← Product Code [NIE](/productcode/NIE) · K020089

# MAHURKAR TRIPLE LUMEN CATHETER, 12 FR (K020089)

_The Kendall Company · NIE · Apr 10, 2002 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K020089

## Device Facts

- **Applicant:** The Kendall Company
- **Product Code:** [NIE](/productcode/NIE.md)
- **Decision Date:** Apr 10, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5540
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

MAHURKAR® Triple Lumen Catheter is intended for short-term central venous access for hemodialysis, apheresis, and infusion.

## Device Story

The MAHURKAR® Triple Lumen Catheter is a 12 Fr radiopaque polyurethane catheter designed for short-term central venous access. It features three longitudinal lumens: two large lumens with curved extensions and one smaller medial lumen with straight extension tubing. Lumens are distinguished by color-coded luer-lock adapters on silicone rubber extensions. The device is used by clinicians in clinical settings for hemodialysis, apheresis, and infusion procedures. It is provided as a standalone catheter or in kits/trays containing insertion accessories. The catheter provides venous access, allowing for the delivery or withdrawal of fluids and blood products, thereby facilitating essential renal replacement therapies and infusion treatments.

## Clinical Evidence

Multi-centered, prospective, randomized clinical study of 485 patients (393 evaluable) undergoing acute hemodialysis or apheresis. Primary endpoint: catheter-related bloodstream infection at time of removal. Results: 7.5% infection rate for the triple lumen catheter vs. 8.3% for the dual lumen predicate. Multivariate logistic regression and Chi-square test (P = 0.77) indicated no statistically significant difference in infection rates between the two devices.

## Technological Characteristics

12 Fr radiopaque polyurethane shaft; three longitudinal lumens (two large, one small medial); color-coded luer-lock adapters; silicone rubber extensions; various implantable lengths. Sterile, single-use device.

## Regulatory Identification

A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device. (2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle). (3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).

## Special Controls

*Classification.* (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
*1* ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
*2* ) Proper care and maintenance of the device and device exit site;(
*3* ) Removal of the device;(
*4* ) Anticoagulation;(
*5* ) Management of obstruction and thrombus formation; and(
*6* ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

## Predicate Devices

- MAHURKAR® Triple Lumen Catheter for Infusion
- MAHURKAR® Dual Lumen Catheter

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Page 1/2

## APR 1 0 2002

15 HAMPSHIRE STREET, MANSFIELD, MASSACHUSETTS 02048 • (508) 261-8000

510(k) Premarket Notification MAHURKAR® Triple Lumen Catheter

## Section H - 510(K) Summary

K020089

Date Summary January 9, 2002 Was Prepared:

The Kendall Company Information: Division of Tyco Healthcare Group, LP 15 Hampshire Street Mansfield, MA 02048 Phone: 508-261-8000 Fax: 508-261-8461

Contact: Regina Yeh Senior Specialist, Regulatory Affairs The Kendall Company Division of Tyco Healthcare Group, LP Telephone: 508-261-8404 Fax: 508-261-8461

Device Trade Name:

Submitter's

MAHURKAR® Triple Lumen Catheter for Hemodialysis, Apheresis, and Infusion

Device Common Name:

Catheter, Hemodialysis, Non-implanted

Classification Panel: Gastroenterology

Legally Marketed Devices To Which Substantial Equivalence Is Claimed To: The MAHURKAR® Triple Lumen Catheter is substantially equivalent to MAHURKAR® Triple Lumen Catheter for Infusion with respect to its design, materials, and physical characteristics, and MAHURKAR® Dual Lumen Catheter with respect to its intended use and performance characteristics.

Device Description: The MAHURKAR® Triple Lumen Catheter features three lumens. The catheter has a 12 Fr radiopaque polyurethane catheter shaft with two large lumens and one smaller medial lumen running longitudinally along the length of the catheter shaft. The lumina can be distinguished by the color-coded luer-lock adapters on the clear silicone rubber extensions. The two large lumens have curved extensions, and the smaller medial lumen has straight extension tubing. The catheter is available in various implantable lengths.

