← Product Code [LDJ](/productcode/LDJ) · K014193

# LIFESIGN DOA 4 (THC/OPI/COC/MET) (K014193)

_Princeton BioMeditech Corp. · LDJ · May 23, 2002 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K014193

## Device Facts

- **Applicant:** Princeton BioMeditech Corp.
- **Product Code:** [LDJ](/productcode/LDJ.md)
- **Decision Date:** May 23, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3870
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology

## Indications for Use

Status Stik™ THC/OPI/COC/MET is designed for the qualitative detection of THC at the cutoff of 50 ng/mL 11-nor-Δ -THC-9-carboxylic acid, opiates at the cutoff of 2000 ng/mL morphine, cocaine at the cutoff of 300 ng/mL benzoylecgonine, and methamphetamine at the cutoff of 1000 ng/mL d-methamphetamine in human urine to assist in screening of drugs of abuse samples. For In vitro Diagnostic Use. LifeSign® Home Drug Test is designed for the qualitative detection of THC at the cutoff of 50 ng/mL 11-nor-A2-THC-9-carboxylic acid, opiates at the cutoff of 2000 ng/mL morphine, cocaine at the cutoff of 300 ng/mL benzoylecgonine, and methamphetamine at the cutoff of 1000 ng/mL d-methamphetamine in human urine to assist in screening of drugs of abuse samples. For In vitro Diagnostic Use. This test is intended for use in the home to assist in preventing drug abuse.

## Device Story

One-step immunochromatographic test; detects THC, opiates, cocaine, and methamphetamine in human urine. Input: urine sample; output: qualitative visual result. Used in clinical, home, or workplace settings; operated by healthcare professionals or lay users. Provides rapid screening to assist in drug abuse detection; aids clinical or personal decision-making regarding substance use.

## Clinical Evidence

Consumer study for LifeSign® Home Drug Test demonstrated >95% overall accuracy. Bench testing showed 100% correlation with predicate devices across >90 specimens per drug.

## Technological Characteristics

Immunochromatographic assay; qualitative detection; lateral flow format (stick, card, or strip). Cutoffs: 50 ng/mL (THC), 2000 ng/mL (opiates), 300 ng/mL (cocaine), 1000 ng/mL (methamphetamine).

## Regulatory Identification

A cannabinoid test system is a device intended to measure any of the cannabinoids, hallucinogenic compounds endogenous to marihuana, in serum, plasma, saliva, and urine. Cannabinoid compounds include delta-9-tetrahydrocannabinol, cannabidiol, cannabinol, and cannabichromene. Measurements obtained by this device are used in the diagnosis and treatment of cannabinoid use or abuse and in monitoring levels of cannabinoids during clinical investigational use.

## Special Controls

*Classification.* Class II (special controls). A cannabinoid test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

## Predicate Devices

- Status DS™ DOA10 ([K990786](/device/K990786.md))
- Status DS™ OPI ([K981771](/device/K981771.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# MAY 2 3 2002

### 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR Part 807.92.

The assigned 510(k) number is: _ K 014193

- 1. Date of Summary: Dec.12, 2001
Princeton BioMeditech Corperation 2. Submitted by: 4242 U.S. Route 1, Monmouth Jct., NJ 08852 PHONE 732-274-1000 FAX 732-274-1010 Contact Person: Jemo Kang, Ph.D., Director

- 3. Device Name
Trade Names: Stick: Status Stik™ THC/OPI/COC/MET, AccuSign®Stik THC/OPI/COC/MET, AccuStik™ DOA4 Card: AccuSign® DOA4, Status DSTM DOA4 Strip: AccuStrip™ DOA4 Common or Usual Name: Immunoassay for detection of THC, opiates, cocaine, and methamphetamine in urine Classification Name: Drugs of Abuse Analysis Systems, Toxicology (91DKE, 91DJG, 91DIO for Enzyme Immunoassay, 91LAG for HPLC)

