INSTACLOT FOR OTC USE, TRAUMASSTAT FOR PRESCRIPTION USE
K013225 · Medafor, Inc. · KMF · Dec 26, 2001 · General Hospital
Device Facts
Record ID
K013225
Device Name
INSTACLOT FOR OTC USE, TRAUMASSTAT FOR PRESCRIPTION USE
Applicant
Medafor, Inc.
Product Code
KMF · General Hospital
Decision Date
Dec 26, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5090
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
Over the Counter, OTC: Bleed-X™ is intended as a topical dressing for the local management of bleeding wounds such as minor cuts, lacerations and abrasions. Prescription, Rx: Traumadex™ is intended as a topical dressing for the local management of bleeding wounds such as cuts, lacerations and abrasions. Under the care of a health care professional, Traumadex™ may be used for the temporary treatment of severely bleeding wounds such as surgical wounds (post-operative, donor sites, dermatological), cuts, lacerations and traumatic injuries.
Device Story
Bleed-X and Traumadex consist of dry, sterile, spherical particles made from purified plant-based polysaccharide. Particles act as molecular sieves, excluding large proteins and cells. Upon contact with blood, particles rapidly dehydrate blood, causing hemoconcentration; serum proteins and cells form a viscous gel. Normal platelet activation and fibrin deposition within this gel create a tenacious clot to limit bleeding. Used in clinical or home settings; applied topically to wounds. Benefits include accelerated bleeding control. Device is provided in single-use containers or applicators, terminally sterilized via gamma irradiation.
Clinical Evidence
Clinical investigation using human subjects in a side-by-side comparison of Bleed-X/Traumadex against a control of normal clotting demonstrated an accelerated rate of bleeding control in the presence of the particles. Animal testing using a porcine model also confirmed efficacy in bleeding control.
Technological Characteristics
Dry, sterile, spherical particles of purified plant-based polysaccharide. Controlled porosity via emulsion cross-linking process. Biocompatibility per ISO 10993-1 and FDA G95-1. Sterilization via Cobalt-60 gamma irradiation per AAMI/ANSI/ISO 11137 and EN552 (SAL 1x10^-6). Packaged in medical-grade plastic applicators or laminate pouches.
Indications for Use
Indicated for local management of bleeding wounds including minor cuts, lacerations, and abrasions (OTC). For prescription use, indicated for temporary treatment of severely bleeding wounds including surgical wounds (post-operative, donor sites, dermatological), cuts, lacerations, and traumatic injuries under professional care.
Regulatory Classification
Identification
A liquid bandage is a sterile device that is a liquid, semiliquid, or powder and liquid combination used to cover an opening in the skin or as a dressing for burns. The device is also used as a topical skin protectant.
Predicate Devices
Calgon Vestal Labs, Kaltostat Wound Dressing
Hemostace
Sorbastace
Kendall Hydrophylic Powder Wound Dressing
Marine Polymer Technology Healtec and Syvek Patch
Dermaphylyx Calcium Alginate Wound Dressing
Closure Medical, Liquiderm Liquid Bandage
Marine Polymer Technology RDH Bandage
Submission Summary (Full Text)
{0}------------------------------------------------
Page 19
013225
1 1/28/2001
# DEC 2 6 2001
#### 510(k) Summary of Safety and Effectiveness h.
1. General Company Information
| Name: | Medafor, Inc. |
|----------------|------------------------------------------------------------|
| Address: | 5201 East River Road<br>Suite 312<br>Minneapolis, MN 55421 |
| Telephone: | 763/571-6300 |
| FAX: | 763/571-1035 |
| Contact: | Philip B. Jarvi, Director of Regulatory Affairs |
| Date Prepared: | September 26, 2001 |
## 2. General Device Information
| Product Name: | Bleed-XTM and TraumadexTM containing HemadexTM Clotting Beads |
|------------------------|---------------------------------------------------------------------|
| Classification:<br>KMF | Liquid Bandage, Unclassified (General Controls), Product Code - KMF |
- 3. Predicate Devices
Calgon Vestal Labs, Kaltostat Wound Dressing
Hemostace, Sorbastace
Kendall Hydrophylic Powder Wound Dressing
Marine Polymer Technology Healtec and Syvek Patch
Dermaphylyx Calcium Alginate Wound Dressing
Closure Medical, Liquiderm Liquid Bandage
Marine Polymer Technology RDH Bandage
{1}------------------------------------------------
#### Page 20
#### 4. Description
Meclafor, Inc.'s Bleed-X™ and Traumadex™ consists of dry, sterile, controlled porosity, spherical particles manufactured from purified plant based polysaccharide. The porosity is controlled such that the particles act as molecular sieves excluding large proteins and cells.
