The Datex-Ohmeda S/5 Network and Central is intended to be used with Datex-Ohmeda devices for displaying, storing, printing and otherwise processing information received from other networked devices. The Datex-Ohmeda S/5TM Network and Central transfers information between networked Datex-Ohmeda devices in the Datex-Ohmeda monitor network. It also allows information transfer between several Centrals. Within one Datex-Ohmeda monitor network it allows a networked device to display, store, print and otherwise process information received from other networked devices. The Datex-Ohmeda S/5TM Central maintains the network connections between the Datex-Ohmeda bedside monitors and other networked devices in Datex-Ohmeda monitor network. Furthermore, it coordinates the transfer of information between devices in the Datex-Ohmeda S/5TM Network as well as between the Datex-Ohmeda Network and Hospital Information Systems (HIS). The Datex-Ohmeda S/5TM Central can be used for remote monitor management, storing, printing, viewing or otherwise processing of information from several bedside monitors or other networked devices. The Datex-Ohmeda S/5TM ViewStation can be used for remote monitor management, printing, viewing or otherwise processing of information from several bedside monitors or other networked devices. The Datex-Ohmeda S/5TM Network will be used for patients in the hospital and it is meant for continuous use. The device is for use by qualified personnel only.
Device Story
System consists of networked Datex-Ohmeda monitors and networking hardware (cabling, repeaters). Central station acts as primary communication maintainer; enables real-time display of waveforms, numeric values, and alarms from bedside monitors. Supports anesthesia record keeping, data management, and transfer to Hospital Information Systems (HIS). Features include remote alarm silencing, alarm limit adjustment, and monitor disconnection warnings. ViewStation provides remote display/printing without data storage. Operated by qualified hospital personnel in clinical settings. Enhances patient care via centralized monitoring, data continuity during patient transport, and integrated reporting.
Clinical Evidence
Bench testing only. Compliance with safety and performance standards including EN60950 (electrical safety), EN55022 (EMC), EN1441 (risk analysis), EN475/ISO9703 (alarm signals), and IEEE 802.3 (networking). No clinical data presented.
Technological Characteristics
Networked system using commercial PC hardware, standard graphics controllers, and PC keyboards. Connectivity via IEEE 802.3 Ethernet standards. Supports real-time data transmission, storage, and printing. Software-based central monitoring and remote management. Complies with IEC 60601-1-4 for software safety.
Indications for Use
Indicated for hospital patients requiring continuous physiological monitoring. For use by qualified clinical personnel only.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
Predicate Devices
Datex-Ohmeda S/5 Network and Central and Information Center (K000647)
Submission Summary (Full Text)
{0}------------------------------------------------
DEC 1 1 2001
Image /page/0/Picture/1 description: The image is a black and white logo. The logo consists of a large circle with a white letter "D" inside. There is a small square in the upper right corner of the circle.
# Premarket Notification 510(k) Summary As required by section 807.92 Datex-Ohmeda S/5 Network and Central '01
### GENERAL COMPANY INFORMATION as required by 807.92(a)(1)
### COMPANY NAME/ADDRESS/PHONE/FAX:
Datex-Ohmeda 86 Pilgrim Road Needham, MA 02492 USA Tel: 781-449-8685 Fax: 781-433-1344
# NAME OF CONTACT:
Mr. Joel Kent
## DATE:
September 25, 2001
### DEVICE NAME as required by 807.92(a)(2)
### TRADE NAME:
Datex-Ohmeda S/5 Network and Central '01
## COMMON NAME:
Clinical network and central station
## CLASSIFICATION NAME:
The following Class III classification appears applicable:
System, network and communication, physiological monitors 870.2910
## NAME OF LEGALLY MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3)
The Datex-Ohmeda S/5 Network and Central '01 is substantially equivalent in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda S/5 Network and Central and Information Center (510(k) number: K000647).
