The Endoscopic Multi-fire Clip Applier was designed to be used with an endoscope to place clips in the gastrointestinal (GI) tract for the purpose of endoscopic marking, hemostasis of mucosal and submucosal defects < 3 cm, bleeding ulcers, arteries < 2 mm, polyps < 1.5 cm in diameter, and anchoring to affix jejunal feeding tubes to the wall of the small bowel. The device is not intended for repair of gastrointestinal tract lumenal perforations.
Device Story
Endoscopic multi-fire clip applier used with endoscope for GI tract procedures; enables marking, hemostasis, and anchoring. Device comprises control handle, insertion tube, and distal jaws; holds up to 5 clips. Physician operates handle to deploy clips individually onto tissue. Used in clinical settings for mucosal/submucosal defects, bleeding ulcers, arteries, and polyps. Provides mechanical tissue closure/marking; benefits patient by facilitating hemostasis and tube fixation. Sterile, single-use device.
Clinical Evidence
Bench testing only. Tests included scope passage, rotation, clip deliverability, vessel sealing, and minimum/average clip removal force.
Technological Characteristics
Materials: Unalloyed Titanium (ASTM F67-00), Ti-6Al-4V ELI (ASTM F136-98), 18Cr-14Ni-2.5Mo Stainless Steel (ASTM F138-97), Stainless Steel (ASTM F899-95), Ni-Ti Shape Memory Alloy (ASTM F3063-00). Mechanical actuation via control handle and jaws. Sterile, single-use, non-powered.
Indications for Use
Indicated for patients requiring endoscopic marking or hemostasis in the GI tract, specifically for mucosal/submucosal defects < 3 cm, bleeding ulcers, arteries < 2 mm, polyps < 1.5 cm, and anchoring jejunal feeding tubes. Contraindicated for repair of GI tract lumenal perforations.
Regulatory Classification
Identification
A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.
Special Controls
*Classification.* Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
# FEB 0 7 2002
K013889
Page 1 of 3
# Summary of Safety and Effectiveness for the InScope Endoscopic Multi-Fire Clip Applier
submitted by
InScope, LLC 8210 NW 27th Street Miami. Fl 33122 Phone: (305) 716-7000 Facsimile: (305) 716-7020
# Identification of a Legally Marketed Predicate Device
The InScope Endoscopic Multi-Fire Clip Applier is substantially equivalent to the legally marketed Olympus HX-5/6-1 Endoscopic Clipping Device manufactured and marketed by the Olympus Optical Company, Ltd pursuant to 510(k) K963160.
### Device Description
The InScope Endoscopic Multi-Fire Clip Applier is designed for gastrointestinal endoscopic clipping for marking and hemostasis. It consists of a control handle, an insertion tube, jaws at the distal end for gripping the tissue, and up to 5 clips that can be attached to the tissue individually with each actuation of the device handle. The device is provided sterile and is for one time use only.
### Intended Use
The Endoscopic Multi-fire Clip Applier was designed to be used with an endoscope to place clips in the gastrointestinal (GI) tract for the purpose of endoscopic marking, hemostasis of mucosal and submucosal defects < 3 cm, bleeding ulcers, arteries < 2 mm, polyps < 1.5 cm in diameter, and anchoring to affix jejunal feeding tubes to the wall of the small bowel. The device is not intended for repair of gastrointestinal tract lumenal perforations.
### Summary of Technological Characteristics
Technological characteristics of the InScope Endoscopic Multi-Fire Clip Applier were compared with the Olympus HX-5/6-1 Endoscopic Clipping Device manufactured and
{1}------------------------------------------------
K013889
Page 2 of 3
It was de-
marketed by the Olympus Optical Company, Ltd pursuant to 510(k) K963160. I termined that the compared devices are technologically similar.