- 4. Identification of legally marketed device to which claims equivalence: For THC, COC, MET: Status DS™ DOA10 (MET/OPI/COC/THC/PCP/BZO/BAR/ MTD/TCA/AMP), k990786 For OPI: Status DS™ OPI, k981771
- Stik™ THC/OPI/COC/MET is simple one step Status 5. Device Description: immunochromatographic test for the rapid, qualitative, simultaneous detection of THC, opiates, cocaine, and methamphetamine.
- 6. Intended Use: Status Stik™ THC/OPI/COC/MET is designed for the qualitative detection of THC at the cutoff of 50 ng/mL 11-nor-Δ -THC-9-carboxylic acid, opiates at the cutoff of 2000 ng/mL morphine, cocaine at the cutoff of 300 ng/mL benzoylecgonine, and methamphetamine at the cutoff of 1000 ng/mL dmethamphetamine in human urine to assist in screening of drugs of abuse samples. For In vitro Diagnostic Use.
- 7. Substantial Equivalence: Status Stik™ THC/OPI/COC/MET is substantially equivalent to the k990786, Status DS" DOA 10 for THC, COC, MET and k981771, Status DS" OPI for OPI. All three products use the same assay principle and are immunochromatographic assays to detect THC, opiates, cocaine and methamphetamine qualitatively. The detecting cutoff levels are the same. The tests demonstrated 100 % correlation when more than 90 specimens for each drug were compared respectively. The difference is Status Stik™

{1}------------------------------------------------

THC/OPI/COC/MET detects four drugs, while Status DS™ DOA 10 detects seven, other 111c/of resolver in addition to THC, cocaine and methamphetamine, and Status DS™ OPI detects morphine only.

- Conclusion: The device is substantially equivalent to the legally marketed devices k990786, -110 at 110 11 - 10 abbankary operature - 11 - 11 - 11 - 11 - 11 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10

{2}------------------------------------------------

#### 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR Part 807.92.

The assigned 510(k) number is:

2. Date of Summary: Dec.12, 2001

- Princeton BioMeditech Corperation 2. Submitted by: 4242 U.S. Route 1, Monmouth Jct., NJ 08852 PHONE 732-274-1000 732-274-1010 FAX Contact Person: Jemo Kang, Ph.D., Director
- 6. Device Name

Trade Names: LifeSign® Home Drug Test (THC/OPI/COC/MET) Common or Usual Name: Immunoassay for detection of THC, opiates, cocaine, and methamphetamine in urine Classification Name: Drugs of Abuse Analysis Systems, Toxicology (91DKE, 91DJG, 91DIO for Enzyme Immunoassay, 91LAG for HPLC)

- 7. Identification of legally marketed device to which claims equivalence: For THC, COC, MET: Status DS™ DOA10 (MET/OPI/COC/THC/PCP/BZO/BAR/ MTD/TCA/AMP), k990786 For OPI: Status DS™ OPI, k981771
- 8. Device Description: LifeSign® Home Drug Test (THC/OPVCOC/MET) is simple one step immunochromatographic test for the rapid, qualitative, simultaneous detection of THC, opiates, cocaine, and methamphetamine.
- 6. Intended Use: LifeSign® Home Drug Test is designed for the qualitative detection of THC at the cutoff of 50 ng/mL 11-nor-A2-THC-9-carboxylic acid, opiates at the cutoff of 2000 ng/mL morphine, cocaine at the cutoff of 300 ng/mL benzoylecgonine, and methamphetamine at the cutoff of 1000 ng/mL d-methamphetamine in human urine to assist in screening of drugs of abuse samples. For In vitro Diagnostic Use. This test is intended for use in the home to assist in preventing drug abuse.
- 7. Substantial Equivalence: LifeSign® Home Drug Test is substantially equivalent to the k990786, Status DS™ DOA 10 for THC, COC, MET and k981771, Status DS™ OPI for OPI. All three products use the same assay principle and are immunochromatographic assays to detect THC, opiates, cocaine and methamphetamine qualitatively. The detecting cutoff levels are the same. The tests demonstrated 100 % correlation when more than 90 specimens for each drug were compared respectively. The difference is that LifeSign® Home Drug Test detects four drugs, while Status DS™ DOA 10 detects seven other drugs of abuse in addition to THC, cocaine and methamphetamine, and Status DS™ OPI detects morphine only.

{3}------------------------------------------------

8. Consumer Study: In a consumer study, LifeSign® Home Drug Test (THC/OPI/ COC/ MET) showed over 95% overall accuracy.

Conclusion: The device is substantially equivalent to the legally marketed devices k990786, Status DS™ DOA 10 (MET/OPI/COC/THC/PCP/BZO/BAR/ MTD/TCA/AMP) and k981771, Status DS™ OPI. The product is safe in the hands of the lay user.

