K013346 · W. L. Gore & Associates, Inc. · LYC · Dec 19, 2001 · Dental
Device Facts
Record ID
K013346
Device Name
GORE DRAPEABLE ST/LT REGENERATIVE MEMBRANE
Applicant
W. L. Gore & Associates, Inc.
Product Code
LYC · Dental
Decision Date
Dec 19, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3930
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
GORE DRAPEABLE ST and LT Regenerative Membranes are bioabsorbable, implantable materials intended to aid in the healing of periodontal and bone defects. They may also be used as a membrane for bone graft containment. When used over a bone graft, GORE DRAPEABLE ST and LT Regenerative Membranes provide a stable barrier to prevent migration of bone graft material and provides a favorable environment for bone regeneration. When used in the treatment of non-spacemaking defects, GORE DRAPEABLE ST and LT Regenerative Membranes should be used in combination with spacemaking bone graft materials (e.g., autogenous bone, bone substitutes).
Device Story
Bioabsorbable, implantable membrane; composed of synthetic resorbable glycolide and trimethylene carbonate copolymer fiber. Surgically placed by clinicians beneath mucoperiosteum in oral cavity. Acts as physical barrier to inhibit epithelial downgrowth; stabilizes wound/blood clot; allows tissue integration; prevents bone graft migration. Used in conjunction with bone grafting/filling materials for spacemaking in non-spacemaking defects. Benefits patient by providing favorable environment for bone and periodontal ligament regeneration.
Clinical Evidence
Preclinical studies in critical size defects of rat mandible demonstrated protection of regenerative environment and tissue tolerance suitable for bone regeneration. Functional integrity and absorption time consistent with clinical goals.
Technological Characteristics
Synthetic resorbable glycolide and trimethylene carbonate copolymer fiber; porous structure; bioabsorbable; cell occlusive; designed for tissue integration.
Indications for Use
Indicated for patients requiring regenerative healing of periodontal and bone defects in the oral cavity. Used as a barrier membrane to aid healing and contain bone graft materials. For non-spacemaking defects, must be used with spacemaking bone graft materials.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
GORE RESOLUT XT Regenerative Membrane
GORE OSSEOQUEST Regenerative Membrane
Submission Summary (Full Text)
{0}------------------------------------------------
# 2 013346
# DEC 1 9 2001
## TAB J
## 510(K) Summary of Safety and Effectiveness
| 1. Applicant Name, Address: | W.L. Gore & Associates, Inc. 3450 W. Kiltie Lane P.O. Box 900 Flagstaff, AZ 86002-0900 |
|-----------------------------|----------------------------------------------------------------------------------------|
| Contact Person: | Jacqueline Kalbach (928)864-3731 |
| Date of Summary: | October 5, 2001 |
| 2. Classification Name: | Resorbable Barrier Membrane |
| Common or Usual Name: | Regenerative Material |
| Proprietary Name: | To be determined |
#### 3. Device Description:
GORE DRAPEABLE ST and LT Regenerative Membranes are composed of a porous structure of synthetic resorbable glycolide and trimethylene carbonate copolymer fiber. The fiber web is designed to allow attachment to surrounding tissues, stabilize the wound and blood clot, inhibit epithelial downgrowth, and isolate the healing defect during the initial phases of wound healing. GORE DRAPEABLE ST and LT Regenerative Membranes have been designed to act in accordance with the accepted principles of wound healing and regenerative therapy. Specifically, the membranes are designed to be biocompatible, separate tissue (cell occlusive), clinically manageable, and allow for tissue integration. The fifth requirement for regenerative therapy is spacemaking, and bone grafting or bone filling materials may be used in conjunction with the membranes to assist in providing the space necessary for regenerative healing. GORE DRAPEABLE ST and LT Regenerative Membranes are surgically placed beneath the mucoperiosteum to aid in the regenerative healing of bone and periodontal ligament defects of the oral cavity or, when placed over bone graft material, to prevent graft material migration.
