K013257 · Nuvasive, Inc. · HAW · Oct 30, 2001 · Neurology
Device Facts
Record ID
K013257
Device Name
HEMI-ARC SURGICAL NAVIGATOR
Applicant
Nuvasive, Inc.
Product Code
HAW · Neurology
Decision Date
Oct 30, 2001
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.4560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NuVasive™ Guided Spinal Arthroscopy System consists of a Surgical Navigator with adjustable cannula guides, and various percutaneous surgical cannulae and instrumentation. This System is intended to assist in gaining controlled percutaneous access to, and visualization of, the spinal nerves, foramina, vertebra, intervertebral disc, and surrounding tissues of the spine via uniportal or biportal posterior, lateral, or posterolateral approach, where anatomical restrictions safely permit. It is intended for use under real-time, or near real-time (i.e., successive static image) radiographic visualization via image-intensified Carm fluoroscopy. The System is designed for use with NuVasive™ spinal arthroscopes and percutaneous surgical instrumentation, but may be used with other arthroscopes and instrumentation having compatible diameter and length, and which are indicated for use in the spine.
Device Story
Hemi-Arc Surgical Navigator acts as instrument holder/guide for spinal arthroscopy; provides controlled percutaneous access to spinal anatomy. Device utilizes real-time or near real-time C-arm fluoroscopy for alignment; adjustable cannula guides allow precise positioning of surgical instrumentation. Used in clinical settings by surgeons; assists in navigating spinal nerves, foramina, vertebrae, and discs. Output is physical guidance for instruments; helps avoid damage to surrounding tissues like nerves or dura. Benefits patient through controlled, minimally invasive access.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Mechanical instrument holder/guide frame; adjustable in multiple planes of motion; compatible with standard spinal arthroscopes and percutaneous instrumentation. No software or electronic components described.
Indications for Use
Indicated for patients requiring percutaneous access to and visualization of spinal nerves, foramina, vertebrae, intervertebral discs, and surrounding spinal tissues via uniportal or biportal posterior, lateral, or posterolateral approaches.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Predicate Devices
Surgical Guide Frame (NuVasive, Inc.)
Submission Summary (Full Text)
{0}------------------------------------------------
OCT 3 0 2001
# NuVasive™, Inc.
KO13257 Special 510(k) Premarket Notification Hemi-Arc™ Surgical Navigator
# VII. 510(k) Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular \$807.92, the following summary of information is provided:
#### Submitted by: A.
Laetitia Bernard Manager of Regulatory Affairs NuVasive™, Incorporated 10065 Old Grove Road San Diego, California 92131 Telephone: (858) 271-7070 Telefacsimile: (858) 271-7101
#### B. Device Name
Trade Name: Hemi-Arc™ Surgical Navigator
Common or Usual Name: Arthroscope, Stereotaxic Frame
Classification Name: Arthroscope, Stereotaxic Instrument
#### C. Predicate Devices
The subject Hemi-Arc™ Surgical Navigator is substantially equivalent to the Surgical Guide Frame currently manufactured and distributed commercially in the U.S. by NuVasive™, Inc.
#### D. Device Description
The Hemi-Arc™ Surgical Navigator essentially serves as an instrument holder and guide which is intended to be aligned with the operative site using real time, or near real-time (i.e. success static images) image-intensified C-arm fluoroscopy. The Hemi-ArcTM guide frame is adjustable in several planes of motion, and their fluoroscopic alignment with spinal anatomical features provides the path of entry of subsequent arthroscopic instrumentation to be precisely controlled, thereby avoiding damage to surrounding tissues such as nerves or dura.
{1}------------------------------------------------
# Intended Use E.
The NuVasive™ Guided Spinal Arthroscopy System consists of a Surgical Navigator with adjustable cannula guides, and various percutaneous surgical cannulae and instrumentation. This System is intended to assist in gaining controlled percutaneous access to, and visualization of, the spinal nerves, foramina, vertebra, intervertebral disc, and surrounding tissues of the spine via uniportal or biportal posterior, lateral, or posterolateral approach, where anatomical restrictions safely permit. It is intended for use under real-time, or near real-time (i.e., successive static image) radiographic visualization via image-intensified Carm fluoroscopy. The System is designed for use with NuVasive™ spinal arthroscopes and percutaneous surgical instrumentation, but may be used with other arthroscopes and instrumentation having compatible diameter and length, and which are indicated for use in the spine.
## Comparison to Predicate Devices F.
The subject device has indications for use identical to those of its predicate, and employs the same principles of operation for alignment with the spinal operative site using real time, or near real-time (i.e. success static images) image-intensified C-arm fluoroscopy to provide access to subsequent arthroscopic instrumentation. Due to this equivalency, the device raises no new safety or effectiveness issues.
#### G. Summary of Design Control Activities
Design control activities employed to control the development of the modification to the Surgical Guide Frame included:
- a comprehensive Risk Analysis to identify potential risks and failures associated with ● operation of the device, any mitigations incorporated to reduce or eliminate those risks and failures, and an assessment of residual risk;
- a comprehensive program of verification and validation activities demonstrating that . acceptance criteria were met, and that design output satisfied design input.
## H. Summary of Clinical Tests
(Not applicable.)
## I. Conclusions
The subject device is substantially equivalent to the currently marketed predicate device, and its development has been adequately and appropriately conducted and validated under a comprehensive design control program complying with Title 21 CFR, §820.30.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
OCT 3 0 2001
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Laetitia Bernard Manager of Regulatory Affairs Nuvasive, Inc. 10065 Old Grove Road San Diego, California 92131
Re: K013257 Trade/Device Name: Hemi-Arc™ Surgical Navigator Regulation Number: 882.4560, 888.1100 Regulation Name: Stereotaxic instrument Arthroscope and accessories Regulatory Class: II Product Code: HAW, HRX Dated: September 28, 2001 Received: October 1, 2001
Dear Ms. Bernard:
We have reviewed your Section 510(k) premarket notification of intent to market the device w one reviews and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for ass sured in the encreases of the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, ateres provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
# Page 2 - Ms. Laetitia Bernard
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Susa Welk, us
fv
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use Statement D.
510(k) Number (if known):
Device Name: Hemi-ArcTM Surgical Navigator
Indications for Use:
The NuVasive™ Guided Spinal Arthroscopy System consists of a Surgical Navigator with adjustable cannula guides, and various percutaneous surgical cannulae and instrumentation. This System is intended to assist in gaining controlled percutaneous access to, and visualization of, the spinal nerves, foramina, vertebra, intervertebral disc, and surrounding tissues of the spine via uniportal or biportal posterior, lateral, or posterolateral approach, where anatomical restrictions safely permit. It is intended for posterelateral time, or near real-time (i.e., successive static image) radiographic visualization via image-intensified C-arm fluoroscopy. The System is designed for use with Nu Vasive™ spinal arthroscopes and percutaneous surgical instrumentation, but may be used with other arthroscopes and instrumentation having compatible diameter and length, and which are indicated for use in the spine.
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | <div> <span></span> </div> |
|----------------------|----------------------------|
| (Per 21 CFR 801.109) | OR |
(Division Sign-Off)
Division of General, Restorative
and Neurological Devices
| 510(k) Number | K013254 |
|---------------|---------|
|---------------|---------|
Page - 10 -
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.