K013627 · Medison America, Inc. · IYN · Nov 16, 2001 · Radiology
Device Facts
Record ID
K013627
Device Name
SONOACE SA-8000 DIAGNOSTIC ULTRASOUND SYSTEM
Applicant
Medison America, Inc.
Product Code
IYN · Radiology
Decision Date
Nov 16, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Fetal (includes infertility monitoring of follicle development), Abdominal, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Cardiac (Adult, Pediatric), Trans-Rectal, Trans-Vaginal, Peripheral-Vascular, Muscular-Skeletal (conventional, superficial), Intra-Operative (Neurological, Abdominal, Peripheral vascular). Typical examinations performed using the system are: General abdominal and pelvic studies including organ surveys, assessment, and retroperitoneal cavity studies. Study of small parts including breasts, shoulders, thyroid, and the abdominal wall. Pediatric scans of organs and bony structures. Peripheral vascular applications including carotid arteries, legs, arms, feet, and penile artery. Monitoring procedures for infertility studies (other than in vitro fertilization). First, second and third trimester pregnancy studies. Prostate, prostate biopsy guidance, and rectal wall studies. Neonatal head studies. Trans-cranial studies of middle cerebral arteries, internal carotid artery, and vertebral arteries. Cardiac studies in adults and children. Biopsy guidance for tissue or fluid sampling. Conventional podiatry scans.
Device Story
SA8000 is a mobile, software-controlled diagnostic ultrasound system. It acquires ultrasound data via piezoelectric transducers (linear, convex, phased array, static) and displays B-mode, M-mode, Color-Flow, Continuous Wave (CW), Pulsed (PW), and Power Doppler modes. The system uses digital beamforming and digital dynamic receive focusing to produce high-resolution images. Operators (physicians/technicians) use a control panel to adjust parameters like transmit focusing, filtering, and image enhancement. The system provides real-time acoustic output (mechanical/thermal indices) and supports biopsy guidance, cine loops (256 frames), and measurement/analysis packages (e.g., fetal biometry, cardiac gradients, vascular indices). Output is displayed on a monitor; images can be filed or exported via DICOM 3.0 for remote viewing. The device assists clinicians in anatomical measurement and diagnosis, potentially improving diagnostic accuracy and patient management through non-invasive imaging.
Clinical Evidence
Bench testing only. The device complies with recognized safety and performance standards, including EN 60601-1, UL 2601-1, and IEC 1157 for acoustic output declaration. No clinical data was required for this 510(k) submission.
Technological Characteristics
Mobile ultrasound system using piezoelectric transducers (1.0–20.0 MHz). Features digital beamforming and digital dynamic receive focusing. Supports B-mode, M-mode, Spectral Doppler (PW/CW), Color Doppler, and Power Doppler. Connectivity via DICOM 3.0. Safety compliance: EN 60601-1, UL 2601-1, C22.2 No. 601.1, IEC 1157, EN 60601-1-2.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in fetal, abdominal, pediatric, small organ, neonatal/adult cephalic, cardiac, trans-rectal, trans-vaginal, peripheral-vascular, musculoskeletal, and intra-operative applications. Prescription use only.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
K013627
# NOV 1 6 2001
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.
- 1)
Submitter's name, address, telephone number, contact person: Bob DePalma (714) 889-3070 Regulatory Affairs Medison America, Inc. 11075 Knott Ave. Cypress, CA 90630 Telephone : 714 - 889 - 3000 Facsimile : 714-889-3079
Email : bdepalma@medison.com
Prepared August 22, 2001
- 2) Name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known:
Common/Usual Name: Diagnostic Ultrasound System and Accessories
Proprietary Name:
SA8000 Diagnostic Ultrasound System and Transducers.
| Classification Names: | FR Number | Product Code |
|------------------------------------------|-----------|--------------|
| Ultrasound Pulsed Echo Imaging System | 892.1560 | 90-IYO |
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYN |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
{1}------------------------------------------------
#### 3) Identification of the predicate or legally marketed device:
Medison America, Inc. believes that SA 8000 ultrasound system is substantially equivalent to the currently marketed SA 9900 ultrasound system (K002185) and SA8800/HDI1500 ultrasound system (K974269)
#### Device Description: 4)
The SA8000 scanner is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as B-mode, M-mode, Color-Flow Doppler, Continuous (CW) Doppler, Pulsed (PW) Doppler and Power Doppler, or as a combination of these modes. The SA8000 also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The SA8000 has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.
