RMNIPULSE HOLMIUM LASER SYSTEM/1210; MAX/1210-VHP; JR. 1230-30; N/A 1500-A
K013974 · Trimedyne, Inc. · GEX · Sep 23, 2002 · General, Plastic Surgery
Device Facts
Record ID
K013974
Device Name
RMNIPULSE HOLMIUM LASER SYSTEM/1210; MAX/1210-VHP; JR. 1230-30; N/A 1500-A
Applicant
Trimedyne, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 23, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K013974 · Sep 23, 2002
RMNIPULSE HOLMIUM LASER SYSTEM/1210; MAX/1210-VHP; JR. 1230-30; N/A 1500-A
Trimedyne, Inc.
Published clinical literature
Published clinical literature was used to support the safety and effectiveness of the Holmium:YAG laser system for percutaneous disc decompression/discectomy applications.
Literature review; Clinical outcomes; Disc decompression
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Published clinical literature; Literature review of clinical data
Patients undergoing percutaneous cervical, lumbar, and thoracic disc decompression/discectomy
Not applicable for this study
Safety and effectiveness for disc decompression/discectomy
Indications for Use
The Trimedyne Holmium:YAG Laser System is intended for incision, resection, ablation, vaporization, coagulation, and hemostasis in multispecialty applications. The applications addressed in this premarket notification include percutaneous cervical. lumbar, and thoracic disc decompression/discectomy.
Device Story
Trimedyne Holmium Laser System (Models 1210, 1210-VHP, 1230-30, 1500-A) is a Class IV, pulsed, solid-state Holmium:YAG laser; emits 2.1 µm infrared energy; 350 µs pulsewidth. Used by physicians in clinical settings for incision, resection, ablation, vaporization, coagulation, and hemostasis. Menu-driven controls allow adjustment of pulse repetition rate, output energy, and lasing duration. Laser energy delivered to target tissue via fiber optics; enables minimally invasive percutaneous disc decompression. Clinical benefit includes tissue removal and hemostasis for spinal disc herniation, reducing need for open surgery.
Clinical Evidence
Clinical evidence provided via published literature and animal studies demonstrating safety and effectiveness for percutaneous disc decompression/discectomy applications.
Technological Characteristics
Holmium:YAG laser; 2.1 µm wavelength; 350 µs pulsewidth; max power 100W. Solid-state laser source. Menu-driven control interface for pulse repetition, energy, and duration. Class IV laser device.
Indications for Use
Indicated for percutaneous lumbar, cervical, and thoracic disc decompression/discectomy in patients with uncomplicated ruptured or herniated discs, intractable pain, sensory/motor deficits, and imaging/electromyography findings consistent with disc herniation, following failure of 12 weeks of conservative therapy.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Standard surgical instruments, such as knives and forceps
Submission Summary (Full Text)
{0}------------------------------------------------
SEP 23 2002
K 013974
# 510(k) Summary of Safety and Effectiveness Information Trimedyne Holmium Laser Systems
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| I. Submitter Information: | Trimedyne, Inc.<br>15091 Bake Parkway<br>Irvine, CA 92618 |
|---------------------------|--------------------------------------------------------------------------------------------------|
| Contact person: | Laurie Cartwright<br>Manager, Regulatory Affairs<br>Or<br>Glenn Yeik<br>Executive Vice President |
| Summary Date: | June 24, 2002 |
#### ll. Device Name
Proprietary: Trimedyne Holmium Laser System, including:
- Trimedyne OmniPulse™ Holmium Laser System (Model 1210) �
- Trimedyne OmniPulse MAX™ Holmium Laser System (Model 1210-� VHP)
- Trimedyne OmniPulse Jr. ™ Holmium Laser System (Model 1230-30) �
- Model 1500-A Holmium Laser System �
- Common: Holmium: Yttrium Aluminum Garnet (Holmium:YAG) Laser
Classification: Laser-Powered Instrument
#### III. Predicate Device
Standard surgical instruments, such as knives and forceps.
### IV. Device Description
The Trimedyne Holmium Laser System is a medical grade, Class IV, pulsed, solid state Holmium:YAG laser system designed to deliver pulsed infrared laser enerqy with a wavelength of 2.1 µm and 350 microseconds pulsewidth. Menu-driven control options allow the users to select pulse repetition rate, output energy, and lasing duration.
