K020128 · Diagnostic Systems Laboratories, Inc. · JHI · Mar 1, 2002 · Clinical Chemistry
Device Facts
Record ID
K020128
Device Name
ACTIVE FREE BETA-HCG ELISA, MODEL DSL-10-8500
Applicant
Diagnostic Systems Laboratories, Inc.
Product Code
JHI · Clinical Chemistry
Decision Date
Mar 1, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1155
Device Class
Class 2
Indications for Use
The DSL-10-8500 ACTIVE™ Free B-hCG ELISA assay is intended for the quantitative determination of Free ß-hCG in human serum. It is intended for in vitro diagnostic use by professional laboratory personnel as an aid in the detection of pregnancy.
Device Story
The DSL ACTIVE™ Free β-hCG ELISA is a two-site sandwich enzyme-linked immunosorbent assay (ELISA) used in clinical laboratories. The device utilizes microtitration wells coated with a capture antibody and a second antibody conjugated to horseradish peroxidase. Patient serum is added to the wells; the free β-hCG analyte binds to both antibodies to form a sandwich complex. After washing to remove unbound materials, the complex is analyzed via spectrophotometry. The resulting absorbance is directly proportional to the concentration of free β-hCG in the sample. The antibodies are specific to sterically remote epitopes on the β-subunit, preventing cross-reactivity with intact hCG, LH, FSH, AFP, TSH, or prolactin. Healthcare providers use the quantitative results to aid in the detection and monitoring of pregnancy and pregnancy-related disorders.
Clinical Evidence
Clinical evaluation included a comparative study of 98 pregnant patient samples analyzed by both the subject device and the predicate (DSL-8300). Linear regression analysis yielded the equation: DSL-10-8500 = 0.007(DSL-8300) - 2.7, with a correlation coefficient of r = 0.82 and p < 0.001. Additionally, 40 samples from non-pregnant individuals (20 males, 20 females) were tested, with 95% of values in both groups measuring < 1 mIU/mL.
Technological Characteristics
Two-site sandwich ELISA; microtitration well format; horseradish peroxidase enzyme conjugate; spectrophotometric detection. Specificity achieved via antibodies raised against sterically remote epitopes on the β-subunit of hCG. In vitro diagnostic use.
Indications for Use
Indicated for the quantitative measurement of Free ß-subunit of human chorionic gonadotropin (hCG) in human serum as an aid in the detection of pregnancy. For professional laboratory use.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
Predicate Devices
DSL-8300 Intact-hCG IRMA
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date "MAR 1 2002" in bold, sans-serif font. The month is abbreviated to "MAR", followed by the number "1", and then the year "2002". The text is black against a white background.
Image /page/0/Picture/1 description: The image shows a logo with the letters "dsl" in white against a black circular background. The letters are stylized and connected, with the "d" and "l" appearing to be mirror images of each other. There is a small registered trademark symbol located between the "s" and "l".
#### SUMMARY OF SAFETY AND EFFECTIVENESS 4.0
Submitter: Diagnostic Systems Laboratories, Inc. 445 Medical Center Boulevard Webster, Texas 77598 Phone: 281-332-9678 FDA Registration # 1628193
Contact: Carroll Potts Manager of Regulatory Affairs Office: (281) 332-9678 Fax: (281) 338-1895 cpotts@dslabs.com
| Date of Summary: | December 31, 2001 |
|------------------------|-------------------------------------------------|
| Device Trade Name: | DSL ACTIVE™ Free β-hCG ELISA Kit |
| Classification Name: | Enzymeimmunoassay, Human Chorionic Gonadotropin |
| Analyte Code and Name: | Human Chorionic Gonadotropin (hCG) |
| Regulatory Class: | II |
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is:________________________________________________________________________________________________________________________________________________
### Test Description:
The DSL-10-8500 ACTIVE™ Free B-hCG ELISA kit was developed for the quantitative measurement of Free ß-subunit of human chorionic gonadotropin (hCG) in human serum. Human chorionic gonadotropin (hCG) is a glycoprotein secreted by the trophoblastic cells of the placenta. hCG is composed of two dissimilar non-covalently linked polypeptides known as the or- and ßsubunits [1]. The or-subunit is very similar to the ox-chain of lutropin (hLH), follitropin (hFSH), and thyrotropin (hTSH). The ß-subunit is responsible for the biological activity of this hormone [1,2].