Seventeen different models of transducers are available and any three may be connected at the same time. In addition to the initial operational settings for each transducer preprogrammed in the system, user-customized parameter settings for each transducer may be inserted by the operator and stored for recall as needed via the system control panel. Customization includes transmit focusing, filtering, image enhancement processing, dynamic window curve selection. Controls are also provided to select display format (single and various combinations), to activate zoom features, and to utilize the cine loop function. More detailed explanations of these functions and controls are included in Chapter 2 of the Operator manual, and in the software/firmware documentation included in this 510(k) Notification.
The SA8000 uses digital beamforming technology, and supports a variety of Linear, Convex, Phased Array and Static probes for a wide variety of applications. It is an ultrasound scanner, which provides high resolution, high penetration performance, and various measurement functions. Probes are supported in frequencies from 1.0 MHz to 20.0 MHz. These probes can be applied to a variety of clinical applications such as fetal, abdominal, pediatric, small organ, neonatal cephalic, adult cephalic, cardiac, trans-rectal, trans-vaginal, peripheral-vascular, muscular-skeletal, and Intra-operative. The same clinical uses were cleared for the predicate devices, SA9900 (K002185) and SA8800/HDI1500 (K974269)
{2}------------------------------------------------
The system can be used to measure distances and calculate areas, circumferences and volumes, as well as calculate the expected date of delivery by using BPD (biparietal diameter), OFD (occipitofrontal diameter), HC (head circumference), AC (abdominal circumference), AD (abdominal diameter), FL (femur length), CRL (crown rump length), APTD (anteroposterior trunk diameter), TTD (transverse trunk diameter), GS (gestational sac), LMP (last menstural period.), Cardiac Analysis (volume by area/length, Simpson biplane and single plane, M-mode analysis, Doppler: peak and mean gradients, pressure half time, E/A ratio and continuity equation) and Vascular Analysis (resistive index, pulsatility index. % stenosis, ICA/CCA ratio, Volume flow).
Biopsy guidelines are provided on screen to assist in the collection of tissue samples, using biopsy guide adapters offered as an optional accessory. M-mode uses the scroll display method which has its images flow from the right to the left on the monitor. The SA8000 supports the Cine function (capable of storing up to 256 sequential images) and real-time zoom function to the region-of-interest. The system provides the ability to perform remote viewing of images, without compression, via a DICOM 3.0 compatible output. Management of patient history is possible by image-filing function. High-resolution images are provided by utilizing a technology called digital dynamic receive focusing.
The SA8000 has been designed to meet the following electromechanical safety standards:
- EN 60601-1 (IEC 601-1,) European Norm, Medical Electrical Equipment ●
- UL 2601-1, Underwriters Laboratories Standards, Medical Electrical . Equipment
- C22.2 No. 601.1, Canadian Standards Association, Medical Electrical . Equipment
- CEVIEC 1157:1992, International Electrotechnical Commission, . Requirements for the declaration of the acoustic output of medical diagnostic ultrasonic equipment
- EN 60601-1-2 (IEC 60601-1-2,) European Norm, Collateral Standard: . Electromagnetic Compatibility
- Compliant with the European Medical Device Directive Certificate issued by . TUV.
{3}------------------------------------------------
#### Intended Use: 5)
SA8000 intended uses as defined FDA guidance documents are:
- Fetal (includes infertility monitoring of follicle development) .
- Abdominal .
- Pediatric .
- Small Organ
- Neonatal Cephalic .
- Adult Cephalic .
- Cardiac (Adult, Pediatric) ●
- Trans-Rectal .
- Trans-Vaginal .
- . Peripheral-Vascular
- Muscular-Skeletal (conventional, superficial)
- . Intra-Operative (Neurological, Abdominal, Peripheral vascular)
Typical examinations performed using the system are:
- General abdominal and pelvic studies including organ surveys, . assessment, and retroperitoneal cavity studies.
