# GE LOGIQ 5 (K014097)

_GE Medical Systems · IYN · Jan 10, 2002 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K014097

## Device Facts

- **Applicant:** GE Medical Systems
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Jan 10, 2002
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transesophageal; Transvaginal; and Intraoperative (abdominal, thoracic, vascular and neurological).

## Device Story

Mobile diagnostic ultrasound console; provides digital acquisition, processing, and display of ultrasound signals. User interface includes keyboard, specialized controls, CRT display, and LCD touch screen. Operates via pulsed echo imaging; transforms acoustic inputs into visual diagnostic images. Used in clinical settings by physicians for evaluation of various anatomical regions and fluid flow. Output allows clinicians to visualize internal structures and hemodynamics to inform diagnostic decision-making and patient management.

## Clinical Evidence

Bench testing only. Evaluated for acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal, electrical, and mechanical safety. No clinical data required.

## Technological Characteristics

Mobile console (52cm x 99cm x 135cm); digital acquisition/processing/display. Modes: B, M, PW/CW Doppler, Color/Power Doppler, Harmonic Imaging. Transducers: 3C, 5C, E8C, 10L, 12L, 3S, 7S, 6T, P6D. Conforms to 21 CFR 820, ISO 9001, EN 46001. Safety standards: thermal, electrical, mechanical, acoustic output.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- GE LOGIQ 7 diagnostic ultrasound system ([K010329](/device/K010329.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K014097

### JAN 1 0 2002

Special 510(k) Premarket Notification GE Medical Systems - LOGIQ 5 Ultrasound System December 11, 2001

# Attachment B:

Summary of Safety and Effectiveness Prepared in accordance with 21 CFR Part 807.92(c).

#### GE Medical Systems General Electric Company P.O. Box 414, Milwaukee, WI 53201 Section a): GE Medical Systems 1. Submitter: PO Box 414 Milwaukee, WI 53201 Allen Schuh. Contact Person: Manager, Safety and Regulatory Engineering Telephone: 414-647-4385; Fax: 414-647-4090 Date Prepared: December 11, 2001

- GE LOGIQ 5 Diagnostic Ultrasound System 2. Device Name: Ultrasonic Pulsed Echo Imaging System, 21 CFR 892.1560, 90-IYO
GE LOGIQ 7 diagnostic ultrasound system, 510(k) Numbers K010329 currently in 3. Marketed Device: commercial distribution.

4. Device Description: The GE LOGIQ 5 is a full featured general purpose diagnostic ultrasound system. It consists of a mobile console approximately 52 cm wide, 99 cm deep and 135 cm high that provides digital acquisition, processing and display capability. The user interface includes a computer keyboard, specialized controls, color video CRT display and LCD touch screen. The modification makes a system with numerous high quality features available to diagnostic ultrasound users in a lower market segment.

5. Indications for Use: The device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transesophageal; Transvaginal; and Intraoperative (abdominal, thoracic, vascular and neurological).

Comparison with Predicate Device: The GE LOGIQ 5 is of a comparable type and substantially ్. equivalent to the current GE LOGIQ 7. It has the same technological characteristics, is comparable in key safety and effectiveness features, it utilizes similar design, construction, and materials, and has the same intended uses and basic operating modes as the predicate device.

### Section b):

1. Non-clinical Tests: The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical and mechanical safety, and has been found to conform with applicable medical device safety standards.

2. Clinical Tests: None required.

3. Conclusion: Intended uses and other key features are consistent with traditional clinical practice, FDA guidelines, and established methods of patient examination. The design and development process of the manufacturer conforms with 21 CFR 820, ISO 9001 and EN 46001 quality systems. The device conforms to applicable medical device safety standards and compliance is verified through independent evaluation with ongoing factory surveillance. Diagnostic ultrasound has accumulated a long history of safe and effective performance. Therefore, it is the opinion of GE Medical Systems that the GE LOGIQ 5 Diagnostic Ultrasound is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# JAN 1 0 2002

Mr. Allen Schuh Manager, GE Ultrasound Safety and Regulatory Engineering GE Medical Systems P.O. Box 414 MILWAUKEE WI 53201

Re: K014097

Trade Name: GE LOGIQ 5 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: 90 IYN Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: 90 IYO Dated: December 11, 2001 Received: December 12, 2001

### Dear Mr. Schuh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the GE LOGIQ 5 Diagnostic System, as described in your premarket notification:

Transducer Model Number

\frac{3C}{5C}
\overline{E8C}

{2}------------------------------------------------

10L
12L
3S
7S
6T
P6D

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.

