebral bodies of the lumbar and thoracic spine (T1-L5) and anterior vertebral bodies of the cervical spine. The indications are degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion.
When used as a nonpedicle, posterior system consisting of hooks, crosslinks, and sacral/iliac screws, the levels of attachment are the lumbar, thoracic, and cervical spine and the sacrum and ilium. The indications are degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion.
When used as a pedicle screw system in the non-cervical spine of skeletally mature patient, the system is indicated for patients: (a) having severe spondylolisthesis (Grades 3 and 4) at L5-S1 joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (d) who are having the device removed after the development of solid fusion mass.
In addition, when used as a pedicle screw system in the non-cervical spine of skeletally mature patients, the system is intended to provide immobilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis).
The Surgical Dynamics Spiral Radius 90-D Rodding System is to be used in conjunction with the SDRS Spinal Rods, SDRS Spinal Rod with Sacral Eye and the SDRS Locking Screw.
{1}------------------------------------------------
#### TESTING
Mechanical testing was performed, including static and dynamic compression testing and static torsion testing.
### SUBSTANTIAL EQUIVALENCE*
The components added to the Spiral Radius 90-D Rodding System were claimed to be substantially The components added to the Sphar Rashas 70 D iterstang System (K970635). The Spiral Radius 90equivalent " to components of the Bargeor was claimed to be substantially equivalent* to the Spinal D Koduling System Rod to Rod Connector and the Spiral Radius 90-D Rodding System Multi-Angle Concepts Daci is Life-to End Combeter ining to these devices was provided in the submission.
* Any claim of substantial equivalence is made exclusively in regard to the U.S. Food, Drug and Cosmetic Act and should not be viewed in any other light.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. The eagle is composed of three curved lines, and the overall design is simple and modern. The logo is likely used to represent the department on official documents and communications.
APR 1 5 2002
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Jenny Schuck Regulatory Affairs Senior Associate United States Surgical 150 Glover Avenue Norwalk, Connecticut 06856
K020134 Re: Spiral Radius 90-D Rodding System Regulation Numbers: 888.3050; 888.3060; and 888.3070 Regulation Names: Spinal Interlaminal Fixation Orthosis; Spinal Intervertebral Body Fixation Orthosis; Spondylolisthesis Spinal Fixation Device System; and Pedicle Screw Spinal System
Regulatory Class: II Product Codes: KWP, KWQ, MNI, MNH Dated: January 14, 2002 Received: January 15, 2002
Dear Ms. Schuck:
We have reviewed your Section 510(k) premarket notification of intent to market the device wt nave reviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreated of the enactment date of the Medical Device Amendments, or to conimered prior to May 20, 1978, are excordance with the provisions of the Federal Food, Drug, de necs that have been recuire approval of a premarket approval application (PMA). and Cosmotion Free (110) that the device, subject to the general controls provisions of the Act. I ou may, therefore, maner and to 150, at include requirements for annual registration, listing of The general controls provision, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is blusbined toor as cotrols. Existing major regulations affecting your device can may be subject to sublications, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean i tease of acreation that i broundevice complies with other requirements of the Act that I Dr Hear Intactions and regulations administered by other Federal agencies. You must or unf 1 catal bather bequirements, including, but not limited to: registration and listing (21 compry with an all the 110 01 CFR Part 801); good manufacturing practice requirements as set OFF Part 8017; and ity systems (21 CFR Part 820); and if applicable, the electronic roluct radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
### Page 2 - Ms. Jenny Schuck
This letter will allow you to begin marketing your device as described in your Section 510(k) rms letter with and your e FDA finding of substantial equivalence of your device to a legally premainer noninate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of compliance at (301) 594-4639 of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Sincerely yours,
Mark N. Milken
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## INDICATIONS ENCLOSURE
### 510(k) Number (if known): K020134
#### Spiral Radius 90-D Rodding System Device Name:
#### Indications For Use:
March 28, 2002
When used as an anterolateral/anterior system, the levels of attachment are the lumbar, thoracic, and cervical spine. The points of attachment are screw fixation into the anterolateral vertebral bodies of the cervical spine. The points of the minent and on vertebral bodies of the cervical spine. The indications are lumbal and inoracie spine (11 L29) and all pain of discogenic origin with degeneration of the disc degenerative dise unsease (connect as bath pes), spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion.
When used as a nonpedicle, posterior system consisting of hooks, crosslinks, and sacral/iliae screws, the When used as a nonpodiers, person thoracic, and cervical spine and the sacrum and ilium. The indications are degenerative disc disease (defined as back pain of discogenic origin with degeneration of Indications are degenciative and radiographic studies), spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, tordosis), tumor, pseudoarthrosis, or failed previous fusion.
When used as a pedicle screw system in the non-cervical spine of skeletally mature patient, the system is when used as a pentents: (a) having severe spondylolisthesis (Grades 3 and 4) at L5-S1 joint; (b) who are indivated for patients. (a) having severe posit only, (c) who are having the device fixed or attached to the receiving rasions using adlegements occa: and (d) who are having the device removed after the development of solid fusion mass.
In addition, when used as a pedicle screw system in the non-cervical spine of skeletally mature patients, In adultion, when used as a pocure convilization and stabilization of spinal segments as an adjunct to the system is intended to provide anims acute and chronic instabilities or deformities of the thoracic, lusion in the treatinent of the long with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudoarthrosis).
The Surgical Dynamics Spiral Radius 90-D Rodding System is to be used in conjunction with the SDRS The Surglear Dymannonial Rod with Sacral Eye and the SDRS Locking Screw.
(Please do not write below this line - continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use: | X | OR Over-The-Counter Use: |
|----------------------|---|--------------------------|
| (Per 21 CFR 801.109) | | |
for Mark N Mellesun
(Division Sign-Off)
Division of General, Restorative and Neurological Devices
| 510(k) Number | K020134 |
|---------------|---------|
|---------------|---------|
United States Surgical, Surgical Dynamics
510(k) K020134
Page D1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.