RELISA MPO-ANCA TEST SYSTEM FOR ANTIBODIES TO MYELOPEROXIDASE, MODEL # 7096-15
K013285 · Immuno Concepts, Inc. · MOB · Nov 14, 2001 · Immunology
Device Facts
Record ID
K013285
Device Name
RELISA MPO-ANCA TEST SYSTEM FOR ANTIBODIES TO MYELOPEROXIDASE, MODEL # 7096-15
Applicant
Immuno Concepts, Inc.
Product Code
MOB · Immunology
Decision Date
Nov 14, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5660
Device Class
Class 2
Indications for Use
This is an enzyme immunoassay test system for the detection of antibodies to myeloperoxidase (MPO) in human serum. This test system is to be used as an aid in the detection of antibodies associated with microscopic polyangiitis, idiopathic necrotizing and crescentic glomerulonephritis, and other vasculitides.
Device Story
Enzyme immunoassay (EIA) test system for detection of anti-myeloperoxidase (MPO) antibodies in human serum; used in clinical laboratory settings. Device utilizes microwell strips to capture antibodies; results interpreted by healthcare providers to aid in diagnosis of vasculitides including microscopic polyangiitis and glomerulonephritis. Subject device features individual pouch packaging for 8-well strips to improve stability and uses five calibrator sera compared to six in the predicate. Output provides qualitative or semi-quantitative detection of MPO antibodies, assisting clinicians in differential diagnosis of autoimmune vasculitis.
Clinical Evidence
Clinical study compared subject device to predicate using 758 human serum samples (including patients with Churg-Strauss, vasculitis, Wegener's granulomatosis, and healthy donors). Results: 96.4% relative sensitivity, 94.6% relative specificity, and 94.7% overall agreement. 38 false positives were noted, with 28 showing P-ANCA patterns by immunofluorescence.
Technological Characteristics
Enzyme immunoassay (EIA) test system. Microwell plate format. Features individual foil pouch packaging for 8-well strips to maintain stability. Uses five calibrator sera. Class II device (21 CFR 866.5660).
Indications for Use
Indicated for the detection of anti-myeloperoxidase (MPO) antibodies in human serum as an aid in diagnosing microscopic polyangiitis, idiopathic necrotizing and crescentic glomerulonephritis, and other vasculitides.
Regulatory Classification
Identification
A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Immuno Concepts. The logo is black and white and features a stylized "IC" with the words "Immuno" and "Concepts" stacked on top of each other. The "I" in "IC" is represented by two horizontal lines, and the "C" is represented by a circle with three smaller circles inside.
NOV 1 4 2001
Date Prepared: July 23, 2001 Contact Person: Eric S. Hoy, Ph.D., SI(ASCP) Name of Device:
- · Trade Name RELISA® MPO-ANCA Test System for Antibodies to Myeloperoxidase
- · Common Name Anti-MPO Antibody Enzyme Immunoassay Test Kit
- · Classification Name · Test System, Antineutrophil Cytoplasmic Antibodies (ANCA) (21 CFR 866.5660)
Legally marketed device with which this device has been shown to be equivalent: Bindazyme™ Anti-MPO Enzyme Immunoassay Kit manufactured by The Binding Site, Ltd., Birmingham, England, UK, K981030.
### Description:
This is an enzyme immunoassay for the detection of antibodies to myeloperoxidase (MPO) in human serum.
### Intended Use:
This is an enzyme immunoassay test system for the detection of antibodies to myeloperoxidase (MPO) in human serum. This test system is to be used as an aid in the detection of antibodies associated with microscopic polyangiitis, idiopathic necrotizing and crescentic glomerulonephritis, and other vasculitides.
### Summary of Technological Characteristics Compared to the Predicate Device:
Technologically, this device is similar to the predicate device with the following exceptions:
a) The predicate device has all 96 microwells sealed in a single pouch. The present device has each eight well strip of microwells sealed in an individual pouch, so that unused wells remain in the pouches, and maintain at least 12 months stability.
b) The predicate device uses six calibrator sera, the present device uses five calibrator sera.
### Description of Laboratory Data That Indicate Substantial Equivalence:
The Immuno Concepts RELISA® MPO Test System was compared to the Bindazyme™ Anti-MPO Enzyme Immunoassay Kit manufactured by The Binding Site, Ltd., Birmingham, England, UK. (K981030). The population studied consisted of 208 samples which were submitted to clinical laboratories for ANCA, MPO and PR3 testing, but without specific diagnoses, 7 samples which were from pateints with a diagnosis of Churg-Strauss syndrome, 18 patients with a diagnosis of vasculitis, 25 patients with a diagnosis of Wegener's granulomatosis, 261 All samples from male blood donors, and 239 samples from female blood donors. samples were tested in parallel on the predicate device and the subject device. Based on this comparison, the following data were obtained:
{1}------------------------------------------------
### Bindazyme™ Antl-MPO Enzyme lmmunoassay Kit Negative Positive
| Immuno Concepts | Positive | 53 | 38 |
|---------------------------------|----------|----|-----|
| RELISA® MPO-ANCA<br>Test System | Negative | 2 | 665 |
These data yield the following statistics: relative sensitivity, 96.4%; relative specificity, 94.6%; and overall agreement, 94.7%
Among the 38 "false positive" samples, 28 showed a P-ANCA pattern by Among the 56 Traise postere banged, by immunofluorescence, 3 showed immanoridorescence, 3 Biowed a Chilon parceace, and 4 did not demonstrate any immunofluorescent pattern.
In accordance with 21 CFR 807.92(b)(3), we conclude from these data that the present device is substantially equivalent to the predicate device.
{2}------------------------------------------------
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Eric Hoy, Ph.D., SI (ASCP) NOV 1 4 2001 Chief Scientific Officer, Regulatory Affairs Immuno Concepts, Inc. 2280 Springlake Road, Suite 106 Dallas, Texas 75234
Re: K013285
Trade/Device Name: Immuno Concepts RELISA® MPO-ANCA Test System for Antibodies to Myeloperoxidase Regulation Number: 21 CFR § 866.5660 Regulation Name: ANCA Test System Regulatory Class: II Product Code: MOB Dated: September 27, 2001 Received: October 2, 2001
Dear Dr. Hoy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
## Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Autman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
NOV 1 4 2001
Page 1 of 1
510(k) number (if known):
Device Name: RELISA® MPO-ANCA Test System for Antibodies to Myeloperoxidase
# Indications for use:
This is an enzyme immunoassay test system for the detection of antibodies to myeloperoxidase (MPO) in human serum. This test system is to be used as an aid in the detection of antibodies associated with microscopic polyangiitis, idiopathic necrotizing and crescentic glomerulonephritis, and other vasculitides.
Sousan Saltaie
510(k) Numb
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use J (Per 21 CFR 801.109)
OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.