### V. Intended Use
The Trimedyne Holmium:YAG Laser System is intended for incision, resection, ablation, vaporization, coagulation, and hemostasis in multispecialty applications. The applications addressed in this premarket notification include percutaneous cervical. lumbar, and thoracic disc decompression/discectomy.
{1}------------------------------------------------
### VI. Technological Characteristics
The laser system is a Holmium:YAG laser that emits light at a wavelength of 2.1 um (near infrared) and a pulsewidth of 350 microseconds. The laser has the capability of attaining a maximum output of 100 watts of power.
### VII. Animal Data and Human Clinical Data
Animal data and human clinical data from published literature were included in this premarket notification to demonstrate that the Trimedyne Holmium Laser Systems are safe and effective for the specified indications.
### VIII. Conclusion
The Trimedyne Holmium:YAG Laser System is substantially equivalent to the predicate device described in this premarket notification. Furthermore, the non-clinical and clinical data submitted demonstrated the safety and effectiveness of the device for the proposed applications. Therefore, upon clearance of this submission, the Trimedyne Holmium:YAG Laser System will be marketed for the proposed expanded indications.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with a bird-like figure incorporated into the design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the central emblem. The seal is presented in black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 2 3 2002
Ms. Laurie Cartwright Manager, Regulatory Affairs Trimedyne, Inc. 15091 Bake Parkway Irvine, CA 92618
Re: K013974
Trade/Device Name: Trimedyne Holmium Laser Systems: OmniPulse™ Holmium Laser System, Model 1210 OmniPulse MAX™ Holmium Laser System, Model 1210-VHP OmniPulse Jr.TM Holmium Laser System, Model 1230-30 Model 1500-A Holmium Laser System Regulation Number: 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: June 24, 2002 Received: June 25, 2002
Dear Ms. Cartwright:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Laurie Cartwright
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Sincerely yours,
Mark N Millkerson
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications Page
- This indications page includes only the indications statement affected by this premarket Note: notification submission.
# 510(k) Number (if known): K013974
### Device Name:
Trimedyne Holmium Laser Systems, including:
- . OmniPulse™ Holmium Laser System, Model 1210
- OmniPulse MAX™ Holmium Laser System, Model 1210-VHP ●
- OmniPulse Jr.™ Holmium Laser System, Model 1230-30 .
- Model 1500-A Holmium Laser System .
### Indications for Use:
Incision, excision, resection, ablation, coagulation, hemostasis, and vaporization, with or without an endoscope, in the following indications:
- Percutaneous Lumbar Disc Decompression/Discectomy in soft, cartilaginous, and bony tissue, . including:
- foraminoplasty .
- Percutaneous Cervical Disc Decompression/Discectomy in soft tissue, in patients with: .
- Uncomplicated ruptured or herniated discs .
- Neck pain with radiation down the arm .
- Symptoms and signs of sensory loss, tingling, numbness, muscle weakness, and/or . decreased deep tendon reflexes
- MRI. CT, myelogram, or discogram findings of disc herniation consistent with patient signs . and symptoms
- Positive electromyography and/or nerve conduction studies .
- No improvement after 12 weeks of conservative therapy (i.e., physical therapy, traction, bed . rest, exercises, and medication)
- Percutaneous Thoracic Disc Decompression/Discectomy in soft tissue, in patients with: .
- Uncomplicated ruptured or herniated discs .
- . Thoracic and intercostal intractable pain
- Paresthesias at levels appropriate to the herniated discs visualized on MRI and CT-. myelography
- . MRI, CT, myelogram, or discogram findings of disc herniation consistent with patient signs and symptoms
- . No improvement after 12 weeks of conservative therapy (i.e., physical therapy, traction, bed rest, exercises, and medication)
(PLEASE DO NOT WRITE BELOW THIS LINE –
CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
*(Division Sign-Off)*
Division of General, Restorative
and Neurological Devices
510(k) Number K013974
Prescription Use ***_***
OR Over-the-Counter Use ***_***
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.