- . Study of small parts including breasts, shoulders, thyroid, and the abdominal wall.
- . Pediatric scans of organs and bony structures.
- Peripheral vascular applications including carotid arteries, legs, arms, . feet, and penile artery.
- Monitoring procedures for infertility studies (other than in vitro . fertilization).
- . First, second and third trimester pregnancy studies.
- Prostate, prostate biopsy guidance, and rectal wall studies. .
- . Neonatal head studies.
- Trans-cranial studies of middle cerebral arteries, internal carotid artery, . and vertebral arteries.
- . Cardiac studies in adults and children.
- Biopsy guidance for tissue or fluid sampling. .
- . Conventional podiatry scans.
{4}------------------------------------------------
#### Technological Characteristics: 6)
This device operates identical to the predicate devices in that piezoelectric material in the transducer is used as an ultrasound source to transmit sound waves into the body. Sound waves are reflected back to the transducer and converted to electrical signals that are processed and displayed as 2D and M-mode, Spectral Doppler, Color Doppler, Power Doppler, or 3D images. Transducer patient contact materials are biocompatible.
The device's acoustic output limits are:
All Applications:
| ISPTA | (Maximum Range<br>720 mW/cm2 |
|-------|------------------------------|
| MI | 1.9 |
The limits are the same as predicate Track 3 devices.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 6 2001
Medison America, Inc. % Mr. Mark Job Program Manager TÜV Product Service Inc. 1775 Old Highway 8 NW, Suite 104 NEW BRIGHTON MN 55112-1891
Re: K013627
Trade Name: SonoAce SA-8000 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: 90 IYN Product Code: 90 IYO Product Code: 90 ITX Dated: November 1, 2001 Received: November 5, 2001
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the SonoAce SA-8000 Diagnostic Ultrasound System , as described in your premarket notification:
{6}------------------------------------------------
# Transducer Model Number
Image /page/6/Figure/2 description: The image shows a list of ultrasound transducers with their specifications. The list includes linear arrays such as L5-9EC, L5-9ER, HL5-9ED, L5-10ED, and LI5-9EV, along with curved arrays like C3-7ED, EC4-9ES, and CL4-8EV. Phased arrays such as P2-4AM, P2-5AC, and P3-7AM are also listed, as well as static CW transducers and volume curved arrays like S-VAW3-5, S-VAW4-7, S-VDW5-8(B), and S-VNA5-8(B).
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
{7}------------------------------------------------
Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.
Sincerely yours,
Nancy C Brogdon
Nancy C. Brogdon Director. Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{8}------------------------------------------------
# NOV 1 6 2001
# Section 4.3 INDICATIONS FOR USE
K013627
### DIAGNOSTIC ULTRASOUND INDICATIONS STATEMENT
510(k) No .: System:
Intended Use:
SA8000 Ultrasound System Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
|--------------------------|---------------------------|------------------------------|---------------------------------------------------|----|-----|-----|-------------------|----------------------|------------------|
| | General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| | Ophthalmic | Ophthalmic | | | | | | | |
| | | Fetal (See Note 3) | P | P | P | | P | Note 1 | Notes 2, 7, 8 |
| | | Abdominal | P | P | P | P1 | P | Note 1 | Notes 2, 7, 8 |
| | | Intra-operative (See Note 6) | P1 | P1 | P1 | | P1 | Note 1 | Notes 8 |
| | | Intra-operative (Neuro.) | P1 | P1 | P1 | P1 | P1 | Note 1 | Notes 8 |
| Fetal Imaging<br>& Other | | Laparoscopic | | | | | | | |
| | | Pediatric | P | P | P | P | P | Note 1 | Notes 2, 7, 8 |
| | | Small Organ (See Note 5) | P | P | P | | P | Note 1 | Notes 2, 8 |
| | | Neonatal Cephalic | P | P | P | P | P | Note 1 | Notes 7, 8 |
| | | Adult Cephalic | P | P | P | P | P | Note 1 | Notes 8 |
| | | Trans-rectal | P | P | P | | P | Note 1 | Notes 2, 8 |
| | | Trans-vaginal | P | P | P | | P | Note 1 | Notes 2, 8 |
| | | Trans-urethral | | | | | | | |
| | | Trans-esoph. (non-Cardiac) | | | | | | | |
