Status Stik™ THC is designed for the qualitative detection of THC metabolite, 11-nor-A -THC-9-carboxylic acid, at the cutoff of 50 ng/mL in human urine to assist in screening of drugs of abuse samples. For In vitro Diagnostic Use. LifeSign®Home Drug Test (THC)is designed for the qualitative detection of THC metabolite. 11-nor-A-THC-9-carboxylic acid. at the cutoff of 50 ng/mL in urine to assist in screening of drugs of abuse samples at home. For In vitro Diagnostic Use.
Device Story
Rapid, one-step immunochromatographic test for qualitative detection of THC metabolite in human urine; utilizes specific antibodies to detect 11-nor-A9-THC-9-carboxylic acid at 50 ng/mL cutoff. Device formats include sticks, cards, and strips. Used by healthcare professionals or consumers (home use) to screen for drug abuse. Visual readout provided by test line presence/absence; assists in clinical or personal screening decisions. Benefits include rapid, point-of-care identification of THC presence.
Clinical Evidence
Bench testing demonstrated 100% correlation with predicate device using 100 specimens (50 positive, 50 negative). Consumer study for LifeSign® Home Drug Test (THC) reported over 95% overall accuracy.
Technological Characteristics
Immunochromatographic assay; lateral flow format (stick, card, or strip). Uses specific antibodies for THC metabolite detection. Qualitative visual readout. No electronic components or external energy source required.
Indications for Use
Indicated for qualitative detection of THC metabolite (11-nor-A9-THC-9-carboxylic acid) in human urine at a 50 ng/mL cutoff. Intended for screening drugs of abuse in professional settings (prescription use) or home settings (OTC use).
Regulatory Classification
Identification
A cannabinoid test system is a device intended to measure any of the cannabinoids, hallucinogenic compounds endogenous to marihuana, in serum, plasma, saliva, and urine. Cannabinoid compounds include delta-9-tetrahydrocannabinol, cannabidiol, cannabinol, and cannabichromene. Measurements obtained by this device are used in the diagnosis and treatment of cannabinoid use or abuse and in monitoring levels of cannabinoids during clinical investigational use.
Special Controls
*Classification.* Class II (special controls). A cannabinoid test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
Status DS™ DOA10 (k990786)
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) Summarv
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR Part 807.92.
The assigned 510(k) number is:
1. Date of Summary: Nov.30, 2001
2. Submitted by: Princeton BioMeditech Corperation 4242 U.S. Route 1, Monmouth Jct., NJ 08852 PHONE 732-274-1000 732-274-1010 FAX Contact Person: Jemo Kang
3. Device Name: Trade Names: Stick Device: Status Stik™ THC, AccuSign® Stik THC, AccuStik™ THC Card Device: AccuSign® THC, Status DS™ THC Strip Test: AccuStrip™ THC Common or Usual Name: Immunoassay for detection of THC in human urine Classification Name: Drugs of Abuse Analysis Systems, Toxicology (91DKE for Enzyme Immunoassay)
- 4. Identification of legally marketed device to which claims equivalence: k990786 Status DS™DOA10 (MET/OPI/COC/THC/PCP/BZO/BAR/MTD/TCA/AMP)
- 5. Device Description: Status Stik™ THC is simple one step immunochromatographic test for the rapid, qualitative detection of THC.
- 6. Intended Use: Status Stik" THC is designed for the qualitative detection of THC metabolite, 11-nor-A -THC-9-carboxylic acid, at the cutoff of 50 ng/mL in human urine to assist in screening of drugs of abuse samples. For In vitro Diagnostic Use.
- 7. Substantial Equivalence: Status Stik™ THC is substantially equivalent to k990786; Status DS "DOA10. Both products use the same assay principle and immunochromatographic assay, and the same antibody to detect THC metabolite qualitatively. The detection cutoff level is the same. The tests demonstrated 100 % correlation when 100 specimens (50 negative and 50 positive) were compared. The difference is that Status Stik" THC detects THC only, while Status DS DOA10 detects nine other drugs of abuse in addition to THC.
- Conclusion: The device is substantially equivalent to the legally marketed device k990786, Status DS ""DOA10 (MET/OPI/COC/THC/PCP/BZO/ BAR/MTD/TCA/AMP).
{1}------------------------------------------------
## 510(k) Summarv
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR Part 807.92.
