K013724 · Gambro Renal Products · KPO · Jan 8, 2002 · Gastroenterology, Urology
Device Facts
Record ID
K013724
Device Name
BICART
Applicant
Gambro Renal Products
Product Code
KPO · Gastroenterology, Urology
Decision Date
Jan 8, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5820
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Gambro BiCart® is indicated for use in a bicarbonate hemodialysis for acute and chronic renal failure, or acute intoxication with dialyzable substances.
Device Story
Gambro BiCart® is a non-refillable, disposable cartridge containing dry sodium bicarbonate concentrate; used in conjunction with acid dialysis concentrate and water to prepare bicarbonate hemodialysis solution on-line. Device integrates with adapted, commercially available hemodialysis machines. Used in clinical settings for patients with acute/chronic renal failure or acute intoxication. Cartridge sizes vary to accommodate different treatment durations; functional principle remains identical to predicate devices. Healthcare providers install cartridge into dialysis machine; machine automatically prepares dialysate for treatment. Benefits include simplified on-line preparation of bicarbonate solution for hemodialysis.
Clinical Evidence
No clinical testing performed. Equivalence established via comparison of technological characteristics and chemical composition to predicate devices.
Technological Characteristics
Non-refillable polypropylene cartridge containing dry sodium bicarbonate (USP/European Pharmacopoeia grade). Designed for on-line preparation of 38 mEq/L bicarbonate dialysate. Storage below 40°C. No electronic components or software.
Indications for Use
Indicated for patients with acute or chronic renal failure, or acute intoxication with dialyzable substances requiring bicarbonate hemodialysis.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
{0}------------------------------------------------
Page 1 of 3
GAMBRO Renal Products
510(k) Premarket Notification
# 510(k) Summary
K013724
8 2002 JAN
# BiCart®
Sodium Bicarbonate for Hemodialysis
Contact Information:
**GAMBRO** Renal Products
1845 Mason Avenue Daytona Beach, FL 32117 Phone: 386-274-2811 Fax: 386-274-2833 Contact: Fei Law
Prepared: November 5, 2001
{1}------------------------------------------------
K15124
Page 2 of 3
Image /page/1/Picture/1 description: The image shows the logo for GAMBRO Renal Products. The logo consists of the word "GAMBRO" in bold, followed by the words "Renal Products" in a smaller, less bold font. There is a small circular graphic to the left of the word "GAMBRO."
510(k) Premarket Notification
### 510(k) Summary
| Name of the device: | Gambro BiCart® | | |
|----------------------|------------------------------------------------------------------------------------------------------------------------------|--|--|
| Common Name: | Sodium Bicarbonate for Hemodialysis | | |
| Classification Name: | Dialysate Concentrate for Hemodialysis (Liquid or Powder)<br>per 21 CFR 876.5820. The Product/Classification Code is<br>KPO. | | |
| Predicate Devices: | Gambro BiCart® System<br>K873155 cleared 10/26/1987<br>Gambro, Inc.<br>Manufactured by Gambro Lundia AB | | |
| | Gambro BiCart® 720g<br>K940601 cleared 03/22/1994<br>C.G.H Medical, Inc.<br>Manufactured by Gambro Lundia AB | | |
#### Device Description:
Gambro BiCart® Sodium Bicarbonate for hemodialysis is a dry concentrate used to prepare bicarbonate concentrate solution for use in conventional hemodialysis.
The Gambro BiCart® is a non-refillable cartridge containing sodium bicarbonate, which enables on-line preparation of bicarbonate hemodialysis solution. It is used in conjunction with appropriate acid dialysis concentrate and water meeting Association for the Advancement of Medical Instrumentation (AAMI) guidelines. The resulting dialysate is used in conventional, commercially available hemodialysis machines/monitors which have been adapted to receive the cartridges.
#### Indications for Use:
Gambro BiCart® is indicated for use in a bicarbonate hemodialysis for acute and chronic renal failure, or acute intoxication with dialyzable substances.
{2}------------------------------------------------
MBRO Renal Products
510(k) Premarket Notification
## 510(k) Summary
#### Comparison to Predicate Devices:
The Gambro BiCart® Sodium Bicarbonate for Hemodialysis cartridges are equivalent to the Gambro BiCart® System (includes 650g BiCart®) and the Gambro BiCart® 720g which have been previously approved by the FDA under Premarket Notifications K873155 and K940601. This Premarket Notification is intended to add cartridge sizes to existing product offerings. The only functional difference between the different sizes is the quantity of sodium bicarbonate contained in the cartridge. The varying quantities of sodium bicarbonate provide for varying lengths of treatment. The resulting bicarbonate concentrations for all cartridge sizes are identical, and the resulting solution is used in exactly the same way.
| | Gambro BiCart™<br>System<br>K873155<br>cleared 10/26/1987 | Gambro BiCart™<br>120g<br>K940601<br>cleared 03/22/1994 | Gambro BiCart®<br>(various sizes,<br>including 650, 720,<br>and 1150g) |
|------------------------------------|---------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Manufacturer | Gambro Lundia AB | Gambro Lundia AB | Gambro Lundia AB |
| Indication for Use | Bicarbonate<br>hemodialysis for<br>acute and chronic<br>treatment of renal<br>failure | Bicarbonate<br>hemodialysis for<br>acute and chronic<br>treatment of renal<br>failure | To be used in<br>bicarbonate<br>hemodialysis<br>treatment for patients<br>suffering from acute |
| Indication included<br>in labeling | In cases of poisoning<br>by substances which<br>can be removed by<br>hemodialysis | In cases of poisoning<br>by substances which<br>can be removed by<br>hemodialysis | renal failure, chronic<br>renal failure, or acute<br>intoxication with<br>dialyzable substances |
| Disposable? | Yes | Yes | Yes |
| Sodium Bicarbonate<br>Grade | USP & European<br>Pharmacopoeia | USP & European<br>Pharmacopoeia | USP & European<br>Pharmacopoeia |
| Sodium Bicarbonate<br>Weight Spec. | Target (± 50g) | Target (± 50g) | Target (+30g) |
| Bicarbonate<br>Concentration | 38 mEq/L | 38 mEq/L | 38 mEq/L |
| Storage Conditions | Store below +40 °C<br>(+104 °F) | Store below +40 °C<br>(+104 °F) | Store below +40 °C<br>(+104 °F) |
| Housing Material | Polypropylene | Polypropylene | Polypropylene |
Comparing the proposed device to the predicate devices, its indication for use, technological characteristics, and chemical composition are identical. We therefore consider this device to be substantially equivalent to existing products in commercial distribution in the United States.
#### Performance Data
No clinical testing was performed.
No testing was necessary to determine equivalence to predicate devices.
{3}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville . MD 20850
8 2002 JAN
Ms. Fei Law Quality Assurance Manager Gambro Renal Products 1845 Mason Avenue . DAYTONA BEACH FL 32117 Re: K013724
Trade/Device Name: Gambro BiCart® Regulation Number: 21 CFR §876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: 78 KPO Dated: November 5, 2001
Received: November 9, 2001
Dear Ms. Law:
We have reviewed your Section 510(k) premarket notification of intent to market the device we forenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
of I Page 1
K013724 510(k) Number (if known):
Gambro BiCart® Device Name:
Indications For Use:
Gambro BiCart® is indicated for use in bicarbonate hemodialysis for acute and chronic renal failure, or acute intoxication with dialyzable substances.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C Bragtim
(Division Sign-Off)
Division of Reproductive, Abidemás,
and Radiological Devices
510(k) Number. Kor
(Optional Format 3-10-98)
*Prescription Use*
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.