| | | Musculo-skel. (Convent.) | P | P | P | | P | Note 1 | Notes 2, 8 |
| | | Musculo-skel. (Superfic.) | P2 | P2 | P2 | | P2 | Note 1 | Notes 2, 8 |
| | | Intra-luminal | | | | | | | |
| | | Other (spec.) | | | | | | | |
| | | Cardiac Adult | P | P | P | P | P | Note 1 | Notes 4, 7, 8 |
| Cardiac | | Cardiac Pediatric | P | P | P | P | P | Note 1 | Notes 4, 7, 8 |
| | | Trans-esophageal (Cardiac) | | | | | | | |
| | | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | | Peripheral vessel | P | P | P | N | P | Note 1 | Note 8 |
| | | Other (spec.) | | | | N | | | Note 5 |
N= new indication; P= previously cleared by FDA in K002185 and K974269; P1= previously cleared by FDA in K974269; P2= previously cleared by FDA in K002185; P3= previously cleared by FDA in K984639; E= added under Appendix E Additional Comments:
Color Doppler includes Power (Amplitude) Doppler (P1, P2, P3)
Note 1: PWD/Color Doppler, PWD/Power Doppler, CWD/Color Doppler, CWD/Power Doppler (P1, P2, P3)
Note 2: Includes imaging for guidance of biopsy (P1, P2)
Note 3: Includes infertility monitoring of follicle development (P1, P2)
Note 4: Color M-mode (P1)
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients (P1, P2)
Note 6: Abdominal organs and peripheral vessel (P1, P2)
Note 7: Tissue Harmonic Imaging (THI) (P1, P2)
Note 8: 3D Imaging (P1, P2, P3)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 24 CFR 801.109)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K013627
Section 4.3, Page 1
{9}------------------------------------------------
| 510(k) No.: | | | | | | | | | | |
|---------------------------|------------------------------------------------------------------------------------|---|---|-----|-----|-------------------|----------------------|---------------------------------------------------|--|--|
| System: | SA8000 Ultrasound System | | | | | | | | | |
| Transducer: | L5-9EC / 5-9MHz / 7.5 MHz / 40mm Linear Array | | | | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | | |
| | Clinical Application | | | | | | | Mode of Operation (*includes simultaneous B-mode) | | |
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | |
| Ophthalmic | Ophthalmic | | | | | | | | | |
| | Fetal (See Note 3) | | | | | | | | | |
| | Abdominal | | | | | | | | | |
| | Intra-operative<br>(Abdominal,<br>vascular) | | | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | | |
| | Pediatric | N | N | N | | N | Note 1 | Note 2, 8 | | |
| | Small Organ (See Note 5) | N | N | N | | N | Note 1 | Note 2, 8 | | |
| | Neonatal Cephalic | | | | | | | | | |
| | Adult Cephalic | | | | | | | | | |
| | Trans-rectal | | | | | | | | | |
| | Trans-vaginal | | | | | | | | | |
| | Trans-urethral | | | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | | | |
| | Musculo-skel. (Convent.) | N | N | N | | N | Note 1 | Note 2, 8 | | |
| | Musculo-skel. (Superfic.) | N | N | N | | N | Note 1 | Note 2, 8 | | |
| | Intra-luminal | | | | | | | | | |
| | Other (spec.) | | | | | | | | | |
| | Cardiac Adult | | | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | | | |
| | Other (spec.) | | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | N | N | N | | N | Note 1 | Note 8 | | |
| | Other (spec.) | | | | | | | | | |
N= new indication; P= previously cleared by FDA in K002185 and K974269; P1= previously cleared by FDA in K974269; P2= previously cleared by FDA in K002185; P3= previously cleared by FDA in K984639; E= added under Appendix E Additional Comments:
#### Color Doppler includes Power (Amplitude) Doppler (P1, P2, P3)
- Note 1: PWD/Color Doppler, PWD/Power Doppler, CWD/Color Doppler, CWD/Power Doppler (P1, P2, P3)
- Note 2: Includes imaging for guidance of biopsy (P1, P2)
- Note 3: Includes infertility monitoring of follicle development (P1, P2)
and Radiologi 510kl N
- Note 4: Color M-mode (P1)
- Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients (P1, P2)
- Note 6: Abdominal organs and peripheral vessel (P1, P2)
- Note 7: Tissue Harmonic Imaging (THI) (P1, P2)
- Note 8: 3D Imaging (P1, P2, P3)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Prescription Use (Per 21 CFR 801.109)
Nancy C. Brogdon
(Division Sign-Off)
Division of Reproductive, Abdominal
{10}------------------------------------------------