The bleeding ccssation is accomplished by the rapid dehydration and subsequent hemoconcentration of blood in contact with the particles. The concentration of serum proteins and cells produces a viscous gel. Normal plateler activation and fibrin deposition within the congealed blood produces a clot that limits further bleeding.
#### 5. Indications
#### Over the Counter, OTC:
Blocd-X™ is intended as a topical dressing for the local management of blocding wounds such as minor cuts, lacerations and abrasions.
#### Prescription, Rx:
Traumadex™ is intendod as a topical dressing for the local management of bleeding wounds such as cuts, laccrations and abrasions. Under the care of a health care professional, Traumadex™ may be used for the temporary treatment of severely bleeding wounds such as surgical wounds (post-operative, donor sites, dermatological), cuts, lacerations and traumatic injuries.
#### 6. Substantial Equivalence
This section summarizes information about the product and outlines the tests and clinical
information we believe will provide sufficient data to support our claims of efficacy and
safety.
Medafor, Inc.'s Hemadcx cloting beads consists of dry, sterile, controlled porosity, spherical particles manufactured from purified plant based polysaccharide using an emulsion cross-liking process.
The particles will be packaged in single use containers with a moisture-protective laminate outer wrap. The final package will be terminally sterilized using gamma irradiation. Contract manufacturers will do both packaging and sterilization, following written contracts with Mcdafor.
{2}------------------------------------------------
#### Page 21
Contract manufacturers will do both packaging and sterilization, following written contracts with Medafor.
Bleed-XTM will be available for OTC use in sizes from .5 to 10 grams using a medical grade plastic applicator package or a "peel and pour" laminate pouch.
Traumadex™ will be available for prescription use in sizes from 1 to 10 grams using a medical grade plastic applicator.
The hemostatic effect of the particles is produced by the rapid dehydration and subsequent hemoconcentration of blood in contact with the particles. The concentration of serum proteins and cells produces a viscous gel. Normal platelet activation and fibrin deposition within the congealed blood produces a tenacious clot that limits further bleeding.
Biological testing per ISO 10993-1 and FDA G95-1 demonstrates the biological compatibility of Bleed-XTM and TraumadexTM.
Sterilization validation of the Cobalt- 60 gamma irradiation process per AAMI / ANSI / ISO 11137 and EN552 demonstrate a sterility assurance level (SAL) of 1 X 10-6.
Animal testing using a porcine model shows the efficacy of Bleed-XTM and Traumadex™ in the control of bleeding.
A non-significant risk clinical investigation using human subjects in a side by side comparison of Bleed-X ™ / Traumadex™ and a control of normal clotting demonstrated an accelerated rare of bleeding control in the presence of the particles.
The initial indications are for topical application as an aid in the control of bleeding. This product will be intended for both prescription and OTC use and labels is worded accordingly.
Medafor, Inc. believes that as a result of the biocompatibility testing, the physical analysis, the animal studics and the human clinical that the Bleed-XTM and Traumadex™ containing Hemadex™ clotung beads are safe and effective for the control of bleeding wounds and perform in a manner equivalent and similar to the predicates. The device has been shown to effectively control bleeding. It has been demonstrated to be biocompatible per an internationally recognized standard and it has demonstrated to be procedurally and physiologically appropriate in medical upplications.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular in shape, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle with three lines representing its wings.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 2 6 2001
Medafor, Inc. Mr. Philip B. Jarvi Director of Clinical and Regulatory Affairs 5201 East River Road, #312 Minneapolis, Minnesota 55421-1035
Re: K013225 Trade Name: Traumadex™ and Bleed-X™ containing Hemadex Clotting Beads Regulation Number: 880.5090 Regulation Name: Liquid Bandage Regulatory Class: I Product Code: KMF Dated: November 28, 2001 Received: December 3, 2001
Dear Mr. Jarvi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
#### Page 2 - Mr. Philip Jarvi
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Neikhr Ogden
Celia M. Witten, Ph.D., M.D. for
Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
12/19/2001
Page 24
#### Indications for Use Statement
Applicant: Medafor, Inc.
510(k) Number (if known): K013225
Device name: Bleed-X TM and TraumadexTM
#### Intended UseAndications for Use:
Over the Counter, OTC:
Bleed-X™ is intended as a topical dressing for the local management of bleeding wounds such as minor cuts, lacerations and abrasions.
Prescription, Rx:
Traumadex™ is intended as a topical dressing for the local management of bleeding wounds such as cuts, lacerations and abrasions. Under the care of a health care professional, Traumadex™ may be used for the temporary treatment of severely bleeding wounds such us surgical wounds (post-operative, donor sites, dermatological), cuts, lacerations and traumatic injuries.
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
NRe for cmw
Division Sign-Off Division of General, Restorative and Neurological Devices
**510(k) Number** K013225
Prescription Use /
OR
Over-The-Counter \$\nearrow\$
(Per 21 CFR 801.109) 1-86)
(Optional Format 1-
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.