K013748 · Candela Corp. · GEX · Feb 11, 2002 · General, Plastic Surgery
Device Facts
Record ID
K013748
Device Name
CANDELA SCLERO LONG PULSE LASER (AKA VBEAM)
Applicant
Candela Corp.
Product Code
GEX · General, Plastic Surgery
Decision Date
Feb 11, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Candela Vbeam Pulse Dye Laser System is indicated for the following uses in: General Surgery: Photocoagulation of benign cutaneous vascular lesions and benign cutaneous lesions. Dermatology/Plastic Surgery: For treatment of benign cutaneous vascular lesions, such as facial and leg telangiectasia, rosacea, port wine stains, hemangiomas, angioma, spider angioma, poikiloderma of Civatte and benign cutaneous lesions, such as warts, scars, striae, and psoriasis. Treatment of periocular wrinkles. Gynecology: Photocoagulation of benign cutaneous lesions and benign vascular lesions in gynecology. Podiatry: Treatment of benign cutaneous lesions, such as warts.
Device Story
Vbeam Pulse Dye Laser is a 595 nm flash lamp-excited pulsed dye laser; uses embedded microprocessor for control. System includes power supply, optical delivery system, handpiece, and Dynamic Cooling Device (DCD). DCD delivers cryogen spray to skin surface to minimize thermal damage and reduce pain. Physician operates device via onboard control panel; laser energy delivered through optical fiber to skin. Safety interlock system protects patients and operators. Used in clinical settings for dermatological, surgical, gynecological, and podiatric procedures. Output affects clinical decision-making by enabling targeted photocoagulation of vascular and cutaneous lesions.
Clinical Evidence
Bench testing only. Device conforms to Laser Performance Standards (21 CFR 1000-1040), UL 2601 Electrical Safety Standard, and EN 60601-1-2.
Technological Characteristics
595 nm flash lamp-excited pulsed dye laser. Includes Dynamic Cooling Device (cryogen spray). Dimensions/form factor: laser enclosure with handpiece and hose. Connectivity: standalone with onboard control panel. Software: embedded microprocessor control. Sterilization: N/A (external use). Energy source: electrical power supply.
Indications for Use
Indicated for patients requiring photocoagulation of benign cutaneous vascular lesions (e.g., telangiectasia, rosacea, port wine stains, hemangiomas, angioma, spider angioma, poikiloderma of Civatte) and benign cutaneous lesions (e.g., warts, scars, striae, psoriasis), and treatment of periocular wrinkles. Applicable to general surgery, dermatology, plastic surgery, gynecology, and podiatry.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Candela Clearbeam Pulse Dye Laser System (K013043)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the Candela logo and the date FEB 11 2002. The Candela logo is a stylized image of a flame. The text "CANDELA" is written in all caps below the logo. The date is written in a simple, sans-serif font.
# 510(k) Summary
## General Information:
Ocuchar Information:
This 510(k) is to provide notification of substantial equivalence for the Candela Vbeam Pulse Dye I ms > Pro vio to provide is substantially equivalent to previously marketed devices intended for the photocoagulation of benign cutaneous lesions, such as warts, scars, striea and psoriasis; benign photocoughanon of oting, such as facial and leg telangiectasia, rosacea, port wine stains, hemangiomas, angioma, spider angioma, poikiloderma of Civatte and in gynecology and treatment of periocular wrinkles.