| 510(k) No.: | | | | | | | | |
|---------------------------|------------------------------------------------------------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| System: | SA8000 Ultrasound System | | | | | | | |
| Transducer: | L5-9ER / 5-9MHz / 7.5 MHz / 50mm Linear Array | | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | |
| Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal (See Note 3) | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Abdominal,<br>vascular) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | N | N | N | | N | Note 1 | Note 2, 8 |
| | Small Organ (See Note 5) | N | N | N | | N | Note 1 | Note 2, 8 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | N | N | N | | N | Note 1 | Note 2, 8 |
| | Musculo-skel. (Superfic.) | N | N | N | | N | Note 1 | Note 2, 8 |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | N | N | N | | N | Note 1 | Note 8 |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA in K002185 and K974269; Pl= previously cleared by FDA in K974269; P2= previously cleared by FDA in K002185; P3= previously cleared by FDA in K984639; E= added under Appendix E Additional Comments:
#### Color Doppler includes Power (Amplitude) Doppler (P1, P2, P3)
Note 1: PWD/Color Doppler, PWD/Power Doppler, CWD/Color Doppler, CWD/Power Doppler (P1, P2, P3)
- Note 2: Includes imaging for guidance of biopsy (P1, P2)
- Note 3: Includes infertility monitoring of follicle development (P1, P2)
Note 4: Color M-mode (P1)
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients (P1, P2)
.
- Note 6: Abdominal organs and peripheral vessel (P1, P2)
- Note 7: Tissue Harmonic Imaging (THI) (P1, P2)
- Note 8: 3D Imaging (P1, P2, P3)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801-109)
| (Division Sign-Off) | |
|---------------------------------------------------------------|---------|
| Division of Reproductive, Abdominal, and Radiological Devices | |
| 510(k) Number | K013627 |
Section 4.3, Page 3
{11}------------------------------------------------
| 510(k) No.: | | | | | | | | |
|---------------------------|------------------------------------------------------------------------------------|---------------------------------------------------|---|-----|-----|----------------|----------------------|------------------|
| System: | SA8000 Ultrasound System | | | | | | | |
| Transducer: | HL5-9ED / 5-9MHz / 7.5 MHz / 40mm Linear Array | | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | |
| General<br>(Track I only) | Clinical Application<br>Specific<br>(Tracks I & III) | Mode of Operation (*includes simultaneous B-mode) | | | | | Combined*<br>(Spec.) | Other<br>(Spec.) |
| | | B | M | PWD | CWD | Color Doppler* | | |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal (See Note 3) | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Abdominal, vascular) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | N | N | N | | N | Note 1 | Note 2, 8 |
| | Small Organ (See Note 5) | N | N | N | | N | Note 1 | Note 2, 8 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | N | N | N | | N | Note 1 | Note 2, 8 |
| | Musculo-skel. (Superfic.) | N | N | N | | N | Note 1 | Note 2, 8 |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | N | N | N | | N | Note 1 | Note 8 |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA in K002185 and K974269; P1= previously cleared by FDA in K974269; P2= previously cleared by FDA in K002185; P3= previously cleared by FDA in K984639; E= added under Appendix E Additional Comments:
# Color Doppler includes Power (Amplitude) Doppler (P1, P2, P3)
- Note 1: PWD/Color Doppler, PWD/Power Doppler, CWD/Color Doppler, CWD/Power Doppler (P1, P2, P3)
- Note 2: Includes imaging for guidance of biopsy (P1, P2)
- Note 3: Includes infertility monitoring of follicle development (P1, P2)
- Note 4: Color M-mode (P1)
- Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients (P1, P2)
- Note 6: Abdominal organs and peripheral vessel (P1, P2)
- Note 7: Tissue Harmonic Imaging (THI) (P1, P2)
- Note 8: 3D Imaging (P1, P2, P3)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
| | Nancy C. Brogdon |
|---------------------------------------------------------------|---------------------|
| (Division Sign-Off) | |
