The SK YLight Imaging System is intended to produce images depicting the anatomical distributions of single photon and positron emitting radioisotopes within the human body for interpretation by medical personnel.
Device Story
SKYLight Imaging System is an open-frame gamma camera for nuclear medicine imaging; utilizes EPIC HP+ detectors to capture single photon and positron emissions. System includes gantry, two detector arms, patient imaging tables, and acquisition computer. Detectors move in longitudinal, vertical, and rotational axes to perform circular/non-circular ECT, SPECT, and planar studies. Modified version adds capability for simultaneous single-head acquisition on two separate patients. Operated by medical personnel in clinical environments; interfaces with Pegasys workstation for image processing, database management, and archiving. Provides anatomical distribution images to assist clinicians in diagnostic decision-making; supports various nuclear medicine procedures including bone, lung, and renal scans.
Clinical Evidence
No clinical data provided; substantial equivalence is based on identical technological characteristics and intended use to the predicate device.
Technological Characteristics
Open-frame gamma camera system; dual-detector configuration; EPIC HP+ detectors; EPIC collimators or MCD shields. Supports circular/non-circular orbits via 4-axis motion (X, Y, Z, rotation). Connectivity via acquisition computer to Pegasys workstation. Software-controlled gantry motions and image acquisition.
Indications for Use
Indicated for patients requiring nuclear medicine imaging, including SPECT, static, dynamic, gated, total body, and coincidence studies, to visualize radioisotope distribution. Used for clinical procedures such as bone, lung, renal, gastric, hepatobiliary, thyroid, and GI studies. For use by medical personnel in a clinical setting.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{0}------------------------------------------------
K013496
## 510(k) SUMMARY
| A. | Submitted By:<br>ADAC Laboratories<br>540 Alder Dr.<br>Milpitas, CA 95035 | Contact: David Kolesar<br>Tel: (408) 468-3455<br>Fax: (408) 468-3050 |
|----|----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| B. | Device Trade Name:<br>Common Name:<br>Classification Name:<br>Device Class:<br>Product Code: | SKYLight Imaging System<br>Gamma Camera Systems<br>Emission Computed Tomography System<br>21 CFR 892.1200, Class II<br>90 KPS |
| C. | Date prepared: | September 28, 2001 |
| D. | Predicate Device: | SKYLight Imaging System (K000908) |
| E. | Intended Use: | |
The SK YLight Imaging System is intended to produce images depicting the anatomical distributions of single photon and positron emitting radioisotopes within the human body for interpretation by medical personnel.
#### F. Device Description:
The modified SKYLight Imaging System offers all the features of the existing SKYLight Imaging System (K000908) while adding an optional feature that provides the used option or ability to acquire a single head acquisition studies on two separate but simultaneous planar patients. The SKYLight is designed to provide extended imaging functionality relative to a ring style gantry. The SKYLight was designed for single or dual detector nuclear imaging accommodating a broad range of emission computed tomography (ECT) studies. The device includes a gantry frame, two detector arms (with detectors), a collimator storage structure with an acquisition computer unit, a patient imaging table, and a remote hand controller. The modified SKYLight is capable of accommodating two separate patient tables. The gantry is "open" so that a high degree of imaging flexibility is available to image patients sitting, standing or lying down, with or without the included patient imaging table. The patient imaging tables are mechanized to allow for patient loading access and then raised to an imaging height. The tables do not move during imaging since the gantry is flexible enough to perform all required motions for noncircular orbits. The imaging pallet includes removable arm, leg, breast, and headrest supports for patient positioning during studies that require support.
SK YLight is an 'open frame' imaging system, the 'open frame' consisting of an overhead rectangular structure that is supported by 4 (or more) columns that are bolted to the imaging room floor. A detector arm support structure is mounted onto the open frame which allows for the detector arm support structure (and both detector arms) to move in the longitudinal motion (X-axis). Arms that can telescope up and down (Z-axis) support the detectors. These detector arms are supported by a structure that allows for each detector
{1}------------------------------------------------
arm to move toward or away from each other (Y-axis). Each detector has the ability to rotate and acquire images independently
The combination of these four motions (Z, Y, X, and detector rotate), permit the detectors to perform all standard non-circular and circular ECT studies in both the 90-degree and 180-degree relative position. The SkyLight uses the EPIC HP+ detectors. The detectors use standard EPIC collimators, or MCD shields, which may be exchanged using an automatic exchange unit..
The modified SKYLight acquisition computer uses the same acquisition software (subsystem) as existing SKYLight to acquire ECT imaging studies and interfaces with a Pegasys computer system (workstation). The acquisition CPU is used to install the camera software and to archive and restore acquisition parameters and correction tables. The acquisition software is used to complete pre-programmed gantry motions for setup of imaging studies and exchange of collimators; to perform quality assurance tests on the SK YLight system; and to program image parameters and gantry for clinical studies.
When using either a single detector or dual detectors placed in a relative 90-degree or relative 180-degree positions (as study appropriate), SKYLight can be used to perform static, dynamic, gated, total body, circular-orbit and non-circular orbit SPECT studies, coincidence studies, gated SPECT (circular and non-circular) studies, computerprogrammed protocol strings, and reference scans (dual detectors only). SPECT and total body acquisitions are routinely acquired with two detectors. There are some planar procedures such as bone statics and lung scan also use two detectors. There are many additional nuclear medicine procedures that only use one detector at a time. These single detector procedures are typically renal, gastric emptying, hepatobiliary, flow studies, GI bleed, thyroid, and delayed static views. This acquisition sub-system interfaces with a Pegasys computer system for entry into the database. The Pegasys workstation also includes software used for image processing, database utilities, and archiving utilities.
#### G. Technological Comparison:
The modified SK YLight Imaging System and existing SKYLight Imaging System have identical intended use and indications for use. The modified SKYLight and the existing SKYLight are technologically identical. They share the identical mechanical and electrical components. No changes or modifications have been made to the mechanical or electrical components. All the features provided on the existing SKYLight are provided on the modified SKYLight. The only change has been to the acquisition software which provides the modified SKYLight Imaging System an optional feature that allows the user the to acquire a single head acquisition on two separate but simultaneous planar patients.
#### H. Conclusion
The modified SKYLight Imaging System is substantially equivalent to the predicate device SK YLight based upon identical indications for use, technological comparison and overall system performance.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three parallel lines that curve upwards and to the right, resembling a stylized bird or wing. The caduceus is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 0 6 2001
ADAC Laboratories % Michael Kwan, Ph.D. Underwriters Laboratories, Inc. 1655 Scott Boulevard SANTA CLARA CA 95050-4169 Re: K013496
Trade/Device Name: SKYLight Imaging System Gamma Camera System Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: 90 KPS Dated: October 19, 2001 Received: October 22, 2001
Dear Dr. Kwan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### NOV 0 6 2001 INDICATIONS FOR USE STATEMENT
KO13496 510 (k) NUMBER (IF KNOWN): _
DEVICE NAME:
SKYLight Imaging System
ADAC Laboratories SPONSOR NAME:
## INDICATIONS FOR USE:
The SK Y Light Imaging System is intended to produce images depicting the anatomical I the UN I Light ining. Boton and positron emitting radioisotopes within the body for interpretation by medical personnel.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter-Use_ (Optional Format 1-2-96)
David G. Syverson
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.