| Classification: | Class II (21 CFR § 878.4810 Laser Surgical Instrument for use<br>in General and Plastic Surgery and in Dermatology) |
|--------------------|---------------------------------------------------------------------------------------------------------------------|
| Common Name: | Dermatology Laser, Vbeam Pulse Dye Laser System |
| Predicate Devices: | Candela Clearbeam Pulse Dye Laser System (K013043) |
| | Candela SPTL-1e Pulse Dye Laser (K011092) |
| | Candela Long Pulsed Dye Laser (K993671) |
| | Candela SPTL Long Pulse/Tunable Pulse Dye Laser (K943292,<br>K954934) |
| | SLS Biophile Ltd. NLite Pulse Dye Laser (K000811) |
#### Description:
The Vbeam Pulse Dye laser is a 595 nm, flash lamp excited pulsed dye medical laser, controlled by an embedded microprocessor. The Vbeam Pulse Dye Laser System is comprised of a power supply, optical delivery system, software control system and Dynamic Cooling Device. The laser output energy is delivered via an optical fiber to a handpiece, which produces circular or elliptical laser beams onto the skin. The Dynamic Cooling Device provides a short burst of cryogen spray during the laser treatment. The cryogen, which is housed within the laser enclosure, is delivered via a hose to a nozzle located in the handpiece. The Dynamic Cooling Device functions to cool the skin during the laser treatment minimizing thermal damage to skin during laser treatment and reducing pain associated with laser treatment.
The Candela Vbeam Pulse Dye Laser is equipped with a safety interlock system to protect patients and operators. Users of the device make selections from an onboard control panel to regulate operation during treatment.
### Testing:
As a laser product, the Vbeam Pulse Dye Laser is required to conform and will conform to the Laser Performance Standards (21 CFR 1000 - 1040). In addition, the device will conform to the UL 2601 Electrical Safety Standard and with the Harmonized Standard EN 60601-1-2, Part 2 established by and required by the European Community (EC).
### Summary of Substantial Equivalence:
The Candela Vbeam Pulse Dye Laser System has the same intended use, utilizes similar operating principles and matches key design aspects, including spot sizes, pulse durations, wavelength and the same maximum delivered energy as the predicate devices.
On the basis of similarities in methods of assembly, method of operation, and intended uses, Candela believes that its Candela Vbeam Pulse Dye Laser is substantially equivalent to the predicate devices.
8
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with outstretched wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 1 2002
Mr. L. Nelson Manager, Regulatory Affairs Candela Corporation 530 Boston Post Road Wayland, Massachusetts 01778
Re: K013748
Ro15746
Trade/Device Name: Candela Vbeam Pulse Dye Laser System Regulation Number: 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: November 12, 2001 Received: November 13, 2001
Dear Mr. Nelson:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becally be device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrobare) to tegains and ment date of the Medical Device American be of the depend commerce prior to May 20, 1978, the excordance with the provisions of the Federal Food, Drug, de necs that have been require approval of a premarket approval application (PMA). alla Cosmette 71ct (71ct) that as not of corporal controls provisions of the Act. The r ou may, therefore, mains of the Act include requirements for annual registration, listing of general controls provisioning practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is elassined (600 a00 otrols. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r loase of acrised a determination that your device complies with other requirements of the Act that I Dr Has Internand regulations administered by other Federal agencies. You must or any I odolar statutes and regionements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set OF It Furt 6077, adoling (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 - Mr. L. Nelson
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally promated predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
for Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Candela. The logo consists of a stylized flame on the left and the word "CANDELA" on the right. The flame is partially filled with horizontal lines, and the word "CANDELA" is in a simple, sans-serif font.
510(k) Number (if known): _ | _ 013748
Device Name: Candela Vbeam Pulse Dye Laser System
Indications For Use:
The Candela Vbeam Pulse Dye Laser System is indicated for the following uses in:
General Surgery: Photocoagulation of benign cutaneous vascular lesions and benign cutaneous lesions.
For treatment of benign cutaneous vascular lesions, such as facial Dermatology/Plastic Surgery: and leg telangiectasia, rosacea, port wine stains, hemangiomas, angioma, spider angioma, poikiloderma of Civatte and benign cutaneous lesions, such as warts, scars, striae, and psoriasis. Treatment of periocular wrinkles.
Gynecology: Photocoagulation of benign cutaneous lesions and benign vascular lesions in gynecology.
Podiatry: Treatment of benign cutaneous lesions, such as warts.
# (DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Musiam C. Provost
(Division Sign-Off)
Division of General, Restorative
and Neurological Devices
Prescription Use
(Per 21 CFR 801.109)
510(k) Number K013748
(Optional format 1-2-96)
9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.