SOLID STATE X-RAY IMAGING DEVICES, MODEL IRIS 20 AND IRIS 41 WITH IRIS VIEW
Applicant
Bio-Scan S.A
Product Code
MQB · Radiology
Decision Date
May 21, 2002
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
Models IRIS 20 or IRIS 41 SSXI detector combined with IRIS-View Image Acquisition control, display viewing, and archiving console is indicated for use as a SSXI Portal Imaging Device used in conjunction with Medical Accelerator Radiotherapy Devices as an alternative to Conventional Portal Film.
Device Story
IRIS 20/41 are solid-state X-ray imagers (SSXI) for radiotherapy portal imaging. System consists of a flat-panel detector (amorphous silicon thin-film transistor photodiode array with Gd2O2S(Tb) luminescent screen) and IRIS-View PC-based console. Detector captures X-ray beam (up to 30 MeV) during treatment; converts to digital signal; transmits to console. Console software processes data for display on standard PC monitor; enables image archiving in DICOM format. Used in radiation oncology clinics; operated by radiation therapists/physicists. Provides real-time visualization of treatment field; replaces conventional film; allows windowing/leveling for improved assessment of target organ lesions; enhances clinical decision-making regarding patient positioning and treatment accuracy.
Clinical Evidence
Clinical study conducted at University of Pennsylvania with 30 patients undergoing radiation therapy. Comparison of IRIS SSXI images vs. conventional X-ray film. 7 specialists in radiation oncology evaluated images. Results: 90% of IRIS images rated better than film; 10% rated equivalent; 0% rated poorer. Study supports clinical effectiveness for portal imaging.
Technological Characteristics
Flat panel X-ray detector; Gd2O2S(Tb) luminescent screen; Amorphous Silicon Thin Film Transistor Photodiode Integrating Storage Matrix Array. Dimensions: IRIS 20 (26x53x7cm), IRIS 41 (69.2x59.9x5cm). Connectivity: PC card interface, DICOM III standard. Power: 110/230VAC. Software: Windows-based image processing. Regulatory: Class II, Product Code 90 MQB.
Indications for Use
Indicated for patients undergoing radiation therapy requiring portal imaging. Used in conjunction with medical accelerator radiotherapy devices as an alternative to conventional portal film.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Generic X-Ray film/screen (preamendment device)
Submission Summary (Full Text)
{0}------------------------------------------------
K01389/s/
January 15, 2002
MAY 2 1 2002
**Bio-Scan**
— See life in real-time
# SUMMARY of SAFETY and EFFECTIVENESS
#### I. GENERAL INFORMATION PER 21 CFR 807.87
| Device Classification Name: | Solid State X-Ray Imager (flat panel/digital imager) |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Submission Type: | New Device |
| Regulation Number: | 892.1650 |
| 510(k) Number: | K013897 filed on November 26, 2001 |
| Device Names: | <i>IRIS 20</i> , <i>IRIS 41</i> , <i>IRIS-View</i> |
| Device Product Code | 90 MQB |
| Device Classification: | Class II, per FDA guidance for the submission of<br>510(k) 's for Solid State Imaging Devices, FDA, CDRH,<br>ODE issue date August 6, 1999 |
| Predicate Device: | Generic X-Ray film/screen (a preamendment device) |
| Registration Action: | Device Listing Document Number 103059 for <i>IRIS</i><br>Device Models submitted to FDA, CDRH, Information<br>Processing and Office Automation Branch (HFZ-308)<br>was filed on 07/24/2001 |
| Applicant: | Bio-Scan SA |
| Manufacturer's Address: | Bio-scan SA,<br>Pre Bouview 27,<br>CH-1217 Meyrin/Geneva, Switzerland |
| United States Contact: | Richard M. Sano<br>Consultant to Bio-Scan SA<br>153 Sanford Lane,<br>Stamford, Connecticut 06905<br>Phone: 212-305-1831<br>Page Number: 917-424-0330<br>FAX: 203-323-3329<br>E-mail: rsano50390@aol.com |
| Proposed Labeling: | Operators Manual, Preliminary Product Data Sheets,<br>CE, ISO, and EN Labels are included in the abbreviated<br>510(k) submission |
{1}------------------------------------------------
#### II. DEVICE DESCRIPTION
### IRIS SOLID STATE X-RAY IMAGER RECTANGULAR SSXI FLAT PANEL X-RAY TRANSDUCER
Applicable elements contained in FDA guidance for the submission of 5100x) 's for Solid State Imaging Devices, FDA, CDRH, ODE issue date August 6, 1999
Indications for Use.
