MODIFICATION TO SYNCHRON LX CLINICAL CHEMISTRY SYSTEMS (LX20 AND LX20 PRO)
K014034 · Beckman Coulter, Inc. · JFL · Dec 18, 2001 · Clinical Chemistry
Device Facts
Record ID
K014034
Device Name
MODIFICATION TO SYNCHRON LX CLINICAL CHEMISTRY SYSTEMS (LX20 AND LX20 PRO)
Applicant
Beckman Coulter, Inc.
Product Code
JFL · Clinical Chemistry
Decision Date
Dec 18, 2001
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1160
Device Class
Class 2
Indications for Use
The SYNCHRON LX Clinical Systems are fully automated, computer controlled, clinical chemistry analyzers intended for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries of clinical interest in biological fluids such as serum, plasma, urine, and cerebral spinal fluid (sample type is chemistry dependant).
Device Story
Fully automated, computer-controlled clinical chemistry analyzers; perform in vitro diagnostic testing on serum, plasma, urine, and CSF. System utilizes bar-coded sample/reagent identification; automated sample dilution; delivery to reaction vessels or ISE flowcells. Modified system features reduced ISE sample volume (40 µL vs 62 µL). Hardware includes reagent compartments, cranes, cartridge/modular chemistry sections, and hydropneumatics. Operated by laboratory personnel in clinical settings. Output provides quantitative analyte concentrations; assists clinicians in diagnosing/monitoring metabolic, electrolyte, and acid-base disorders. Benefits include improved diagnostic efficiency and patient management through precise electrolyte monitoring.
Clinical Evidence
Bench testing only. Performance data from validation testing supports equivalency of the modified ISE sample volume (40 µL) to the predicate system.
Technological Characteristics
Fully automated clinical chemistry analyzer; utilizes Ion Selective Electrode (ISE) technology for electrolyte measurement (Na, K, Cl, CO2, Calcium). Features bar-code identification, automated dilution, and modular chemistry flowcells. Software-controlled operation. No specific materials or connectivity standards provided.
Indications for Use
Indicated for quantitative determination of CO2, Calcium, Chloride, Potassium, and Sodium in serum, plasma, urine, or CSF (analyte dependent) to assist in diagnosis/treatment of acid-base balance, parathyroid/bone/renal disease, electrolyte/metabolic disorders (e.g., cystic fibrosis, diabetic acidosis), and conditions involving electrolyte imbalance (e.g., aldosteronism, diabetes insipidus, Addison's disease).
Regulatory Classification
Identification
A bicarbonate/carbon dioxide test system is a device intended to measure bicarbonate/carbon dioxide in plasma, serum, and whole blood. Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance.
{0}------------------------------------------------
# DEC 1 82001
## 510(k) SUMMARY
## 1.0 Submitted By:
K014034
Annette Hellie Regulatory Affairs Manager Beckman Coulter, Inc. 200 S. Kraemer Blvd. W-104 Brea, CA 92822-8000 Telephone: (714) 993-8767 FAX: (714) 961-4123
### 2.0 Date Submitted
December 6, 2001
### 3.0 Device Name(s):
- Proprietary Names 3.1
SYNCHRON LX® Clinical Systems (ISE Module Chemistries)
- Classification Names 3.2
862.1160, pH Rate Measurement, Carbon Dioxide 862.1665 Electrode, Ion Specific, Sodium 862.1600 Electrode, Ion Specific, Potassium 862.1170 Electrode, lon Specific, Chloride 862.1145 Electrode, Ion Specific, Calcium
### 4.0 Legally Marketed Device
The SYNCHRON LX® Systems (ISE Module Chemistries) claim substantial equivalence to the SYNCHRON LX® Systems currently in commercial distribution. FDA 510(k) Numbers K965240 (LX20 System) and K011213 (LX20 PRO System).
### Device Description 5.0
The SYNCHRON LX® Clinical Systems are manufactured by Beckman Coulter, Inc.
The SYNCHRON LX Clinical Systems are fully automated, computer controlled, clinical chemistry analyzers intended for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries of clinical interest in biological fluids such as serum, plasma, urine, and cerebral spinal fluid
{1}------------------------------------------------
(sample type is chemistry dependent). The analyzers operate in conjunction with reagents, calibrators, and controls designed for use with the system. The instruments feature bar code identification of samples and reagents. Thev automatically dilute samples and deliver them to the associated reaction vessel for each system module (cuvette, cup, or ISE flowcell) along with reagents and reaction constituents. The systems analyze up to 41 analytes per sample.
Major hardware components include a reagent compartment, sample and reagent cranes, cartridge chemistry section, modular chemistry section, sample carousel and crane, hydropneumatics, electronics, and power supplies.
The LX20 PRO is differentiated from the standard LX20 system with the following hardware: LPIA (Large Particle Immunoassay) Module and TS-CTS (Thick Stopper-Closed Tube Sampling) Module
The fixed menu ISE flow cell module is contained in the modular chemistry area. The ISE flow cell contains ISE electrodes for the measurement of Sodium (NA), Potassium (K), Chloride (CL), carbon dioxide (CO2), and Calcium (CALC).
### 6.0 Intended Use
The SYNCHRON LX Clinical Systems are fully automated, computer controlled, clinical chemistry analyzers intended for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries of clinical interest in biological fluids such as serum, plasma, urine, and cerebral spinal fluid (sample type is chemistry dependant).
