K013521 · Adac Laboratories · JAK · Nov 7, 2001 · Radiology
Device Facts
Record ID
K013521
Device Name
GEMINI IMAGING SYSTEM
Applicant
Adac Laboratories
Product Code
JAK · Radiology
Decision Date
Nov 7, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
GEMINI is an imaging system that combines positron emission tomography (PET) and X-ray computed tomography. The GEMINI produces attenuation and non-attenuation corrected images of the distribution of PET radiopharmaceuticals in the head, body and total body as well as x-ray transmission images of these areas. The PET and CT images are registered and displayed in a fused image, providing both functional and anatomical data at different angles for interpretation by trained health professionals. The PET and CT portions of the system can be used either as an integrated system or as a stand-alone PET or CT system. GEMINI can provide CT data suitable for use in attenuation correction.
Device Story
GEMINI is a hybrid PET/CT imaging system integrating an ADAC ALLEGRO PET scanner and a Marconi MX8000 CT scanner into a single housing. The system acquires PET metabolic data and CT anatomical data, which are registered and fused for clinical interpretation. It operates in three modes: CT-only, PET-only, or combined PET/CT. The system features a separation mechanism to increase distance between PET and CT gantries for improved patient access and a dedicated table to minimize room requirements. Operated by trained health professionals in a clinical setting, the device provides attenuation-corrected PET images using CT-derived transmission maps. This fusion of functional and anatomical data facilitates faster diagnostic decision-making compared to separate imaging procedures.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing, including performance measurements of the ALLEGRO PET system according to NEMA-NU2 standards, evaluation of image quality using clinical phantoms, and verification of safety standards (IEC 60601-1, IEC 60601-2, IEC 60825-1, UL 2601-1, CAN/CSA-C22.2).
Technological Characteristics
Hybrid PET/CT system. PET component: GSO-based detectors (ALLEGRO). CT component: Dual-slice CT (MX8000). Connectivity: Integrated workstation for image registration and fusion. Standards: 21 CFR 1020.30-33 (CT radiation), IEC 60601-1, IEC 60601-2, IEC 60825-1, UL 2601-1, CAN/CSA-C22.2. Operation: Three modes (CT, PET, combined).
Indications for Use
Indicated for whole-body oncology imaging in patients requiring PET and CT diagnostic scans. Used for functional and anatomical localization of radiopharmaceutical distribution. Intended for use by trained health professionals.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
(012 510(k) SUMMARY OF SAFETY AND EFFECTIVENES
General Information
| A. | Submitted By:<br>ADAC Laboratories<br>540 Alder Dr.<br>Milpitas, CA 95035 | Contact: Coleen Coleman<br>Tel: (408) 468-3051<br>Fax: (408) 468-3050 |
|----|---------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| B. | Device Trade Name:<br>Common Name: | GEMINI<br>Positron Emission Tomography |
| | Classification Name: | Computed Tomography X-Ray<br>System, Emission Computed Tomography, (892.1200)<br>System, Computed Tomography X-Ray, (892.1750) |
| | Device Class: | 21CFR 892.1200, Class II<br>21 CFR 892.1750, Class II |
| | Product Code: | 90 KPS and 90 JAK |
| C. | Date prepared: | October 5, 2000 |
| D. | Predicate Device: | GE CT-PET (K010641) ADAC Allegro PET, K003434<br>EXACT CT PET (K002715) Marconi MX 8000 CT, K010817 |
| E. | | Performance Standards 21 CFR 1020.30 – 1020.33 as applicable |
for Computed Tomography X-Ray, (892.1750)
Intended Use: E.
GEMINI is an imaging system that combines positron emission tomography (PET) and X-ray OLMINI is an imaging System that other attenuation and non-attenuation corrected images computed toniography. - The OEMIX in the head, body and total body as well as x-ray of the distribution of I ET radiophalmater The PET and CT images are registered and displayed in a transmission mages of these arcas: "XX" and anatomical data at different angles for interpretation by trained health professionals. The PET and CT portions of the system can be used either as an by transd health professionals: "The RET or CT system. GEMINI can provide CT data suitable for use in attenuation correction.