| Division of Reproductive, Abdominal, and Radiological Devices | Section 4.3, Page 4 |
| 510(k) Number | K013627 |
{12}------------------------------------------------
510(k) No.: SA8000 Ultrasound System System: L5-10ED / 5-10 / 7.5 MHz / 40mm Linear Array Transducer: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use: Mode of Operation (*includes simultaneous B-mode) Clinical Application Specific B M PWD CWD Color Combined* Other General (Track I only) (Tracks I & III) Doppler* (Spec.) (Spec.) Ophthalmic Ophthalmic Fetal (See Note 3) Abdominal Intra-operative (Abdominal, vascular) Intra-operative (Neuro.) Laparoscopic Fetal Imaging & Other Pediatric N N N Note 1 Note 2, 8 N Small Organ (See Note 5) N N N N Note 1 Note 2, 8 Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph. (non-Cardiac) Musculo-skel. (Convent.) N N N N Note l Note 2, 8 Musculo-skel. (Superfic.) Note 1 Note 2, 8 N N N N Intra-luminal Other (spec.) Cardiac Adult Cardiac Pediatric Cardiac Trans-esophageal (Cardiac) Other (spec.) Peripheral Peripheral vessel N N N N Note 1 Note 8 Vessel Other (spec.)
N= new indication; P= previously cleared by FDA in K002185 and K974269; P1= previously cleared by FDA in K974269; P2= previously cleared by FDA in K002185; P3= previously cleared by FDA in K984639; E=added under Appendix E Additional Comments:
#### Color Doppler includes Power (Amplitude) Doppler (P1, P2, P3)
Note 1: PWD/Color Doppler, PWD/Power Doppler, CWD/Color Doppler, CWD/Power Doppler (P1, P2, P3)
Note 2: Includes imaging for guidance of biopsy (P1, P2)
Note 3: Includes infertility monitoring of follicle development (P1, P2)
Note 4: Color M-mode (P1)
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients (P1, P2)
Note 6: Abdominal organs and peripheral vessel (P1, P2)
Note 7: Tissue Harmonic Imaging (THI) (P1, P2)
Note 8: 3D Imaging (P1, P2, P3)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)

| (Division Sign-Off) | |
|--------------------------------------|---------|
| Division of Reproductive, Abdominal, | |
| and Radiological Devices | |
| 510(k) Number | K013627 |
Section 4.3, Page 5
{13}------------------------------------------------
| 510(k) No.: | | | | | | | | |
|---------------------------|------------------------------------------------------------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| System: | SA8000 Ultrasound System | | | | | | | |
| Transducer: | LI5-9EV / 5-9MHz / 7.5 MHz / 40mm Linear Array | | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | |
| | Clinical Application | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal (See Note 3) | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Abdominal, vascular) | N | N | N | | N | Note 1 | Notes 8 |
| | Intra-operative (Neuro.) | N | N | N | N | N | Note 1 | Notes 8 |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (See Note 5) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA in K002185 and K974269; P1= previously cleared by FDA in K974269; P2= previously cleared by FDA in K002185; P3= previously cleared by FDA in K984639; E= added under Appendix E Additional Comments:
#### Color Doppler includes Power (Amplitude) Doppler (P1, P2, P3)
- Note 1: PWD/Color Doppler, PWD/Power Doppler, CWD/Color Doppler, CWD/Power Doppler (P1, P2, P3)
Note 2: Includes imaging for guidance of biopsy (P1, P2)
Note 3: Includes infertility monitoring of follicle development (P1, P2)
Note 4: Color M-mode (P1)
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients (P1, P2)
- Note 6: Abdominal organs and peripheral vessel (P1, P2)
Note 7: Tissue Harmonic Imaging (THI) (P1, P2)
- Note 8: 3D Imaging (P1, P2, P3)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 80
Nancy C. Hogdon
Basic Information
{14}------------------------------------------------
| 510(k) No.:<br>System:<br>Transducer:<br>Intended Use: | SA8000 Ultrasound System<br>C3-7ED / 3-7MHz / 4.5 MHz / 50R Curved Array<br>Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|---|---|-----|-----|-------------------|---------------------------------------------------|------------------|
| | Clinical Application | | | | | | Mode of Operation (*includes simultaneous B-mode) | |