The assigned 510(k) number is:
1. Date of Summary: Nov.30, 2001
- 2. Submitted by: Princeton BioMeditech Corperation 4242 U.S. Route 1. Monmouth Jct., NJ 08852 PHONE 732-274-1000 FAX 732-274-1010 Contact Person: Jemo Kang
3. Device Name: Trade Names: LifeSign®Home Drug Test (THC) Common or Usual Name: Immunoassay for detection of THC in human urine Classification Name: Drugs of Abuse Analysis Systems, Toxicology (91DKE for Enzyme Immunoassay)
- 4. Identification of legally marketed device to which claims equivalence: k990786 Status DS ""DOA10(MET/OPI/COC/THC/PCP/BZO/BAR/MTD/TCA/AMP)
- LifeSign®Home Drug Test (THC) is simple one 5. Device Description: sten immunochromatographic test for the rapid, qualitative detection of THC.
- 6. Intended Use: LifeSign®Home Drug Test (THC)is designed for the qualitative detection of THC metabolite. 11-nor-A-THC-9-carboxylic acid. at the cutoff of 50 ng/mL in human urine to assist in screening of drugs of abuse samples at home. For In vitro Diagnostic Use.
- 7. Substantial Equivalence: LifeSign®Home Drug Test (THC) is substantially equivalent to k990786; Status DS DOA10. Both products use the same assay principle and immunochromatographic assay, and the same antibody to detect THC metabolite qualitatively. The detection cutoff level is the same. The tests demonstrated 100 % correlation when 100 specimens (50 negative and 50 positive) were compared. The difference is that LifeSign Home Drug Test (THC) detects THC only, while Status DS "DOA10 detects nine other drugs of abuse in addition to THC.
- 8. Consumer Study: In a consumer study, LifeSign® Home Drug Test (THC) showed over 95% overall accuracy.
Conclusion: The device is substantially equivalent to the legally marketed device k990786. Status DS ""DOA10 (MET/OPI/COC/THC/PCP/BZO/ BAR/MTD/TCA/AMP).
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
JUN 1 0 2002
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Jemo Kang, Ph.D. Director Princeton BioMeditech Corporation 4242 U.S. Route 1 Monmouth Junction, NJ 08852-1905
Re: k014066
> Trade/Device Name: LifeSign®Home Drug Test (THC) Status Stik™ THC. AccuSign®Stik THC. AccuStik™ THC. AccuSign® THC, Status DSTM THC, Strip: AccuStrip™ THC Regulation Number: 21 CFR 862.3870 Regulation Name: Cannabinoid test system Regulatory Class: Class II Product Code: LDJ Dated: April 9, 2002 Received: April 10, 2002
Dear Dr. Kang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and ' additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
| T | く<br>A<br>1<br>1 |
|---|------------------|
| | |
510(k) Number (if known): _ | ص ب ب 06
Device Name: __ LifeSign Home Drug Test (THC)
Indications For Use:
Immunoassay for the qualitative detection of THC metabolite, 11-nor-A2-THC-9-carboxylic acid, at the cut-off of 50 ng/mL in urine to assist in screening of drugs of abuse samples at home. For In vitro Diagnostic Use
(Division S-Off)
Division.mical Laboratory Devices
510(k) Number Koi14066
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH Office of Device Evaluation (ODE)
Professional Use:
Prescription Use:_____________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109)
OR
Over The Counter Use: _X
(Optional Format 1-2-96)
{5}------------------------------------------------
f
510(k) Number (if known): k014066
Device Name: Stick: Status Stik™ THC, AccuSign®Stik THC, AccuStik™ THC Card: AccuSign® THC, Status DSTM THC Strip: AccuStrip™ THC
510(k) Number "
Indications For Use:
Immunoassay for the qualitative detection of THC metabolite, 11-nor-A -THC-9-carboxylic acid, at the cut-off of 50 ng/mL in urine to assist in screening of drugs of abuse samples. For In vitro Diagnostic Use
Dan Cooper
(Division Sign-Off)
Division of Clinical Laboratory/Devices
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH Office of Device Evaluation (ODE)
| Professional Use: | |
|-------------------|---|
| Prescription Use: | X |
(Per 21 CFR 801.109)
OR
Over The Counter Use: ________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.