K013955 · J. Morita USA, Inc. · EHD · Feb 4, 2002 · Dental
Device Facts
Record ID
K013955
Device Name
VERAVIEWEPOCS, MODEL VE
Applicant
J. Morita USA, Inc.
Product Code
EHD · Dental
Decision Date
Feb 4, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.1800
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Veraviewepocs is an extraoral source X-ray unit that is used for dental radiographic examination and diagnosis of teeth, jaw, and oral structure by exposing an X-ray image receptor to ionizing radiation.
Device Story
Veraviewepocs is an extraoral panoramic X-ray unit for dental imaging. System comprises base, support column, patient frame with chin rest/temple stabilizers, C-arm assembly, and remote control box. Operator (dentist, physician, or technician) positions patient within C-arm; initiates exposure via remote button. C-arm rotates around patient head, emitting X-rays toward a cassette containing intensifying screen and film. Device offers standard, pediatric, maxillary sinus, and TMJ quadruple imaging modes with varying magnifications. High-speed mode reduces radiation exposure time by 50% using high-sensitivity intensifying screens. Automatic exposure mode utilizes a sensor to detect X-rays passing through the patient, adjusting tube voltage and current to ensure consistent image density across varying patient sizes. Output is an X-ray film image used for clinical diagnosis of dental and oral structures.
Clinical Evidence
No clinical data. Substantial equivalence is based on technological characteristics and performance comparisons to the predicate device.
Technological Characteristics
Extraoral X-ray system; 184 kg; 235.5 x 100 x 122 cm. Components: C-arm, support column, patient positioning frame. Energy source: Ionizing radiation. Imaging: X-ray film with intensifying screens (e.g., Fuji Grenex HR-12, Kodak Lanex MEDIUM). Modes: High-speed and normal-speed radiographic modes. Automatic exposure control via sensor-based feedback loop adjusting tube voltage and current. Manual and automatic exposure control.
Indications for Use
Indicated for dental radiographic examination and diagnosis of teeth, jaw, and oral structures in patients requiring dental imaging.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
Predicate Devices
Orthopantomograph OP100
Submission Summary (Full Text)
{0}------------------------------------------------
J. Morita Manufacturing Corporation's Veraviewepocs Panoramic X-ray Unit K
### 510(k) SUMMARY
#### FEB 0 4 2002 J. Morita Manufacturing Corporation's Veraviewepocs Panoramic X-ray Unit
# Name of Device and Name/Address of Sponsor
| Trade or Proprietary Name: | Veraviewepocs |
|----------------------------|-------------------------------------|
| Common Name: | Panoramic X-ray Unit |
| Classification Name: | Extraoral Source X-ray System |
| Product Code: | EHD (Extraoral Source X-ray System) |
J. Morita USA, Inc. 9 Mason Irvine, California USA 92618 Telephone: 949-581-9600 949-581-9688 Facsimile: Contact Person: Mr. Junichi Miyata, President Date Prepared: November 20, 2001
#### Intended Use
The Veraviewepocs is an extraoral source X-ray unit that it used for dental radiographic examination and diagnosis of teeth, jaw, and oral structure by exposing an X-ray image receptor to ionizing radiation.
## Technological Characteristics and Substantial Equivalence
The Veraviewepocs includes a base, a support column, a patient frame mounted to the support column, a C-arm assembly, and a control box. The unit weighs approximately 184 kilograms and has dimensions of 235.5 cm (height), 100 cm (width), and 122 cm (depth). The patient frame and the C-arm assembly are mounted to the support column. The control box, which includes operational lights and an emission button, is typically placed in a separate room from the rest of the system. The patient frame includes patient handles for the patient to hold, a chin rest on which to position the patient's head, and temple stabilizers to position the patient's head during the X-ray session. A lip and nose positioner can optionally be used with the chin rest. The C-arm assembly includes a cassette holder to hold an X-ray cassette on one side of the patient's head, an X-ray head on the other side of the patient's head, a Frankfurt plane light beam, a Frankfurt plane light beam knob, a temple stabilizer knob, an image layer light beam, and a control panel. The Frankfurt plane beam, Frankfurt light beam knob, the temple stabilizer knob, and the image layer light beam are used to position the patient's head for optimal X-ray performance. The control panel also is used to position the patient with respect to the Carm. With the patient positioned within the C-arm, the operator leaves the room and uses the emission button mounted to the control box to expose an X-ray. Pressing the emission button causes the C-arm to rotate around the patient's head while the X-ray
{1}------------------------------------------------
J. Morita Manufacturing Corporation's Veraviewepocs Panoramic X-ray Unit K
head emits X-rays in the direction of the X-ray cassette, which contains an intensifying screen and X-ray film, to expose the X-ray film.