Sincerely yours.

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure(s)

{3}------------------------------------------------

K014097

### Diagnostic Ultrasound Indications for Use Form

### GE LOGIQ 5 Ultrasound System

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application<br>Anatomy/Region of Interest | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
|                                                    | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Fetal / Obstetrics                                 | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |
| Abdominal[1]                                       | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |
| Pediatric                                          | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |
| Small Organ[2]                                     | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |
| Neonatal Cephalic                                  | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |
| Adult Cephalic                                     | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |
| Cardiac[3]                                         | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |
| Peripheral Vascular                                | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |
| Musculo-skeletal Conventional                      | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |
| Musculo-skeletal Superficial                       | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |
| Other[4]                                           | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transesophageal                                    | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |
| Transrectal                                        | P                 | P | P             | P             | P                | P                  | P                | P                 |                     |                |       |
| Transvaginal                                       | P                 | P | P             | P             | P                | P                  | P                | P                 |                     |                |       |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative[5]                                  | P                 | P | P             | P             | P                | P                  | P                | P                 |                     |                |       |
| Intraoperative Neurological                        | P                 | P | P             | P             | P                | P                  | P                | P                 |                     |                |       |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes renal, GYN/Pelvic

[2] Small organ includes breast, testes, thyroid.

[3] Cardiac is Adult and Pediatric.

[4] Other use includes Urology/Prostate

[5] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV).

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nancy C. Broadon

and Rs 510(k) Nuo

{4}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form

## GE LOGIQ 5 with 3C Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Fetal / Obstetrics                                 | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Abdominal[1]                                       | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Pediatric                                          | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Small Organ (specify)                              |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Cardiac                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Peripheral Vascular                                |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Other[4]                                           | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative (specify)                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN;

[4] Other use includes Urology;

[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nancy Brogdon

of Reprod and Radiological Device 510(k) Number

{5}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form

### GE LOGIQ 5 with 5C Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Fetal / Obstetrics                                 | P                 | P | P             |               | P                |                    | P                | P                 |                     |                |       |
| Abdominal[1]                                       | P                 | P | P             |               | P                |                    | P                | P                 |                     |                |       |
| Pediatric                                          | P                 | P | P             |               | P                |                    | P                | P                 |                     |                |       |
| Small Organ (specify)                              |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Cardiac                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Peripheral Vascular                                |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Other[4]                                           | P                 | P | P             |               | P                |                    | P                | P                 |                     |                |       |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative (specify)                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN/Pelvic, Renal and Aortailliac artery;

[4] Other includes urology

[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Nancy C. Brogdon

(Division Sign Division of Reproductive, A and Radiological Devices 510(k) Number .

{6}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form

# GE LOGIQ 5 with E8C Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                     | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |
|-----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|
| Clinical Application<br>Anatomy/ Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse |
| Ophthalmic                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Fetal / Obstetrics                                  | P                 | P | P             |               | P                |                    | P                | P                 |                     |                |
| Abdominal [1]                                       | P                 | P | P             |               | P                |                    | P                | P                 |                     |                |
| Pediatric                                           |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Small Organ (specify)                               |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Neonatal Cephalic                                   |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Adult Cephalic                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Cardiac                                             |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Peripheral Vascular                                 |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Musculo-skeletal Conventional                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Musculo-skeletal Superficial                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Other [4]                                           | P                 | P | P             |               | P                |                    | P                | P                 |                     |                |
| Exam Type, Means of Access                          |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transesophageal                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transrectal                                         | P                 | P | P             |               | P                |                    | P                | P                 |                     |                |
| Transvaginal                                        | P                 | P | P             |               | P                |                    | P                | P                 |                     |                |
| Transuretheral                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative (specify)                            |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative Neurological                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intravascular                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Laparoscopic                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN/Pelvic;

[4] Other use includes Urology/Prostate;

[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nancy C Brogdon

(Division Sic Division of Reproduct live. A and Radiological Devices 510(k) Number