The primary utility of hCG measurement is for the early detection and monitoring of pregnancy, and pregnancy related disorders. Healthy, non-pregnant individuals have low (< 5 IU/L) to undetectable hCG concentrations in serum. During pregnancy, hCG concentrations increase to about 50 IU/L in the first week after conception and double every 1.5 to 3 days for the first six weeks. Levels continue to rise until the end of the first trimester, then gradually fall to a lower level for the remainder of the pregnancy. After delivery, hCG returns to < 5 IU/L and is usually undetectable several days postpartum.
Initial immunoassays for hCG, although capable of achieving sufficient sensitivity for early detection of pregnancy, suffered from a significant degree of cross-reactivity with pituitary glycoproteins, particularly hLH.
Page 12 of 19
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image is a black and white logo. The logo consists of a black circle with the letters "dsl" in white inside the circle. The letters are stylized and connected, with the "d" and "l" being taller than the "s". There is a small registered trademark symbol next to the "s".
The DSL ACTIVE® Free B-hCG ELISA is a two-site ELISA assay in which the free ß-hCG to be measured is "sandwiched" between two antibodies. The first antibody is immobilized to the inside wall of the microtitration well, the other antibody is conjugated to the enzyme horseradish peroxidase for detection. The analyte present is bound by both the antibodies to form a "sandwiched" complex. Unbound materials are removed by washing the wells. The resultant is analyzed in a spectrophotometer for absorbance. The amount of bound Free B-hCG is directly proportional to the concentration of the Free ß-hCG present in the sample. The antibodies were raised against sterically remote epitopes on the beta subunit of hCG, such that in intact hCG molecules, these antibody recognition sites are not available. The DSL ACTIVE® Free M-RCG ELISA is very specific for free ß-hCG without cross-reactivity or interference due to LH, FSH, AFP, TSH or prolactin.
# Intended Use:
The DSL-10-8500 ACTIVE™ Free B-hCG ELISA assay is intended for the quantitative determination of Free ß-hCG in human serum. It is intended for in vitro diagnostic use by professional laboratory personnel as an aid in the detection of pregnancy.
## Substantial Equivalence:
The DSL-10-8500 ACTIVE™ Free B-hCG ELISA is substantially equivalent to the DSL-8300 Intact-hCG IRMA.
To demonstrate substantial equivalence between the two assays, pregnant patient samples (n=98) were collected and assayed using both methods. Samples were chosen based on expected hCG levels so that samples with low, intermediate and high levels would be evaluated. Linear regression analysis of the results obtained for the comparison gave the equation: DSL-10-8500 = 0.007(DSL-8300) - 2.7, with a correlation coefficient of r = 0.82 and p < 0.001.
Additionally, serum samples from 20 normal males and 20 non-pregnant females were analyzed for Free B-hCG with the DSL-10-8500 ACTIVE™ Free B-hCG ELISA, where 95% of male values were < 1 mIU / mL and 95% of female values were < 1 mIU / mL. Therefore, the physiological profiles for intact hCG and Free ß-hCG remain parallel throughout pregnancy.
Page 13 of 19
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the department's name encircling an image of an abstract caduceus. The caduceus is a symbol often associated with healthcare and medicine. The text reads, "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
MAR 1 2002
Ms. Carroll Potts, M.S. Manager of Regulatory Affairs Diagnostic Systems Laboratories, Inc. 445 Medical Center Blvd Webster, TX 77598
Re: k020128
Trade/Device Name: DSL ACTIVE™ Free β-hCG ELISA Regulation Number: 21 CFR 862.1155 Regulation Name: Human chorionic gonadotropin (HCG) test system Regulatory Class: Class II Product Code: JHI Dated: February 8, 2002 Received: February 11, 2002
Dear Ms.Potts:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and ' additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo consisting of a black circle with white letters inside. The letters appear to be "dsl" in a stylized, interconnected font. The letters are designed to flow into each other, creating a unified and abstract design within the circular frame.
#### 1.0 INDICATIONS FOR USE
510 (k) Number (if known): K. (201288
Device Name: DSL ACTIVE™ Free B-hCG ELISA
Indications for Use:
The DSL-10-8500 ACTIVE™ Free 3-hCG Enzyme-Linked Immunosorbent Assay (ELISA) kit provides materials for the quantitative measurement of Free ß-subunit of human chorionic gonadotropin (hCG) in serum. This assay is intended for in vitro diagnostic use as an aid in the detection of pregnancy.
Siecin Coopin
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K02128
(Please Do Not Write Below This Line. Continue On Another Page If Needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The Counter Use _
Page 7 of 19
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.