HSG CATHETER SET, 5F MODEL RGS0947, HSG CATHETER SET, 7F, MODEL R65-948
K013491 · A & A Medical, Inc. · LKF · Nov 19, 2001 · Obstetrics/Gynecology
Device Facts
Record ID
K013491
Device Name
HSG CATHETER SET, 5F MODEL RGS0947, HSG CATHETER SET, 7F, MODEL R65-948
Applicant
A & A Medical, Inc.
Product Code
LKF · Obstetrics/Gynecology
Decision Date
Nov 19, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4530
Device Class
Class 2
Indications for Use
The 5F and 7F HSG Catheter set is a device used to fill the uterus with sterile saline to facilitate the ultrasound examination of the uterus
Device Story
HSG Catheter Set (5F and 7F) is a physician-operated device used in hospital settings for hysterosalpingography. The device consists of a double-lumen catheter and a semi-rigid introducing sheath. The catheter features a distal non-latex Kraton balloon (1.5 cc capacity for 5F; 3 cc for 7F) and a side port. The proximal end includes a halkey clamp, female luer lock hub, one-way stopcock, and a 3cc syringe. The device is inserted to fill the uterus with sterile saline, providing contrast for ultrasound imaging. This allows physicians to visualize the uterine cavity, aiding in clinical diagnosis and decision-making regarding uterine health.
Clinical Evidence
No clinical data provided; substantial equivalence is based on descriptive information and comparison of technological characteristics.
Technological Characteristics
Double-lumen catheter and semi-rigid introducing sheath; 5F and 7F diameters; 30 cm length; non-latex Kraton balloon; female luer lock hub; one-way stopcock; 3cc syringe. Sterilization via Ethylene Oxide.
Indications for Use
Indicated for filling the uterus with sterile saline to facilitate ultrasound examination of the uterus in patients undergoing such procedures.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
Predicate Devices
HSG Catheter Set for Hysterosalpingography (Lyco Enterprises, Inc.)
Submission Summary (Full Text)
{0}------------------------------------------------
### 510(k) Summary NOV 1 9 2001 As Required by 21 section 807.92 ( c )
- 1-Submitter Name: A & A Medical, Inc.
- 2-Address: 9370 Industrial Trace
- Alpharetta, GA 30004 3-Phone:
- (770) 343- 8400 4-Fax: (770) 343- 8985
- Jay Mansour 5-Contact Person:
- 6-Date summary prepared: October 17th, 2001
- 7-Device Trade or Proprietary Name: HSG Catheter Set (5F and 7F)
- 8-Device Common or usual name: Catheter for Hysterosalpingography
- 9-Device Classification Name: Cannula, Manipulator/Injector. Uterine
- 10-Substantial Equivalency is claimed against the following device:
- . HSG Catheter Set for Hysterosalpingography, from Lyco Enterprises. Inc.
- 11-Description of the Device:
- The device is to be used by physicians in hospitals
The HSG catheter set consists of two parts: a double lumen catheter and a semi-rigid introducing sheath.
The catheter is available in two different sizes 5 and 7 French diameter with a length of 30 cm. The distal end has a side port and a non-latex Kraton balloon whose capacity is 1.5 cc for 5F and 3cc for 7F.
The proximal end that ends in side port is fitted with a halkey clamp and female luer lock hub. The other channel is fitted with one way stop cock and a 3cc syringe.
#### 12-Intended use of the device:
The 5F and 7F HSG Catheter set is a device used to fill the uterus with sterile saline to facilitate the ultrasound examination of the uterus
#### 13-Safety and Effectiveness of the device:
This device is safe and effective as the other predicate device cited above. This is better expressed in the tabulated comparison (Paragraph 14 below)
{1}------------------------------------------------
## 14-Summary comparing technological characteristics with other predicate device:
Please find below a tabulated comparison supporting that this device is substantially equivalent to other medical devices in commercial distribution. Also, Equivalency overview chart path is attached.
| FDA file reference number | 510k K943494 |
|-----------------------------------------------------|------------------------------------------------|
| Attachments inside notification<br>submission file | REFER TO TABLE ON PAGE 11 OF 12 FOR<br>DETAILS |
| TECHNOLOGICAL<br>CHARACTERISTICS | Comparison result |
| Indications for use | Identical |
| Target population | Identical |
| Design | Identical |
| Materials | Identical |
| Performance | Identical |
| Sterility | Similar (Ethylene Oxide instead of Gamma) |
| Biocompatibility | Identical |
| Mechanical safety | Identical |
| Chemical safety | Identical |
| Anatomical sites | Identical |
| Human factors | Identical |
| Energy used and/or delivered | Identical |
| Compatibility with environment<br>and other devices | Identical |
| Where used | Identical |
| Standards met | Identical |
| Electrical safety | Identical (not applicable) |
| Thermal safety | Identical (not applicable) |
| Radiation safety | Identical (not applicable) |
{2}------------------------------------------------
510(k) "SUBSTANTIAL EQUIVALENCE" DECISION-MAKING PROCESS (DITAILED)
Image /page/2/Figure/2 description: This image is a flowchart that outlines the process for determining if a new device is substantially equivalent to a marketed device. The flowchart starts with the question, "New Device is Compared to Marketed Device" and proceeds through a series of questions about the device's intended use, technological characteristics, and performance data. The flowchart ends with a determination of whether the new device is "Substantially Equivalent" or "Not Substantially Equivalent".
SIO(k) submissions compare new devices to marketed devices. FDA requests additional inforclarions in the relations big between marketed and "predicate" (pre-Amendments or reclassified post-Amendments) devices is unders.
· This decision is normally based on descriptive information alone, but limited testing information is sometimes required.
*** Data may be in the 310(k), other S10(k)s, the Center's classification files, or the linerature.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around an emblem. The emblem is a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The overall design is simple and conveys a sense of official government affiliation.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# NOV 1 9 2001
Mr. Jay Mansour, MSQA, BE, LA, RAC Quality and Regulatory Manager A & A Medical, Inc. 9370 Industrial Trace ALPHARETTA GA 30004
Re: K013491
Trade/Device Name: HSG Catheter Set (5F and7F) Model R65-947 and R65-948 Regulation Number: None Regulatory Class: Unclassified Product Code: 85 LKF Dated: October 17, 2001 Received: October 22, 2001
Dear Mr. Mansour:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket I ills letter will anow you to ocgin nationing your device of your device to a legally marketed noutication. The I Dri- inding of succiantial for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please IT you desire specific advice for your de of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, 101 questions on the premes . Also, please note the regulation entitled, "Misbranding Other of Comphanes at (2017 077 100 (21 CFR Part 807.97). Other general information on by relefence to promation of the Act may be obtained from the Division of Small Manufacturers, your lesponsibilities under the rice may of its toll-free number (800) 638-2041 or (301) 443-6597 niccilational and Oolisamer : Itsitive.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
NOV 1 9 2001
of Page
510(k) Number (if known): ___K013491
Device Name: HSG Catheter Set (5F & 7F)
Indications For Use:
HSG Catheter Set (5F & 7F) is indicated for use to fill the uterus with sterile saline to facilitate the ultrasound examination of the uterus.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use v (Per 21 CFR 801.109) Over-the-Counter Use _________________________________________________________________________________________________________________________________________________________
(Optional Format 3-10-98)
Ylancy C. Cropton
(Division Sten-Off)
Division of aoducthminal.
and Rediefental Devic
510kl N
K013491
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.