{7}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form

# GE LOGIQ 5 with 10L Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application<br>Anatomy/Region of Interest | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|
|                                                    | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Fetal / Obstetrics                                 | P                 | P | P             |               | P                |                    | P                | P                 |                     |                |
| Abdominal                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Pediatric                                          | P                 | P | P             |               | P                |                    | P                | P                 |                     |                |
| Small Organ[2]                                     | P                 | P | P             |               | P                |                    | P                | P                 |                     |                |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Cardiac                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Peripheral Vascular                                | P                 | P | P             |               | P                |                    | P                | P                 |                     |                |
| Musculo-skeletal Conventional                      | P                 | P | P             |               | P                |                    | P                | P                 |                     |                |
| Musculo-skeletal Superficial                       | P                 | P | P             |               | P                |                    | P                | P                 |                     |                |
| Other (specify)                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative[5]                                  | P                 | P | P             |               | P                |                    | P                | P                 |                     |                |
| Intraoperative Neurological                        | P                 | P | P             |               | P                |                    | P                | P                 |                     |                |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [2] Small organ includes breast, testes, thyroid.

[5] Intraoperative includes abdominal, thoracic, and vascular.

[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nancy C. Brogdon
(Division Sign-Off)

Division of Reproduc and Radiological Devices 510(k) Number _

{8}------------------------------------------------

# Diagnostic Ultrasound Indications for Use Form

# GE LOGIQ 5 with 12L Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Fetal / Obstetrics                                 |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Abdominal                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Pediatric                                          | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |
| Small Organ[2]                                     | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Cardiac                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Peripheral Vascular                                | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |
| Musculo-skeletal Conventional                      | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |
| Musculo-skeletal Superficial                       | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |
| Other (specify)                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative [5] (specify)                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [2] Small organ includes breast, testes, thyroid.

[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Nancy C. Brandon

(Division Sign-Off) Division of Reproduct and Radiological Devices 510(k) Number _

{9}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form

## GE LOGIQ 5 with 3S Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application<br>Anatomy/Region of Interest | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|
|                                                    | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Fetal / Obstetrics                                 | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |
| Abdominal[1]                                       | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |
| Pediatric                                          | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |
| Small Organ (specify)                              |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Adult Cephalic                                     | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |
| Cardiac[3]                                         | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |
| Peripheral Vascular                                |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Other[4]                                           | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative (specify)                           |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN;

[3] Cardiac is Adult and Pediatric;

[4] Other use includes Urology;

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Nancy C. Hargdon

(Division Si Division of Reprodu and Radiological Device 510(k) Number .

{10}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form

## GE LOGIQ 5 with 7S Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |            |            |               |                 |               |                |                  |             |
|----------------------------------------------------|-------------------|---|------------|------------|---------------|-----------------|---------------|----------------|------------------|-------------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes | Harmonic Imaging | Coded Pulse |
| Ophthalmic                                         |                   |   |            |            |               |                 |               |                |                  |             |
| Fetal / Obstetrics                                 | P                 | P | P          | P          | P             | P               | P             | P              | P                |             |
| Abdominal[1]                                       | P                 | P | P          | P          | P             | P               | P             | P              | P                |             |
| Pediatric                                          | P                 | P | P          | P          | P             | P               | P             | P              | P                |             |
| Small Organ (specify)                              |                   |   |            |            |               |                 |               |                |                  |             |
| Neonatal Cephalic                                  | P                 | P | P          | P          | P             | P               | P             | P              | P                |             |
| Adult Cephalic                                     | P                 | P | P          | P          | P             | P               | P             | P              | P                |             |
| Cardiac[3]                                         | P                 | P | P          | P          | P             | P               | P             | P              | P                |             |
| Peripheral Vascular                                |                   |   |            |            |               |                 |               |                |                  |             |
| Musculo-skeletal Conventional                      |                   |   |            |            |               |                 |               |                |                  |             |
| Musculo-skeletal Superficial                       |                   |   |            |            |               |                 |               |                |                  |             |
| Other (specify)                                    |                   |   |            |            |               |                 |               |                |                  |             |
| Exam Type, Means of Access                         |                   |   |            |            |               |                 |               |                |                  |             |
| Transesophageal                                    |                   |   |            |            |               |                 |               |                |                  |             |
| Transrectal                                        |                   |   |            |            |               |                 |               |                |                  |             |
| Transvaginal                                       |                   |   |            |            |               |                 |               |                |                  |             |
| Transuretheral                                     |                   |   |            |            |               |                 |               |                |                  |             |
| Intraoperative (specify)                           |                   |   |            |            |               |                 |               |                |                  |             |
| Intraoperative Neurological                        |                   |   |            |            |               |                 |               |                |                  |             |
| Intravascular                                      |                   |   |            |            |               |                 |               |                |                  |             |
| Laparoscopic                                       |                   |   |            |            |               |                 |               |                |                  |             |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN;

[3] Cardiac is Adult and Pediatric.