The Veraviewepocs can be used to take standard panoramic images, pediatric panoramic images, maxillary sinus panoramic images, and TMJ quadruple images (i.e., four views of the TMJ, one each of both sides of the mouth with the mouth open and closed). A different button on the control panel is used to set the parameters for that type of image. The standard panoramic images can be magnified at 1.3X or 1.7X. The pediatric panoramic image has a magnification of 1.3X, reduces the X-ray image emission by 10 to 15 percent relative to the standard panoramic image, and reduces the angle of the C-arm's rotation to better coincide with the smaller size of a child's dental arch. The maxillary sinus panoramic image has a magnification of 1.5X and is used to examine the maxillary sinus and facial injuries. The TMJ quadruple image has a magnification of 1.3X and causes the C-arm to rotate twice - once with the patient's mouth open and once with the patient's mouth closed.
The Veraviewepocs operates in either a high speed radiographic mode or a normal speed radiographic mode, and with automatic or manual exposure. Use in the high speed radiographic mode causes the C-arm to rotate approximately twice as fast as many conventional extraoral X-ray source units, such as the Veraviewepocs, while still using conventional exposure rate factors, such as tube voltage and tube current (amperage). The benefit of operating in this mode is that the radiation exposure time is reduced by approximately 50% and the amount of ionizing radiation received by the image receptor, is approximately 50% less than during operation in the normal mode of operation. This mode is made possible by using a conventional high sensitivity intensifying screen, and results in the image density remaining the same as in conventional or normal modes. The intensifying screen fluoresces in response to the X-ray radiation. A high sensitivity intensifying screen merely fluoresces more, relative to a low sensitivity intensifying screen, when exposed to the same level of X-ray radiation. The X-ray film is exposed to the fluorescence from the intensifying screen to impart the image on the film.
One example of a screen and film combination is a combination of a conventional Fuji Photo Film Grenex HR-12 Intensifying Screen with conventional Kodak T-Mat G film. The Veraviewepocs also can be operated in the normal speed mode and use a normal speed radiographic screens, such as the Kodak Lanex MEDIUM.
When used in the automatic exposure mode, the X-rays passing through the patient are detected by a sensor. The device uses the signal received to control the tube voltage and current with respect to the speed of the film's movement. The objective of operating in this mode is to ensure a high quality image with a consistent level of density for patients varying in size from small children to large adults.
The Veraviewepocs is substantially equivalent to Instrumentarium Corporation Imaging Division's Orthopantomograph OP100 ("OP100"). The Veraviewepocs and its predicate device are both panoramic X-ray units. The Veraviewepocs is substantially equivalent to the OP100 because both devices have the same general intended use, and similar technological characteristics and operating principles. Specifically, the intended use of
2
{2}------------------------------------------------
J. Morita Manufacturing Corporation's Veraviewepocs Panoramic X-ray Unit K
the Veraviewepocs is for dental radiographic examination of teeth, jaw, and oral structures by exposing an X-ray image receptor to ionizing radiation. The Veraviewepocs is intended to be operated by physicians, dentists, and X-ray technicians. The intended use of the OP100 is for producing diagnostic X-ray radiographs of dentition, TM-joints, and skull, and has an ortho trans option for producing crosssectional and lateral images of the dental arch using linear tomographic methods. Although the wording of the two intended uses is not word-for-word the same, the general intended use is the same.
The principal differences in the technological characteristics between the Veraviewepocs and the OP100 are: (1) the availability of operation of the Veraviewepocs at a high speed radiograph mode and the resulting differences in exposure time; (2) the availability of the preprogrammed exposure control on the OP100, which is not available on the Veraviewepocs; (3) the differences in nominal magnifications available to the operator: and (4) the unavailability of the TMJ postero anterior mode on the Veraviewepocs. These differences do not raise any new issues of safety or effectiveness. As such, the Veraviewepocs raises no new issues of safety or effectiveness.
40075243.doc
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol featuring three horizontal bars above a shape resembling a human figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 0 4 2002
J. Morita USA % Mr. William D. Hare Fish & Richardson, P.C. 601 Thirteenth Street N.W. WASHINGTON DC 20005 Re: K013955
Trade/Device Name: Veraviewepocs, Model VE Dental Panoramic X-Ray Unit Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: 90 EHD Dated: November 29, 2001 Received: November 30, 2001
Dear Mr. Hare:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Snogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
U.S. Food and Drug Administration - Center for Devices and Radiological Health
Page 1 of 1
510(k) Number (if known): Køl3955
Device Name: Veraviewepocs Panoramic X-ray Unit
Indications for Use:
The Veraviewepocs is an extraoral source X-ray unit that is used for dental radiographic examination and diagnosis of teeth, jaw, and oral structure by exposing an X-ray image receptor to ionizing radiation.
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Optional Format 3-10-98)
Prescription Use
✓
Vinit A. Segmon
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.