IGG ANTI-NDNA FLUORESCENT TEST SYSTEM; MODEL # 3040G
K013432 · Immuno Concepts, Inc. · KTL · Nov 29, 2001 · Immunology
Device Facts
Record ID
K013432
Device Name
IGG ANTI-NDNA FLUORESCENT TEST SYSTEM; MODEL # 3040G
Applicant
Immuno Concepts, Inc.
Product Code
KTL · Immunology
Decision Date
Nov 29, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5100
Device Class
Class 2
Indications for Use
This is an indirect fluorescent antibody test for screening and semiquantitative detection of IgG anti-nDNA antibody in human serum. This test system is to be used as an aid in the diagnosis of systemic lupus erythematosus.
Device Story
Indirect fluorescent antibody test system for detection of IgG anti-nDNA antibodies in human serum; used as an aid in diagnosing systemic lupus erythematosus. Device utilizes substrate slides with concentric circles to minimize cross-contamination and integrated positive/negative/PBS control wells. Procedure involves manual application of patient serum to slides; no blotting required after washing steps. Results are interpreted by clinical laboratory personnel via fluorescence microscopy. Output is semiquantitative titer value. Device provides diagnostic support for clinicians evaluating patients for systemic lupus erythematosus.
Clinical Evidence
Clinical study compared subject device to predicate using 223 samples (121 clinical, 100 donors, 2 WHO standards). Results: 93.3% relative sensitivity, 100% relative specificity, 100% positive predictive value, 99.0% negative predictive value, 99.1% overall agreement. Reproducibility testing: intra-assay (30 replicates) and inter-assay (15 lots) showed Geometric Coefficient of Variation of 0.78% and 0.82% respectively; all titers within one dilution of median.
Technological Characteristics
Indirect immunofluorescence assay; substrate slide with concentric wells; manual processing; no blotting required. Performance validated against WHO standards. Class II device (21 CFR 866.5100).
Indications for Use
Indicated for screening and semiquantitative detection of IgG anti-nDNA antibodies in human serum as an aid in the diagnosis of systemic lupus erythematosus.
Regulatory Classification
Identification
An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).
Predicate Devices
Crithidia lucilliae DS DNA Kit (Diagnostic Use) (K930987)
Submission Summary (Full Text)
{0}------------------------------------------------
NOV 2 9 2001
Immuno
concepts
*510(k) SUMMARY*
**STORY BOMB KIT!**
---
Date Prepared: November 8, 2001 Contact Person: Eric S. Hoy, Ph.D., SI(ASCP) Name of Device:
- · Trade Name IgG Anti-nDNA Fluorescent Test System
- · Common Name IgG Anti-nDNA Fluorescent Test System
- Classification Name Anti-DNA Indirect Immunofluorescent Solid Phase (21 CFR 866.5100)
Legally marketed device with which this device has been shown to be equivalent: "Crithidia lucilliae DS DNA Kit (Diagnostic Use)" distributed by The Binding Site, Ltd., Los Angeles, CA 90064 (K930987, 8/10/93).
### Description:
This is an indirect fluorescent antibody test for screening and semiquantitative detection of IgG anti-nDNA antibody in human serum.
### Intended Use:
This is an indirect fluorescent antibody test for screening and semiquantitative detection of IgG anti-nDNA antibody in human serum. This test system is to be used as an aid in the diagnosis of systemic lupus erythematosus.
Summary of Technological Characteristics Compared to the Predicate Device: Technologically, this device is similar to the predicate device with the following exceptions:
- 1. The substrate slide wells concentric circles which decrease potential cross contamination of patient sera.
- 2. The slides include three additional wells designated positive (+), negative (-), and phosphate buffered saline (PBS) to insure proper performance of controls on slides.
- 3. Unlike the predicate device and other similar kits, the Immuno Concepts slides are not blotted after each washing step.
### Description of Laboratory Data That Indicate Substantial Equivalence:
The Immuno Concepts IgG Anti-nDNA Fluorescent Test System was compared to the "Crithidia lucilliae DS DNA Kit´ (Diagnostic Use)" distributed by The Binding Site, Ltd., Los Angeles, CA 90064 (K930987, 8/10/93). The population studied consisted of 121 samples which were submitted to clinical laboratories for anti-nDNA testing, 100 blood donors, and 2 WHO standards that are known to contain ant-nDNA antibodies. All samples were tested in parallel on the predicate device and the subject device. Based on this comparison, the following data were obtained:
{1}------------------------------------------------
#### Predicate IgG anti-nDNA Test Positive Negative
| Immuno<br>Concepts | Positive | 28 | 0 |
|--------------------|----------|----|-----|
| IgG anti-nDNA | Negative | 2 | 193 |
These data yield the following statistics: relative sensitivity, 93.3%;
100%; mediation 1000 memitive mrodictive value. 100%; negative These data yield the following scaller predictive value, 100%; negative
relative specificity, 100%; positive predictive value, 100 % relative specificity, 1000, poblem. I
predictive value, 99.0%; and overall agreement, 99.1%
One of the two "false negative" samples was negative for antinuclear one of the two "false negative" samples was negation in most clinical antibodies by indirect immunoriuorescence using hip work.
have met the screening criteria for anti-nDNA testing in most clinical laboratories.
To determine intra-assay reproducibility, a single anti-mDNA positive serum To determine intra-assay a single technologist in a single run. The was tested in 30 replicates by a single technologics ha december in Coefficient
median and mode values for these data were 1:160, with a Getemining median and mode values for these acta were fr.ov. Metro by examining
of Variation of 0.78%. Inter-assay variability was it was run in of Variation of 0.78%. Incel-assay variability namma it was run in
the titer values for a single anti-mDNA positive serum as it was data the titer values for a single and mode walves for these data were
fifteen different lots of kirs officion of Variation of O.82%. In both the fifteen different lots of kits. The median and mode variation of 0.828. In both the or also 1:16, with a Geomeric Coefficient of Variation of Vol. 1:16.
inter-assay and intra-assay tests, all titer values fell within plus or minus
tion intra-assay tests, all ti one dilution of the median.
In accordance with 21 CFR 807.92(b)(3), we conclude from these data that the present device is substantially equivalent to the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a black and white logo for the Department of Health & Human Services. The logo features a stylized eagle with three curved lines representing its wings. The words "DEPARTMENT OF HEALTH & HL" are arranged in a semi-circle around the left side of the eagle, and the letters "SA" are on the right side.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
NOV 2 9 2001
Eric S. Hoy, Ph.D., SI(ASCP) Chief, Scientific Officer Immuno Concepts Incorporated 9779 "D" Business Park Drive Sacramento, CA 95827
Re: k013432 Trade/Device Name: IgG Anti-nDNA Fluorescent Test System Regulation Number: 21 CFR 866.5100 Regulation Name: Antinuclear antibody immunological test system Regulatory Class: Class II Product Code: KTL Dated: October 1, 2001 Received: October 16, 2001
Dear Dr. Hoy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket I has letter will and my of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# 510(k) number (if known): ко13432
IgG Anti-nDNA Fluorescent Test System Device Name:
## Indications for use:
This is an indirect fluorescent antibody test for screening and semiquantitative detection of IgG anti-nDNA antibody in human serum. This quantitacive dececcion of igo anor head mo.
test system is to be used as an aid in the diagnosis of systemic lupus erythematosus .
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sousan S, Altule
510(k) Number k013432
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.