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Nancy C. Brogdon
(Division Sign-Off)

Division of Reproduct ive. A and Radiological Device 510(k) Number

{11}------------------------------------------------

# Diagnostic Ultrasound Indications for Use Form

# GE LOGIQ 5 with 6T Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application<br>Anatomy/ Region of Interest | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |
|-----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|
|                                                     | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse |
| Ophthalmic                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Fetal / Obstetrics                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Abdominal                                           |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Pediatric                                           |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Small Organ (specify)                               |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Neonatal Cephalic                                   |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Adult Cephalic                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Cardiac                                             | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |
| Peripheral Vascular                                 |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Musculo-skeletal Conventional                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Musculo-skeletal Superficial                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Other (specify)                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Exam Type, Means of Access                          |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transesophageal                                     | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |
| Transrectal                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transvaginal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transuretheral                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative (specify)                            |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative Neurological                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intravascular                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Laparoscopic                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |

N = new indication; P = previously cleared by FDA; E = added under Appendix E Notes: [*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

> (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Nancy Obroados

(Division Sign-Off Division of Reproductive, A and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________

{12}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form

## GE LOGIQ 5 with P6D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                     | Mode of Operation |   |            |            |               |                 |               |                |                  |             |
|-----------------------------------------------------|-------------------|---|------------|------------|---------------|-----------------|---------------|----------------|------------------|-------------|
| Clinical Application<br>Anatomy/ Region of Interest | B                 | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes | Harmonic Imaging | Coded Pulse |
| Ophthalmic                                          |                   |   |            |            |               |                 |               |                |                  |             |
| Fetal / Obstetrics                                  |                   |   |            |            |               |                 |               |                |                  |             |
| Abdominal                                           |                   |   |            |            |               |                 |               |                |                  |             |
| Pediatric                                           |                   |   |            |            |               |                 |               |                |                  |             |
| Small Organ (specify)                               |                   |   |            |            |               |                 |               |                |                  |             |
| Neonatal Cephalic                                   |                   |   |            |            |               |                 |               |                |                  |             |
| Adult Cephalic                                      |                   |   |            |            |               |                 |               |                |                  |             |
| Cardiac [3]                                         |                   |   |            | P          |               |                 |               |                |                  |             |
| Peripheral Vascular                                 |                   |   |            | P          |               |                 |               |                |                  |             |
| Musculo-skeletal Conventional                       |                   |   |            |            |               |                 |               |                |                  |             |
| Musculo-skeletal Superficial                        |                   |   |            |            |               |                 |               |                |                  |             |
| Other (specify)                                     |                   |   |            |            |               |                 |               |                |                  |             |
| Exam Type, Means of Access                          |                   |   |            |            |               |                 |               |                |                  |             |
| Transesophageal                                     |                   |   |            |            |               |                 |               |                |                  |             |
| Transrectal                                         |                   |   |            |            |               |                 |               |                |                  |             |
| Transvaginal                                        |                   |   |            |            |               |                 |               |                |                  |             |
| Transuretheral                                      |                   |   |            |            |               |                 |               |                |                  |             |
| Intraoperative (specify)                            |                   |   |            |            |               |                 |               |                |                  |             |
| Intraoperative Neurological                         |                   |   |            |            |               |                 |               |                |                  |             |
| Intravascular                                       |                   |   |            |            |               |                 |               |                |                  |             |
| Laparoscopic                                        |                   |   |            |            |               |                 |               |                |                  |             |

N = new indication; P = previously cleared by FDA; E = added under Appendix E Notes: [3] Cardiac is Adult and Pediatric.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nancy C. Croydon

(Division Sigh-Off Division of Reproduc and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________

Prescription User (Per 21 CFR 801.109)

E-11

---

**Source:** [https://fda-staging.innolitics.com/device/K014097](https://fda-staging.innolitics.com/device/K014097)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).