## Comparison to the Predicate 7.0 (Description of the Modification to the Legally Marketed Device)
The modified SYNCHRON LX Systems utilize a new ISE (Ion Selective Electrode) sample volume of 40 µL. The prior sample volume was 62µL. These SYNCHRON LX ISE methods include:
Sodium (NA) Potassium (K) Chloride (CI) Carbon Dioxide (CO2) Calcium (CALC)
### 8.0 Summary of Performance Data
Performance data from validation testing supports equivalency.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Annette Hellie Regulatory Affairs Manager Beckman Coulter, Inc. 200 S. Kraemer Blvd M/S W-104 Box 8000 Brea, CA 92822-8000
Re:
DEC 1 3 2001
k014034 Trade/Device Name: Synchron LX® Systems Carbon Dioxide (CO2) Assay; Synchron LX® Systems Calcium (CALC) Assay; Synchron LX® Systems Chloride (CL) Assay; Synchron LX® Systems Potassium (K) Assay; Synchron LX® Systems Sodium (NA) Assay; Regulation Number: 21 CFR 862.1160; 21 CFR 862.1665; 21 CFR.1600; 21 CFR 862.1170; 21 CFR 862.1145 Regulation Name: Bicarbonate/carbon dioxide test system; Sodium test system; Potassium test system; Chloride test system; Calcium test system Regulatory Class: Class II; Class II; Class II; Class II; Class II Product Code: JFL; JGS; CGM; CGZ; JFP Dated: December 7, 2001 Received: December 7, 2001
Dear Ms. Hellie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
## Page 2
be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594- . Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): KOI4034
Device Name:
# SYNCHRON LX® Systems Carbon Dioxide (CO2) Assay
Indications for Use:
SYNCHRON LX ISE Electrolyte Buffer Reagent, SYNCHRON LX ISE Electrolyte Reference Reagent, and CO2 Alkaline Buffer and Acid Reagent, in conjunction with SYNCHRON LX AQUA CAL 1 and 3, are intended for quantitative determination of carbon dioxide (CO2) in serum or plasma on SYNCHRON LX Systems.
Clinical Significance:
A bicarbonate/carbon dioxide test system is a device intended to measure bicarbonate/carbon dioxide in plasma, serum, and whole blood. Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance.
Ours for Jean Commer
(Division Sign-Off)
Division of Clinical Laboratory Devices
1014034 510(k) Number _
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (per 21 CFR 801.109)
OR
{5}------------------------------------------------
page 2 of 5
510(k) Number (if known):
Device Name:
# SYNCHRON LX® Systems Calcium (CALC) Assay
Indications for Use:
SYNCHRON LX ISE Electrolyte Buffer Reagent and SYNCHRON LX ISE Electrolyte Reference Reagent, in conjunction with SYNCHRON LX AQUA CAL 1 and 2, are intended for quantitative determination of calcium (CALC) in serum, plasma or urine on SYNCHRON LX Systems.
Clinical Significance:
A calcium test system is a device intended to measure the total calcium level in serum. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).
Thomas Plata for Jean Cooper
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number 1014034
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (per 21 CFR 801.199)
OR
{6}------------------------------------------------
Kol4034 510(k) Number (if known):
Device Name:
# SYNCHRON LX® Systems Chloride (CL) Assay
Indications for Use:
SYNCHRON LX ISE Electrolyte Buffer Reagent and SYNCHRON LX ISE Electrolyte Reference Reagent, in conjunction with SYNCHRON LX AQUA CAL 1 and 2, are intended for quantitative determination of chloride (CL) in serum, plasma, urine or cerebrospinal fluid (CSF) on SYNCHRON LX Systems.
Clinical Significance:
A chloride test system is a device intended to measure the level of chloride in plasma, serum, sweat, and urine. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K014034
-
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (per 21 CFR 801.109)
OR
{7}------------------------------------------------
page 4 of 5
K014034 510(k) Number (if known):
Device Name:
## SYNCHRON LX® Systems Potassium (K) Assay
Indications for Use:
SYNCHRON LX ISE Electrolyte Buffer Reagent and SYNCHRON LX ISE Electrolyte Reference Reagent, in conjunction with SYNCHRON LX AQUA CAL 1, 2 and 3, are intended for the quantitative determination of potassium (K) in serum, plasma or urine on the SYNCHRON LX System.
Clinical Significance:
A potassium test system is a device intended to measure potassium in serum, plasma, and urine. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.
Rannel Kust for Jean Cooper
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number k014034
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (per 21 CFR 801.109)
OR
{8}------------------------------------------------
page 5 of 5
510(k) Number (if known):
Device Name:
## SYNCHRON LX® Systems Sodium (NA) Assay
Indications for Use:
SYNCHRON LX ISE Electrolyte Buffer Reagent and SYNCHRON LX ISE Electrolyte Reference Reagent, in conjunction with SYNCHRON LX AQUA CAL 1, 2 and 3, are intended for the quantitative determination of sodium (NA) in serum, plasma or urine on the SYNCHRON LX System.
Clinical Significance:
A sodium test system is a device intended to measure sodium in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.
Moment Thate for Sean Coxen
Division Sign Off
Division of Clinical Laboratory Devices
510(k) Number K014039
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (per 21 CFR 801.109)
OR
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.