{1}------------------------------------------------
#### Intended Use: 4.
GORE DRAPEABLE ST and LT Regenerative Membranes are bioabsorbable, implantable materials intended to aid in the healing of periodontal and bone defects. They may also be used as a membrane for bone graft containment. When used over a bone graft, GORE DRAPEABLE ST and LT Regenerative Membranes provide a stable barrier to prevent migration of bone graft material and provides a favorable environment for bone regeneration. gran material the treatment of non-spacemaking defects, GORE DRAPEABLE ST and LT Regenerative Membranes should be used in combination with spacemaking bone graft materials (e.g., autogenous bone, bone substitutes).
#### Technological Characteristics: 5.
GORE DRAPEABLE ST and LT Regenerative Membranes have been designed to act in accordance with the accepted principles of wound healing and regenerative therapy. Specifically, the membranes are designed to be biocompatible, separate tissue (cell occlusive), clinically manageable, and allow for tissue integration. The fifth requirement for regenerative therapy is spacemaking, and bone grafting or bone filling materials may be used to provide the space necessary for regeneration.
#### Assessment of Performance Data: 6.
Preclinical studies conducted in critical size defects of the rat mandible indicate that GORE DRAPEABLE ST and LT Regenerative Membranes provide protection of the regenerative environment and tissue tolerance suitable for regeneration of bone. The functional integrity and time to absorption of this material are consistent with the clinical goals of regenerative therapy.
#### 7. Conclusion:
The modified GORE DRAPEABLE ST and LT Regenerative Membranes have an indication statement that is similar to the current GORE RESOLUT XT and GORE OSSEOQUEST Regenerative Membranes. Their intended use is the same: the modified GORE DRAPEABLE ST and LT Regenerative Membranes act as a barrier intended to allow controlled tissue regeneration and thus allow repair of periodontal and bone defects in the oral cavity. They may also be used as membranes for bone graft containment, and bone grafting or bone filling materials may be used in conjunction with the membranes to assist in providing the space necessary for regenerative healing. Thus, the modified GORE DRAPEABLE ST and LT Regenerative Membranes are substantially equivalent to the current GORE RESOLUT XT and GORE OSSEOQUEST Regenerative Membranes. The indication statement has been modified, and the added
{2}------------------------------------------------
wording is consistent with that of other legally marketed resorbable regenerative membranes that also recommend the use of bone filling materials when nonspacemaking defects are treated.
{3}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 9 2001
Ms. Jacqueline Kalbach Regulatory Affairs Associate W. L. Gore & Associates Medical Products Division 3450 Kiltie Lane P.O. Box 500 Flaggstaff, Arizona 86002-0500
Re: K013346
Trade/Device Name: Gore Drapeable St/Lt Regenerative Membrane Regulation Number: None Regulation Name: Regenerative Material Regulatory Class: Unclassified Product Code: LYC Dated: October 5, 2001 Received: October 9, 2001
Dear Ms. Kalbach:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave rottewou your betermined the device is substantially equivalent (for the relerenced above and have acteriminoure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Include comments, or to devices that have been reclassified in accordance with the provisions of Amondinente, or to arrestic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include controls providents or an al registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 Mrt), it may of such of Federal Regulations, Title 21, Parts 800 to 898. In your device can be roundsh further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2 - Ms. Kalbach
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Pat 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Timothy . Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
# KO13346
Page __ 1_ of __ 1_
510(k) Number (if known) ____________________________________________________________________________________________________________________________________________________
Device Name: GORE DRAPEABLE ST and LT Regenerative Membranes
### INDICATIONS FOR USE:
GORE DRAPEABLE ST and LT Regenerative Membranes are bioabsorbable, GURE DRAF EADLE OF and ET Regoneraling of periodontal and bone
implantable materials intended to aid in the healing of periodontal and bone implantable materials intended to and membrane for bone graft containment.
defects. They may also be used as a membrane of a contralian delects. They may also be assult as a moments of and LT Regenerative When used over a bone grant, OORLE over in migration of bone graft material Membranes provide a stable barner to proveneration. When used in the and provides a favorable environment . GORE DRAPEABLE ST and LT
treatment of non-spacemaking defects, GORE DRAPEABLE SC and LT treathent of non-spacemaking donoter of one ination with spacemaking bone Regencrative Monibrancemous bone, bone substitutes).
## (Please do not write below this line - continue on another PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (per 21 CFR 801.109)
OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Susan Ryan
IDivision Sign-Off) Prision of Dental, Infection Control, are General Hospital Devices
Number __________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.