Reference Check is recommended for use in controlling the accuracy of quantitative hemostasis assays in the normal range.
Device Story
Reference Check is a normal control plasma derived from pooled, citrated human plasma collected from 20 or more screened donors. The plasma is buffered with 0.01M HEPES, aliquoted, and frozen. It is used in clinical laboratory settings to verify the accuracy of quantitative hemostasis assays. Healthcare professionals use the device to establish reference ranges for various coagulation and fibrinolytic parameters, ensuring the reliability of patient test results. By providing a standardized control, it assists in the quality control of diagnostic testing for coagulation disorders.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Pooled citrated human plasma; buffered with 0.01M HEPES; frozen format; 1 mL per vial. Functions as an in vitro diagnostic control for hemostasis assays.
Indications for Use
Indicated for use as a control plasma to monitor the accuracy of quantitative hemostasis assays within the normal range for patients undergoing coagulation testing.
Regulatory Classification
Identification
A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.
Special Controls
*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
{0}------------------------------------------------
DEC 1 0 2001
# 510(K) Summary Reference Check
### This summary of 510(k) safety and effectiveness information is being submitted in accordance
11 accordance of 510(k) safety and effective a 100 accord and CCCR 007 02 ro(K) Salety and Chock. Choos. Choose of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: _______________________________________________________________________________________________________________________________________________ ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
| Submitters Name & Address: | Precision BioLogic Incorporated<br>900 Windmill Road, Suite 100<br>Dartmouth, Nova Scotia B3B 1P7<br>Canada |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Name: | Stephen L. Duff – Director of New Business Development<br>Phone: 902-468-6422 ext. 224<br>Fax: 902-468-6421<br>Email: sduff@precisionbiologic.com |
| Preparation Date: | November 7, 2001 |
| Device Name & Classification: | Reference Check<br>Common Name: Normal Control Plasma<br>Classification Name: Plasma, Control, Normal<br>Regulatory Class II, 81 GIZ |
| Predicate Device: | Control Plasma N (K001256)<br>Dade Behring Inc.<br>Newark, Delaware 19714<br>USA |
| Device Description: | Reference Check is citrated human plasma collected from 20 or<br>more carefully screened normal donors. Plasmas are pooled,<br>buffered using 0.01M HEPES buffer, aliquoted and rapidly frozen. |
| Device Intended Use: | Reference Check is recommended for use in controlling the<br>accuracy of quantitative hemostasis assays in the normal range. |
{1}------------------------------------------------
### Comparison to Predicate Device:
| Parameter | Reference Check | Control Plasma N (K001256) |
|--------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Assayed Control - Normal | Assayed Control - Normal |
| Analytes | Fibrinogen, Coagulation factors II, V,<br>VII, VIII, vWF: Antigen, vWF: Ristocetin<br>Cofactor, IX, X, XI, XII, XIII,<br>prekallikrein, Protein C (antigen and<br>activity), Protein S (activity, total and<br>free), antithrombin (antigen and<br>activity), Alpha-2 antiplasmin,<br>plasminogen | PT, aPTT, TT, Batroxobin time,<br>Fibrinogen, Coagulation factors, II, V,<br>VII, VIII, IX, X, XI, XII, XIII*, VWF<br>Ristocetin Cofactor, Antithrombin III,<br>Protein C, Protein S*, Alpha-2<br>antiplasmin, C1-inhibitor*,<br>Plasminogen, Lupus anticoagulants<br>* not available in the US |
| Matrix | Reagent from human plasma (citrated<br>human plasma collected from 20 or<br>more carefully screened donors that is<br>pooled and buffered using HEPES<br>buffer). | Reagent from human plasma<br>(obtained from pooled plasma<br>collected from selected healthy blood<br>donors and stabilized with HEPES<br>buffer solution). |
| Format | Frozen | Lyophilized |
| Volume | 1 mL per vial | 1 mL per vial |
#### Comments on Substantial Equivalence:
It is the opinion of Precision BioLogic Inc. that Reference Check is substantially equivalent to Control Plasma N (K001256), manufactured by Dade Behring Marburg GmbH, and currently marketed in the United States by Dade Behring Inc. This opinion is based on the following:
- . Both products consist of pooled citrated normal human plasma
- Both products are assayed and provide reference ranges for fibrinolytic and . coagulation parameters
- Both products are intended for use in the accuracy and control of assays for . coagulation and fibrinolysis (i.e. hemostasis) in the normal range
#### Conclusion:
Reference Check is substantially equivalent to Control Plasma N.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features an abstract image of an eagle with three curved lines representing its wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. Stephen Duff Director of New Business Development Precision Biologic 900 Windmill Road, Suite 100 Dartmouth, N.S., Canada
DEC 1 0 2001
Re: k013708
Trade/Device Name: Reference Check Regulation Number: 21 CFR 864.5425 Regulation Name: Multipurpose system for in vitro coagulation studies Regulatory Class: Class II Product Code: GIZ Dated: November 7, 2001 Received: November 8, 2001
Dear Mr. Duff:
We have reviewed your Section 510(k) premarket notification of intent to market the device w o no re ro rowe and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use Statement
510(k) Number:
K013708
Device Name:
Reference Check
#### Indications for Use:
Reference Check is recommended for use in controlling the accuracy of quantitative hemostasis assays Reference Check is recommended for use in controlling the following parameters in the normal range:
- Fibrinogen .
- Factor II .
- Factor V .
- Factor VII .
- Factor VIII:C �
- vWF: antigen ●
- vWF: Ristocetin Cofactor ●
- Factor IX ●
- Factor X .
- Factor XI .
- Factor XII .
- Factor XIII ●
- Prekallikrein .
- Protein C: antigen ●
- Protein C: activity ●
- Protein S: activity .
- Protein S: total .
- Protein S: free .
- Antithrombin: activity ●
- Antithrombin: antigen .
- Alpha-2-antiplasmin .
- Plasminogen .
Josephine Buettner
(Division Syln-Off) (Division of Clinical Laboratory Deving of 510(k) Number ________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.