K013246 Page 1 of 5
{1}------------------------------------------------
### DEVICE DESCRIPTION as required by 807.92(a)(4)
The Datex-Ohmeda S/5 Network (also referred as D-O Network in the related documentation) is a system, which consists of networked devices (which have separate 510k clearance) and the actual networking hardware. The networked devices are Datex-Ohmeda products containing a network adapter for physical access to the D-O Network as well as software modules supporting network access.
Examples of currently available networked devices are:
1. Datex-Ohmeda S/5 Anesthesia Monitor
2. Datex-Ohmeda S/5 Compact Anesthesia Monitor
3. Datex-Ohmeda S/5 Critical Care Monitor
4. Datex-Ohmeda S/5 Compact Critical Care Monitor
5. Datex-Ohmeda S/5 Light Monitor
6. Datex-Ohmeda S/5 Cardiocap 5 Monitor
7. Datex-Ohmeda S/5 Network and Central, included in this 510(k)
8. Datex-Ohmeda S/5 Telemetry System
The DeioRecorder for Anesthesia (formerly named as Datex-Ohmeda AS/3 Record Keeper.) is also related to the D-O Network as an application using the services provided by the D-O Network.
No changes must be made to the Datex-Ohmeda S/5 Network and Central itself due to a new type of networked device.
The Datex-Ohmeda S/5 Central (also referred to as D-O Central in the related documentation) is the primary maintainer of communication between other networked devices and is, thus, an essential part of the network. The structure and functionality of the revised network corresponds to the structure and functionality of the substantially equivalent predicate device Datex-Ohmeda Network and Central (510(k) number: K000647).
The Datex-Ohmeda Network will be used for real-time communication between devices, for record keeping and for data management in a hospital. Practical examples of currently available features are:
O Transmission and display of measured values and alarms in the Datex-Ohmeda S/5 Central screen (central monitoring) and on the screen of another networked monitor (monitor-to-monitor communication).
O Anesthesia record keeping.
[ Storing and transferring of trend and record keeping data in the network. When the patient is moved from one monitor to another, the data can be transferred with the patient. This feature includes also transferring data from/to an external system (HIS, laboratory, etc.) to/from Datex-Ohmeda S/5 Network.
D Printing of anesthesia records, ICU reports, trend printouts, spirometry loop printouts, waveform snapshot printouts, etc.
The actual networking hardware consists of cabling, patch panels, racks, connectors, repeaters, etc. The networking hardware is similar to the networking hardware of the substantially equivalent predicate device Datex-Ohmeda Network and Central (510(k) number: K000647).
The Datex-Ohmeda S/5 ViewStation is a D-0 Central version that can show real-time curves and numeric information from any monitor residing in the Datex-Ohmeda Network. It also allows printing to laser printer or recording to a strip-chart recorder. The Datex-Ohmeda S/5 ViewStation does not store patient data, or provide any other network services
{2}------------------------------------------------
than display and printing services. The ViewStation uses the same hardware and a subset of the software used by the main Central.
The Datex-Ohmeda S/5 Telemetry System Network is a computer-based system for monitoring patients using telemetry. It consists of a PC based Central Station including moments patients assurence as and up to 16 telemetry transmitters per station. The central station supports arrhythmia monitoring, and measuring and trending of ST changes. An Station Jupports arres has been developed to enable connection of Telemetry central to D-O STS TeleNot I ackage enables transfer of ECG waveforms, ECG related parameters and archythmia information to the D-O Network. Also bed-to-bed services from telemetry sessions to bedside monitors are made available.
The new version of Datex-Ohmeda S/5 Network and Central adds the following features: Commercial hardware: The display driver is a commercial board instead of D-O proprietary boards. The proprietary keyboard is not used anymore, instead there is a commercial keyboard with dedicated "EasyAccess" keys
O User Interface is now mouse controlled in addition to keyboard controlled O Numerical and graphical trends have been added
O Alarm management functionality has been extended: remotely silencing of bedside alarms and changing alarm limits from Datex-Ohmeda S/5 Central or ViewStation have been added
□ Monitor disconnection warning on Central has been added
□ Possibility to start a network recorder from a bedside monitor has been added D TeleNet Package to enable connection of Datex-Ohmeda S/5 Telemetry System to D-O
Network
### INTENDED USE as required by 807.92(a)(5)
The Datex-Ohmeda S/5 Network and Central is intended to be used with Datex-Ohmeda devices for displaying, storing, printing and otherwise processing information received from other networked devices.
