K013548 · Standard Imaging, Inc. · KPT · Nov 9, 2001 · Radiology
Device Facts
Record ID
K013548
Device Name
STANDARD IMAGING IVB 1000 WELL CHAMBER
Applicant
Standard Imaging, Inc.
Product Code
KPT · Radiology
Decision Date
Nov 9, 2001
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1360
Device Class
Class 2
Indications for Use
The Standard Imaging IVB 1000 Well Chamber is a well-type chamber. It is specifically designed to measure the amount of radiation of high-dose-rate (HDR), low-dose-rate (LDR) and intravascular (IVB) brachytherapy (gamma and beta) sources, with the appropriate calibration from an accredited dosimetry calibration laboratory. Sources must be measured using the appropriate and specific source holder as described in the IVB 1000 labeling. To check the bremstralung component of a source, the X-Ray Contamination Test Tool, an accessory to the IVB 1000, should be used.
Device Story
Well-type ionization chamber; measures radiation output of HDR, LDR, and IVB brachytherapy sources (gamma/beta). Requires external electrometer (10-12 A to 10-7 A range) via triax connector; 300V bias applied. User inserts source into specific holder; chamber vents to ambient pressure; readings corrected for temperature/pressure. Used by medical physicists/dosimetrists in clinical radiotherapy environments to ensure accurate source calibration before patient treatment. Benefits patient by ensuring precise radiation delivery.
Clinical Evidence
No clinical data. Device validated through calibration testing conducted by the University of Wisconsin - Madison, Department of Medical Physics Accredited Dosimetry Calibration Laboratory.
Technological Characteristics
Well-type ionization chamber; triax connector; 300V bias; ambient pressure venting. Complies with IEC 601-1 (1988) and IEC 60731 (1997).
Indications for Use
Indicated for measuring radiation output of HDR, LDR, and IVB (gamma and beta) brachytherapy sources in clinical dosimetry settings.
Regulatory Classification
Identification
A radionuclide dose calibrator is a radiation detection device intended to assay radionuclides before their administration to patients.
Predicate Devices
Standard Imaging HDR 1000 Plus/ IVB 1000 Well Chambers (K001825)
Submission Summary (Full Text)
{0}------------------------------------------------
NOV 0 9 2001
October 24, 2001
K013548
Page 1 of 2
510(k) Summary of Safety and Effectiveness Information for the Subject: Standard Imaging IVB 1000 Well Chamber
Standard Imaging IVB 1000 Well Chamber Proprietary Name:
Ion Chamber Common Name:
Class II - 21CFR892.5700, 90JAQ or Classification: Class I - 21CFR892.5650, 90IWJ
Radiology Panel:
Raymond Riddle, Vice President, Regulatory Affairs Contact Person:
Class I - 21CFR892.1940, 90LHO
The 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 1992.
The Standard Imaging IVB 1000 Well Chamber is substantially equivalent to the Standard Imaging HDR 1000 Plus/ IVB 1000 Well Chambers, which were cleared by FDA with 510(k) premarket notification K001825.
The Standard Imaging IVB 1000 Well Chamber is a well-type chamber. It is specifically designed to measure the amount of radiation of high-dose-rate (HDR), low-dose-rate (LDR) and intravascular (IVB) brachytherapy (gamma and beta) sources, with the appropriate calibration from an accredited dosimetry calibration laboratory. Sources must be measured using the appropriate and specific source holder as described in the IVB 1000 labeling. To check the bremstralung component of a source, the X-Ray Contamination Test Tool, an accessory to the IVB 1000, should be used.
It is recommended that the chamber be calibrated every two years, as is standard practice for other ionization chambers. Initially, the calibration factor is given in the calibration report from an Accredited Dosimetry Calibration Laboratory (ADCL).
The measurement of brachytherapy sources requires an electrometer with a calibrated scale for measuring currents in the range from 10-12 A to 10-7A. Alternatively, a calibrated charge scale may be used with timed runs. If integral charge techniques are used with the time determined by the HDR irradiator timer, the contribution from the source transit-time should be taken into account.
