← Product Code [GEX](/productcode/GEX) · K013201

# CARDIOFOCUS DIODE LASER SYSTEM, MODELS 60R, 60NR, 30NR (K013201)

_Cardiofocus, Inc. · GEX · Mar 18, 2002 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K013201

## Device Facts

- **Applicant:** Cardiofocus, Inc.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Mar 18, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The CardioFocus Laser Diode System is intended for the delivery of 980nm laser light to soft tissue in contact or noncontact mode during surgical procedures including via endoscopes, introducers or catheters. Indications include the incision, excision, dissection, vaporization, ablation or coagulation of soft tissue,

## Device Story

CardioFocus Diode Laser System is a surgical generator delivering 980nm laser light for soft tissue procedures. System includes diode laser module, power supply, and thermal safety system (TSS). TSS monitors black body radiation from disposable delivery devices (catheters/endoscopes) to detect excessive heating and disable laser output. Operated by clinicians in surgical settings; controlled via foot switch, keypad, and LCD interface. Provides therapeutic energy for hemostasis, incision, ablation, coagulation, and vaporization. Benefits include precise tissue interaction via endoscopic or catheter-based delivery.

## Clinical Evidence

Bench testing only. Safety and performance verified through conformance to UL 2601-1, IEC 60601-2-22, IEC 60825-1, IEC 60601-1-2, IEC 1000, and 21 CFR 1040.10/.11.

## Technological Characteristics

980nm diode laser generator; 30W (Model 30NR) or 60W (Model 60NR) output. Includes thermal safety system (TSS) with IR detector for black body radiation monitoring. Interfaces: LCD display, keypad, foot switch. Standards: UL 2601-1, IEC 60601-2-22, IEC 60825-1, IEC 60601-1-2, IEC 1000, 21 CFR 1040.10/.11. Internal AC/DC power supply.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Ceralas Diode Laser System CeramOptec, Inc. ([K001975](/device/K001975.md))

## Reference Devices

- CardioFocus Lightstic 180L ([K993834](/device/K993834.md))
- CardioFocus Lightstic 180C ([K011988](/device/K011988.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

MAR 1 8 2002

K013201 1/4

## 510(k) SUMMARY FOR CARDIOFOCUS DIODE LASER SYSTEM

#### Submitter's Name, Address, Telephone Number, and Contact Person:

CardioFocus Norton Commerce Center 10 Commerce Way Norton, MA 02766

Contact: Joe Curtis Phone: (508) 285-1700 Fax: (508) 285-7579 Email: joec@cardiofocus.com

#### Name of Device:

CardioFocus Laser System.

#### Common or Usual Name:

Surgical Laser Instrument

#### Classification Name:

Laser, Powered, Surgical Instrument. CFR ref: 878.4810, product Code GEX

#### Predicate Devices:

Ceralas Diode Laser System CeramOptec, Inc., East Longmeadow, MA K001975; S.E. 8-4-2000

#### Intended Use:

The CardioFocus Laser Diode System is intended for the delivery of 980nm laser light to soft tissue in contact or noncontact mode during surgical procedures including via endoscopes, introducers or catheters. Indications include the incision, excision, dissection, vaporization, ablation or coagulation of soft tissue,

{1}------------------------------------------------

#### General Description:

The CardioFocus Diode Laser Ablation System is a generator designed for the delivery of 980nm laser light and may be used in conjunction with surgical treatment for hemostasis, incision, ablation, coagulation and vaporization of tissue as required by the clinician. The instrument includes a diode laser module as the therapeutic energy source, as well as an internal thermal safety system for monitoring and disabling laser power output.

#### Device Characteristics:

The CardioFocus Laser Diode System is provided in two Models. The Model 30NR provides up to 30 watts, laser light and the Model 60NR provides up to 60 watts, laser light., The CardioFocus Diode Laser System may be used with the CardioFocus Lightstic 180L or Lightstic 180C delivery catheter (REF: 510K# K993834 and K011988), and successor delivery catheters from CardioFocus, Inc.