The Datex-Ohmeda S/5TM Network and Central transfers information between networked Datex-Ohmeda devices in the Datex-Ohmeda monitor network. It also allows information transfer between several Centrals. Within one Datex-Ohmeda monitor network it allows a networked device to display, store, print and otherwise process information received from other networked devices.
The Datex-Ohmeda S/5TM Central maintains the network connections between the Datex-Ohmeda bedside monitors and other networked devices in Datex-Ohmeda monitor network. Furthermore, it coordinates the transfer of information between devices in the Datex-Ohmeda S/5TM Network as well as between the Datex-Ohmeda Network and Hospital Information Systems (HIS).
The Datex-Ohmeda S/5TM Central can be used for remote monitor management, storing, printing, viewing or otherwise processing of information from several bedside monitors or other networked devices.
The Datex-Ohmeda S/5TM ViewStation can be used for remote monitor management, printing, viewing or otherwise processing of information from several bedside monitors or other networked devices.
The Datex-Ohmeda S/5TM Network will be used for patients in the hospital and it is meant for continuous use.
The device is for use by qualified personnel only.
{3}------------------------------------------------
### SUMMARY OF TECHNOLOGICAL CHARACTERITICS OF DEVICE COMPARED TO THE PREDICATE DEVICE as required by 807.92(a)(6)
The Datex-Ohmeda S/5 Network and Central is substantially equivalent in safety and effectiveness to the Datex-Ohmeda S/5 Network and Central and Information Center (510(k) number: K000647).
Similarities:
The indications for use is the same as in predicate except that "remote monitor management" has been expressed in the statement.
The structure and functionality of the Datex-Ohmeda S/5 Network and Central '01 closely corresponds to the structure and functionality of the Datex-Ohmeda Network and Central '99 (predicate). The basic architecture of Datex- Ohmeda S/5 Network and Central '01 is the same as that of Datex-Ohmeda Network and Central '99 (predicate).
The user interface uses the same basic functionality: Single patient view and multiple patient views are similar to the predicate. Alarm indications are similar.
The ability to have audible alarm indications only on the Central is similar to the predicate.
Differences:
The hardware platform now uses commercial graphics controller and normal PC keyboard instead of D-O proprietary versions.
The user interface can now also be controlled with a mouse. The Central uses a PC keyboard that has labeled special "Easy Access" keys. The earlier keyboard was proprietary with special keys.
Remote monitor management adds the option to silence bedside monitor alarms from the Central and to adjust monitor alarm limits. These functions are by default disabled. They can be activated by the system administrator and activation is password protected.
Numerical and graphical trend display has been added to the Central.
Monitor disconnection warning has been added since Central may in some cases be used for monitoring alarms.
Possibility to start a network recorder from monitor has been added to support work methods of e.g. some intensive care units.
A SW package to enable S/5 Telemetry system integration to the D-O Network has been developed. It uses similar connection to the network as the Central.
Summary:
In summary, the new Datex-Ohmeda S/5 Network and Central, described in this submission is substantially equivalent to the predicate (K000647) device.