Models IRIS 20 or IRIS 41 SSXI detector combined with IRIS-View Image Acquisition control, display viewing, and archiving console is indicated for use as a SSXI Portal Imaging Device used in conjunction with Medical Accelerator Radiotherapy Devices as an alternative to Conventional Portal Film.
### Detectors
The IRIS flat panel X-ray detector is available in two sizes, 20 x 20 cm (IRIS 20) and 41 x 41 cm (IRIS 41), with an Element Matrix of 256 x 256 and 1024 x 1024 respectively. X-rays are converted to light using a Gd202S(Tb) opto-mechanical luminescent screen coupled to an Amorphous Silicon Thin Transistor Photodiode Integrating Storage Matrix Array.
### Viewing Console
The IRIS-View console accepts the digital signals from the detectors through a PC card interface into a standard PC with a Microsoft Windows operating system. Software processing of the data is applied to enhance the visibility of the viewed image seen on a standard PC Color Monitor.
#### REGULATORY REQUIREMENTS III.
Under MDA, the IRIS Device in this abbreviated 510(k) submission is being shown to be substantially equivalent to a legally marketed predicate device, Generic X-ray film/screen (a preamendment device). Even though the preamendment device is exempt from 510(k) requirements, FDA guidance for the submission of 510(k) 's for Solid State Imaging Devices, FDA, CDRH, ODE issue date August 6, 1999, assigns a classification of Class II and a Product Code "90 MQB" for these SSXI devices.
#### IV. NONCLINICAL CONSIDERATIONS
A. Physical Characteristics:
- 1. Overall Dimension: IRIS 20, 26 x 53 x 7 cm. IRIS 41, 69.2 x 59.9 x 5 cm
- 2 Active Area: IRIS 20, 192 x 192 mm. IRIS 41 409.6 x 409.6mm
- 3 Matrix : IRIS 20, 256 x 256 (750 x 750um), IRIS 41, 1024 x 1024 (400 x 400μm)
- 4 Fill Factor: 80%
- 5 Drawings: Cutaway of the SSXI panel, System Components and Connections
- 6 Power: 110vac ±10% 60 cycle or 230vac ±10% 50 cycle
{2}------------------------------------------------
- B. Operational Functions:
- 1. Exposure Characteristics: There is no physical or electrical control of the Medical Accelerator possible through any manual or automatic IRIS or IRIS-View control interactions. The Medical Accelerator which is used with IRIS is not manufactured or marketed by Bio-Scan SA. When used in conjunction with a Medical Accelerator, the IRIS device is as passive as X-ray film, and does not control output of Energy from the Medical Accelerator Device.
Frame rate: Typically single static images, IRIS is not intended for fluoroscopic applications
Dynamic Mode: dependent on field of view, from 10 fps for 10 x 10cm partial area to 4 fps for 41 x 41cm
Integration time: 285mecs to 10,000msec (user selected)
- 2. X-ray Absorber: 25mm thick aluminum absorber
Detection properties: Up to 30 MeV accelerator beam source Measured linear from 50kV to 117kV
- Energy Conversion: Gd2O2S(Tb) opto-mechanical luminescent screen 3.