System safety characteristic are assured through conformance with applicable U.S. and International standards, including :

- UL 2601-1 Medical Electrical Equipment PART 1: General t Requirements For Safety, Medical Electrical Equipment
- IEC 60601-2-22 Medical Electrical Equipment PART2: Particular . requirements for the safety of diagnostic and therapeutic laser equipment
- IEC 60825-1 Safety of laser products Part 1: Equipment . classification, requirements and user's guide
- IEC 60601-1-2 (EN 60601-1-2) General Requirements for Safety, 2. . Collateral Standard: Electromagnetic Compatibility
- IEC 1000 Electromagnetic Compatibility .
- 21 CFR 1040.10/.11 Performance Standard for Light Emitting ● Products

K013201 2/

{2}------------------------------------------------

# K013201 3/4

#### Technical Characteristics

- System Controller: the system controller represents the hardware . and software that monitors and controls the functioning of the laser instrument.
- Thermal Safety System (TSS): the thermal safety system is a safety 0 circuit, intended to monitor the black body radiation associated with excessive heating in the disposable device and associated optics and to disable the laser in the event that excessive black body radiation is detected.
- Optics Module: the optics module is an optical sub-system which . provides means for delivering laser energy and collecting white light energy from the delivery device and also provides means for directing black body radiation from the disposable device to the safety system IR detector.
- User Interface: the user interface provides means for information . display to the user as well as a mechanism for user input. The user interface for the laser instrument will likely include an LCD display and a keypad.
- Foot Switch: the laser instrument will interface to a foot switch for . on/off control of the laser emission.
- Diode Driver: the diode driver block represents the electronics ● required to control power supplied to the laser module
- . Diode Module: the diode module is a purchased component, comprised of a series of laser diodes, which produces the output laser energy.
- . Power Supply: an AC/DC power supply provides DC power necessary to run the system electronic components. The supply is internal to the laser instrument.

{3}------------------------------------------------

K0132014/4

## Predicate information is support of Substantial equivalence

The CardioFocus Diode Laser System is substantially equivalent to the CeramOptec Celaras D25 Fiber-Coupled Diode Laser System. Both devices ocruth optor energy at similar power levels and similar wavelengths, have similar intended uses and provide for safety though conformance to recognized standards.

#### Conclusion

There are more similarities than differences between the predicate device and the CardioFocus Diode laser System. Both the predicate devices use the same technology and theory of operation. Both the CardioFocus Laser System and the Ceralas Diode laser System use three configurations for different power levels and both devices have similar Indications for Use When used in accordance with the directions for use, by qualified personnel, the CardioFocus Diode Laser System is safe and effective, as indicated, for its intended use,

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with flowing lines, representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. The logo is presented in black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 1 8 2002

CardioFocus, Inc. c/o Mr. John Greenbaum Generic Devices Consulting 20310 SW 48th Street Ft. Lauderdale, FL 33332

Re: K013201

Trade/Device Name: CardioFocus Diode Laser System Regulation Number: 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: December 20, 2001 Received: December 21, 2001

Dear Mr. Greenbaum:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave teviewed your becamed the device is substantially equivalent (for the indications felerenced above and nave acterimes and marketed predicate devices marketed in interstate 10 tuse stated in the encrosale) to tegains and the Medical Device Amendments, or to Connected pror to May 20, 1978, as eacordance with the provisions of the Federal Food, Drug, de necs that have been resuire approval of a premarket approval application (PMA). and Cosmetic Hot (110) that as nov request to the general controls provisions of the Act. The I ou may, mereleve, manis of the Act include requirements for annual registration, listing of Economic controls profictive, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device to such additional controls. Existing major regulations affecting your device can may be subject to basil as a suese as Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I Drinas intatutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Orth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2 - Mr. John Greenbaum

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301):443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Muriam C. Provost

for Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Page 1 of 1

510(k)Number(if known): KOJ320

Device Name: CardioFocus Diode Laser System

Indications For Use:

The CardioFocus Diode Laser System is intended for the delivery of 980nm laser light to soft tissue in contact or non-contact mode during surgical procedures including via endoscopes, introducers or catheters. Indications include the incision, excision, dissection, vaporization, ablation or coagulation of soft tissue.

Miriam C. Provost

(Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number K013201

### PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use_X (Per 21 CRF 801.109) OR

Over-The-Counter Use

(Optional Format 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K013201](https://fda-staging.innolitics.com/device/K013201)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com