#### Summary of Performance Data
The InScope Endoscopic Multi-Fire Clip Applier complies with the following standards, practices, and guidances:
- Sterilization of health care products-Radiation sterilization-Substantiation . of 25 kGy for radiation sterilization as a sterilization Does, AAMI TIR 27 -Method VDmax: 1996, Approved 9 May 1996
- Sterilization of health care products-- Requirements for validation and routine . control-Radiation sterilization, ANSI/AAMI/ISO 11137-1994, Approved 25 May 1994 by Association for the Advancement of Medical Instrumentation, Approved 11 July 1994 by American National Standards Institute, Inc.
- ISO 10993-7, Biological Evaluation of Medical Devices-Part 7: Ethylene ● oxide. Second Edition 1996-08
- ISO 10993-1, Biological evaluation of medical devices-Part 1: Guidance on . the selection of tests: International Standards Organization
- ASTM F67-00, Standard Specification for Unalloyed Titanium, for Surgical ● Implant Applications (UNS R50250, UNS R50400, UNS R50550, UNS R50700)
- ASTM F136-98. Specification for Wrought Titanium-6 Aluminum-4 Vanadium . ELI (Extra Low Interstitial) Alloy (UNS R56401) for Surgical Implant Applications:
- ASTM F138-97, Standard Specification for Wrought 18 Chromium-14 Nickel-● 2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants
- ASTM F899-95, Standard Specification for Stainless Steel Billet, Bar, and Wire . for Surgical Instruments
- ASTM F3063-00, Standard Specification for Wrought Nickel-Titanium Shape ● Memory Alloys for Medical Devices and Surgical Implants
The InScope Endoscopic Multi-Fire Clip Applier is substantially equivalent to the legally marketed Olympus HX-5/6-1 Endoscopic Clipping Device manufactured and marketed by the Olympus Optical Company, Ltd pursuant to 510(k) K963160. This has been demonstrated through bench testing and comparative analysis of features. Bench tests included
{2}------------------------------------------------
R013889
Page 3 of
num and average clip
scope passage, rotation, clip deliverability, vessel sealing, and minimum and average clip removal force.
The tissue contact materials used to fabricate the InScope Endoscopic Multi-Fire Clip Appliers have a long history of safe usage in medical devices. Since the InScope Endoscopic Multi-Fire Clip Applier meets the requirements of the stated standards and embodies technological characteristics essentially identical to the predicate device, we believe the device is safe and effective and performs as well as or better than the predicate device. The In-Scope Endoscopic Multi-Fire Clip Applier will be manufactured per specifications and good manufacturing practices that ensure the device is safe and effective for its intended use.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is a stylized design of an eagle with three stripes above it and two wavy lines below it.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 0 7 2002
Mr. Bruce Weber Vice President, Product Assurance Syntheon, LLC 8210 NW 27th Street MIAMI FL 33122
Re: K013889 Trade/Device Name: Endoscopic Multi-Fire Clip Applier Regulation Number: 21 CFR 876.4400 Regulation Name: Hemorrhoidal ligator Regulatory Class: II Product Code: 78 FHN, MND Dated: November 21, 2001 Received: November 23, 2001
Dear Mr. Weber:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for ass black in the May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
8013889
#### Indications for Use
Page_1_of_1
8013889 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: InScope Endoscopic Multi-Fire Clip Applier
Indications for Use:
The Endoscopic Multi-fire Clip Applier was designed to be used with an endoscope to place clips in the gastrointestinal (GI) tract for the purpose of endoscopic marking, hemostasis of mucosal and submucosal defects < 3 cm, bleeding ulcers, arteries < 2 mm, polyps < 1.5 cm in diameter, and anchoring to affix jejunal feeding tubes to the wall of the small bowel. The device is not intended for repair of gastrointestinal tract lumenal perforations.
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | |
|----------------------|--|
| (Per 21 CFR 801.109) | |
OR
| Over-The-Counter Use | |
|--------------------------|--|
| (Optional Format 1-2-96) | |
*Nancy Brogdon*
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices
510(k) Number: K013889
InScope, LLC
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.