MAHURKAR® Triple Lumen Catheter is available in singles, kit or tray configuration which contains the sterile catheter as well as accessory items needed for insertion.

{1}------------------------------------------------

510(k) Premarket Notification MAHURKAR® Triple Lumen Catheter

K620089

# Section H - 510(K) Summary

Intended Use: MAHURKAR® Triple Lumen Catheter is intended for short-term central venous access for hemodialysis, apheresis, and infusion.

Performance Data: Performance data for MAHURKAR® Triple Lumen Catheter were remormation battle of the predicate devices identified in this 510(K) summary. These test results demonstrate that the proposed device is substantially equivalent to the legally marketed predicate devices.

#### Clinical Summary

In a multi-centered, prospective, randomized clinical study, 485 patients underwent an infection evaluation, based on a comparison between the MAHURKAR® Triple Lumen Catheter and the evaluation, baoou on a com Catheter. Patients received either acute hemodialysis or apheresis treatments were randomized to receive either the proposed or the predicate device. Performance data, treatment outcomes and complications were collected during these procedures. The primary outcome of this study is whether there was a catheter related blood stream infection for each patient at the time of catheter removal.

Based on 393 evaluable patients studied, multivariate logistic regression analysis was performed to compare the infection rate of the two devices. The results show that the infection rate was 7.5% for the MAHURKAR® Triple Lumen Catheter and 8.3% for the MAHURKAR® Dual Lumen Catheter. The MAHURKAR® Triple Lumen Catheter infection rate was thus regarded as acceptable as there was no indication that the infection rates of the two catheters were different (P = 0.77 by the Chi-square test).

Regina S. Yeh

Regina S. Yeh Senior Regulatory Affairs Specialist The Kendall Company

1/9/02

Date

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. The bottom half of the circle contains a stylized image of three human profiles facing to the right, with wavy lines below them. The profiles and lines are black, and the background is white.

APR 10 2002

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Regina S. Yeh Sr. Regulatory Affairs Specialist The Kendall Company Division Tyco Healthcare Group, LP 15 Hampshire Street MANSFIELD MA 02048

Re: K020089

Trade/Device Name: MAHURKAR® Triple Lumen Catheter Regulation Number: 21 CFR §876.5540 Regulation Name: Blood access device and accessories Regulatory Class: II Product Code: 78 NIE Dated: January 9, 2002 Received: January 10, 2002

Dear Ms. Yeh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market the device, subject to the general controls provisions of Act. However, you are responsible to determine that the medical devices you use as components in the kit and tray have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit and tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807);

{3}------------------------------------------------

Page 2 - Ms. Regina Yeh

labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality. systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

In addition, we have determined that your device tray contains 1% Lidocaine, which is subject to regulation as a drug.

Our substantially equivalent determination does not apply to the drug component of your device. We recommend you first contact the Center for Drug Evaluation and Research before marketing your device with the drug component. For information on applicable Agency requirements for marketing this drug, we suggest you contact:

Director, Division of Drug Labeling Compliance (HFD-310) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857 (301) 594-0101

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR Part 807.97). Other general information on your responsibilities under the Act, may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/dsma/dsmamain.html

Sincerely yours,

Nancy C Snogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use Statement

#### Device Name:

Mahurkar® Triple Lumen Catheter

### Indications for Use:

Mahurkar® Triple Lumen Catheter is intended for short-term central venous access for hemodialysis, apheresis, and infusion.

Please Do Not Write Below This Line – Continue On Another Page If Needed

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)
OR

Over-The-Counter Use

(Division Sign-Om) 0

Concurrence of CDRH, Office of Device Evaluation (ODE)

---

**Source:** [https://fda-staging.innolitics.com/device/K020089](https://fda-staging.innolitics.com/device/K020089)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).