- 4. Readout Mechanism: Amorphous Silicon Thin Film Transistor Photodiode Integrating Storage Matrix Array, 35us signal charge transfer time, minimum integration time 285msec
- Output Signal: Internal between SSXI Detector and IRIS-View PC Computer ర్ is 16 bit intensity data and serial matrix addressing output signal for images into a network or onto CD-R or CD-RW are DICOM III standard format. (Bitmap format is also available for standard PC graphic presentation archiving)
- C. Functional Characteristics:
- No video Signal output, Output to Monitor is SVGA PC display standard 1.
- DQE: over 65% at 0.3 mm 1 frequency for 10μGy exposure 2.
- 3. SNR: proportional to square root of exposure, SNR vs. Entrance Kerma is measured and graph submitted from 0 to 25 µGy for 50kV to 117kV showing SNR from 25 to 275
- MTF: measured and graph submitted from 0.1 to 1.275mm1 frequency 4. showing MTF from 0.98 to 0.01
- ડ. Aliasing: Matrix doubling algorithm is utilized to eliminate aliasing
- Dynamic Range: Slice Profile of a 45° wedge filter image shows input energy 6. from 18MeV vs. 16 bit digital intensity output. Dose Response is plotted from 0 uGy to 40 uGy for 50 to 117 kV, showing linear response from 0 to 65,535 intensity units.
- 7. Lag Time: Lag time includes 35usec signal charge transfer and reset to zero
- Underscanning: Field of View is user selected from the IRIS-View Console 8.
{3}------------------------------------------------
- 9. X-ray Beam Alignment: Equipment installation function, Detector is mounted in alignment directly behind X-ray film cassette housing on the Medical Linear Accelerator Arm.
- 10. Pixel Defects: up to 5% may be noisy or inefficient and are ignored by a calibration protocol. These defects are corrected by adjacent pixel averaging algorithm
- 11. Device Ready: After one hour after power turn-on, IRIS Device is "Ready" and after an exposure in 280 msec the IRIS Device can acquire another image.
- 12. Latent image: Part of the readout cycle is the reset to zero. After 35 usec, the integrator is reset to zero with in a few milliseconds. A time of 280 msec total is set before a new integration
- D. Exposure Characteristics:
- 1. Dose Requirement: Acceptable SNR or 100 can be achieved at Entrance Air Kerma of 2uGy from 50 to 117kV
- Stability: After one hour of power on, the IRIS system is stable. 2
- 3. Uniformity: Weekly Quality assurance protocol produces a calibration matrix, which corrects for any non uniformity of the gain of individual pixels and also corrects for X-Ray beam non uniformities.
- 4. Frame Rate: This device is not indicated for Fluoroscopic applications.
- 5. Reuse Rate: Serial static images can be acquired from 4 fps to 10 fps depending on field of view selected. The limitation is related to readout rate from the photosensors.
- V. SAFETY
- A. Ready Indicator: An indicator with the word "READY" is seen on the monitor when the system is prepared to accept and process an X-ray input
- B. Passive System: The IRIS detector is passive as is x-ray film, and does not control the x-ray beam or movements of the Therapy accelerator. Bio-Scan does not manufacturer or distribute the Medical Accelerator Therapy devices.
- C. Leakage Currents: The detector is mounted under the treatment table and not in contact with the patient. However, as with UL testing. conformity with all provisions of the directive 93/42/EEC certifies through testing by an authorized body that the system passes ISO 13485/EN46001 Quality system for medical devices.
- D. RF Emissions: The FCC Class A Verification testing by an authorized body was performed to assure that EMC emissions are within levels that do not interfere with other devices.
{4}------------------------------------------------
## VI. CLINICAL CONSIDERATIONS - EFFECTIVENESS
- A. Concurrence Effectiveness Study: A clinical trial under Local IRB approval was sponsored and conducted by University of Pennsylvania Radiation Oncology Department, Division of Medical Physics with 30 patients scheduled to undergo radiation therapy. A table of reader responses was developed comparing the results from x-ray film with simultaneously acquired IRIS SSXI Images. Reading was done by specialists in the radiation treatment of the specific target organ lesions to be irradiated.