{4}------------------------------------------------
# SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3)
The Datex-Ohmeda S/5 Network and Central '01 complies with the safety standards below and is therefore safe and effective for the intended use. The device has been thoroughly and is therefore safe and orietr. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . environmental tolerance, software validation and verification of specifications. environmental toloration, solling following mandatory and voluntary standards has been made:
- EN60950: 1995 (IEC950 2nd edition) Safety of information technology equipment, 그 including electrical business equipment
- EN 55022: 1994 (IEC-CISPR 22) -- Information technology equipment Radio 내 disturbance characteristics. Limits and methods of measurement
- distance characteristics is not 2011-3, IEC 801-4) -- Electromagnetic compatibility -. Generic immunity standard
- EN 1441, Medical devices Risk analysis l
- EN 475, Medical devices Electrically-generated alarm signals l
- ISO 9703-1, ISO 9703-2, Anesthesia and respiratory care alarm signals l
- . IEC 60601-1-4
- CAN/CSA-C22.2 No 950: Information Technology Equipment Including Electrical 미 Business Equipment
- UL1950: Information Technology Equipment Including Electrical Business Equipment 1
- ISO/IEC 8802-3 (ANSI/IEEE 802.3), EIA/TIA-568, EIA/TIA-TSB40, international 내 network cabling standards
### Conclusion:
- The summary above shows that there are no new questions of safety and effectiveness for the Datex-Ohmeda S/5 Network and Central as compared to the predicate device.
{5}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
: 上一篇
DEC 1 1 2001
Mr. Joel C. Kent Manager, Quality and Regulatory Affairs Datex Ohmeda 86 Pilgrim Road Needham, MA 02492
Re: K013246
Trade Name: Datex-Ohmeda S/5 Network and Central '01 Regulation Number: 21 CFR 870.2300 Regulation Name: Network and Communication Physiological System Regulatory Class: Class II (two) Product Code: MSX Dated: September 26, 2001 Received: September 28, 2001
Dear Mr. Kent:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave to to wowed your becally is the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 101 use stated in the encreatinent date of the Medical Device Amendments, or to conninered pror to that 2011-17-11, in accordance with the provisions of the Federal Food, Drug, devices that have been recuired in assee approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, dicrorolo, mance of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is olassified (600 we controls. Existing major regulations affecting your device can may be subject to sach additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
## Page 2 - Mr. Joel C. Kent
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I odetar band the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Of A rat 0077, accesses (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally premarket notineated. "The sification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 11 Jou desire specific and 10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Doulla Teth
Bram D. Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
510(k) Number (if known): __
Device Name: Datex-Ohmeda S/5 Network and Central '01
K013246
Indications For Use:
The Datex-Ohmeda S/5™ Network and Central transfers information The Datex-Ohmeda 3/5 - Network dha Sossia Costex-Ohmeda monitor
between networked Datex-Ohmeda devices in the Datex-Ohmeda between lietworked Dutox Onation transfer between several Centrals.
It also allows information transfer between severalsed des network. It also anows information namses it allows a networked device Within one DateX-Onmedia monitor news.
to display, store, print and otherwise process information received from other networked devices.
The Datex-Ohmeda S/5™ Central maintains the network connections The Datex-Onmeda S/3 - Gentral manitors and other networked devices
between the Datex-Ohmeda bedside monitors and other networth o between the Datex-Ohmeda society. Furthermore, it coordinates the in Datex-Olimeda monitor noen of ices in the Datex-Olimeda S/5™
transfer of information between devices in the Datex-Olimed Hooped Hoose transfer of Information between the Datex-Ohmeda Network and Hospital Information Systems (HIS).
The Datex-Ohmeda S/5™ Central can be used for remote monitor The Datex-Onlineuu 0/0 - - Ong, viewing or otherwise processing of management, storing, printing, viowing or other networked devices.
information from several bedside monitors or other networked devices.
The Datex-Ohmeda S/5™ ViewStation can be used for remote monitor The Datex-Onlineda 0/3 - - V101 One of otherwise processing of information management, printing, "citions" or other networked devices.
. The Datex-Olimeda S/5™ Network will be used for patients in the hospital and it is meant for continuous use.
The device is for use by qualified personnel only.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | <div> <span style="text-decoration: overline;"></span> <span style="float: right;">X</span> </div> |
|----------------------|----------------------------------------------------------------------------------------------------|
| (Per 21 CFR 801.109) | |
OR
Over-The-Counter Use _
(Optional Format 1-2-96)
**Division of Cardiovascular & Respiratory Devices**
510(k) Number K013246
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.