SATARI POWDERED LATEX EXAMINATION GLOVES WITH PROTEIN CONTENT LABELING CLAIM (50 MICROGRAMS OR LESS) POLYMER COATED WHIT

K014110 · Siam Sempermed Corp., Ltd. · LYY · Feb 28, 2002 · General Hospital

Device Facts

Record IDK014110
Device NameSATARI POWDERED LATEX EXAMINATION GLOVES WITH PROTEIN CONTENT LABELING CLAIM (50 MICROGRAMS OR LESS) POLYMER COATED WHIT
ApplicantSiam Sempermed Corp., Ltd.
Product CodeLYY · General Hospital
Decision DateFeb 28, 2002
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

This device is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner

Device Story

SATARI® latex patient examination glove; disposable, non-sterile, powdered, single-side polymer-coated. Worn on examiner's hands in clinical settings to provide barrier protection against cross-contamination between patient and examiner. Device features protein content labeling (50 µg or less per gram) and residual powder limits (10 mg/dm² or less). Available in white, pink, and blue.

Clinical Evidence

No clinical data. Substantial equivalence is based on bench testing and comparison of technological characteristics to the predicate device.

Technological Characteristics

Latex patient examination glove; single-side polymer coated; non-sterile; powdered. Protein content ≤ 50 µg/g; residual powder ≤ 10 mg/dm². Class I device (Product Code LYY).

Indications for Use

Indicated for use by medical examiners to prevent cross-contamination between patient and examiner during medical procedures.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # KO14110 510(k) Summary As Required by 21 section 807.92 ( c ) FEB 2 8 2002 1-Submitter Name: Siam Sempermed Corp., Ltd 110 Moo 8 Kanjanavanit Road. Pathong Hatyai 2-Address: Songkhla. Thailand 90230 (++66) 074 291 648 3-Phone: 3-Phone: 4-Fax: 5-Contact Person: Dr Poonsuk Cherdkiatgumchai (Chief Quality Officer) 6-Date summary prepared: December 7th, 2001 7- Official Correspondent: Mansour Consulting (++66) 074 291 650 8- Address: 1308 Morningside Park Dr Alpharetta. GA 30022 USA - (678) 908-8180 9- Phone: 10- Fax: 11- Contact person: Jay Mansour, president 12-Device Trade or Proprietary Name: SATARI® latex patient examination powdered glove single side polymer coated, non sterile, 50 µg or less of total water extractable protein per gram, 10 mg/dm2 or less of residual powder (425) 795-9341 13-Device Common or usual name: Examination glove 14-Device Classification Name: Glove, Patient Examination, Latex 15-Substantial Equivalency is claimed against the following device: Siam Sempermed Latex Patient Examination Glove Polymer powder free, 510k #k981096 (refer to Appendix 2 for FDA website printout. This notification for the SATARI® latex examination glove is of the ABBREVIATED type as per the declaration of conformity on page 4 of this summary #### 11-Description of the Device: SATARI® latex patient examination glove, is a powder glove single side polymer coated, non sterile, 50 ug or less of total water extractable protein per gram, 10 mg/dm2 or less of residual powder ## 12-Intended use of the device: (Indications for use typed on a separate FDA form) This device is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner #### 13-Safety and effectiveness of the device: This device is safe and effective as the predicate device Siam Sempermed Latex Patient Examination Glove polymer, powder-free. Indeed, it is equivalent. This is better expressed in the tabulated comparison (Paragraph 14 below) ### 14-Summary comparing technological characteristics with other predicate device: {1}------------------------------------------------ General comparison result between SATARI® latex examination glove and the predicate device (Siam Sempermed Latex Patient Examination Glove polymer, powder-free) is tabulated below. Technical comparison of specific elements is attached in the main submission | FDA file reference number | 510k 970794 | |--------------------------------------------------|-------------------------------------------------------------------------------------------------------| | Attachments inside notification submission file | REFER TO APPENDIX 2 | | <b>TECHNOLOGICAL CHARACTERISTICS</b> | <i>Comparison result</i><br>REFER TO ADDITIONAL TECHNICAL<br>COMPARATIVE TABLE WITHIN 510K SUBMISSION | | Indications for use | Identical | | Target population | Identical | | Design | Similar | | Materials | Identical | | Performance | Identical | | Sterility | Identical | | Biocompatibility | Identical | | Mechanical safety | Identical | | Chemical safety | Identical | | Anatomical sites | Identical | | Human factors | Identical | | Energy used and/or delivered | Identical (not applicable) | | Compatibility with environment and other devices | Identical | | Where used | Identical | | Standards met | Identical | | Electrical safety | Identical (not applicable) | | Thermal safety | Identical (not applicable) | | Radiation safety | Identical (not applicable) | {2}------------------------------------------------ Image /page/2/Picture/2 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three wavy lines, which are likely meant to represent a symbol or logo for the department. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # FEB 2 8 2002 Siam Sempermed Corporation Limited C/O Mr. Jay Mansour Mansour Consulting 1308 Morningside Park Drive Alpharetta, Georgia 30022 Re: K014110 Trade/Device Name: Satari Powdered Latex Examination Gloves with Protein Content Labeling Claim ( 50 Micrograms or Less) Polymer Coated White, Pink and Blue Regulation Number: 880.6250 Regulation Name: Patient Examination Gloves Regulatory Class: I Product Code: LYY Dated: December 10, 2001 Received: December 14, 2001 Dear Mr. Mansour: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Mansour Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements moun that I Dr man mass statutes and regulations administered by other Federal agencies. or the receingly with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618 . Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and the receinal be sense at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, A. Ulatowski Timt Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ P I | 510(k) Number (if known): | K014110 | |---------------------------|---------------------------------------------------------------------------------------------------------------| | Device Name: | SATARI® LATEX PATIENT EXAMINATION POWDERED GLOVE,<br>SINGLE SIDE POLYMER COATED, NON STERILE, 50mg OR LESS of | | Indications For Use: | TOTAL WATER EXTRACTABLE PROTEIN PER GRAM, 10mg/dm² OR LESS<br>OF RESIDUAL POWDER (White, Pink AND BLUE) | THIS DEVICE IS A DISPOSABLE DEVICE INTENDED EXAMINER'S PURPOSES THAT IS WORN ON THE for MEDICAL PATIENT AND PREVENT CONTAMINATION BETWEEN ಗೆ HAND EXAMINER (PLEASE DO Concurrence of CDRH, Office of Device Evaluation (ODE) *Olin S. Lin* (Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 510(k) Number: K014110 Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use X (Optional Format 1-2-96) AGC C3
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...