← Product Code [KGO](/productcode/KGO) · K013560

# SEMPERMED SUPREME, SIZES 5.5, 6, 6.5, 7, 7.5, 8, 8.5, AND 9 (K013560)

_Semperit Technische Produkte Gesellshaft · KGO · Jan 16, 2002 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K013560

## Device Facts

- **Applicant:** Semperit Technische Produkte Gesellshaft
- **Product Code:** [KGO](/productcode/KGO.md)
- **Decision Date:** Jan 16, 2002
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 878.4460
- **Device Class:** Class 1
- **Review Panel:** General, Plastic Surgery

## Indications for Use

A surgeon's glove is a device made of natural or synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination. (21CFR 878.4460)

## Device Story

Sempermed Supreme is a sterile, powder-free surgeon's glove; constructed from natural rubber; interior coated with polyurethane derivatives and butadiene-based copolymer cross-linked by polyacrylates; siliconized. Used in surgical settings by operating room personnel as a barrier against infectious materials and contaminants. Provides protection for surgical wounds. Device is disposable.

## Clinical Evidence

Bench testing only. No clinical data provided. Substantial equivalence is based on comparative technical characteristics and material safety assessments.

## Technological Characteristics

Natural rubber surgeon's glove; interior coating of polyurethane derivatives and butadiene-based copolymer cross-linked by polyacrylates; siliconized. Water-soluble protein content < 50 ug/dm2. Sterile, powder-free, disposable.

## Regulatory Identification

A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Supreme Sempermed Latex Surgeon's gloves, powder-free ([K981096](/device/K981096.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## KO13560

# 510(k) Summary As Required by 21 section 807.92 ( c )

JAN 1 6 2002

| 1-Submitter Name:                                                   | Semperit Technische Produkte GmbH & Co KG                                                                                                                                                                                                                                   |
|---------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2-Address:                                                          | Modecentraße 22<br>A-1031 Vienna AUSTRIA                                                                                                                                                                                                                                    |
| 3-Phone:                                                            | (++43) 2630 310 216                                                                                                                                                                                                                                                         |
| 4-Fax:                                                              | (++43) 2630 310 479                                                                                                                                                                                                                                                         |
| 5-Contact Person:                                                   | Dipl.-Ing. Richard Ehrenfeldner                                                                                                                                                                                                                                             |
| 6-Date summary prepared:                                            | October 15th, 2001                                                                                                                                                                                                                                                          |
| 7- Official Correspondent:                                          | Mansour Consulting                                                                                                                                                                                                                                                          |
| 8- Address:                                                         | 1308 Morningside Park Dr<br>Alpharetta, GA 30022 USA                                                                                                                                                                                                                        |
| 9- Phone:                                                           | (678) 908-8180                                                                                                                                                                                                                                                              |
| 10- Fax:                                                            | (425) 795-9341                                                                                                                                                                                                                                                              |
| 11- Contact person:                                                 | Jay Mansour, president                                                                                                                                                                                                                                                      |
| 12-Device Trade or Proprietary Name:                                | sempermed® SUPREME                                                                                                                                                                                                                                                          |
| 13-Device Common or usual name:                                     | Surgeon's glove                                                                                                                                                                                                                                                             |
| 9-Device Classification Name:                                       | Surgeon's glove                                                                                                                                                                                                                                                             |
| 10-Substantial Equivalency is claimed against the following device: | "Supreme" Sempermed® Latex Surgeon's gloves, powder-free<br>510k #k981096 (refer to Appendix 2 for FDA notification letter<br>This notification for the sempermed® SUPREME is of the<br>ABBREVIATED type as per the declaration of conformity on page -<br>of this summary. |
|                                                                     | THIS DEVICE (SUPREME) UNDERWENT A CHANGE IN MATERIALS FOR<br>WHICH THIS NEW 510K IS NEEDED (refer to section D3 in CDRH Submissio                                                                                                                                           |