- B. Effectiveness Results:
- 1. In no situation was any quality of the Bio-Scan Image rated of poorer quality than the X-ray film image
- 2. In 27 out of 30 comparison images, the Bio-Scan Images were rated better than the X-ray images
- 3. In three out of the total of 30 image comparisons by 7 physicians, the rating was considered equivalent
- C. Effectiveness Conclusion: Physicians with expertise in the Speciality of Radiation Oncology for the target organ, rated the Bio-Scan IRIS results better in 90% of the comparison images or equivalent in 10% of the studies to the practice standard used, X-ray film/screen cassette acquired images.
- D. Sample Images: Copies of all 60 comparative images are printed in the submission and samples of films are provided in the submission. Copies of the IRIS images are provided in CD for review as bitmap images.
- E. Other findings: Written comments by Oncology readers of the comparative images
- 1. "Windowing (on IRIS-View) allows much more dynamic range so that each aspect of the assessment can be individually optimized."
- 2. "Increased dynamic range (of IRIS) helps assessment"
- 3. "Table edge interferes with assessment of both modalities (IRIS and film)."
- "Film assessment is very difficult. Very simple with IRIS." র্ব
- "Study is good, but technique not as good as other films (study # 9010)." న్న
## VII. LABELING
Indication for use, promotional PowerPoint presentation, promotional product data sheets, user manual, installation instructions, and scientific papers (bibliography) are provided in this submission.
{5}------------------------------------------------
## VIII. QUALITY ASSURANCE PROGRAM
The ISO 9001 "Quality Documentation including the CE 0120 "Quality Manual Master File" is available on-site for inspection. Upon request, a copy can be made and submitted for review by FDA ODE.
The declaration of Conformity is included in the submission per directive 93/42 CE:
| 0 | ISO 9001 | Quality system |
|--------------------------|--------------------|----------------------------------------------------------------------------------|
| 0 | ISO 13585/EN 46001 | Quality system, medical devices |
| ロ | EN 1441 | Risk Analysis |
| 0 | EN1041 | Information supplied by the manufacturer with<br>medical devices |
| 0 | CEI 61223-3-1 | Evaluation & routine testing in |
| ם | CEI 61 223-3-3 | medical imaging departments |
| D | EN 980 | Graphical Symbol for use in the labeling of medical<br>devices |
| D | EN 540 | Clinical investigation of medical devices for human<br>subjects |
| 0 | EN 60601 | Electromagnetic Compatibility (FCC Class A) |
| 0 | EN 60950-A4 | Safety on information technology equipment |
| The Testing was done by: | | SGS Yarsley International Certification Services Ltd<br>as Notified Body CE 0120 |
| | | INTERTest Systems, GMBH, international<br>Zulassungen und Testsysteme |
Copies of the Certificates of Compliance are included in the submission.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle. The logo is black and white.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Bio-Scan, SA % Mr. Richard M. Sano, Consultant Columbia University The Neurological Institute Mail Box #48 710 West 168th Street NEW YORK NY 10032
AUG 23 2013
Re: K013897
Trade/Device Name: Solid State X-Ray Imaging Systems (IRIS) Model 20 & 41 Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB and LHN Dated: February 16, 2002 Received: February 20, 2002
Dear Mr. Sano:
This letter corrects our substantially equivalent letter of May 21, 2002.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{7}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
510(k) Number (it known): Kll13897
Devica Namo:_IRIS 20 and 41 with IRIS-View
Models IRIS 20 or IRIS 41 SSX1 detector combined with IRIS-View Indications For Use: image acquisition control, display viewing, and archiving console is indicated for use as a SSXI Portal Imaging Device used in conjunction with Medical Accelerator Radiotherapy Devices as an alternative to Conventional Portal Film.
(PLEASE DO NOT WRITE: BEI OW THIS LINE-CONTINUE (IN ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use J (Por 21 CFR 801. 104)
ાર
Over - The-Counter Use . . ..
Page,
ol
(Optional Formal 1-2-96)
David A. Seguin
(Division Sign-Off) Division of Repre and Radiological Da 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.