Image /page/0/Picture/14 description: The image shows the logo for Standard Imaging. The logo features the words "Standard Imaging" in a stylized font, with the word "Standard" stacked on top of the word "Imaging". A trademark symbol is present to the right of the word "Imaging". Below the text is a graphic of lines radiating outward from a central point.
7601 Murphy Drive • Middleton, WI 53562-2532 • www.standardimaging.com
Tel (608) 831-0025 • (800) 261-4446 • Fax (608) 8
{1}------------------------------------------------
K013548
Page 2 of 2
The IVB 1000 Well Chamber has a vent hole to maintain the internal air at ambient atmospheric pressure. Thus, the readings obtained must be corrected for ambient temperature and pressure to the temperature and pressure of calibration (22º C and 760 mm Hg) at "normal" relative humidity (50% ± 25%non-condensing) in the usual accepted manner. The IVB 1000 has available different inserts for IVB, HDR, LDR and X-Ray contamination measurements.
The IVB 1000 Well Chamber has a conventional triax connector and cable to be connected to a suitable electrometer. A bias of 300 volts must be applied to the electrometer low-impedance connection relative to chassis ground. The voltage polarity effect is less than 0.1%. If desired, a second bias level of 150 volts can also be used to determine the ionic recombination loss at 300 V.
The Standard Imaging IVB 1000 Well Chamber was designed to comply with the limited applicable portions of the following voluntary standards:
- 1. IEC 601-1: 1988 - Medical Electrical Equipment
- IEC 60731: 1997 Medical Electrical Equipment Dosimeters with ionization chambers 2. used in radiotherapy.
The Standard Imaging IVB 1000 Well Chamber and the predicate Standard Imaging HDR 1000 Plus and IVB 1000 Ion Chambers are substantially equivalent in design concepts, technologies, materials and intended uses. The Standard Imaging IVB 1000 Well Chamber has been validated through calibration testing conducted by the University of Wisconsin - Madison, Department of Medical Physics Accredited Dosimetry Calibration Laboratory.
Image /page/1/Picture/7 description: The image shows the logo for Standard Imaging. The logo features the words "Standard Imaging" in a stylized font, with the word "Standard" stacked above the word "Imaging". A graphic of radiating lines is to the left of the text. A horizontal line is below the word "Imaging".
7601 Murphy Drive • Middleton, WI 53562-2532 • www.standardimaging.com
Tel (608) 831-0025 • (800) 261-4446 • Fax (60
EXRADIN.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services (HHS). The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged around the perimeter. Inside the circle is a stylized representation of three wavy lines, which symbolize the department's mission to protect and promote the health and well-being of Americans.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 0 9 2001
Raymond T. Riddle, PE Vice President, Regulatory Affairs Standard Imaging™, Inc. 7601 Murphy Drive MIDDLETON WI 53562-2532
Re: K013548
Trade/Device Name: Standard Imaging IVB 1000 Well Chamber Regulation Number: 21 CFR 892.1360 Regulation Name: Radionuclide dose calibrator Regulatory Class: II Product Code: 90 KPT Dated: October 24, 2001 Received: October 24, 2001
Dear Mr. Riddle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed notheation. The I Driving of Scation for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you don't specific at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, 10. questions of (301) 594-4639. Also, please note the regulation entitled, "Misbranding Other or Compination and (21 CFR Part 807.97). Other general information on by responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
NOV 0 9 2001
510(k) Number (if known):
K013548
Device Name: Standard Imaging IVB 1000 Well Chamber
Indications For Use:
The Standard Imaging IVB 1000 Well Chamber is a well-type chamber. It is specifically designed to measure the amount of radiation of high-dose-rate (HDR), low-dose-rate (LDR) and intravascular (IVB) brachytherapy (gamma and beta) sources, with the appropriate (calibration from an accredited dosimetry calibration laboratory. Sources must be measured using the appropriate and specific source holder as described in the IVB 1000 labeling. To check the bremstralung component of a source, the X-Ray Contamination Test Tool, an accessory to the IVB 1000, should be used.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21CFR801.109)
OR
Nancy broghon
Division of Repredustive, Abdominal,
and Radiological Devices
510(k) Number K013548
Over-The-Counter Use (Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.