I и 1 cove IT INCURRED THE CHANGE IN MATERIAL

#### 11-Description of the Device:

sempermed® SUPREME is a sterile powder-free surgeon's glove. It is a disposable device made of natural rubber inside coated with a layer of polyurethane derivatives, and butadiene-based copolymer cross-linked by
polyacrylates- free of process, siliconized. This glove is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants. Water soluble protein less than 50 ug/dm2

### 12-Intended use of the device: (Indications for use typed on a separate FDA form)

sempermed® DERMA plus is a powdered surgeon's glove and it is a disposable device made of natural rubber inside coated with synthetic latex material that bears powder to facilitate donning and it is intended to be worn on the hands usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants. The glove is powdered with absorbable dusting powder USP Grade

### 13-Safety and effectiveness of the device:

The sempermed® SUPREME is safe and effective as the predicate device as only minor change in materials occurred. Indeed, it is equivalent. This is better expressed in the tabulated comparison (Paragraph 14 below)

{1}------------------------------------------------

### 14-Summary comparing technological characteristics with other predicate device:

General comparison result between sempermed® SUPREME gloves and the predicate device (sempermed® SUPREME ("PREVIOUS MATERIALS") is tabulated below.

Technical comparison of specific elements is attached in the main submission

| FDA file reference number                        | 510k 981096                                                                                    |
|--------------------------------------------------|------------------------------------------------------------------------------------------------|
| Attachments inside notification submission file  | REFER TO APPENDIX 2                                                                            |
| TECHNOLOGICAL CHARACTERISTICS                    | Comparison result<br>REFER TO ADDITIONAL TECHNICAL<br>COMPARATIVE TABLE WITHIN 510K SUBMISSION |
| Indications for use                              | Identical                                                                                      |
| Target population                                | Identical                                                                                      |
| Design                                           | Identical                                                                                      |
| Materials                                        | Similar                                                                                        |
| Performance                                      | Identical                                                                                      |
| Sterility                                        | Identical                                                                                      |
| Biocompatibility                                 | Identical                                                                                      |
| Mechanical safety                                | Identical                                                                                      |
| Chemical safety                                  | Identical                                                                                      |
| Anatomical sites                                 | Identical                                                                                      |
| Human factors                                    | Identical                                                                                      |
| Energy used and/or delivered                     | Identical (not applicable)                                                                     |
| Compatibility with environment and other devices | Identical                                                                                      |
| Where used                                       | Identical                                                                                      |
| Standards met                                    | Identical                                                                                      |
| Electrical safety                                | Identical (not applicable)                                                                     |
| Thermal safety                                   | Identical (not applicable)                                                                     |
| Radiation safety                                 | Identical (not applicable)                                                                     |

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

### JAN 1 6 2002

Semperit Technische Produkte Gesellshaft C/O Mr. Jay Mansour Mansour Consulting 1308 Morningstar Park Drive Apharetta, Georgia 30022

Re: K013560

Trade/Device Name: Sempermed Supreme Powder Free Latex Surgeon's Gloves with Protein Content Labeling Claim (50 Micrograms or Less) Regulation Number: 878.4460 Regulation Name: Surgeon's Gloves Regulatory Class: I Product Code: KGO Dated: October 15, 2001 Received: October 26, 2001

Dear Mr. Mansour:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements

{3}------------------------------------------------

#### Page 2 - Mr. Mansour

of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Timothy A. Ulatowski

Timo Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

### Indication for Use Statement

### Applicant:

510(k) Number (if known): Koll 3560

Device Name: Sempermed Supreme Powder Free Latex Surgeon's Gloves With Protein Content Labeling Claim (50 Microgram or Less)

### Indications For Use:

A surgeon's glove is a device made of natural or synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination. (21CFR 878.4460)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Olin S. Lim

(Division Sign-Off) Diston of Dental, Infection Control, ann General Hospital Devices F . O(k) Number _

(Optional Format 3-10-98)

---

**Source:** [https://fda-staging.innolitics.com/device/K013560](https://fda-staging.innolitics.com/device/K013560)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).