← Product Code [ITI](/productcode/ITI) · K013361

# WIN-2 TANGO ELITE NP (K013361)

_Durable Medical Equipment, Inc. · ITI · Feb 26, 2002 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K013361

## Device Facts

- **Applicant:** Durable Medical Equipment, Inc.
- **Product Code:** [ITI](/productcode/ITI.md)
- **Decision Date:** Feb 26, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3860
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine

## Indications for Use

The intended use of the WIN-2 Tango Elite is to provide mobility for a person restricted to a sitting position.

## Device Story

WIN-2 Tango Elite NP is a battery-powered wheelchair for indoor/outdoor use; features two 0.25 horsepower motors and a joystick-controlled unit. Designed for single-rider operation; allows user to control speed and direction. Device is transportable via disassembly; includes battery charger accessory. Operated by the patient; intended for mobility support. Output is physical movement of the chair. No clinical data or complex algorithms involved.

## Clinical Evidence

No clinical data submitted. Evidence consists of bench testing, including ANSI/RESNA Wheelchairs Vol. 2 standards, electromagnetic compatibility, and functional testing of upholstery and battery charger.

## Technological Characteristics

Powered wheelchair; battery-operated; two 0.25 HP motors; joystick controller. Dimensions/form factor: transportable/disassemblable. Standards: ANSI/RESNA Wheelchairs Vol. 2. No software algorithm or complex connectivity described.

## Regulatory Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- WIN-1 Tango ([K000204](/device/K000204.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K013361

WIN-2TM Tango Elite NP 510(k) Modification Page 4

FEB 2 6 2002

## 510(k) SUMMARY

Duracare Medical Equipment, L.C. Submitter's name: 6182 Idlewild St., Fort Myers, FL 33912 (941) 791-4000

Date summary prepared:

October 5, 2001

Device name:

Proprietary name: Common or usual name: Classification name:

WIN-2TM Tango Elite NP Power chair Powered wheelchair, Class II, 21 CFR 890.3860.

Legally marketed device for substantial equivalence comparison:

WIN-1 Tango submitted by Duracare Medical Equipment, L.C. and cleared for marketing under 510(k) *K000204.

Description of the device:

The WIN-2 Tango Elite NP is a powered wheelchair for use both indoors and outdoors. It is battery powered, has two 0.25 horsepower motors, and a controller unit with joystick. It is designed for a single rider who controls the speed and direction of the chair's movement with the joystick. It can be disassembled for transport, and a battery charger is supplied as an accessory. The only difference between the WIN-1 and the WIN-2 wheelchairs is the controller.

Intended use of device:

The intended use of the WIN-2 Tango Elite NP is to provide mobility for a person restricted to a sitting position.

Technological characteristics:

The device features of the WIN-1 and the WIN-2 are identical except for the difference in controllers. The use parameters vary only in minor ways such as maximum speeds.

Testing conducted:

Results were reported from testing of relevant portions of the ANSI/RESNA Wheelchairs Vol. 2 Standard. Additional tests were conducted on electromagnetic compatibility, functional testing reported on the Summary Matrix, the upholstery material, and the battery charger.

Performance testing:

Comparative performance testing and clinical evaluations were not submitted as part of this 510(k).

4

{1}------------------------------------------------

## SECTION 3 - INTENDED USE

Intended use of device:

The intended use of the WIN-2 Tango Elite is to provide mobility for a person restricted to a sitting position. This is consistent with the intended use of other powered wheelchairs.

Intended use of predicate device:

The intended use of the WIN-1 Tango is to provide mobility for a person restricted to a sitting position.

## Comparison:

These two models have the identical intended use statements.

Labeling:

The intended use of the WIN-2 is implied in the User Manual by the use of phrases such as "allowing you the freedom to go places you have always wanted to go" and "designed for maximum maneuverability even in narrow and tight spaces." A draft copy of the Users Manual can be found in Appendix II.

A copy of the Users Manual for the WIN-1 model can be found in Appendix VI. The intended use sheet from K*000204 is provided in Appendix IV.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of a bird or eagle in flight, with three distinct lines forming the body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the bird image.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 2 6 2002

Durable Medical Equipment, Inc. c/o Mr. Robert S. McQuate R. S. McQuate & Associates, Inc. 3636 E. Columbine Drive Phoenix, AZ 85032

Re: K013361

Trade/Device Name: WIN-2™ Tango Elite NP Regulation Number: 890.3860 Regulation Name: Powered wheelchair Regulatory Class: II Product Code: ITI Dated: February 4, 2002 Received: February 5, 2002

Dear Mr. McQuate:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. Robert S. McQuate

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

L. Merle N. Millerson

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

WIN-2™ Tango Elite NP 510(k) Modification Page 3

## Indications for Use Statement

510(k) Number (if known): __KO1336 |

WIN-2™ Tango Elite NP Device name:

Indications for Use:

To provide mobility for a person restricted to a sitting position.

(Please do not write below this line)

Concurrence of CDRH, Office of Device Evaluation (ODE) ion Sign-Off) Division of General, Restorative and Neurological Devices 510(k) Number _ OR Over-The-Counter Use Prescription Use _

---

**Source:** [https://fda-staging.innolitics.com/device/K013361](https://fda-staging.innolitics.com/device/K013361)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