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal (See Note 3) | N | N | N | | N | Note 1 | Note 2, 7, 8 |
| | Abdominal | N | N | N | | N | Note 1 | Note 2, 7, 8 |
| | Intra-operative<br>(Abdominal,<br>vascular) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | N | N | N | | N | Note 1 | Note 2, 7, 8 |
| | Small Organ (See Note 5) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA in K002185 and K974269; P1= previously cleared by FDA in K974269; P2= previously cleared by FDA in K002185; P3= previously cleared by FDA in K984639; E=added under Appendix E Additional Comments:
Color Doppler includes Power (Amplitude) Doppler (P1, P2, P3)
Note 1: PWD/Color Doppler, PWD/Power Doppler, CWD/Color Doppler, CWD/Power Doppler (P1, P2, P3)
Note 2: Includes imaging for guidance of biopsy (P1, P2)
Note 3: Includes infertility monitoring of follicle development (P1, P2)
Note 4: Color M-mode (P1)
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients (P1, P2)
Note 6: Abdominal organs and peripheral vessel (P1, P2)
Note 7: Tissue Harmonic Imaging (THI) (P1, P2)
Note 8: 3D Imaging (P1, P2, P3)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 80
Mancy C. bhogdm
(Division Sign-Off
Section 4.3, Page 7
Division of Reproductive, Abdominal,
and Radiological Devices
KO
{15}------------------------------------------------
| 510(k) No.: | | | | | | | | | |
|---------------------------|------------------------------------------------------------------------------------|--|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| System: | SA8000 Ultrasound System | | | | | | | | |
| Transducer: | EC4-9ES / 4-9MHz / 6.5 MHz / 10R Curved Array | | | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | |
| | Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
| General<br>(Track I only) | Specific<br>(Tracks I & III) | | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal (See Note 3) | | | | | | | | |
| | Abdominal | | | | | | | | |
| | Intra-operative (Abdominal, vascular) | | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (See Note 5) | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | N | N | N | | N | Note 1 | Note 2, 8 |
| | Trans-vaginal | | N | N | N | | N | Note 1 | Note 2, 8 |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | | |
| | Intra-luminal | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| | Cardiac Adult | | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | | |
| | Other (spec.) | | | | | | | | |
N= new indication; P= previously cleared by FDA in K002185 and K974269; P1= previously cleared by FDA in K974269; P2= previously cleared by FDA in K002185; P3= previously cleared by FDA in K984639; E= added under Appendix E Additional Comments:
#### Color Doppler includes Power (Amplitude) Doppler (P1, P2, P3)
- Note 1: PWD/Color Doppler, PWD/Power Doppler, CWD/Color Doppler, CWD/Power Doppler (P1, P2, P3)
- Note 2: Includes imaging for guidance of biopsy (P1, P2)
- Note 3: Includes infertility monitoring of follicle development (P1, P2)
- Note 4: Color M-mode (P1)
- Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients (P1, P2)
- Note 6: Abdominal organs and peripheral vessel (P1, P2)
- Note 7: Tissue Harmonic Imaging (THI) (P1, P2)
- Note 8: 3D Imaging (P1, P2, P3)
Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)
Nancy C. hogdon
(Division Sign-Om
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number
U
K01362
Section 4.3, Page 8
{16}------------------------------------------------
| 510(k) No.: | |
|---------------|------------------------------------------------------------------------------------|
| System: | SA8000 Ultrasound System |
| Transducer: | CL4-8EV / 4-8MHz / 6.5 MHz / 40R Curved Array |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: |
| | Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | | |
|--------------------------|---------------------------|---------------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|--|
| | General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | |
| | Ophthalmic | Ophthalmic | | | | | | | | |
| | | Fetal ( <i>See Note 3</i> ) | | | | | | | | |
| | | Abdominal | | | | | | | | |
| | | Intra-operative (Abdominal, vascular) | N |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.