F. Device Description:
The GEMINI hybrid PET/CT system is a combined positron emission tomography and X-100 The OLMIN Hyorathy scanner. This system uses the ADAC ALLEGRO PET system, K003434, Computed tonlogiaply Seather. "This Systems MX8000-v5.0 CT, K010817. The GEMINI integrates the two and the Marcon Meanen of on a single workstation to allow straightforward planning and system system operator consores mo the MX8000 CT system gantries remain intact as major operation. The PEEDOricated within a common integrated housing. It can be used in any subsydent components rocedures, which have been clinically conducted in a separate CT system end/or a PET system. No modifications have been made to either system, which would affect system performance.
performanos:
GEMINI is intended for use primarily as a clinical whole body oncology scanner with high-end dual slice CT capability and high-end GSO based PET performance. It mechanically might one data. Biros of capreater flexibility for above three operation modes. The availability of
GEMINI 510(k) Notification ADAC Laboratories
Section X.-1-1
{1}------------------------------------------------
K013521
-Page 2 of 2
low dose tomographic data sets implies that a natural extension of the ALLEGRO based cesium source attenuation correction to a CT transmission map attenuation correction exists. The primary purpose, however, of the CT is to provide precise anatomical localization for the metabolically significant positron emission distributions imaged on the PET system. These images when properly registered and displayed as a fused image, provide both functional and anatomical information with reliable spatial correspondence in a single image.
Comparison to Predicate Device: G.
The predicate devices, GE CT-PET (Discovery LS) and the ECAT CT-PET (Biograph) are similar to the GEMINI in that all of the devices consist of a full functional CT and PET. The patient may have a diagnostic CT and PET scan performed consecutively without having to move the patient. The GEMINI provides a mean to reach the diagnostic decision faster than the conventional way of imaging patients with both CT and PET systems in separate locations. The differences are the area of overall system dimensions, room size requirements, scanning length, and ease of use.
The GEMINI CT-PET is designed so that the system can operate in three modes: CT only, PET only and combined CT/PET. The major difference is that the GEMINI imaging system has a separation system that allows the distance of the CT and PET units to be increased. This feature provides easy access to the patient. GEMINI provides a newly developed dedicated PET/CT table to minimize the system room requirements
H. System Performance Test:
- The MX 8000 v.5.0 is manufactured in accordance with Performance Standards in 21 ● CFR 1020.30 -- 33.
- ALLEGRO system performance was measured according to the NEMA-NU2 standard. . In addition, clinical phantoms with clinical protocols were used to evaluate ALLEGRO image quality in terms of the noise texture and contrast of the image.
- The GEMINI complies with voluntary standards for safety and effectiveness (IEC 60601-. 1, IEC 60601-2, IEC 60825-1, UL 2601-1, & CAN/CSA-C22.2 and is tested to demonstrate the hazards, i.e., electrical, mechanical, and radiation have been minimized.
## I. Conclusion:
The GEMINI Imaging System is substantially equivalent to the predicate devices, the GE CT-PET (K010641) and the EXACT CT-PET (K002715) based upon similar intended use, technological comparison, and system performance.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. The bottom half of the circle contains a stylized image of an eagle with its wings spread. The eagle is facing to the right.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## NOV 0 7 2001
ADAC Laboratories % Michael Kwan, Ph.D. Principal Reviewer/Office Coordinator Underwriters Laboratories, Inc. 1655 Scott Boulevard SANTA CLARA CA 95050-4169
Re: K013521
Trade/Device Name: GEMINI PET-CT Device Combined PET/CT System Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: 90 KPS Product Code: 90 JAK Dated: October 22, 2001 Received: October 23, 2001
Dear Dr. Kwan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## NOV 0 7 2001
INDICATIONS FOR USE STATEMENT
510 (k) NUMBER (if KNOWN): _ KO1352 / 1
DEVICE NAME:
GEMINI
ADAC Laboratories SPONSOR NAME:
## INDICATIONS FOR USE:
GEMINI is an imaging system that combines positron emission tomography (PET) and CEMINI is an inflaging system that como produces attenuation and non-attenuation A-1ay computed tonlography. corrected images of the distribution of I D F ranspition images of the PET and CT images tolal body as well as x-ray transmission. Integories combined PET and anatomical all registered and displayed in a etation by trained health professionals. The PET and data at different angles for merpretained by ther as an integrated system or as a stand-alone CT portions on the system can or arouvide CT data suitable for use in attenuation correction.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use (Per 21 CFR 801.109) | <span style="text-decoration: overline;">✓</span> |
|---------------------------------------|---------------------------------------------------|
|---------------------------------------|---------------------------------------------------|
OR
| Over-The-Counter-Use (Optional Format 1-2-96) | _________________ |
|-----------------------------------------------|-------------------|
|-----------------------------------------------|-------------------|
✓
Nancy C brigdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.