← Product Code [EBF](/productcode/EBF) · K014111

# REBILDA SC (K014111)

_Voco GmbH · EBF · Mar 6, 2002 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K014111

## Device Facts

- **Applicant:** Voco GmbH
- **Product Code:** [EBF](/productcode/EBF.md)
- **Decision Date:** Mar 6, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3690
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

Rebilda SC is intended to be used for core build-up of vital and non-vital teeth.

## Device Story

Rebilda SC is a self-curing, radiopaque, flowable composite material used by dental professionals for core build-up procedures in vital and non-vital teeth. The device is supplied as a two-component system. It is applied in conjunction with a light-curing dentin/enamel bonding agent (Solobond Plus). The material is available in two shades: light blue (for preparation margins) and dentine (for use under translucent crowns and bridges). The composite provides mechanical support for subsequent restorative procedures. It is intended for prescription use only.

## Clinical Evidence

Bench testing only. The device conforms to EN 24049/ISO 4049 standards for dental restorative materials. No clinical data provided.

## Technological Characteristics

Two-component, self-curing, flowable composite. Composition: silicate glass, methacrylate polymers, curing catalysts. Radiopaque. Conforms to EN 24049/ISO 4049. No software or electronic components.

## Regulatory Identification

Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.

## Predicate Devices

- Luxacore ([K982201](/device/K982201.md), Foremost Dental Mfg.)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Figure/27 description: The image shows four black shapes arranged horizontally. The first shape is a triangle with a small rectangle cut out of the top. The second shape is a circle with a smaller circle cut out of the center. The third shape is a circle with a cone-shaped section removed. The fourth shape is a circle with a smaller circle cut out of the center.

K014111

(TVOCO GmbH · P.O.B./ Postfach 767 · D-27457 Cuxhaven (Germany)

6 2002 MAR

Anton-Flettner-Straße 1-3 D-27472 Cuxhaven (Germany)

Telefon: (0 47 21) 7 19 - 0 intl.: +49 (47 21) 7 19-0

Telefax: (0 47 21) 7 19 - 1 09 intl.: +49 (47 21) 7 19 - 1 40

Internet: http://www.voco.de e-mail: info@voco.de

Inr Zeichen yr. ref.

ﺎ.

Ihre Nachricht vom did.

Unser Zeichen our ref.

ﻟـــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ

Datum date

## 510(k) SUMMARY

### REBILDA® SC (76 EBF)

- Submitter's Name 1.
- Contact Person For VOCO GmbH 2.
- Date that 510(k) Summary Was Prepared 3.
- Date that 5 ro(k) Sunnaly viao Free (Classification / Common / Proprietary)
Name of the Medical Device (Classification / Common / Proprietany)
- 4. Name of the Medical Device (Olasonoution) Solassinoution is Claimed 5.
- Description of the Device હ.
- Intended Use of the Device 7.
- Thended Use of the Dovice 8.
- Technological Companson Domoon Starformance Studies and Conclusions From 9.
- Preclinical Performance Studies

| 1. SUBMITTER'S NAME                                               |
|-------------------------------------------------------------------|
| VOCO GmbH<br>Anton-Flettner-Str. 1-3<br>27472 Cuxhaven<br>GERMANY |
| Tel: 011-49-47 21 719 0<br>Fax: 011-49 47 21 719 140              |

{1}------------------------------------------------

#### U.S. REGULATORY CONTACT PERSON FOR VOCO GmbH 2.

Evan Dick, Ph.D E.G. Dick & Associates 7527 Westmoreland Avenue St. Louis, MO 63105

(314) 721-0112 Tel: Fax: (314) 721-7591

#### DATE THAT510(k) SUMMARY WAS PREPARED 3.

December 13, 2001

| 4. NAME OF THE MEDICAL DEVICE |                                                 |
|-------------------------------|-------------------------------------------------|
| Classification name           | Material, tooth shade, resin<br>(Dental 76 EBF) |
| Common / usual name           | Self-curing dental restorative material         |
| Proprietary name              | Rebilda SC                                      |

### 5. LEGALLY MARKETED DEVICES TO WHICH SUBSTANTIAL EQUIVALENCE IS CLAIMED

Luxacore (K982201, Foremost Dental Mfg.)

{2}------------------------------------------------

7.

#### DESCRIPTION OF THE DEVICE ਵ

Rebilda SC is a self-curing, highly radiopaque, flowable composite for core buildup. Rebilda SC has excellent mechanical properties. Rebilda SC is available in two shades, light blue and dentine. The light blue shade shows the preparation two onado, light blue tin shade is for use under translucent crowns and bridges. margins, while the achieved by using Rebilda SC in conjunction with the light-curing dentin/enamal bonding agent Solobond Plus.

Rebilda SC corresponds to EN 24049/ISO 4049

### INTENDED USE OF THE DEVICE

Rebilda SC is intended to be used for core build-up of vital and non-vital teeth.

### ECHNOLOGICAL COMPARISON BETWEEN SUBJECT AND PREDICATE DEVICES

Rebilda SC and Luxacore (K982201) are both two-component, self-curing, flowable composites. Both products are primarily composed of silicate glass, methacrylate polymers, and curing catalysts. Both products are substantially equivalent with regard to such physical properties as compressive strength, flexural strength, working time, and hardness.

### SUMMARY OF PRECLINICAL PERFORMANCE STUDIES AND 9. CONCLUSIONS FROM PRECLINICAL PERFORMANCE STUDIES

Rebilda SC is formulated from chemical components that are commonly associated with currently marketed dental composite materials.

The chemistry of Rebilda SC raises no new issues or questions that effect safety, effectiveness, or biocompatibility for dental composite product.

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a bird or a human figure with outstretched arms, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

6 2002 MAR

Voco GmbH C/O Dr. Evan Dick E. G. Dick & Associates 7527 Westmoreland Avenue Saint Louis, Missouri 63105

Re: K014111

Trade/Device Name: Rebilda SC Regulation Number: 872.3690 Regulation Name: Tooth Shade, Resin Material Regulatory Class: II Product Code: EBF Dated: December 14, 2001 Received: December 14, 2001

Dear Dr. Dick:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device mendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 11 1), 1 million in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{4}------------------------------------------------

Page 2 - Dr. Dick

You must comply with all the Act's requirements, including, but not limited to: registration 1 od intiles compry wart 807); labeling (21 CFR Part 801); good manufacturing practice and ilsting (21 CF Part 2017), systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 11 mis letter will and in Jourie Cognit The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your-device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 If you desire up of CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and ulle Act may be obtained its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Timothy A. Ulatowski

Timo Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# INDICATIONS FOR USE STATEMENT

510(k) Number (if known): K014/////'

Rebilda SC Device Name:

Indications For Use:

Rebilda SC is a self-curing, radiopaque, flowable composite intended to be used for core build-up in vital and non-vital teeth.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation

Prescription Use______________________________________________________________________________________________________________________________________________________________

OR

Over-The-Counter______________________________________________________________________________________________________________________________________________________________

**(as per 21 CFR 801.109)**

unne ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------(Division Sign-Off) Division of Dental, Infection Control, - General Hospital Devices · · Number_Kolall Status and States of Concession Comments of

---

**Source:** [https://fda-staging.innolitics.com/device/K014111](https://fda-staging.innolitics.com/device/K014111)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com