{4}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of three abstract shapes that resemble birds in flight, stacked on top of each other.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Jemo Kang, Ph.D. Director . Princeton BioMeditech Corporation 4242 U.S. Route 1 Monmouth Junction, NJ 08852-1905

Re: k014193

> Trade/Device Names: LifeSign®Home Drug Test (THC/OPI/COC/MET) Status Stik™ THC/OPI/COC/MET, AccuSign®Stik THC/OPI/COC/MET, AccuStik"1d DOA4.AccuSign® DOA4, Status DSTM DOA4, AccuStrip™ DOA4 Regulation Numbers: 21 CFR 862.3610; 21 CFR 862.3870; 21 CFR 862.3250; 21 CFR 862.3650 Regulation Names: Methamphetamine test system; Cannabinoid test system; Cocaine and cocaine metabolite test system: Opiate test system Regulatory Class: Class II Product Codes: MVO; DJC; LDJ; DIO; DIO; DJG Dated: April 9, 2002 Received: April 10, 2002

Dear Dr. Kang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

#### Page 2

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or mry with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) rms letter notification. The FDA finding of substantial equivalence of your device to a legally premaince nedicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

ely yours,
Steven Autman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

{6}------------------------------------------------

| Page |  |
|------|--|
|------|--|

# 510(k) Number (if known):

Device Name: Status Stik™ THC/OPI/COC/MET, AccuSign® Stik THC/OPI/COC/MET, IC. Status Dilk - Tric. AccuSign® DOA4, Status DS™ DOA4, AccuStrip™ DOA4

Indications For Use:

Immunoassay for the qualitative detection of THC metabolite, opiates, cocaine metabolite, methamphetamine in urine to assist in screening of drugs of abuse. For in vitro Diagnostic Use

The detection cutoff concentrations are as follows:

| THC | 11-nor-Δ9-THC-9-carboxylic acid | 50 ng/ml   |
|-----|---------------------------------|------------|
| OPI | Morphine                        | 2000 ng/ml |
| COC | Benzoylecgonine                 | 300 ng/ml  |
| MET | D-Methamphetamine               | 1000 ng/ml |

Trade Names for each device format

| Stick: | Status Stik <sup>TM</sup> THC/OPI/COC/MET, AccuSign <sup>®</sup> Stik THC/OPI/COC/MET,  AccuStik <sup>TM</sup> DOA4 |
|--------|---------------------------------------------------------------------------------------------------------------------|
| Card:  | AccuSign <sup>®</sup> DOA4, Status DS <sup>TM</sup> DOA4                                                            |
| Strip: | AccuStrip <sup>TM</sup> DOA4                                                                                        |

| Division Sign-Off                      |         |
|----------------------------------------|---------|
| Division of Clinical Laboratory Device |         |
| 510(k) Number                          | K014193 |

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF
NEEDED)# Concurrence of CDRH Office of Device Evaluation (ODE)

Professional Use: Prescription Use: X (Per 21 CFR 801.109)

OR

Over The Counter Use: __

(Optional Format 1-2-96)

{7}------------------------------------------------

2014193 510(k) Number (if known):

Device Name: LifeSign® Home Drug Test (THC/OPI/COC/MET)

Indications For Use:

Immunoassay for the qualitative detection of THC metabolite, opiates, cocaine metabolite, methamphetamine in urine to assist in screening of drugs of abuse samples at home or The detection cutoff concentrations are as work places. For in vitro Diagnostic Use. follows:

| THC | 11-nor-Δ⁹-THC-9-carboxylic acid | 50 ng/ml   |
|-----|---------------------------------|------------|
| OPI | Morphine                        | 2000 ng/ml |
| COC | Benzoylecgonine                 | 300 ng/ml  |
| MET | D-Methamphetamine               | 1000 ng/ml |

Liam Cooper

Division Sign-Off) Ilnical Laboratory Cavices Division o 510(k) Number

# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

#### Concurrence of CDRH Office of Device Evaluation (ODE)

Professional Use:_

Prescription Use: (Per 21 CFR 801.109) OR

Over The Counter Use: _X

(Optional Format 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K014193](https://fda-staging.innolitics.com/device/K014193)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
