HILL-ROM PRIMAVIEW MONITOR SYSTEM

K013709 · Hill-Rom, Inc. · FET · Nov 16, 2001 · Gastroenterology, Urology

Device Facts

Record IDK013709
Device NameHILL-ROM PRIMAVIEW MONITOR SYSTEM
ApplicantHill-Rom, Inc.
Product CodeFET · Gastroenterology, Urology
Decision DateNov 16, 2001
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The Hill-Rom PrimaView Mouitor System is indicated for use in providing the clinician with a color video display of images that are generated from endoscopic, laproscopic or similar equipment (not supplied by Hill-Rom) during surgical procedures. This system may also be used to display radiographic images for reference only.

Device Story

The Hill-Rom PrimaView Monitoring System is a fixed, surgical-mounted LCD flat-screen display used in operating rooms or medical facilities. It receives analog or digital video signals from third-party endoscopic or laparoscopic camera systems. The device routes these signals to the display for real-time visualization by clinicians during surgical procedures. It also supports the display of radiographic images for reference purposes only. The system functions as an accessory to endoscopic/laparoscopic equipment, providing a visual interface for the surgeon to monitor the procedure.

Clinical Evidence

Bench testing only. Compliance with IEC, EN, CISPR, UL, and CSA standards (including IEC 601-1-2, EN 60601-1/-2, EN 61000-4-3/4/5, CISPR 11, UL 2601-1, and CSA C22.2 601.1-M90) confirmed through design testing.

Technological Characteristics

LCD flat-screen display; fixed surgical-mount form factor; accepts analog and digital video signals; connectivity via interconnecting hardware and circuitry. Standards: IEC 601-1-2, EN 60601-1, EN 60601-2, EN 61000-4-3, EN 61000-4-4, EN 61000-4-5, CISPR 11, UL 2601-1, CSA C22.2 601.1-M90.

Indications for Use

Indicated for clinicians requiring color video display of images generated from endoscopic, laparoscopic, or similar surgical equipment; also for reference display of radiographic images.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 2 7 2015 .............................................................................................................................................................................. Hill-Rom, Inc. Ms. Chantel Carson Underwriters Laboratories, Inc. 333 Pfingsten Road Northbrook, IL 60062 > K013709 Trade/Device Name: Hill-Rom Prima View Monitoring System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FET Dated (Date on orig SE ltr): November 6, 2001 Received (Date on orig SE Itr): November 8, 2001 Dear Ms. Carson, Re: This letter corrects our substantially equivalent letter of November 16, 2001. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be {1}------------------------------------------------ Page 2 - found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, ## Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## "Indications for Use Statement" NOV 1 6 2001 510(k) Number (if known): _ KO13709 Device Name: Hill-Rom PrimaView Monitoring System Indications For Use: The Hill-Rom PrimaView Mouitor System is indicated for use in providing the clinician with a color video display of images that are generated from endoscopic, laproscopic or similar equipment (not supplied by Hill-Rom) during surgical procedures. This system may also be used to display radiographic images for reference only. (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF REQUIRED) Concurrence of CDRH, Office of Device Evaluation (ODE) Or Prescription Use: (21 CFR 801.109) Over-The Counter Use: ________________________________________________________________________________________________________________________________________________________ Susan Walk (Division Si -n-Off) (Division Sisti-chil) Division of --- reral, Restorative Division of Devices Division of " 16.15." and Neurok 21. al Devices Neuro 510(k) Number K013709 510(k)5a {3}------------------------------------------------ K013 709 Image /page/3/Picture/1 description: The image shows the logo for Hill-Rom, a Hillenbrand Industry. The text "Hill-Rom" is in a bold, sans-serif font, with a slightly distressed or textured appearance. Below the main logo, in a smaller font, is the text "A HILLENBRAND INDUSTRY", indicating that Hill-Rom is part of the Hillenbrand group of companies. 510(k) Summary Date: September 12, 2001 NOV 1 6 2001 - 1. Establishment Information: | Submitter: | Hill-Rom<br>A Hillenbrand Industry<br>1069 State Route 46 East<br>Batesville, IN 47006-9167 | |------------------|---------------------------------------------------------------------------------------------| | Registration #: | 1824206 (Owner/Operator)<br>1836145 (Manufacturing Site) | | Contact Name: | Timothy M. Davis | | Contact Phone #: | 812-931-3825 or 1-800-445-2114 (x-13825) | | Contact Fax #: | 812-934-1675 | - 2. General Device Information: | Common Name: | Video Display (No Audio) | |------------------------|----------------------------------------------------------------------------------------------------| | Trade Name: | Hill-Rom PrimaView Monitor System | | Classification Name: | Endoscope and accessories | | Classification Number: | 876.1500 (accessory item - monitor) | | Device Classification: | Class II | | Performance Std: | Performance Standards for this device have not been established under section 514 of the FD&C Act. | - 3. Substantial Equivalence: Hill-Rom PrimaView Monitor System is substantially equivalent in the intended use (display video images) to the Sony Trinitron Color Video Monitor PVM-1343MD cleared under 510(k) number K885042. - 4. Technology Differences: The primary difference between the predicate device and the Hill-Rom PrimaView Monitor System is that the display on the Sony Trinitron Monitor is CRT based and the PrimaView uses LCD flat screen technology. The Sony system also is a portable device (stand-alone), whereas the PrimaView is a fixed surgical mounted display that uses interconnecting hardware and circuitry to route the video signal. However both devices are designed to accept various types of signals (analog and digital) from Endoscopic, Laproscopic or other similar sources for display. - 5. Device Description: The Hill-Rom PrimaView Monitor System is considered and accessory item to an endoscopic or laproscopic system (not supplied by Hill-Rom) in that it provides a means to view images generated by an endoscopic or laparoscopic camera and their associated components. The PrimaView flat screen display is mounted in the operating room or other area of a medical facility where endoscopic or similar procedures are done. Output video signals from the camera and processing unit are fed into the appropriate input connector of the PrimaView System. The signal 000. 1069 STATE ROUTE 46E, BATESVILLE, INDUNA 47006-9167 12-934-7777 Telefox International: 112-934-1191 Telefox Domestic: 812-93-4189 {4}------------------------------------------------ is then routed to the display for presentation to the clinician. The system can also be used to display reference quality (not for diagnosis) radiographic images - 6. Intended Use: The Hill-Rom PrimaView Monitor System is indicated for use in providing the clinician with a color video display of images that are generated from endoscopic, laparoscopic or similar equipment (not supplied by Hill-Rom) during surgical procedures. This system may also be used to display radiographic images for reference only. - 7. Performance Data: Compliance to all or applicable parts of IEC, EN, CISPR, UL and CSA standards (IEC 601-1-2, EN 60601-1 and -2, EN 61000-4-3,4,5, CISPR 11, UL 2601-1 and CSA C22.2 601.1-M90) will be confirmed through design testing and documented. Documentation will be retained in the design and development file as part of the Design History File (DHF). - 8. Conclusion: The Hill-Rom PrimaView Monitor System is designed to be safe and effective for its intended use. Verification and Validation Testing will be completed prior to commercial distribution.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...